Image quality of a novel light-emitting diode (LED)-illuminated colonoscope
Publikationsverlauf
submitted 18. November 2015
accepted after revision 06. Juni 2016
Publikationsdatum:
29. Juli 2016 (online)

Background and study aims: Light-emitting diodes (LEDs) are used widely for their high luminous efficiency and durability. We developed a novel prototype high definition endoscope with white LEDs and evaluated the image quality it produced against a commercial endoscope with conventional light source.
Patients and methods: The specifications of both colonoscopes were identical, except for the LED light source at the tip of the prototype. We examined 20 patients with rectal or sigmoid colon lesions and the image quality was evaluated in 40 images (one image from the LED colonoscope and one from the conventional colonoscope for each lesion) by three endoscopists. We additionally evaluated the 17 videos recorded with the LED colonoscope that were available. Image quality, mucosal and vascular color, and luminous distribution and intensity were scored on a 5-point scale.
Results: The mean score for vascular color given by one evaluator was significantly higher using the LED colonoscope than using the conventional colonoscope. The mean scores for mucosal color and luminous intensity from another evaluator were significantly lower with the LED colonoscope than with the conventional colonoscope. There were no significant differences in the luminous distribution scores for any of the evaluators. The image quality of the videos was evaluated as being similar with both colonoscopes.
Conclusions: Image quality from the LED and conventional colonoscopes were similar, although the luminous intensity of the LEDs is inferior to that of the conventional light source at the present time.
-
References
- 1 Taguchi T. Present status of white LED lighting technology in Japan. J Light Vis Env 2003; 27: 1-6
- 2 Sakuta H, Fukui T, Miyachi T et al. Near-ultraviolet LED of the external quantum efficiency over 45% and its application to high-color rendering phosphor conversion white LEDs. J Light Vis Env 2008; 32: 39-42
- 3 Yanai H, Okita K, Kikuchi K et al. Preliminary experience with a gastrointestinal endoscope using a white light-emitting diode. Endoscopy 2006; 38: 290-291
- 4 Nishikawa J, Yanai H, Okamoto T et al. A novel colonoscope with high color-rendering white light-emitting diodes. Gastrointest Endosc 2011; 73: 598-602
- 5 Yanai H, Nishikawa J, Okamoto T et al. Preliminary clinical experience with high-definition colonoscope illuminated by light-emitting diode. Endoscopy 2013; 45: E1-E26
- 6 Lantz CA, Nebenzahl E. Behavior and interpretation of the k statistic: resolution of the two paradoxes. J Clin Epidemiol 1996; 49: 431-434
- 7 Byrt T, Bishop J, Carlin JB. Bias, prevalence and kappa. J Clin Epidemiol 1993; 46: 423-429
- 8 Gralnek IM, Carr-Locke DL, Segol O et al. Comparison of standard forward-viewing mode versus ultrawide-viewing mode of a novel colonoscopy platform: a prospective, multicenter study in the detection of simulated polyps in an in vitro colon model (with video). Gastrointest Endosc 2013; 77: 472-499
- 9 Sugiyama T, Naka H, Yabana T et al. Is Helicobacter pylori infection responsible for postendoscopic acute gastric mucosal lesions?. Eur J Gastroenterol Hepatol 1992; 4: 93-S96
- 10 Nelson DB. Infectious disease complication of GI endoscopy: part II, exogenous infections. Gastrointest Endosc 2003; 57: 695-711
- 11 Design of endoscopic retrograde cholangiopancreatography (ERCP) duodenoscopes may impede effective cleaning: FDA Safety Communication. Updated: 04.03.2015 Available from http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm434871.htm Accessed: August 2015.