J Knee Surg 2021; 34(13): 1454-1462
DOI: 10.1055/s-0040-1710360
Original Article

All-Polyethylene Tibial Component Does Not Affect Survivorship of Medial Unicompartmental Knee Arthroplasty at Mid-Term Follow-Up

Michele Gagliardi
1   Department of Orthopaedic Surgery, Università degli Studi di Modena e Reggio Emilia, Modena, Emilia-Romagna, Italy
,
1   Department of Orthopaedic Surgery, Università degli Studi di Modena e Reggio Emilia, Modena, Emilia-Romagna, Italy
,
Alois Franz
2   Department of Orthopedic Surgery and Sports Traumatology, St. Marien-Krankenhaus - Siegen, Germany
,
Vitantonio Digennaro
3   1st Orthopedic Clinic, Istituto Ortopedico Rizzoli, Bologna, Emilia-Romagna, Italy
,
Fabio Catani
1   Department of Orthopaedic Surgery, Università degli Studi di Modena e Reggio Emilia, Modena, Emilia-Romagna, Italy
› Institutsangaben
 

Abstract

The present study's primary aim was to determine the survivorship of a large cohort of patients implanted with a single design all-polyethylene tibial component medial unicompartmental knee arthroplasty (UKA). Its secondary purpose was to investigate the reasons underlying implant failure, with specific attention to component positioning and limb alignment. Between 2007 and 2013, 166 patients underwent medial UKA with a single design all-polyethylene tibial component at two centers. Preoperatively and postoperatively, patients were administered clinical outcome scores and radiographic information were collected. Postoperative complications and causes of revision were recorded. A total of 140 patients (80 in Center A and 60 in Center B) who underwent all-polyethylene tibial component medial UKA (82 cases in Center A and 60 in Center B) were taken into account. Kaplan–Meier cumulative survivorship of implants was 96.5% (confidence interval [CI]: 91.7–98.6%) at an average follow-up of 61.1 months. Tibial aseptic loosening was accounted for failure in one case, while no correlation was found between implant positioning and failure. Two revisions were performed in Center A and three in Center B. Slight correction of the preoperative varus deformity was performed at both centers. All-polyethylene tibial component UKA provided satisfactory clinical and functional outcome, with excellent survival rate in the early and mid-term follow-up. Continued patient follow-up is needed to determine long-term survivorship of the examined UKA model.


The superiority of an all-polyethylene or a metal-backed tibial component in unicompartmental knee arthroplasty (UKA) for the treatment of knee osteoarthritis (OA) or osteonecrosis continues to be a point of debate. Some authors have reported reduced survival rates, poorer functional outcomes, and increased difficulties in intraoperative alignment with all-polyethylene UKAs compared with metal-backed implants,[1] [2] [3] whereas others[4] [5] [6] have demonstrated satisfactory outcomes and excellent survivorship with all-polyethylene prostheses. Despite the reduced costs and the theoretically improved wear characteristics favoring the use of all-polyethylene tibial components in UKA, it remains unclear whether the clinical outcomes and survivorship rates are comparable to metal-backed modular components.[7]

Only a few authors have assessed survivorship rates and clinical performance of UKA using an all-polyethylene tibial component at long-term follow-up. Heyse et al[8] reported a survivorship of approximately 93% at 10.8 years in a nonhomogeneous series of UKAs including medial and lateral components with both all-polyethylene and metal-backed tibial implants. At a mean follow-up of 10 years, Lustig et al[6] reported an overall survivorship of approximately 93% in all-polyethylene implants including medial and lateral UKAs. Recent papers report survival probability higher than 95% at a mean follow-up of 8 years, which is comparable to reports from studies using metal-backed UKAs.[9] Other studies have highlighted early failures in all-polyethylene UKAs, particularly in obese patients and in those with significant bony defects.[10] [11]

There are still differing interpretations about the correct placement and alignment of the prosthetic components relative to the design.[12] [13] Nevertheless, adequate limb and prosthetic components' alignment are considered to lead to satisfactory outcomes, no matter what tibial component is selected.[6] In general, a slight hypocorrection of the preoperative varus limb deformity and some residual degrees of varus in tibial component coronal alignment are advised.[4] [6] [10]

The primary aim of the present multicenter retrospective study was to determine the survivorship of a large cohort of patients implanted with a single design all-polyethylene tibial component medial UKA at a mid-term follow-up. Secondary purposes were to evaluate clinical outcomes and to investigate the reasons underlying implant failures, with specific attention to component placement and limb alignment. It was hypothesized that an all-polyethylene UKA would be a reliable prosthetic implant with superior or comparable overall survivorship to literature reports of other medial all-polyethylene implants and for metal-backed UKAs at mid-term follow-up.

