Background: Previous studies have demonstrated that cortical auditory-evoked potentials (CAEPs)
can be reliably elicited in response to speech stimuli in listeners wearing hearing
aids. It is unclear, however, how close to the aided behavioral threshold (i.e., at
what behavioral sensation level) a sound must be before a cortical response can reliably
be detected.
Purpose: The purpose of this study was to systematically examine the relationship between
CAEP detection and the audibility of speech sounds (as measured behaviorally), when
the listener is wearing a hearing aid fitted to prescriptive targets. A secondary
aim was to investigate whether CAEP detection is affected by varying the frequency
emphasis of stimuli, so as to simulate variations to the prescribed gain-frequency
response of a hearing aid. The results have direct implications for the evaluation
of hearing aid fittings in nonresponsive adult clients, and indirect implications
for the evaluation of hearing aid fittings in infants.
Research Design: Participants wore hearing aids while listening to speech sounds presented in a sound
field. Aided thresholds were measured, and cortical responses evoked, under a range
of stimulus conditions. The presence or absence of CAEPs was determined by an automated
statistic.
Study Sample: Participants were adults (6 females and 4 males). Participants had sensorineural
hearing loss ranging from mild to severe-profound in degree.
Data Collection and Analysis: Participants' own hearing aids were replaced with a test hearing aid, with linear
processing, during assessments. Pure-tone thresholds and hearing aid gain measurements
were obtained, and a theoretical prediction of speech stimulus audibility for each
participant (similar to those used for audibility predictions in infant hearing aid
fittings) was calculated. Three speech stimuli, (/m/, /t/, and /g/) were presented
aided (monaurally, nontest ear occluded), free field, under three conditions (+4 dB/octave,
−4 dB/octave, and without filtering), at levels of 40, 50, and 60 dB SPL (measured
for the unfiltered condition). Behavioral thresholds were obtained, and CAEP recordings
were made using these stimuli. The interaction of hearing loss, presentation levels,
and filtering conditions resulted in a range of CAEP test behavioral sensation levels
(SLs), from −25 to +40 dB.
Results: Statistically significant CAEPs (p < .05) were obtained for virtually every presentation where the behavioral sensation
level was >10 dB, and for only 5% of occasions when the sensation level was negative.
In these (“false-positive”) cases, the greatest (negative) sensation level at which
a CAEP was judged to be present was −6 dB SL.
Conclusions: CAEPs are a sensitive tool for directly evaluating the audibility of speech sounds,
at least for adult listeners. CAEP evaluation was found to be more accurate than audibility
predictions, based on threshold and hearing aid response measures.
Key Words
Evoked potentials - auditory; hearing aids - adults