The European Concerted Action on Anticoagulation (ECAA): Field Studies of Coagulometer Effects on the ISI of ECAA Thromboplastins

L. Poller
1   From the Central Facility, ECAA, The University of Manchester, UK
,
A. M. H. P. van den Besselaar
2   From the Haemostasis and Thrombosis Research Centre, Leiden University Medical Centre, NL
,
J. Jespersen
3   From the Department of Clinical Biochemistry, Centralsygehuset, Esbjerg, DK
,
A. Tripodi
4   From the A Bianchi Bonomi, Hemophilia and Thrombosis Centre, University & IRCCS Maggiore Hospital, Milan, IT
,
D. Houghton
1   From the Central Facility, ECAA, The University of Manchester, UK
,
› Author Affiliations
Further Information

Publication History

Received 27 January 1998

Accepted after revision 05 June 1998

Publication Date:
08 December 2017 (online)

Preview

Summary

Local calibration studies have been performed with lyophilized plasmas at 155 European laboratories to assess coagulometer effects on manual ISI of ECAA thromboplastins (low ISI human recombinant and high ISI rabbit brain). Common sets of 7 normal and 20 artificially depleted lyophilized plasmas were tested with the thromboplastins in the routine local coagulometers. With the human reagent, marked lowering of the manual ISI resulted with most coagulometers, which was associated with disproportionate shortening of the normals. Where this shortening did not occur, there was little coagulometer effect on the ISI. With the rabbit reagent, proportionate shortening with both normal and abnormal plasmas occurred in most instruments with little effect on ISI. INR correction by local ISI assignment appeared successful with the ECAA human reagent. There was negligible INR correction from local calibration of the ECAA rabbit reagent in coagulometers where thromboplastin ISI were unchanged from the ECAA established manual values.