Materials and Methods

This multicenter retrospective study included all patients undergoing medial all-polyethylene UKA from January 2007 to January 2013 at two different orthopaedic centers. All the operations were performed by two high-UKA volume surgeons, both of whom had at least 5 years' experience in this surgical procedure prior to the start of the case series. The present study was performed in accordance with the ethical standards of the 1964 Declaration of Helsinki, as revised in 2000. Patients were eligible for study inclusion if they had a diagnosis of primary unicompartmental OA, posttraumatic medial OA, or osteonecrosis with no prior history of joint replacement on the affected side and would comply with the follow-up schedule, with at least 1 year of follow-up. Surgical indications were homogeneous for both centers and were as follows: pain and tenderness localized to the medial joint line, active and passive knee flexion >85 degrees, extension lag <10 degrees, varus deformity <15 degrees as measured on long-leg standing radiographs, isolated medial compartment OA (Ahlback's grade 3–4), with no clinical evidence of patello-femoral and lateral compartment involvement. Anterior and posterior cruciate ligaments integrity was clinically checked. Exclusion criteria were as follows: prior history of osteotomy on the affected knee, coronal radiographic deformity >15 degrees, flexion contracture >15 degrees; functional incompetency of the anterior cruciate ligament (ACL) and peripheral ligaments, as well as the presence of inflammatory arthropathies or conditions determining inability to adhere to the study protocols. A total of 168 all-polyethylene tibial component UKAs (104 in Center A and 64 in Center B) were performed in 166 subjects. A total of 140 patients (142 knees, 84.5%) were included for clinical outcome analysis ([Table 1]).

Table 1

Patients' demographics and characteristics after application of inclusion/exclusion criteria

Parameter

Center A

Center B

Overall

Number of patients

80 (82 knees)

60 (60 knees)

140 (142 knees)

Males/females

47/33

26/34

67/73

Right/left

53/29

31/29

84/58

Age (y)

67.0

(min. 46.3, max. 85.1)

68.1

(min. 49.2, max. 86.2)

67.1(min. 46.3, max. 86.2)

Follow-up (mo)

61.8 (SD: 10.5, min. 14, max. 82)

64.1(SD: 11.8, min. 15, max. 86)

62.4

(SD: 11.2, min. 14., max. 86)

Abbreviations: max., maximum; min., minimum; SD, standard deviation.


Note: Values are reported as mean (standard deviations, min., and max.).


All-polyethylene tibial component implants of the JOURNEY UNI Unicompartmental Knee System (Smith & Nephew Inc., Memphis, TN) were implanted in all cases performing a minimally invasive mid-vastus approach. Surgery was performed according to the proposed surgical technique, as suggested by the manufacturer. The tibial coronal cut was performed respecting the tibial epiphyseal anatomy. The tibial cutting guide was adjusted in order to reproduce with the cut the angle subtended by the tibial joint line and the tibial mechanical axis on the coronal plane and the native tibial slope on the sagittal plane, as measured on the preoperative radiographs. After removal of tibial resection, the adequate trial polyethylene component was positioned, trying to use the trial component with minimum thickness (7 mm) and checking for flexion and extension gaps. In all knees, the femur cuts were performed aligning to the tibial cut surface. Postoperative care protocols were homogeneous for both centers. One day after surgery, blood drainage was removed, passive knee motion started, and partial weight bearing was allowed in all patients.

Data analysis was performed by one independent observer in Center A and one in Center B. Observers were not involved in surgical activity. Clinical assessment was performed preoperatively at 3 and 6 months postoperatively and then yearly thereafter. Those patients who did not attend two or more consecutive clinical assessments were contacted via telephone call and asked if they had undergone revision or reoperation for any reason with tick-box options.

Pre- and postoperative visual analog (0–10 cm) scale (VAS) for pain assessment was collected from all patients' medical records. Knee range of motion was measured with a goniometer pre- and postoperatively at every follow-up visit. Patients were administered the five subscales (pain, symptoms, function in activities of daily living, function in sports, and quality of life) of the Knee Injury and Osteoarthritis Outcome Score[14] and the Knee Society Score[15] pre- and postoperatively. Implant revision for any reason was considered as failure.

