Vet Comp Orthop Traumatol 2014; 27(02): 97-101
DOI: 10.3415/VCOT-13-03-0037
Original Research
Schattauer GmbH

Sterilization by gamma radiation of antibiotic impregnated polymethylmethacrylate and plaster of Paris beads

A pilot study
P. Trencart
1   University of Montreal, Department of Clinical Sciences, Faculty of Veterinary Medicine, St Hyacinthe, Quebec, Canada
,
Y. A. Elce
1   University of Montreal, Department of Clinical Sciences, Faculty of Veterinary Medicine, St Hyacinthe, Quebec, Canada
2   University of Montreal, Department of Pathology and Microbiology, Faculty of Veterinary Medicine, St Hyacinthe, Quebec, Canada
,
E. Rodriguez Batista
1   University of Montreal, Department of Clinical Sciences, Faculty of Veterinary Medicine, St Hyacinthe, Quebec, Canada
,
G. Michaud
1   University of Montreal, Department of Clinical Sciences, Faculty of Veterinary Medicine, St Hyacinthe, Quebec, Canada
› Author Affiliations
Our thanks to Bruno Laventure and the Centre de Recherche of Agriculture and Agroalimentaire Canada for the sterilization by radiation. This project was supported by the Fonds de Recherche Clinique de Pfizer Santé Animale.
Further Information

Publication History

Received: 15 March 2013

Accepted: 03 November 2013

Publication Date:
20 December 2017 (online)

Summary

Introduction

Ethylene oxide is currently recommended for sterilization of antibiotic impregnated beads; however this method carries health risks to personnel and is becoming less available.

Objective

To perform a pilot study of the effect of radiation for sterilization of polymethylmethacrylate (PMMA) and plaster of Paris (POP) beads impregnated with amikacin, enrofloxacin, and ceftiofur.

Hypothesis

Radiation would effectively sterilize the beads without affecting the efficacy of the antibiotic.

Materials and methods

Beads of PMMA and POP were prepared in a clean but nonsterile manner with one of the three antibiotics (amikacin, enrofloxacin, ceftiofur) or no antibiotic. Beads were then exposed to radiation for a total dose of 0 kiloGray (kGy), 10 kGy and 25 kGy. Beads were incubated on Mueller-Hinton agar plates seeded with Escherichia coli, Staphylococcus aureus or Pseudomonas aeruginosa for 24 hours or cultured in brain-heart infusion broth for 48 hours. Zones of inhibition were measured on the agar plates and statistics were performed on the diameters of the zones of inhibition using an analysis of variance.

Results

There were no differences in the diameters of inhibition for all levels of radiation for all PMMA beads. The same was true with POP beads with the exception of enrofloxacin which had a significantly decreased zone of inhibition with increased levels of radiation, though the clinical significance of this finding was not assessed. Only beads without antibiotics and not exposed to radiation had bacterial growth.

Clinical significance

Radiation may be an effective method of sterilization for antibiotic impregnated beads.

