Endoscopy 2016; 48(02): 141-148
DOI: 10.1055/s-0034-1393331
Original article
© Georg Thieme Verlag KG Stuttgart · New York

Clinical outcomes following stent placement in refractory benign esophageal stricture: a systematic review and meta-analysis

Lorenzo Fuccio
1   Department of Medical and Surgical Sciences, S. Orsola-Malpighi Hospital, University of Bologna, Bologna, Italy
,
Cesare Hassan
2   Gastroenterology Unit, Nuovo Regina Margherita Hospital, Rome, Italy
,
Leonardo Frazzoni
1   Department of Medical and Surgical Sciences, S. Orsola-Malpighi Hospital, University of Bologna, Bologna, Italy
,
Rossella Miglio
3   Department of Statistical Sciences “Paolo Fortunati”, University of Bologna, Bologna, Italy
,
Alessandro Repici
4   Digestive Endoscopy Unit, Division of Gastroenterology, Humanitas Research Hospital, Rozzano (MI), Italy
› Author Affiliations
Further Information

Publication History

submitted: 29 April 2015

accepted after revision: 30 August 2015

Publication Date:
03 November 2015 (online)

Background and aims: The management of refractory benign esophageal strictures (RBESs) is challenging. Stent placement has been proposed as a possible rescue strategy. We performed a systematic review and meta-analysis to examine the efficacy of this strategy in the long-term resolution of dysphagia.

Methods: PubMed, SCOPUS, and Google Scholar were searched (up to January 2015). Studies recruiting adults with RBES treated with stent placement were eligible. The success, adverse event, and migration rates were pooled by means of a random-effect model to obtain odds with a 95 % confidence interval (CI).

Results: Eighteen studies (444 patients) were eligible for inclusion. The pooled clinical success rate was 40.5 % (95 %CI 31.5 % – 49.5 %), yielding odds of 0.68 (95 %CI 0.46 – 0.98) with high heterogeneity (I2 = 65.0 %). The meta-regression analysis showed stricture etiology as the only influencing factor. Treatment with self-expanding plastic (SEPSs) and metal stents (SEMSs) did not result in significantly higher success rates than treatment with biodegradable stents (BDSs) (46.2 % vs. 40.1 % vs. 32.9 %, respectively). The migration rate was 28.6 % (95 %CI 21.9 % – 37.1 %), yielding odds of 0.40 (95 %CI 0.28 – 0.59), with migration rates for SEPSs and SEMSs reported as being not significantly higher than BDSs (33.3 % vs. 31.5 % vs. 15.3 %, respectively). The adverse event rate was 20.6 % (95 %CI 15.3 % – 28.1 %), yielding odds of 0.26 (95 %CI 0.18 – 0.39) with no significant difference between stent types (19.4 % vs. 21.9 % vs. 21.9 %, respectively).

Conclusions: Stent placement in patients with RBES is effective in about 40 % of cases. Further studies should investigate whether the clinical success rate varies according to the stricture etiology.

(PROSPERO registration number: CRD42015019017.)

