Key words
safety - magnetic resonance imaging - cardiac - MRI, pacemaker - MRI, implants - MRI,
specific absorption rate
Introduction
Due to excellent soft tissue contrast and absence of radiation exposure, magnetic
resonance imaging (MRI) is currently recommended by European and American guidelines
as the imaging method of choice in many clinical scenarios [1]
[2]
[3]
[4]. Parallel to this, the number of patients with implanted cardiac pacemakers (PM)
continues to increase; in Germany alone, the new implantation rate is about 75 000
annually, with approximately 600 000 patients currently living with the device [5]. It is estimated that due to existing comorbidities, approximately 70 % of these
typically older patients will exhibit a clinical indication for an MR examination
at least once in the course of their lives [6].
However, there are numerous possible interactions between the static and pulsed electromagnetic
fields of a magnetic resonance imaging system and a pacemaker system [6]
[7]
[8]
[9]
[10]
[11]
[12]
[13]
[14]
[15]
[16]
[17]
[18]
[19]. Essentially these interactions pose three main risks – each potentially life-threatening
– for a pacemaker patient during an MR examination:
-
False perception of pulsed MRI gradient fields as intrinsic cardiac activity with
inhibition of the pacemaker stimulation function and consecutive asystole in pacemaker-dependent
patients,
-
voltage induction into the pacemaker leads through coupling of the magnetic pulsating
fields in a metallic conductor with asynchronous myocardial stimulation and the potential
risk of induction of ventricular tachycardia/ventricular fibrillation, as well as
-
strong local concentration of the energy of the high frequency field due to the antenna
effect of the pacemaker leads with potential thermal damage in the myocardium/endocardium
at the lead tips and consecutive ineffective PM stimulation due to increase of the
pacing capture threshold.
With this in mind, two developments should be mentioned which have high clinical relevance
for the management of patients with pacemakers scheduled for MR imaging:
-
Based on scientific studies and clinical experience over the past 15 years [7]
[8]
[9]
[10]
[11]
[12]
[13]
[14]
[20]
[21]
[22]
[23]
[24]
[25]
[26]
[27]
[28], the presence of a conventional pacemaker system – according to the overwhelming
majority of experts and relevant national and international professional associations
[9]
[29] – no longer represents an absolute, but rather a relative contraindication for an
MR exam. MRI can be performed taking into consideration the individual risk/benefit
profile as a single-case decision and as “off-label” use applying dedicated safety
measures.
-
In 2008 a new technology, so-called “MR conditional” pacemaker systems were introduced,
initially by Medtronic, and are now provided by all pacemaker manufacturers. These
PM systems – defined as a functional unit composed of pacemaker pulse generator and
leads – are tested and approved for MRI under certain conditions (“in-label” use)
[24]. The problem with these conditions is that they can vary significantly depending
on the manufacturer and specific pacemaker model. Decisive for patient safety during
an MR examination is precise understanding of, and compliance with, these conditions
and the terms of use for the specific pacemaker system.
Recent experience in Germany has shown that there are currently a few specialized
centers that perform MRI on patients with MR conditional pacemaker systems. Nevertheless,
an MR exam is still withheld from many patients with pacemakers, even when the pacemaker
is conditionally MRI-safe. One reason for this is a significant lack of information
and legal uncertainty on the part of both radiology and cardiology with respect to
dealing concretely in the clinical setting with these new developments and the technically
complex and interdisciplinary issue.
The aim of these clinical practice guidelines of the German Radiology Society (DRG),
Working Group on Cardiovascular Imaging is to present, from the radiological viewpoint,
necessary background knowledge and to express specific recommendations for performing
MR imaging on patients with conventional and MR conditional cardiac pacemaker systems.
In the medium term, development of a sufficient number of centers with relevant expertise
should be encouraged in order to guarantee enough clinical MRI facilities for the
large patient population with pacemakers in Germany.
Further, these clinical practice guidelines should substantiate and supplement from
the radiological perspective the abbreviated description of this complex topic in
the current guidelines of the European Society of Cardiology [29], and provide corrections of some points.
MRI of patients with conventional pacemakers
MRI of patients with conventional pacemakers
The recommended procedural management of patients with conventional pacemakers is
shown in Algorithm 1 ([Fig. 1]). The individual steps are explained below.
