Abstract

The determination of CRP in serum was performed by the fluorescence polarization immunoassay (FPIA) and the kinetic nephelometry (KNM). Both methods showed a good precision and accuracy. The investigation was carried out in 190 sera with a CRP concentration up to 49 mg/dl. The comparison of the two methods in 100 sera with a CRP concentration between 0.6 and 5 mg/dl showed a good correlation (correlation coefficient r = 0.9753). Close correlation (correlation coefficient r - 0.9947) was also found in 90 sera with a CRP concentration between 5 and 49 mg/dl. Discussion of the FPIA and KNM methods.

Zusammenfassung

Die Bestimmung der CRP-Konzentration im Serum erfolgte mit dem Fluoreszenz-Polansations-Immunoassay (FPIA) und mit der Methode der kinetischen Nephelometrie (KNM). Beide Methoden zeigten eine gute Präzision und Richtigkeit. Als Untersuchungsmaterial dienten 190 Seren mit einer CRP-Konzentration bis 49 mg/dl. Bei 100 untersuchten Seren mit einer CRP-Konzentration zwischen 0,6 und 5 mg/dl ergab der Vergleich der FPIA- und KNM-Methode eine gute Korrelation (KorrelationskoeiTizient r = 0,9753). Ebenfalls eine sehr gute Korrelation (Korrelationskoeffizient r = 0,9947) ergab sich beim Vergleich von 90 Seren mit einer CRP-Konzentration zwischen 5 und 49 mg/dl. Diskussion der Methoden FPIA und KNM.

    >