Endoscopy 2018; 50(05): E113-E114
DOI: 10.1055/s-0044-101019
E-Videos
© Georg Thieme Verlag KG Stuttgart · New York

Successful digital cholangioscopy removal of a stent-retriever tip migrated into the periphery of the bile duct

Takeshi Ogura
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Atsushi Okuda
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Akira Miyano
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Nobu Nishioka
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
,
Kazuhide Higuchi
2nd Department of Internal Medicine, Osaka Medical College, Osaka, Japan
› Author Affiliations
Further Information

Corresponding author

Takeshi Ogura, MD
2nd Department of Internal Medicine
Osaka Medical College
2-7 Daigakuchou, Takatsukishi
Osaka 569-8686
Japan   
Fax: +81-726-846532   

Publication History

Publication Date:
16 February 2018 (online)

 

Various techniques to dilate occluded uncovered metal stents (UMSs) have been reported [1]. A stent retriever (Soehendra stent retriever, SSR-7; Cook, Tokyo, Japan) can also be used as a dilation device in endoscopic retrograde cholangiopancreatography (ERCP) [2] [3]. Herein, we describe a case in which the tip of a stent retriever migrated into the periphery of the bile duct and was successfully removed using digital cholangioscopy.

An 85-year-old man had undergone placement of multiple UMSs for hepatic hilar biliary stricture 1 year previously. Since the stent placement, re-intervention had been performed three times because of stent occlusion; however, he was admitted to our hospital because of a recurrence of obstructive jaundice. Computed tomography (CT) scanning showed anterior bile duct dilatation, and re-intervention was attempted.

First, the guidewire was placed into the anterior bile duct. Neither an ERCP catheter nor a balloon catheter could be advanced into the anterior bile duct through the mesh of the UMS, so recanalization using a stent retriever was attempted. However, during this procedure, the tip of the stent retriever became detached and migrated into the periphery of the bile duct ([Fig. 1]). To remove this migrated stent-retriever tip, several techniques including the use of biopsy forceps, basket catheter, and balloon catheter were performed, but without successful removal being achieved.

Zoom
Fig. 1 Radiographic image showing the migrated tip of the Soehendra stent retriever (arrow) in the periphery of the bile duct.

Removal under direct visualization was therefore attempted using a digital single-operator intraductal cholangioscope (IDC; SPY-DS; Boston Scientific). The cholangioscope was advanced into the periphery of the anterior bile duct through the UMS until the migrated tip of the stent retriever could be seen ([Fig. 2]; [Video 1]). Next, the migrated tip was grasped by a SPY-Bite device ([Fig. 3 a, b]). The cholangioscope was then itself carefully withdrawn from the bile duct into the intestine ([Fig. 3 c]), successfully removing the migrated tip without any adverse events ([Fig. 4]).

Zoom
Fig. 2 Cholangioscopic view showing the tip of the stent retriever using a digital intraductal cholangioscope (SPY-DS).

Video 1 The intraductal cholangioscope (SPY-DS) is inserted into the periphery of the bile duct through the uncovered metal stents and the migrated tip of the stent retriever is observed. The migrated tip is then grasped by the SPY-Bite device under direct visualization, allowing it to be successfully removed.

Zoom
Fig. 3 The tip of the stent retriever is grasped using the SPY-Bite device: a on radiographic image; b on cholangioscopic image; and c is carefully removed.
Zoom
Fig. 4 The tip of the stent retriever following its successful removal.

In conclusion, this case may be the first to show this adverse event associated with a stent retriever, along with the subsequent successful removal. Direct visualization with the digital IDC was helpful in this case.

Endoscopy_UCTN_Code_CPL_1AK_2AD

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos


Competing interests

None


Corresponding author

Takeshi Ogura, MD
2nd Department of Internal Medicine
Osaka Medical College
2-7 Daigakuchou, Takatsukishi
Osaka 569-8686
Japan   
Fax: +81-726-846532   


Zoom
Fig. 1 Radiographic image showing the migrated tip of the Soehendra stent retriever (arrow) in the periphery of the bile duct.
Zoom
Fig. 2 Cholangioscopic view showing the tip of the stent retriever using a digital intraductal cholangioscope (SPY-DS).
Zoom
Fig. 3 The tip of the stent retriever is grasped using the SPY-Bite device: a on radiographic image; b on cholangioscopic image; and c is carefully removed.
Zoom
Fig. 4 The tip of the stent retriever following its successful removal.