Fig. 1 Algorithm 1.
Critical review of indications for MR imaging
The clinical urgency and therapeutic consequences of a requested MR examination should
be documented by the referring clinical physician; the lack of adequate imaging alternatives
should be indicated by the attending radiologist. An interdisciplinary decision regarding
the performance or withholding of the examination should be made on a case-by-case
basis by the referring physician, cardiologist and radiologist.
Critical for this decision is the estimation of the individual risk/benefit ratio
taking into account the electrophysiological risk profile ([Table 1]), the specific risk of the respective region to be examined ([Table 2]), as well as local circumstances (expertise and collaboration among attending radiological
and cardiological personnel, guarantee of qualified patient monitoring during the
MR examination, as well as adequate management of any possible emergency situation).
Table 1
High risk electrophysiological parameters for MRI of PM patients.
electrophysiological parameters
|
potential hazards
|
pacemaker-dependence of patient
|
ineffective stimulation due to RF-induced tissue heating at the tips of the leads
with increase of the pacing capture thresholds
inhibition of stimulation due to gradient fields (false sensing)
|
vulnerable myocardium (acute/subacute myocardial infarction, acute myocarditis), arrhythmogenic
substrate/structural cardiac disease
|
increased risk for induction of ventricular fibrillation/ventricular tachycardia via
voltage induction in the leads due to gradient fields or rectified RF fields
|
abandoned pacemaker leads
|
elevated risk of RF-induced tissue heating on the tips of abandoned leads compared
to leads connected to the pacemaker system
|
additional leads (e. g. coronary sinus/epicardial leads)
lead extensions, adapters
|
elevated risk of RF-induced heating effects at the lead tips due to addition of the
individual antenna effects and/or elongation of the antenna route of the leads
|
lead defect
|
elevated risk of RF-induced heating effects at the lead tips
|
metallic cardiac or extracardiac implants (length > 5 cm) directly adjacent (< 4 cm)
to the leads
|
elevated risk of RF-induced heating effects at the lead tips due to addition of the
individual antenna effects
|
primarily increased pacing capture thresholds
|
insufficient stimulation due to RF-induced tissue heating at the tips of the leads
with further increase of the pacing capture thresholds
|
low battery voltage (ERI, EOL criteria)
|
elevated risk of switching into emergency mode (electrical reset), which is usually
VVI
|
implantation time < 6 weeks
|
unstable pacing capture thresholds in the healing phase after lead implantation
|
Table 2
Risk assessment of the MR examination region of patients with pacemakers.
MR examination region: higher risk
|
potential hazards
|
thoracic spine, heart, chest, breast, shoulder
|
leads located completely in RF transmitter coil with increased coupling of RF energy
and potentially increased heating effects at the lead tips
|
MR examination region: lower risk
|
potential hazards
|
brain, pelvis, hip joint, knee, foot
|
leads outside/largely outside RF transmitter coil with limited RF energy coupling
and only minimal heating effects at the lead tips
|
Due to its primary significance in determining the indication for examination and
review in the case of a patient with a conventional pacemaker, this risk/benefit analysis
is illustrated using a few examples.
The expected benefit of an MRI scan of the brain when planning surgery of an intracerebral
tumor in a patient with a pacemaker should usually be rated as high. The relatively
low risk profile of an MR examination of the brain ([Table 2]) – particularly in the absence of additional risk-elevating electrophysiological
parameters ([Table 1]) – would essentially support performing an MRI scan, based on the risk/benefit ratio.
On the other hand, an MR examination of the thoracic spine of a patient with a conventional
pacemaker should be undertaken with extreme caution due to the significantly increased
risk of HF-induced heating effects of the leads in this anatomical region ([Table 2]). However, in a case of urgent clinical indication such as incipient paraparesis
related to decompensated spondylodiscitis, the anticipated clinical benefit may be
so great that this increased risk can be considered acceptable, particularly if the
patient is not pacemaker-dependent, so that the primary risk of ineffective pacemaker
stimulation due to heating-induced increased pacing capture thresholds is of lower
clinical importance.