Appendix e1, Fig. e3 – e5, e8 and Table e3 – e6

 
  • References

  • 1 Siersema PD. Treatment options for esophageal strictures. Nat Clin Pract Gastroenterol Hepatol 2008; 5: 142-152
  • 2 Spechler SJ. American Gastroenterological Association medical position statement on treatment of patients with dysphagia caused by benign disorders of the distal esophagus. Gastroenterology 1999; 117: 229-233
  • 3 Repici A, Hassan C, Sharma P et al. Systematic review: the role of self-expanding plastic stents for benign oesophageal strictures. Aliment Pharmacol Ther 2010; 31: 1268-1275
  • 4 Thomas T, Abrams KR, Subramanian V et al. Esophageal stents for benign refractory strictures: a meta-analysis. Endoscopy 2011; 43: 386-393
  • 5 van Boeckel PG, Vleggaar FP, Siersema PD. A comparison of temporary self-expanding plastic and biodegradable stents for refractory benign esophageal strictures. Clin Gastroenterol Hepatol 2011; 9: 653-659
  • 6 Repici A, Vleggaar FP, Hassan C et al. Efficacy and safety of biodegradable stents for refractory benign esophageal strictures: the BEST (Biodegradable Esophageal Stent) study. Gastrointest Endosc 2010; 72: 927-934
  • 7 Dan DT, Gannavarapu B, Lee JG et al. Removable esophageal stents have poor efficacy for the treatment of refractory benign esophageal strictures (RBES). Dis Esophagus 2014; 27: 511-517
  • 8 Saito Y, Tanaka T, Andoh A et al. Novel biodegradable stents for benign esophageal strictures following endoscopic submucosal dissection. Dig Dis Sci 2008; 53: 330-333
  • 9 Shamseer L, Moher D, Clarke M the PRISMA-P Group. Preferred reporting items for systematic review and meta-analysis protocols (PRISMA-P) 2015: elaboration and explanation. BMJ 2015; 349: g76471
  • 10 Well G, Shea B, O’Connell J et al. The Newcastle-Ottawa scale (NOS) for assessing the quality of nonrandomized studies in meta-analysis. Available at: 2013 http://www.ohri.ca/programs/clinical_epidemiology/oxford.asp Accessed: 8 September 2015
  • 11 Well G, Shea B, O’Connell J et al. The Newcastle-Ottawa scale (NOS) for assessing the quality of nonrandomized studies in meta-analysis [abstract]. Proceedings of the 3rd Symposium on Systematic Reviews: Beyond the Basics; 2000 July 3–5. United Kingdom: Oxford; 2000
  • 12 Higgins JP, Thompson SG, Deeks JJ et al. Measuring inconsistency in meta-analyses. BMJ 2003; 327: 557-560
  • 13 Kochman ML, McClave SA, Boyce HW. The refractory and the recurrent esophageal stricture: a definition. Gastrointest Endosc 2005; 62: 474-475
  • 14 Egger M, Davey SmithG, Schneider M et al. Bias in meta-analysis detected by a simple, graphical test. BMJ 1997; 315: 629-634
  • 15 Canena JM, Liberato MJ, Rio-Tinto RA et al. A comparison of the temporary placement of 3 different self-expanding stents for the treatment of refractory benign esophageal strictures: a prospective multicentre study. BMC Gastroenterol 2012; 12: 70-81
  • 16 Chaput U, Heresbach D, Audureau E et al. Comparison of a standard fully covered stent with a super-thick silicone-covered stent for the treatment of refractory esophageal benign strictures: A prospective multicenter study. United European Gastroenterol J 2013; 1: 93-102
  • 17 Dua KS, Vleggaar FP, Santharam R et al. Removable self-expanding plastic esophageal stent as a continuous, non-permanent dilator in treating refractory benign esophageal strictures: a prospective two-center study. Am J Gastroenterol 2008; 103: 2988-2994
  • 18 Evrard S, Le Moine O, Lazaraki G et al. Self-expanding plastic stents for benign esophageal lesions. Gastrointest Endosc 2004; 60: 894-900
  • 19 Hirdes MM, Siersema PD, van Boeckel PG et al. Single and sequential biodegradable stent placement for refractory benign esophageal strictures: a prospective follow-up study. Endoscopy 2012; 44: 649-654
  • 20 Hirdes MM, Siersema PD, Vleggaar FP. A new fully covered metal stent for the treatment of benign and malignant dysphagia: a prospective follow-up study. Gastrointest Endosc 2012; 75: 712-718
  • 21 Repici A, Conio M, De Angelis C et al. Temporary placement of an expandable polyester silicone-covered stent for treatment of refractory benign esophageal strictures. Gastrointest Endosc 2004; 60: 513-519
  • 22 Song HY, Jung HY, Park SI et al. Covered retrievable expandable nitinol stents in patients with benign esophageal strictures: initial experience. Radiology 2000; 217: 551-557
  • 23 Bakken JC, Wong Kee Song LM, de Groen PC et al. Use of a fully covered self-expandable metal stent for the treatment of benign esophageal diseases. Gastrointest Endosc 2010; 72: 712-720
  • 24 Eloubeidi MA, Talreja JP, Lopes TL et al. Success and complications associated with placement of fully covered removable self-expandable metal stents for benign esophageal diseases (with videos). Gastrointest Endosc 2011; 73: 673-681
  • 25 Karbowski M, Schembre D, Kozarek R et al. Polyflex self-expanding, removable plastic stents: assessment of treatment efficacy and safety in a variety of benign and malignant conditions of the esophagus. Surg Endosc 2008; 22: 1326-1333
  • 26 Kim JH, Song HY, Choi EK et al. Temporary metallic stent placement in the treatment of refractory benign esophageal strictures: results and factors associated with outcome in 55 patients. Eur Radiol 2009; 19: 384-390
  • 27 Liu J, Hu Y, Cui C et al. Removable, fully covered, self-expandable metal stents for the treatment of refractory benign esophagogastric anastomotic strictures. Dysphagia 2012; 27: 260-264
  • 28 Martin RC, Woodall C, Duvall R et al. The use of self-expanding silicone stents in esophagectomy strictures: less cost and more efficiency. Ann Thorac Surg 2008; 86: 436-440
  • 29 Oh YS, Kochman ML, Ahmad NA et al. Clinical outcomes after self-expanding plastic stent placement for refractory benign esophageal strictures. Dig Dis Sci 2010; 55: 1344-1348
  • 30 Barthel JS, Kelley ST, Klapman JB. Management of persistent gastroesophageal anastomotic strictures with removable self-expandable polyester silicon-covered (Polyflex) stents: an alternative to serial dilation. Gastrointest Endosc 2008; 67: 546-552
  • 31 García-Cano J. Dilation of benign strictures in the esophagus and colon with the polyflex stent: a case series study. Dig Dis Sci 2008; 53: 341-346
  • 32 Triester SL, Fleischer DE, Sharma VK. Failure of self-expanding plastic stents in treatment of refractory benign esophageal strictures. Endoscopy 2006; 38: 533-537
  • 33 Broto J, Asensio M, Vernet JM. Results of a new technique in the treatment of severe esophageal stenosis in children: poliflex stents. J Pediatr Gastroenterol Nutr 2003; 37: 203-206
  • 34 Ott C, Ratiu N, Endlicher E et al. Self-expanding Polyflex plastic stents in esophageal disease: various indications, complications, and outcomes. Surg Endosc 2007; 21: 889-896
  • 35 Schünemann H, Brożek J, Guyatt G, Oxman A et al. eds. GRADE handbook for grading quality of evidence and strength of recommendations. Updated October 2013. The GRADE Working Group, 2013. Available from: www.guidelinedevelopment.org/handbook Accessed: 10 September 2015