Full leg anteroposterior (AP) and knee lateral weight-bearing radiographs were taken preoperatively in all patients. Postoperatively, AP and lateral weight-bearing knee radiographs were taken 1 month, at 3, 6, and 12 months postoperatively, and then yearly. Radiographic measurements were performed on the last follow-up radiographs. On the preoperative radiographs, the following angles were measured: (1) mechanical femoro-tibial angle (mFTA; [Figure 1]). (2) Cartier's angle (CA), (3) anatomical femoro-tibial angle (aFTA), (4) tibial posterior slope (PS) ([Figure 2]). On the postoperative knee radiographs, the following angles were measured: (1) varus/valgus alignment of the tibial component, (2) aFTA, and (3) tibial component PS ([Figure 3]).

Zoom
Fig. 1 Preoperative full leg standing radiograph. (A) Mechanical femoro-tibial angle, angle between femoral and tibial mechanical axes.
Zoom
Fig. 2 Preoperative anteroposterior (1) and lateral (2) knee radiographs. (A) Cartier's angle, angle subtended between tibial joint line and tibial mechanical axis. (B) Anatomical femoro-tibial angle, angle between the femoral and tibial anatomical axes. (C) Tibial posterior slope, angle between the line perpendicular to the sagittal tibial axis and the posterior inclination of the medial tibial plateau.
Zoom
Fig. 3 Postoperative anteroposterior (1) and lateral (2) knee radiographs. (A) Varus/valgus alignment of the tibial component, angle between the tangent line to the tibial component and tibial anatomical axis. (B) Anatomical femoro-tibial angle, angle between the femoral and tibial anatomical axes. (C) Tibial posterior slope, angle between the line perpendicular to the sagittal tibial axis and the posterior inclination of the tibial component.

Statistical Analysis

Absolute frequency and percentage were calculated for categorical data, while means standard deviations and minimal and maximal values were used for continuous data. Statistical comparisons between paired pre- and postoperative parametric data and failed and revised radiographic outcome measures were performed using the Student's t-test. All tests were paired and two-sided and a p-value <0.05 was used to determine statistical significance. Survivorship analyses were performed using the Kaplan–Meier method, considering UKA revision surgery as endpoint for survivorship. The results were graphed and reported with 95% confidence interval. Statistical analyses were performed using Wolfram Mathematica (Wolfram Research, Inc., 100 Trade Center Drive, Champaign, IL).


Results

A total of 140 patients (142 knees, 84.5%) was included for clinical outcome analysis ([Table 1]). The remaining 26 patients were excluded due to missing postoperative outcome data. All the 166 patients (168 knees) included for study assessment were taken into account into the survival and radiographic analysis. The average follow-up was 61.1 months (standard deviation: 11.2, minimum 14 and maximum 86 months).

Two patients underwent UKA revision in Center A and three patients in Center B, resulting in an overall revision-free survival probability from implant of 96.5% (CI: 91.7–98.6%) at an average follow-up of 61.1 months (minimum 14 and maximum 86) of follow-up ([Figure 4]).

Zoom
Fig. 4 Kaplan–Meier survival probability of all-polyethylene tibial component unicompartmental knee arthroplasty, considering revision as endpoint.

Radiographic alignment assessment of failed implants was performed ([Table 2]). Case 1 was revised 14 months postoperatively for persistent lateral compartment pain, with radiographic examination revealing osteoarthritis progression of the lateral compartment. Case 2 underwent revision for tibial aseptic loosening 61 months postoperatively. Case 3 underwent revision for medial compartment instability at 21 months postoperatively. In case 4, revision was performed for incoming of medial osteonecrosis of the tibia, 30 months after surgery. Case 5 underwent revision in an external center for unexplained pain 48 months postoperatively. There were no cases of infection observed in either center. Radiographic outcome measures showed statistically significant differences in average preoperative CA (p < 0.001) and postoperative tibial posterior slope (p < 0.001) between failed and not-revised UKAs ([Table 3]).

