J Wrist Surg 2020; 09(03): 214-218
DOI: 10.1055/s-0040-1702930
Scientific Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Diaphyseal Screw Prominence in Distal Radius Volar Plating

1   Barwon Health, Deakin University Geelong, Geelong, Australia
2   Deakin University Geelong, Geelong, Australia
,
1   Barwon Health, Deakin University Geelong, Geelong, Australia
2   Deakin University Geelong, Geelong, Australia
,
1   Barwon Health, Deakin University Geelong, Geelong, Australia
2   Deakin University Geelong, Geelong, Australia
› Author Affiliations
Further Information

Publication History

11 July 2019

07 January 2020

Publication Date:
18 March 2020 (online)

Abstract

Background Volar plating for distal radius fractures has become common. Screw prominence on the dorsal side from long screws can lead to tendon injury. Methods for detecting screws that penetrate the far cortex involve X-ray or ultrasound. These have focused on the distal row of screws. No studies have addressed screw penetration in the diaphysis. We describe two cases where diaphyseal screws caused symptoms. We then insert screws in the diaphysis of synbones 2 mm longer than measured and determine what angle of pronation or supination was best to detect this on X-ray.

Methods Three synbones were plated using Synthes volar plate. The three diaphyseal screws were drilled perpendicular to the plate, and the depth measured. Cortical 2.4-mm screws were inserted, 2 mm longer than measured. The three synbones were then placed in a custom clamp to measure rotation. Lateral X-rays were taken at 0 degree rotation, and 5, 10, and 15 degrees of supination and pronation. The prominence of each screw was measured using the synapse digital ruler.

Results For the screws that were placed at a neutral angle (perpendicular to the plate) the maximum visualization of the prominent tips occurred around 0 degree rotation. With screws angled 15 degrees ulna, maximum visualization was between 5 and 10 degrees of pronation. With screws angled 15 degrees radial, maximum visualization was between 5 and 10 degrees of supination. Every 5 degrees of rotation changes the profile of the screw by 0.4 mm.

Discussion The diaphysis of the radius becomes approximately trapezoidal distally. Prominent screws that are placed below the “peak” of the trapezoid may appear to be the correct length. Rotating the wrist into pronation or supination to bring the relevant cortex as parallel to the X-ray beam as possible will help to identify if screws are of the correct length. Screws that are prominent in the second compartment may be particularly symptomatic as the tendons here are closely opposed to the bone. We recommend screening for 2 mm diaphyseal screw prominence in neutral, with 10 degrees of pronation and supination.

 
  • References

  • 1 Arora R, Lutz M, Hennerbichler A, Krappinger D, Espen D, Gabl M. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma 2007; 21 (05) 316-322
  • 2 Benson EC, DeCarvalho A, Mikola EA, Veitch JM, Moneim MS. Two potential causes of EPL rupture after distal radius volar plate fixation. Clin Orthop Relat Res 2006; 451 (451) 218-222
  • 3 Boyer MI, Korcek KJ, Gelberman RH, Gilula LA, Ditsios K, Evanoff BA. Anatomic tilt x-rays of the distal radius: an ex vivo analysis of surgical fixation. J Hand Surg Am 2004; 29 (01) 116-122
  • 4 Oc Y, Kilinc BE, Gulcu A, Varol A, Ertugrul R, Kara A. Ultrasonography or direct radiography? A comparison of two techniques to detect dorsal screw penetration after volar plate fixation. J Orthop Surg Res 2018; 13 (01) 70
  • 5 Cha SM, Shin HD. Detections of screw penetration during volar plating for distal radius fractures. J Wrist Surg 2017; 6 (04) 340-348
  • 6 Dolce D, Goodwin D, Ludwig M, Edwards S. Intraoperative evaluation of dorsal screw prominence after polyaxial volar plate fixation of distal radius fractures utilizing the Hoya view: a cadaveric study. Hand (N Y) 2014; 9 (04) 511-515
  • 7 Giugale JM, Fourman MS, Bielicka DL, Fowler JR. Comparing dorsal tangential and lateral views of the wrist for detecting dorsal screw penetration after volar plating of distal radius fractures. Adv Orthop 2017; 2017: 1402517
  • 8 Hill BW, Shakir I, Cannada LK. Dorsal screw penetration with the use of volar plating of distal radius fractures: how can you best detect?. J Orthop Trauma 2015; 29 (10) e408-e413
  • 9 Joseph SJ, Harvey JN. The dorsal horizon view: detecting screw protrusion at the distal radius. J Hand Surg Am 2011; 36 (10) 1691-1693
  • 10 Klammer G, Dietrich M, Farshad M, Iselin L, Nagy L, Schweizer A. Intraoperative imaging of the distal radioulnar joint using a modified skyline view. J Hand Surg Am 2012; 37 (03) 503-508
  • 11 Lee SK, Bae KW, Choy WS. Use of the radial groove view intra-operatively to prevent damage to the extensor pollicis longus tendon by protruding screws during volar plating of a distal radial fracture. Bone Joint J 2013; 95-B (10) 1372-1376
  • 12 Maschke SD, Evans PJ, Schub D, Drake R, Lawton JN. Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study. Hand (N Y) 2007; 2 (03) 144-150
  • 13 Ozer K, Wolf JM, Watkins B, Hak DJ. Comparison of 4 fluoroscopic views for dorsal cortex screw penetration after volar plating of the distal radius. J Hand Surg Am 2012; 37 (05) 963-967
  • 14 Sügün TS, Karabay N, Gürbüz Y, Ozaksar K, Toros T, Kayalar M. Screw prominences related to palmar locking plating of distal radius. J Hand Surg Eur Vol 2011; 36 (04) 320-324
  • 15 Baumbach SF, Synek A, Traxler H, Mutschler W, Pahr D, Chevalier Y. The influence of distal screw length on the primary stability of volar plate osteosynthesis--a biomechanical study. J Orthop Surg Res 2015; 10: 139
  • 16 Overturf SJ, Morris RP, Gugala Z, Lindsey RW. Biomechanical comparison of bicortical locking versus unicortical far-cortex-abutting locking screw-plate fixation for comminuted radial shaft fractures. J Hand Surg Am 2014; 39 (10) 1907-1913
  • 17 Looft JM, Corrêa L, Patel M, Rawlings M, Ackland DC. Unicortical and bicortical plating in the fixation of comminuted fractures of the clavicle: a biomechanical study. ANZ J Surg 2017; 87 (11) 915-920
  • 18 Hart AJ, Seepaul T, Hewitt RJ, Ang S, Hansen U, Amis AA. The palmar locking compression plate is biomechanically comparable to the dorsal pi plate for dorsally comminuted, intraarticular wrist fractures. J Hand Surg Eur Vol 2007; 32 (04) 388-393
  • 19 Leong JJ, Leff DR, Das A. , et al. Validation of orthopaedic bench models for trauma surgery. J Bone Joint Surg Br 2008; 90 (07) 958-965
  • 20 Poole WEC, Marsland D, Durani P. , et al. Detecting dorsal screw penetration following volar plate fixation of the distal radius – a validation study in a saw bone model. Trauma 2016; 19: 46-53