Case Report
A 59-year-old female presented with a dysfunctional port. She also noticed that the
subcutaneous segment of the port catheter on the chest wall was also impalpable. The
patient denied any chest pain or palpitations. A chest radiograph ([Fig. 1]) revealed the catheter separated from the chamber and migrated into the heart straddling
the atrioventricular (AV) valve; the central tip extended into the right ventricle
(RV), while its peripheral end was in the superior vena cava (SVC). On ultrasound
examination, the catheter tip was just visible in the low right internal jugular vein
(IJV) with thrombus surrounding the tip and extending to the vessel wall. Urgent percutaneous
retrieval of the dislodged port catheter was planned. In the angiography suite using
ultrasound guidance, right IJV and right common femoral venous (CFV) access were established.
Initially capture of the peripheral tip of the catheter in the IJV was contemplated.
However, the tip embedded in thrombus made it difficult for simple loop snare technique.
Because of its location, other techniques were thought either challenging or to increase
the risk of its dislodgement and embolization further into the RV and pulmonary artery
(PA). Via the CFV access, a 14mm×40mm Armada angioplasty balloon (Abbot Vascular,
Santa Clara, CA) was introduced over a guidewire and advanced into the SVC alongside
the port catheter taking care not to dislodge it. The balloon was gently inflated
to “tack” the port catheter to the SVC wall and fix it ([Fig. 2A]). Once its position was secured, a 6 to 10mm trilobed snare device (EN Snare, Merit
medical Systems) was introduced from the IJV access. The tip of the stabilized catheter
was released free of the wall thrombus. It was then captured and extracted from the
neck access, while simultaneously deflating the balloon in the SVC ([Fig. 2B]). Following this the port chamber was also removed from the chest wall. The patient
made an uneventful recovery.
Fig. 1 Frontal chest radiograph depicts the chest wall port chamber with its catheter detached
and dislodged, one end of which is seen in the SVC and the other projected over the
proximal right ventricle.
Fig. 2 (A) A balloon is inflated in the SVC to “fix” the position of the loose end of the
dislodged catheter, (B) The free end of the “fixed” catheter is captured using a snare
from jugular venous access.
Discussion
Intravascular foreign bodies (IFBs) occur commonly because of complication of endovascular
therapy or due to failure of endovascular devices. Detached and embolized venous lines
are among the commonest IFBs; others include guidewires, coils, and stents.[1] Although in the majority instances these remain asymptomatic, it is recommended
that the retained IFBs be removed to prevent serious adverse consequences.[2]
Several retrieval techniques with high technical success rates and low complication
rates have been described to capture the IFBs.[1]
[3]
[5] The optimum technique depends largely on the location and configuration of the IFB.[1] Typically, a snare is used to capture to the free tip of a dislodged port catheter
often from its proximal end closest to the access.[4] In our case, the peripheral (proximal) tip was noted to be surrounded by thrombus
and hence not free. The farther (central) tip across the AV valve extended into the
RV and hence was not easily accessible either. Alternative options included lateral
grasp with a loop snare or using a forceps or using a pigtail catheter to loop around
the midsegment of the catheter to withdraw into IVC followed by its retrieval.[1]
[3]
[5] However, all these techniques required manipulation of these tools and the port
catheter within the heart, with the risk of trauma, arrhythmias, and increased risk
of further migration of the catheter itself during the manipulation.
Fixing the dislodged catheter in position to reduce the risk of its movement was achieved
using an angioplasty balloon inflated alongside the port catheter. This allowed for
a trilobed snare to be introduced from the neck access to manipulate the proximal
tip free of the thrombus and then to capture and retrieve it. Thus, the balloon minimized
the risk of pushing the port catheter as well as the associated clot further into
the RV/PA during the process.
The use of an angioplasty balloon has been described to assist in IFB retrieval, but
this is usually in the context of the balloon inflated within the IFB (typically a
migrated stent, sometimes a catheter) to capture it and retrieve it.[1]
[3] In our case, we used the balloon outside the lumen of the IFB to stabilize it and
to retain its position within the vessel. This facilitated the use of another retrieval
device from a separate access to capture the IFB.
The technique can be particularly useful to minimize migration especially when additional
manipulation of the IFB is anticipated during capture. We believe that this novel
technique adds one more trick to the interventional radiology repertoire for successful
IFB retrieval.