Keywords
osteosarcoma - osteoma - 3D printing - skull reconstruction - tumour
Introduction
Additive manufacturing (AM) has been introduced in the medical field in the last decade,
gaining popularity specifically in the field of maxillofacial, orthopaedic, oncological
and spine surgery.[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
[9]
[10] The availability of materials and printers has led to lower costs of patient-specific
three-dimensional (3D) implants in human medicine bringing them also within the scope
of the veterinary field.[11]
The use of individualized implants is the latest development in implant technology
and offers a wide variety of opportunities and applications. To date, only few studies
in veterinary medicine have used 3D individualized implants. Case reports have described
3D-printed implants for dogs with atlantoaxial subluxation, angular limb deformity,
in limb-sparing surgery and in mandibular and maxillary surgery.[1]
[2]
[3]
[4]
[5]
[6]
[7] Three-dimensional-printed customized cages have also been used for tibial tuberosity
advancement and for the treatment of cervical spondylomyelopathy.[8]
[9]
In contrast to fractures and defects in the appendicular skeleton, reconstruction
of the skull and spine is often complicated due to the individual variety in anatomy
or the lack of veterinary implants. The development of patient-specific 3D printing
will improve surgical planning and enable successful surgical results in veterinary
patients with extensive tumours of the skull.[12] The use of AM offers new surgical possibilities for patients who cannot benefit
from existing surgical treatments. Practical experience from cases treated using this
technique is valuable to accelerate the development of patient-specific 3D implants
in the future.[13] The aim of the present case report is to describe the use of individualized 3D-printed
titanium implants for skull reconstruction following oncologic surgery.
Case Description
Case 1
A 2-year-old female neutered Siberian Husky was presented for depression and a mass
on the right rostral calvarium. For further evaluation of the extension of the mass
and metastasis control, a pre- and postcontrast computed tomography (CT) of the head
and thorax was performed. Computed tomography showed an ossifying mass originating
from the right frontal sinus and adjacent frontal bone ([Fig. 1]). The mass measured approximately 4.2 × 4.4 × 5.1 cm and had infiltrated the entire
right frontal sinus and frontal neurocranium compressing the brain on the right side
([Fig. 1]). Suspected CT diagnosis was a multilobular osteochondrosarcoma, osteosarcoma or
osteoma. There was no evidence of metastases in the regional lymph nodes or the lungs.
Fig. 1 Case 1: Three-dimensional (3D) computed tomography (CT) reconstruction (A, C, E, G) and transverse CT image (B, D, F, H) of a 2-year-old Siberian Husky with a calvarian osteoma. Before surgery (A, B), at 3 months (C, D), and 6 months (E, F) after surgical resection, and after revision surgical resection followed by reconstruction
with a 3D-printed titanium implant (G, H).
The dog was anaesthetized using a protocol for intracranial surgery, a right trans
sinus frontalis craniectomy was performed with 5 mm bone margins and the tumour was
removed as one solid bony mass. Recovery was uneventful and the patient was discharged
2 days after surgery with antibiotic medications and analgesics. The solid bony mass
needed extensive decalcification and histopathological examination showed proliferation
of well-defined bony tissue, consistent with an osteoma ([Fig. 2]). A follow-up CT scan at 3 months after surgery showed possible regrowth of the
tumour which was confirmed 6 months post-surgery at four locations around the original
craniectomy ([Fig. 1]) without evidence of metastasis to the regional lymph nodes and lungs. Because total
excision of the tumour recurrences with 10 mm margins would lead to removal of more
than half of the calvarium, the option of AM to close the defect with an individualized
implant was explored.
Fig. 2 Case 1: Sequential steps in surgical treatment of a 2-year-old Siberian Husky with
a calvarian osteoma (A). Intraoperative views at moment of revision surgery, quadruple recurrences (arrows) are present at the first surgical rim (B). Placement of the saw guide (C) for the three-dimensional-printed titanium implant (D) extending over the right frontal sinus (*) and application of stay sutures for the
attachment of the temporal muscle to the dorsal implant ridge (E). Histology of the mass showed well-differentiated, mainly immature (woven) bone
and a cell-poor fibrous component most consistent with an osteoma. Haematoxylin and
eosin stain, size bar = 200 µm (F).
To design an AM individualized implant, first the digital imaging and communications
in medicine files of the most recent CT scan (250 mAs, 120 kV, 0.6 mm slice thickness)
were exported from the imaging archive system to Mimics v21 (Materialise, Leuven,
Belgium) for anatomical segmentation. Standard bone threshold values (226 HU – upper
threshold) were taken to outline the bone. Furthermore, the tumour recurrences were
manually outlined together with a digitally drawn resection margin that was used to
simulate the minimal needed resection area.
Thereafter, the anatomical models were transferred using stereolithography files to
3-matic software (v.13, Medical, NV, Leuven, Belgium) in which the design took place.
The resection guide was designed around the tumour recurrences, including the digitally
simulated 10 mm resection margin, and reviewed by a board-certified veterinary surgeon.
