J Reconstr Microsurg 2006; 22(7): 519-524
DOI: 10.1055/s-2006-951317
REVIEW

Copyright © 2006 by Thieme Medical Publishers, Inc., 333 Seventh Avenue, New York, NY 10001, USA.

Validation of a 1-mm Transit Time Flow Probe and the Potential for Use in Microsurgery

Christian T. Bonde1 , Niels-Henrik Holstein-Rathlou2 , Jens J. Elberg1
  • 1Department of Plastic Surgery and Treatment of Burns, Center of Hand and Orthopaedics, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark
  • 2Department of Medical Physiology, Panum Institute, University of Copenhagen, Denmark
Further Information

Publication History

Accepted: April 24, 2006

Publication Date:
17 October 2006 (online)

ABSTRACT

The objective of this study was to validate a prototype transit time ultrasound flow probe (Medi-Stim Butterfly Flowmeter BF 2004) of only 1 mm and to evaluate its potential for use in microsurgery. Knowledge of the ability of this kind of system to measure flow in small vessels and at small flow rates did not previously exist. The experimental design was an in vitro study using the median artery of 15-kg pigs. The vessel was perfused by the pigs' own heparinated blood, at a known flow rate, using a Gemini infusion pump. The infusion rate was increased stepwise, starting at 1 ml/min until 16 ml/min. Transit time flow was measured continuously. The results showed very accurate flow measurements at all flow rates when compared to the known infusion rate of the Gemini pump. However, at flow rates below 4 ml/min, variation became a problem.

REFERENCES

  • 1 Lorenzetti F, Ahovuo J, Suominen S, Salmi A, Asko-Seljavaara S. Colour Doppler ultrasound evaluation of haemodynamic changes in free TRAM flaps and their donor sites.  Scand J Plast Reconstr Surg Hand Surg. 2002;  36 202-206
  • 2 Tuominen H P, Asko-Seljavaara S, Svartling N E. Cutaneous blood flow in the free TRAM flap.  Br J Plast Surg. 1993;  46 665-669
  • 3 Nabawi A, Gurlek A, Patrick Jr C W et al.. Measurement of blood flow and oxygen tension in adjacent tissues in pedicled and free flap head and neck reconstruction.  Microsurgery. 1999;  19 254-257
  • 4 Segadal L, Matre K, Engedal H, Resch F, Grip A. Estimation of flow in aortocoronary grafts with a pulsed ultrasound Doppler meter.  Thorac Cardiovasc Surg. 1982;  30 265-268
  • 5 Drost C J. Vessel diameter-independent volume flow measurements using ultrasound.  San Diego Biomedical Symposium. 1978;  17 299-302
  • 6 Lundell A, Bergqvist D, Mattsson E, Nilsson B. Volume blood flow measurements with a transit time flowmeter: an in vivo and in vitro variability and validation study.  Clin Physiol. 1993;  13 547-557
  • 7 Walpoth B H, Bosshard A, Genyk I et al.. Transit-time flow measurement for detection of early graft failure during myocardial revascularization.  Ann Thorac Surg. 1998;  66 1097-1100
  • 8 Lorenzetti F, Suominen S, Tukiainen E et al.. Evaluation of blood flow in free microvascular flaps.  J Reconstr Microsurg. 2001;  17 163-167
  • 9 Lorenzetti F, Salmi A, Ahovuo J, Tukiainen E, Asko-Seljavaara S. Postoperative changes in blood flow in free muscle flaps: a prospective study.  Microsurgery. 1999;  19 196-199
  • 10 Lorenzetti F, Tukiainen E, Alback A, Kallio M, Asko-Seljavaara S, Lepantalo M. Blood flow in a pedal bypass combined with a free muscle flap.  Eur J Vasc Endovasc Surg. 2001;  22 161-164
  • 11 Franklin D L, Baker D W, Rushmer R F. Pulsed ultrasonic transit time flowmeter.  IRE Transac Bio-med Electronics. 1962;  9 44-49
  • 12 Plass K G. A new ultrasonic flowmeter for intravascular application.  IEEE Trans Biomed Eng. 1964;  11 154-156
  • 13 Department of Health MDD .Evaluation 126. Volumetric Infusion Pump Imed Gemini PC-1. Department of Health, Medical Devices Directorate 2004 1992: 1-17
  • 14 Laustsen J, Pedersen E M, Terp K et al.. Validation of a new transit time ultrasound flowmeter in man.  Eur J Vasc Endovasc Surg. 1996;  12 91-96
  • 15 Matre K, Birkeland S, Hessevik I, Segadal L. Comparison of transit-time and Doppler ultrasound methods for measurement of flow in aortocoronary bypass grafts during cardiac surgery.  Thorac Cardiovasc Surg. 1994;  42 170-174
  • 16 Hartman J C, Olszanski D A, Hullinger T G, Brunden M N. In vivo validation of a transit-time ultrasonic volume flow meter.  J Pharmacol Toxicol Methods. 1994;  31 153-160
  • 17 Alback A, Makisalo H, Nordin A, Lepantalo M. Validity and reproducibility of transit time flowmetry.  Ann Chir Gynaecol. 1996;  85 325-331
  • 18 Lorenzetti F, Kuokkanen H, von Smitten K, Asko-Seljavaara S. Intraoperative evaluation of blood flow in the internal mammary or thoracodorsal artery as a recipient vessel for a free TRAM flap.  Ann Plast Surg. 2001;  46 590-593
  • 19 Mortensen F V, Rasmussen J S, Viborg O, Laurberg S, Pedersen E M. Validation of a new transit time ultrasound flowmeter for measuring blood flow in colonic mesenteric arteries.  Eur J Surg. 1998;  164 599-604
  • 20 D'Almeida M S, Gaudin C, Lebrec D. Validation of 1- and 2-mm transit-time ultrasound flow probes on mesenteric artery and aorta of rats.  Am J Physiol. 1995;  268(3 Pt 2) H1368-H1372
  • 21 Suominen S, Svartling N, Silvasti M, Niemi T, Kuokkanen H, Asko-Seljavaara S. The effect of intravenous dopamine and dobutamine on blood circulation during a microvascular TRAM flap operation.  Ann Plast Surg. 2004;  53 425-431
  • 22 Cordeiro P G, Santamaria E, Hu Q Y, Heerdt P. Effects of vasoactive medications on the blood flow of island musculocutaneous flaps in swine.  Ann Plast Surg. 1997;  39 524-531

Christian T BondeM.D. Ph.D. 

Department of Plastic Surgery and Treatment of Burns 2102, Copenhagen University Hospital

Rigshospitalet, Blegdamsvej 9, DK-2100 Copenhagen, Denmark

    >