Table 2

Cases of unicompartmental knee arthroplasty revision in Center A and B

Case

Radiographic parameter

Preoperatively

Postoperatively

∆ (preoperative –postoperative aFTA)

Case 1

(revision at 14 months postoperatively)

mFTA

–8.8 degrees

/

/

aFTA

–4.5 degrees

–2.0

2.5

Posterior tibial slope

4.4 degrees

6.0

1.6

Cartier angle/tibial component varus/valgus

–1.8 degrees

1.5

3.3

Case 2

(revision at 61 months postoperatively)

mFTA

–8.0

/

/

aFTA

–2.0

6.5

8.5

Posterior tibial slope

6.0

10.0

4.0

Cartier angle/tibial component varus/valgus

–10.0

0.0

10.0

Case 3

(revision at 21 months postoperatively)

mFTA

–9.6 degrees

/

/

aFTA

–6.0 degrees

–2.5

3.5

Posterior tibial slope

5.0 degrees

5.0

0.0

Cartier angle/tibial component varus/valgus

–2.0 degrees

–1.5 degrees

0.5

Case 4

(revision at 30 months postoperatively)

mFTA

–8.5

/

/

aFTA

–4.0

–1.0

3.0

Posterior tibial slope

3.0

3.0

0.0

Cartier angle/tibial component varus/valgus

–1.0

–1.0

0.0

Case 5

(revision at 48 months postoperatively

mFTA

–8.3

/

/

aFTA

–4.0

–2.5

1.5

Posterior tibial slope

5.0

4.5

0.5

Cartier angle/tibial component varus/valgus

0.0

–2.0

2.0

Abbreviations: aFTA, anatomical femoro-tibial angle; mFTA, mechanical femoro-tibial angle.


Note: Positive values are suggestive of valgus alignment, negative values of varus alignment.


Table 3

Preoperative and postoperative radiographic alignment of failed and surviving implants

Failed implants

(n = 5)

Surviving implants

(n = 135)

p-Value

Preoperative aFTA

–4.1 degrees (SD: 1.4)

0.7 degrees (SD: 2.8)

n.s.

Postoperative aFTA

–1.6 degrees (SD: 1.1)

2.8 degrees (SD: 2.9)

n.s.

Cartier's angle

–3.0 degrees (SD: 4.0)

–2.1 degrees (SD: 2.2)

<0.001

Tibial component varus/valgus

0.0 degrees (SD: 1.5)

–1.7 degrees (SD: 1.3)

n.s.

Preoperative tibial posterior slope

4.7 degrees (SD: 1.1)

7.6 degrees (SD: 1.4)

n.s.

Postoperative tibial posterior slope

5.7 degrees (SD: 2.6)

7.2 degrees (SD: 1.9)

<0.001

Abbreviations: aFTA, anatomical femoro-tibial angle; n.s., not statistically significant; SD, standard deviation.


Note: Values are reported as mean and standard deviations. Positive values are suggestive of valgus alignment, negative values of varus alignment.


The clinical and functional outcomes of KOOS, KSS, ROM, and VAS all showed significant improvements over preoperative values for patients treated both at Center A and Center B ([Table 4]).

Table 4

Patients' reported outcome measures in Center A and B

Parameter

Preoperative

Postoperative

p-Value

Range of motion

99.3 degrees (SD: 12.2, min. 90, max. 125)

128.2 degrees (SD: 10.3, min. 90, max. 135)

28.9 degrees

 < 0.0001

VAS (0–10)

5.3 (SD: 1.2, min. 3, max. 8)

1.3 (SD: 1.6, min. 0, max. 6)

4.0

< 0.0001

KOOS

 Pain

39.0 (SD: 17.0)

85.7 (SD: 17.3)

46.7

< 0.001

 Symptoms

47.3 (SD: 20.9)

89.5 (SD: 14.3)

42.2

< 0.001

 Function in activities of daily living

40.8 (SD: 19.3)

87.2 (SD: 7.4)

46.4

< 0.001

 Function in sports and recreation

26.0 (SD: 17.0)

67.0 (SD: 27.0)

41.0

< 0.001

 Quality of Life

24.0 (SD: 10.0)

77.6 (SD: 26.0)

53.6

< 0.001

 Total

35.4 (SD: 17.3)

81.4 (SD: 19.4)

46.0

< 0.001

KSS

 Knee

51.0 (SD: 8.5)

90.3 (SD: 12.3)

39.3

< 0.00001

 Function

41.7 (SD: 12.2)

89.2 (SD: 15.2)

47.5

< 0.00001

Abbreviations: KOOS, Knee Injury and Osteoarthritis Outcome Score; KSS, Knee Society Score; max., maximum; min., minimum; SD, standard deviation; VAS, Visual Analogue Scale.