Then the remaining skull was digitally reconstructed. The cranioplasty implant was
designed as press fit and contained five extensions overlaying intact bone with screw
holes for fixation to the skull and a porous mesh border (70% porous, 500–600 µm pore
size, Diamond unit cell) to allow bony ingrowth at the implant bone interface ([Fig. 2]).[14]
[15] Additionally, the midline ridge of the implant contained designated holes for suture
attachment of the temporal muscle fascia and suture anchors through which muscles
could be reattached. The surgical saw guide was 3D-printed in Nylon (PA12) on an EOS
P110 printer (EOS, Krailling, Germany) and the implant was 3D-printed in medical grade
titanium alloy Ti-6Al-4V ELI grade 23 using direct metal printing on a ProX DMP320
printer (3D Systems, Leuven, Belgium). Post-processing included polishing, cleaning
and sterilization.Two months after the last CT scan, the dog underwent a second extended
craniectomy with resection of the tumour recurrences and the 10 mm margin using the
saw guide ([Fig. 2]). The implant was fitted and secured into place using six 2.0 mm self-tapping titanium
cortical screws (Unilock, DePuy Synthes, Johnson-Johnson, Oberdorf, Switzerland) ranging
from 6 to 8 mm in length. The fascia of the temporal muscle was sutured through the
4 holes on the sagittal ridge of the implant. The subcutis and cutis were routinely
closed. A postoperative CT was performed to assess the position of the implant. The
dog received postoperative analgesia, was monitored and was discharged 2 days after
surgery with antibiotic medications and routine analgesics. A follow-up CT was performed
4 months postoperatively and showed no signs of tumour regrowth and intact implant
positioning identical to the immediate postoperative CT ([Fig. 1]). The dog showed no clinical signs and the survival time at time of writing was
943 days.
Case 2
A 11-year-old female neutered Labrador Retriever was presented for the removal of
a mass of the zygomatic bone causing problems with mastication. Computed tomography
was performed for surgical planning and the mass measured 3 cm in diameter ([Fig. 3]). Suspected CT diagnosis was a multilobular osteochondrosarcoma, osteoma or osteosarcoma.
There was no evidence of metastases to the regional lymph nodes or the lungs. Using
the same AM workflow as in case 1, a patient-specific 3D implant was designed based
on the mirrored contralateral skull after complete tumour removal with 10 mm bone
margins on the zygomatic, palatine and adjacent orbital bones. Because of clear bony
landmarks, there was no need for a saw guide in this patient. The implant consisted
of polished titanium with an unpolished titanium mesh around the edges to facilitate
bone ingrowth and was manufactured using the same parameters as in case 1. Six weeks
after the most recent CT scan, the dog was anaesthetized and the tumour was removed
by approaching the zygomatic bone and cutting the bone with an oscillating saw at
the predetermined land marks ([Fig. 4]). The mass was resected completely including the zygomatic bone, maxilla including
molars 109 and 110, palatine and adjacent orbital bones ([Fig. 4]). The implant was secured using six 2.0 mm self-tapping titanium cortical screws
(Unilock, DePuy Synthes, Johnson-Johnson, Oberdorf, Switzerland) ranging from 7 to
10 mm in length. The subcutis and cutis were routinely closed. A postoperative CT
scan showed anatomical placement of the 3D-printed titanium implant ([Fig. 3]). The patient was discharged 1 day after surgery with antibiotic medications and
routine analgesics. The dog was offered soft food and was not allowed to chew on toys.
One month after surgery, the patient showed no clinical signs and normal pellet food
was reintroduced. Histopathological diagnosis of the tumour was consistent with a
parosteal osteosarcoma. A follow-up CT scan 4 months after surgery showed no signs
of recurrence or implant failure and some activity of bony ingrowth on the porous
implant borders. The dog died 670 days after surgery of age-related problems, until
the day of death the dog showed no clinical signs associated with the tumour or implant.
Fig. 3 Case 2: Three-dimensional computed tomography (3D CT) skull reconstruction (A, C) and transverse CT image (B, D) of a 11-year-old Labrador Retriever with a right zygomatic arch parosteal osteosarcoma
involving the maxillary and orbital bones. Before surgery (A, B) and after surgical resection of the mass and reconstruction with a 3D-printed titanium
implant (C, D).
Fig. 4 Case 2: Sequential skull views during surgical treatment of an 11-year-old Labrador
Retriever with a right zygomatic bone parosteal osteosarcoma (A). Intra-operative view showing the right zygomatic bone (*) with the forceps indicating
the neoplasm (B), followed by surgical resection of part of the zygomatic bone and maxilla together
with the neoplasm (C) and placement of the three-dimensional-printed titanium implant (D). Tumour after excision next to canine skull (E). Histology of the mass showed a moderately cellular spindle cell proliferation with
formation of woven bone and a cartilaginous component most consistent with a parosteal
osteosarcoma. Haematoxylin and eosin stain, size bar = 200 µm (F).