 
  • References

  • 1 Ramos JR, Howard RD, Pleasant RS. et al. Elution of metronidazole and gentamicin from polymethylmethacrylate beads. Vet Surg 2003; 32: 251-261.
  • 2 Atilla A, Boothe HW, Tollett M. et al. In vitro elution of amikican and vancomycin from impregnated plaster of Paris beads. Vet Surg 2010; 39: 715-721.
  • 3 Phillips H, Boothe DM, Shofer F. et al. In vitro elution studies of amikacin and cefazolin from polymethylmethacrylate. Vet Surg 2007; 36: 272-278.
  • 4 Cariou M, Boulouis HJ, Moissonnier P. Inclusion of marbofloxacin in PMMA orthopedic cement. An in vitro experimental study. Vet Comp Orthop Traumatol 2006; 6: 106-109.
  • 5 Baez LA, Langston C, Givaruangsawat S. et al. Evaluation of in vitro serial antibiotic elution from meropenem-impregnated polymethylmethacrylate beads after ethylene oxide gas and autoclave sterilization. Vet Comp Orthop Traumatol 2011; 1: 39-44.
  • 6 Schneider RK. Synovial and osseous infections. In: Auer JA, Stick JA. editors. Equine Surgery. St Louis, Missouri: Saunders; 2006. pg. 1127.
  • 7 Steenland K, Whelan E, Deddens J. et al. Ethylene oxide and breast cancer incidence in a cohort study of 7576 women (United States). Cancer Causes Control 2003; 14: 531-539.
  • 8 Steenland K, Stayner L, Deddens J. Mortality analyses in a cohort of 18235 ethylene oxide exposed workers: follow-up extended from 1987-1998. Occup Environ Med 2004; 61: 2-7.
  • 9 Franklin SP, Stoker AM, Cockrell MK. et al. Effects of low-temperature hydrogen peroxide gas plasma sterilization on in vitro cytotoxicity of poly (ε-caprolactone) (PCL). J Biomater Sci Plym Ed; 2012; 23: 2197-2206.
  • 10 Ferreira SD, Dernell WS, Powers BE. et al. Effect of gas-plasma sterilization on the osteoinductive capacity of demineralised bone matrix. Clin Orthop Relat Res 2001; 388: 233-239.
  • 11 Singh BK, Parwate DV, Dassarma IB. et al. Radiation sterilization of flouroquinolones in solid state: investigation of effect of gamma radiation and electron beam. Appl Radiat Isot 2010; 68: 1627-1635.
  • 12 Basly JP, Longy I, Bernard M. Radiation sterilization of formoterol. Pharm Res 1997; 14: 810-814.
  • 13 Katusin-Razem B, Hamitouche K, Maltar-Strmecki N. et al. Radiation sterilization of ketoprofen. Radiation physics and chemistry 2005; 73: 111-116.
  • 14 Basly JP, Duroux JL, Bernard M. The effect of gamma radiation on the degradation of salbutamol. J Pharm Biomed Anal 1997; 15: 1137-1142.
  • 15 Varshney L, Dodke PB. Radiation studies on anticancer drugs, cyclophosphamide and doxorubicin for radiation sterilization. Rad Phys Chem 2004; 72: 1103-1111.
  • 16 Jacobs GP. A review of the effects of gamma radiation on pharmaceutical materials. J Biomater Appl 1995; 10: 59-96.
  • 17 Goldman M, Pruitt L. Comparison of the effects of gamma radiation and low temperature hydrogen peroxide gas plasma sterilization on the molecular structure, fatigue resistance, and wear behaviour of UHMWPE. J Biomed Mater Res 1998; 5: 378-384.
  • 18 Bartolotta A, D’Oca MC, Campisi M. et al. Effects of gamma-irradiation on trehalose-hydroxyethylcellulose microspheres loaded with vancomycin. Eur J Pharm Biopharm 2005; 59: 139-146.
  • 19 Krasko MY, Golenser J, Nyska A. et al. Gentamicin extended release from an injectable polymeric implant. J Control Release 2007; 117: 90-96.
  • 20 Athanasiou KA, Niederauer GG, Agrawal CM. Sterilization, toxicity, biocompatibility and clinical applications of polylactic acid/ polyglycolic acid copolymers. Biomaterials 1996; 17: 92-102.
  • 21 Gatineau M, El-Warrak AO, Bolliger C. et al. Effects of sterilization with hydrogen peroxide gas plasma, ethylene oxide, and steam on bioadhesive properties of nylon and polyethylene lines used for stabilization of canine stifle joints. Am J Vet Res 2012; 73: 1665-1669.
  • 22 Meyer JD, Falk RF, Kelly RM. et al. Preparation and in vitro characterization of gentamycin-impregnated biodegradable beads suitable for treatment of osteomyelitis. J Pharm Sci 1998; 87: 1149-1153.