Note: Values are reported as mean (standard deviations, min., and max.).


All patients had an average radiographic varus preoperative alignment of the operated knee. No statistical significance was revealed between pre- and postoperative CA and tibial posterior slope ([Table 5]).

Table 5

Mean pre- and postoperative radiographic alignment of patients in Center A, Center B, and cumulative data

Center A

(n = 82)

Center B

(n = 60)

Overall

(n = 142)

p-Value

Preoperative mFTA

–6.3 degrees (SD: 3.2)

–6.9 degrees (SD: 2.6)

–6.6 degrees (SD: 2.9)

Preoperative aFTA

0.4 degrees (SD: 2.8)

0.9 degrees (SD: 3.0)

0.6 degrees (SD: 2.9)

Postoperative aFTA

3.1 degrees (SD: 3.3)

2.5 degrees (SD: 2.8)

2.8 degrees (SD: 3.0)

∆ (preoperative–postoperative aFTA)

2.2 degrees (SD: 5.9)

< 0.001

Cartier's angle

–2.4 degrees (SD: 3.0)

–1.4 degrees (SD: 1.1)

–2.1 degrees (SD: 2.1)

Tibial component varus/valgus

–2.1 degrees (SD: 1.8)

–1.1 degrees (SD: 1.0)

–1.7 degrees (SD: 1.4)

∆ (Cartier's angle–tibial component varus/valgus)

0.4 (SD: 3.5)

n.s.

Preoperative posterior tibial slope

7.0 degrees (SD: 3.0)

8.2 degrees (SD: 1.9)

7.5 degrees (SD: 1.6)

Postoperative posterior tibial slope

6.4 degrees (SD: 2.0)

8.3 degrees (SD: 1.9)

7.2 degrees (SD: 2.0)

∆ (preoperative–postoperative posterior tibial slope)

0.3 degrees (SD: 3.6)

n.s.

Abbreviations: aFTA, anatomical femoro-tibial angle; mFTA, mechanical femoro-tibial angle, n.s., not statistically significant; SD, standard deviation.


Note: Values are reported as mean and standard deviations. Positive values are suggestive of valgus alignment, negative values of varus alignment.



Discussion

The most important finding of the present multicenter, retrospective study is the report of 96.5% Kaplan–Meier survival probability at just over 5 years of average follow-up of a single design all-polyethylene tibial component medial UKA at two different centers. Reports of 5-year survival rates of medial all-polyethylene and metal-backed UKAs range from 86.0 to 96.1%, and identify loosening of the tibial component, polyethylene wear, and progression of OA as major problems.[9] [11] [13] [16] [17] Therefore, to the best of the authors' knowledge, the survivorship reported in the present study is the highest in literature for all-polyethylene tibial component UKAs so far.

A recent review analyzed causes of UKA failure, reporting that tibial component loosening (36%) and progression of OA in the contralateral compartment (20%) are the most common causes of UKA failure.[11] Aseptic loosening has been more frequently described as an early cause of failure, whereas progression of OA has been more frequently reported in mid- and long-term follow-ups. Concerns have also been raised regarding the incidence of tibial loosening and subsidence in all-polyethylene tibial components,[18] and the role of stiffness in unexplained pain.[3] [19] Different survival rates[16] [18] in all-polyethylene designs suggest that not all all-polyethylene designs are the same, and this may reflect component thickness.[20] Components of 6 mm thickness have been significantly associated with increased wear and osteolysis.[21] Pathological cancellous bone overload and tibial subsidence may affect thinner implants more so than thicker implants.

In the present study, five patients underwent UKA revision, two in Center A (Case 1 and Case 2) and three in Center B (Case 3, Case 4, and Case 5). One early revision (Case 1) at 14 months was performed for persistent pain on the contralateral compartment. Radiographic evaluation revealed progression of OA in the contralateral and patello-femoral compartment; thus, total knee arthroplasty (TKA) was performed. Another implant (Case 2) was revised at 61 months for persistent pain localized on the medial compartment. Radiographs did not demonstrate progressive radiolucency lines, but during surgical revision to TKA, tibial component aseptic loosening was noted and determined as the cause of failure. Causes of UKA failure in Center B included medial instability, which made revision necessary 21 months after surgery, and medial osteonecrosis of the tibia, which occurred 30 months after treatment. Persistent and unexplained pain was reported in one patient, making UKA revision necessary 48 months postoperatively. Only in one case (Case 2), the authors reported deviations from optimal coronal and sagittal prosthetic alignment, with a significant correction of the preoperative CA (–10 degrees) to a neutral alignment of the tibial component. No other alignment defects were noted in revised patients in any center.