Discussion
These two cases describe the feasibility and the use of patient-specific customized
implants in extensive reconstructive surgery of the skull. Both implants were designed
to precisely press fit into the craniectomy or ostectomy defect. This differs from
the implant described by Oblak and Hayes who created a titanium implant that overlaid
the frontal and temporal bones in a dog with reconstruction of the neurocranium after
removal of a frontal bone tumor.[16] The press fit nature of the design of the implants created the need for a perfect
osteotomy. Because clear anatomical margins were lacking for the location of the tumour
in case 1, a saw guide was designed. The design needed to align with the skull perfectly
to create the perfect osteotomy in this patient. To diminish costs, the saw guide
was printed in Nylon (PA12). Previous work has shown that CT images provide accurate
models of the skull.[17] The accuracy of the designing process, the quality of the program, the collaboration
with the manufacturer and surgeon are all important to achieve a high-quality end
result.[13]
The borders of both implants were made of a porous titanium mesh allowing bone ingrowth
and therefore permanently embedding the implant into the skull, adding to the long-term
durability. Although we have no histological prove that bone ingrowth occurred in
our cases, both in vivo and in vitro push-out tests, micro-CT and histopathology have
proven that bone ingrowth occurred in Ti6-Al4-V scaffolds in dogs and that the strength
in implant embedding was increased.[18]
[19]
[20]
[21]
The solid part of the implant was polished, the mesh surface was unpolished. Polishing
has proven to be beneficial in the reduction in the risk of biofilm formation compared
with unpolished implants.[22] To obtain a smooth surface, implants can also be coated. Calcium-phosphate-coated
porous titanium implants enhanced tissue ingrowth compared with porous implants without
coating.[23] Preliminary results of an in vitro study showed promising results on the prevention
of growth of tumour cells on the margins of selenium-coated implants.[24] For future cases, this could be of added benefit in bone reconstruction after tumour
removal.
When using metal implants in orthopaedic surgery under strict asepsis, there is always
a risk for implant-related infections. In an in vivo study, significantly more bacteria
were cultured from implants with rough surfaces than from those with a smooth surface.[22] However, in non-coated implants with a porous mesh border that stimulated ingrowth
of bone, an in vitro study showed that mesh also increased the risk of bacterial adhesion.[25] In the present case series, there were no clinical or radiological signs of implant-related
infections, despite the use of a porous mesh at the borders of the implants.
Histologic examination of the tumour in case 1 showed well-differentiated, mainly
immature bone and a cell-poor fibrous component most consistent with osteoma. Osteomas
are usually locally invasive benign bone tumours where complete surgical excision
is curative in most cases.[26] In this case, the histological diagnosis was not consistent with the clinical behaviour
of the tumour which showed regrowth of the mass several months after removal and required
extensive second-step excision surgery. This is suggestive of a more malignant tumour
like multilobular osteochondrosarcoma. This tumour may show varying levels of malignancy
in different histological sections.[26]
[27] Nevertheless, the final histopathological diagnosis confirmed an osteoma. The recurrence
of the osteoma can be explained by insufficient margins during the initial resection.
In case 1 metastasis was not identified until the day of writing. The appearance of
the tumour on CT and consistency of the tumour after removal (solid bone) were in
agreement with the histological diagnosis of osteoma. The reason for the recurrences
was most likely due to the limited surgical margins (5 mm) during the first surgery
and histopathologic examination was not able to differentiate between clear or dirty
margins. The typical recurrence of the tumour in four small bone proliferations on
the surgical margin of the first craniectomy is more consistent with an osteoma that
was not completely resected.
As these were the first patients that received patient-specific 3D-printed implants
in our hospital, the manufacturing and printing of the implants took more than 1 month.[1] The commercial company that printed the implants usually does not have a priority
lane for veterinary implants which resulted in a long lead time between CT and the
surgery date. This similar problem has been experienced by other surgeons and presents
a problem in oncological patients in which the tumour continues to grow after CT imaging.[24] Further developments in AM for veterinary use and start-up companies with a veterinary
focus on patient-specific 3D printing should result in future decreased lead times
for 3D-printed implants.
Little is known about the long-term effects and durability of 3D-printed implants.
Although titanium has been used safely in osteosynthesis with long-term follow-up
times, there is currently no long-term data on patient-specific implants in veterinary
medicine. More research is needed to provide information on long-term results and
possible side effects of the implants and the specific designs. It may be expected
that intensified collaboration between human medicine and veterinary medicine accelerates
development and broadens the knowledge on AM in veterinary surgery, thereby adding
more applications for patient-specific implants in dogs and cats.
This case report describes two cases of extensive reconstructions after tumour excision
of the skull with patient-specific customized titanium implants with porous edges.
Three months postoperatively, both dogs were free of clinical signs and CT showed
correct placement of implants without signs of tumour regrowth, implant loosening
or infection. The use of partly porous titanium implants in craniomaxillary surgery
in two dogs resulted in excellent clinical outcome with long-term survival and therefore
may be considered as a treatment option in similar cases.