Thromb Haemost 2001; 86(02): 575-583
DOI: 10.1055/s-0037-1616089
Review Article
Schattauer GmbH

Anticardiolipin Antibody Assay: a Methodological Analysis for a better Consensus in Routine Determinations

A Cooperative Project of the European Antiphospholipid Forum
Angela Tincani
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Flavio Allegri
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Marielle Sanmarco
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Massimo Cinquini
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Marco Taglietti
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Genesio Balestrieri
1   Clinical Immunology Unit, Brescia Hospital, Italy, Laboratoire d’Immunologie, Hopital de la Conception, Marseille, France
,
Takao Koike
2   Department of Medicine II, Hokkaido University School of Medicine, Sapporo, Japan
,
Kenji Ichikawa
2   Department of Medicine II, Hokkaido University School of Medicine, Sapporo, Japan
,
Pierluigi Meroni
3   IRCCS Istituto Auxologico Italiano, University of Milan, Italy
,
Marie Claire Boffa
4   Department of Internal Medicine JC, Piette, Hôpital de la Pitié, Paris, France
› Author Affiliations
Further Information

Publication History

Received 28 July 2000

Accepted after resubmission 06 March 2001

Publication Date:
12 December 2017 (online)

Preview

Summary

Despite the widely recognized practical importance of anticardiolipin (aCL) ELISA, the reliability of this test has been recently discussed. In order to investigate this area on European scale, we sent to 30 experienced centers a questionnaire focusing on the diagnostic procedures applied to patients with antiphospholipid syndrome (APS) and on the detailed protocols used to perform aCL. Anticardiolipin ELISA was found to be the most frequently performed test in patients with suspected APS, but significant difference was shown among the various protocols. The cross-laboratory multiple examination of ten serum samples evaluated independently by the 24 centers pointed out the difficulty in getting comparable results. Therefore a “consensus” protocol was derived from the aCL methods giving the best performance. The materials and reagents necessary to perform the “consensus” method, including, as putative standards, one IgG and one IgM monoclonal antibody (HCAL and EY2C9) were distributed to 19 Centers. The results of one IgG and one IgM aCL high positive sera measured in serial dilutions were compared. A progressive decrease in the variability of the values obtained for a given sample appeared evident when all the laboratories used the same standard, in their own in-house ELISA and even more in the “consensus” ELISA.

Our data show that aCL ELISA standardization is necessary in order to obtain comparable results in different laboratories.