The C-Port xV® Vascular Anastomosis System: Results from an Animal Trial
Publication History
received Sept. 22, 2010
resubmitted March 7, 2011
accepted March 8, 2011
Publication Date:
25 March 2011 (online)

Abstract
Background: In this study, facilitated anastomosis using an anastomotic device was compared to conventional hand-sewn (HS) vascular anastomosis in an animal model. Methods: A pig carotid bypass model was employed. C-Port xV® (xV) and HS anastomoses were compared by evaluating intraoperative performance, midterm graft patency, and histology. Results: All animals survived; none developed early/late neurological deficits. Mean graft blood flow was comparable between groups (HS group: 161 ± 61 ml/min; xV group: 143 ± 44 ml/min). All anastomoses were patent at necropsy (at 111 ± 6 postoperative days). Histologically, no significant inflammation was found around the fasteners or in the vessel wall. Neointimal overgrowth on the lumen surface appeared organized and covered with endothelium. There was no adherence of fibrin, platelets, or inflammatory cells to the surface. The neointimal tissue appeared normal without any inflammation, hemorrhage, calcification, or necrosis. Conclusion: Facilitated vascular anastomosis using the xV anastomotic device is safe and effective in the pig carotid bypass model. Further studies should evaluate the efficacy of this device when used in confined spaces to define its potential role in minimally invasive procedures.
Key words
coronary artery bypass grafts surgery - CABG - pathology - off‐pump surgery - surgical equipment (instruments, sutures, etc.) - coronary anastomosis - new technology device - histology
References
- 1 Carrel A. The surgery of blood vessels. Bull Johns Hopk Hosp. 1906; 19 18-27
- 2 Matschke K E, Gummert J F, Demertzis S et al. The Cardica C-Port System: clinical and angiographic evaluation of a new device for automated, compliant distal anastomoses in coronary artery bypass grafting surgery–a multicenter prospective clinical trial. J Thorac Cardiovasc Surg. 2005; 130 1645-1652
- 3 Kappert U, Tugtekin S M, Cichon R, Braun M, Matschke K. Robotic totally endoscopic coronary artery bypass: a word of caution implicated by a five-year follow-up. J Thorac Cardiovasc Surg. 2008; 135 857-862
- 4 Srivastava S, Gadasalli S, Agusala M et al. Beating heart totally endoscopic coronary artery bypass. Ann Thorac Surg. 2010; 89 1873-1879
- 5 Suyker W J, Borst C. Coronary connector devices: analysis of 1,469 anastomoses in 1,216 patients. Ann Thorac Surg. 2008; 85 1828-1836
- 6 Hanggi D, Reinert M, Steiger H J. C-Port Flex-A-assisted automated anastomosis for high-flow extracranial-intracranial bypass surgery in patients with symptomatic carotid artery occlusion: a feasibility study. Clinical article. J Neurosurg. 2009; 111 181-187
- 7 Trunfio R, Demertzis S, van den Berg J C, Siclari F. A new surgical approach for exclusion of renal artery aneurysms avoiding organ ischemia. Eur J Vasc Endovasc Surg. 2008; 36 559-561
Dr. Stefanos Demertzis
Department of Cardiac Surgery
Cardiocentro Ticino
Via Tesserete 48
6900 Lugano
Switzerland
Phone: +41 9 18 05 31 44
Fax: +41 9 18 05 31 48
Email: demertzis@cardiocentro.org