J Knee Surg 2016; 29(08): 658-663
DOI: 10.1055/s-0036-1571429
Original Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Correlations between Navigation and Radiographic Measures of Alignment

Seung-Beom Han
1   Department of Orthopaedics, Korea University Anam Hospital, Seoul, The Republic of Korea
,
Dae-Hee Lee
2   Department of Orthopaedics, Samsung Medical Center, Seoul, The Republic of Korea
› Author Affiliations
Further Information

Publication History

06 September 2015

16 December 2015

Publication Date:
02 February 2016 (online)

Abstract

This study was designed to compare the correlations between intraoperative navigation and radiographic alignment in patients who underwent total knee arthroplasty (TKA) and open wedge high tibial osteotomy (HTO). The reliability of navigation and radiographic alignment before and after surgery was compared in 82 prospectively enrolled patients who underwent open wedge HTO and 91 knees that underwent TKA. The absolute difference in preoperative mechanical axis (MA) between navigation and radiography was 1.1 degrees greater in the TKA than in the HTO group (2.6 vs. 1.5 degrees, p < 0.001). Postoperatively, however, the absolute difference between navigation and radiographic measurements of MA was 1 degree greater in the HTO than in the TKA group (2.4 vs. 1.4 degrees, p < 0.001). The percentage of outliers, defined as a >3 degrees difference between navigation and radiographic MA, was significantly greater preoperatively in the TKA group (37 vs. 10%, p < 0.001), but significantly greater postoperatively in the HTO group (28 vs. 10%, p < 0.001). The discrepancy between navigation and radiographic measurement of coronal alignment was significantly greater preoperatively in the TKA than in the HTO group, whereas the discrepancy was significantly greater postoperatively in the HTO group.

 
  • References

  • 1 Cheng T, Zhao S, Peng X, Zhang X. Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials?. Knee Surg Sports Traumatol Arthrosc 2012; 20 (7) 1307-1322
  • 2 Choi WC, Lee S, An JH, Kim D, Seong SC, Lee MC. Plain radiograph fails to reflect the alignment and advantages of navigation in total knee arthroplasty. J Arthroplasty 2011; 26 (5) 756-764
  • 3 Moon YW, Kim JG, Han JH, Do KH, Seo JG, Lim HC. Factors correlated with the reducibility of varus deformity in knee osteoarthritis: an analysis using navigation guided TKA. Clin Orthop Surg 2013; 5 (1) 36-43
  • 4 Lee DH, Park JH, Song DI, Padhy D, Jeong WK, Han SB. Accuracy of soft tissue balancing in TKA: comparison between navigation-assisted gap balancing and conventional measured resection. Knee Surg Sports Traumatol Arthrosc 2010; 18 (3) 381-387
  • 5 Han SB, Bae JH, Lee SJ , et al. Biomechanical properties of a new anatomical locking metal block plate for opening wedge high tibial osteotomy: uniplane osteotomy. Knee Surg Relat Res 2014; 26 (3) 155-161
  • 6 Kyung BS, Kim JG, Jang K-M , et al. Are navigation systems accurate enough to predict the correction angle during high tibial osteotomy? Comparison of navigation systems with 3-dimensional computed tomography and standing radiographs. Am J Sports Med 2013; 41 (10) 2368-2374
  • 7 Kyung HS, Lee BJ, Kim JW, Yoon SD. Biplanar open wedge high tibial osteotomy in the medial compartment osteoarthritis of the knee joint: comparison between the Aescula and TomoFix plate. Clin Orthop Surg 2015; 7 (2) 185-190
  • 8 Hankemeier S, Hufner T, Wang G , et al. Navigated open-wedge high tibial osteotomy: advantages and disadvantages compared to the conventional technique in a cadaver study. Knee Surg Sports Traumatol Arthrosc 2006; 14 (10) 917-921
  • 9 Lee DH, Nha KW, Park SJ, Han SB. Preoperative and postoperative comparisons of navigation and radiologic limb alignment measurements after high tibial osteotomy. Arthroscopy 2012; 28 (12) 1842-1850
  • 10 Dexel J, Kirschner S, Günther K-P, Lützner J. Agreement between radiological and computer navigation measurement of lower limb alignment. Knee Surg Sports Traumatol Arthrosc 2013; 22 (11) 2721-2727
  • 11 Goleski P, Warkentine B, Lo D, Gyuricza C, Kendoff D, Pearle AD. Reliability of navigated lower limb alignment in high tibial osteotomies. Am J Sports Med 2008; 36 (11) 2179-2186
  • 12 Rhee SJ, Seo CH, Suh JT. Navigation-assisted total knee arthroplasty for patients with extra-articular deformity. Knee Surg Relat Res 2013; 25 (4) 194-201
  • 13 Lee DH, Han SB, Oh KJ , et al. The weight-bearing scanogram technique provides better coronal limb alignment than the navigation technique in open high tibial osteotomy. Knee 2014; 21 (2) 451-455
  • 14 Chung BJ, Dileep I, Chang CB, Kang YG, Park YB, Kim TK. Novel approach to reducing discrepancies in radiographic and navigational limb alignments in computer-assisted TKA. Orthopedics 2010; 33 (10, Suppl): 62-67
  • 15 Oberst M, Bertsch C, Lahm A, Wuerstlin S, Holz U. Regression and correlation analysis of preoperative versus intraoperative assessment of axes during navigated total knee arthroplasty. Comput Aided Surg 2006; 11 (2) 87-91
  • 16 Yau WP, Leung A, Chiu KY, Tang WM, Ng TP. Intraobserver errors in obtaining visually selected anatomic landmarks during registration process in nonimage-based navigation-assisted total knee arthroplasty: a cadaveric experiment. J Arthroplasty 2005; 20 (5) 591-601
  • 17 Specogna AV, Birmingham TB, Hunt MA , et al. Radiographic measures of knee alignment in patients with varus gonarthrosis: effect of weightbearing status and associations with dynamic joint load. Am J Sports Med 2007; 35 (1) 65-70
  • 18 McGrory JE, Trousdale RT, Pagnano MW, Nigbur M. Preoperative hip to ankle radiographs in total knee arthroplasty. Clin Orthop Relat Res 2002; (404) 196-202
  • 19 Yaffe MA, Koo SS, Stulberg SD. Radiographic and navigation measurements of TKA limb alignment do not correlate. Clin Orthop Relat Res 2008; 466 (11) 2736-2744
  • 20 Kendoff D, Board TN, Citak M , et al. Navigated lower limb axis measurements: Influence of mechanical weight-bearing simulation. J Orthop Res 2008; 26 (4) 553-561
  • 21 Kawakami H, Sugano N, Yonenobu K , et al. Effects of rotation on measurement of lower limb alignment for knee osteotomy. J Orthop Res 2004; 22 (6) 1248-1253
  • 22 Kendoff D, Lo D, Goleski P, Warkentine B, O'Loughlin PF, Pearle AD. Open wedge tibial osteotomies influence on axial rotation and tibial slope. Knee Surg Sports Traumatol Arthrosc 2008; 16 (10) 904-910
  • 23 Sabharwal S, Zhao C. Assessment of lower limb alignment: supine fluoroscopy compared with a standing full-length radiograph. J Bone Joint Surg Am 2008; 90 (1) 43-51