Although there are no direct correlations between alignment deviations and clinical results, current literature underlines that deviations in coronal tibial component alignment can lead to polyethylene wear and early loosening.[10] [22] [23] [24] As noted by Hernigou et al,[21] a slight undercorrection of the preoperative tibial varus deformity is likely to be performed. Others recommend avoiding tibial coronal alignment higher than 5 to 6 degrees of varus and sagittal slope of more than 7 degrees,[4] [6] [13] [25] in particular when performing all-polyethylene tibial component UKAs.[5] In the present study, medial tibial plateau alignment on both coronal and sagittal plane was preserved, with no statistically significant differences reported between radiographic preoperative CA and postoperative tibial component angle, nor for preoperative and postoperative tibial posterior slope.

Postoperative aFTA was less than 6 degrees of valgus in all patients, with the exception of a single case that underwent revision to TKA after 61 months (case 2, Center A).

The excellent mean postoperative KOOS scores and the statistically significant difference between pre- and postoperative mean clinical subscales, highlight the efficacy of the examined all-polyethylene tibial component UKA in restoring knee function. The mean “Function in Sports and Recreation” KOOS subscale did not reach excellent results. This specific result is explained by patient comorbidities, including OA on the contralateral knee and other age-related physical impairment.

The present study has several limitations. First, postoperative knee mechanical alignment was not available for most of the patients, as weight-bearing radiographs of the lower limbs are frequently not requested for postoperative implant evaluation. However, studies have demonstrated that standard-view knee radiographs are reliable instruments to evaluate postoperative knee alignment.[26] Second, the sample of patients is rather small, and only five cases of failure were available for assessment. Third, body mass index (BMI) and its effect on UKA outcome were not considered in the present investigation. Increased BMI has recently been correlated with increased risks of failure in all-polyethylene tibial components UKA; however, none of the revised patients had BMI higher than 30. Follow-up length is heterogeneous, but it is the authors' hope that further studies with longer follow-ups would better evaluate effects of all-polyethylene tibial component positioning on UKA failure.


Conclusion

All-polyethylene tibial component UKAs provided satisfactory clinical and functional outcome, with a low number of UKA revision in the early- and mid-term follow-up. Aseptic tibial loosening, which UKA failure has often been accounted for in all-polyethylene tibial components UKA, was reported only in one case. This demonstrates that appropriate positioning and polyethylene insert thickness are factors to be taken into account when choosing an all-polyethylene tibial component UKA.



Conflict of Interest

None declared.


Address for correspondence

Francesco Zambianchi, MD
Department of Orthopaedic Surgery
Università degli Studi di Modena e Reggio Emilia, Modena, Emilia-Romagna
Italy   

Publikationsverlauf

Eingereicht: 15. Januar 2019

Angenommen: 21. März 2020

Artikel online veröffentlicht:
25. Mai 2020

© 2020. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA


Zoom
Fig. 1 Preoperative full leg standing radiograph. (A) Mechanical femoro-tibial angle, angle between femoral and tibial mechanical axes.
Zoom
Fig. 2 Preoperative anteroposterior (1) and lateral (2) knee radiographs. (A) Cartier's angle, angle subtended between tibial joint line and tibial mechanical axis. (B) Anatomical femoro-tibial angle, angle between the femoral and tibial anatomical axes. (C) Tibial posterior slope, angle between the line perpendicular to the sagittal tibial axis and the posterior inclination of the medial tibial plateau.
Zoom
Fig. 3 Postoperative anteroposterior (1) and lateral (2) knee radiographs. (A) Varus/valgus alignment of the tibial component, angle between the tangent line to the tibial component and tibial anatomical axis. (B) Anatomical femoro-tibial angle, angle between the femoral and tibial anatomical axes. (C) Tibial posterior slope, angle between the line perpendicular to the sagittal tibial axis and the posterior inclination of the tibial component.
Zoom
Fig. 4 Kaplan–Meier survival probability of all-polyethylene tibial component unicompartmental knee arthroplasty, considering revision as endpoint.