Of the electrophysiological risk parameters illustrated in [Table 1], the first three items – 1. pacemaker dependency, 2. vulnerable myocardium/increased
disposition for high-grade arrhythmias, as well as 3. abandoned pacemaker leads –
should be emphasized as main risks during an MR examination of a patient with a conventional
pacemaker. However, none of these three points necessarily imply withholding an MR
examination if – taking into account the urgent clinical need for MR imaging – the
specific risk/benefit ratio is considered positive for the patient on the whole.
Patient information / informed consent
During an appropriate time frame prior to the MR examination, the patient should be
specifically informed regarding the following circumstances and risks.
-
MRI on a patient with a conventional pacemaker is a non-approved (“off-label”) procedure
based on a case-by-case decision on the part of the attending physicians. The relevant
authorities have not issued certification for the MR examination, and neither the
pacemaker manufacturer nor the manufacturer of the MRI equipment will be liable for
any damage or complications.
-
An MR examination of a patient with a conventional pacemaker system poses the following
risks and complications:
-
damage to the pacemaker generator with the necessity of replacement
-
malfunction of the pacemaker leads, e. g. due to heating-related increase of pacing
capture thresholds, necessitating revision/new implantation of leads
-
heating of the pacemaker leads resulting in thermal damage to the cardiac muscle (acute
or chronic) and ineffective pacemaker stimulation as a potentially life-threatening
complication in the case of absolute PM dependency of the patient
-
induction of potentially life-threatening tachycardia arrhythmia
-
inhibition of pacemaker therapy of spontaneous bradyarrhythmic episodes occurring
during the MRI scan with potentially life-threatening consequences in patients with
absolute PM dependency
Occurrence of the risks described above can be minimized by employing specific safety
precautions; however these risks cannot be absolutely eliminated and are not quantifiable
in individual cases.
Monitoring during MRI
A decisive element of safety in the procedural management of patients with conventional
pacemakers is adequate monitoring of vital functions in order to detect and treat
potentially life-threatening complications early – in particular the occurrence of
high-grade ventricular arrhythmias, ventricular fibrillation or spontaneous episodes
of bradycardia. Continuous pulse oximetry monitoring is mandatory, displaying the
peripheral pulse wave and oxygen saturation using a pulse oximeter certified for patient
monitoring in the MR environment. It should be noted that the pulse wave display integrated
into most MRI systems is not sufficient on its own. The advantage of pulse oximetry
monitoring is that it is not affected by either the static magnetic field or the pulsed
radio frequency (RF) and gradient fields, thus allowing accurate assessment of the
heart rate as well as oxygen saturation of the patient in the MR environment. Accurate
rhythm analysis (differentiation between atrial and ventricular tachycardia) is not
possible, but also not mandatory during an MR examination. In the event of a hemodynamically
significant bradycardiac or tachycardiac arrhythmia indicated by a decrease of oxygen
saturation, the examination must be immediately discontinued and the patient brought
out of the MRI room into an area in which standard emergency equipment can be employed.
In this context it should be noted that defibrillation in the static magnetic field
or within the 5-Gauss line is absolutely contraindicated due to magnetic attraction
forces acting on the defibrillator. Furthermore, in a strong static magnetic field
– such as in the direct vicinity of a 1.5 T MRI system – charging the capacitor of
a defibrillator and thus generating a therapeutic shock, is technically not possible
with currently available equipment.
ECG monitoring as the sole control modality during MRI of patients with conventional
pacemakers, is definitely insufficient due to artifacts induced by pulsed RF and gradient
fields. However, it is recommended as a monitoring modality supplementary to pulse
oximetry, particularly when the examination has a high risk profile. In this instance
as well, an MRI-compatible ECG unit certified for patient monitoring in the MR environment
should be used.
The monitoring phase starts immediately upon entry into the MRI room. It is recommended
to attach the pulse oximeter as well as to establish i. v. access outside the MRI
room. During the MR examination, continuous verbal and visual contact with the patient
should be guaranteed, and the patient should be instructed to immediately report any
symptoms such as palpitations, dizziness, heat sensation or movement in the pacemaker
pocket.
In order to respond immediately to high-grade ventricular arrhythmias as well as episodes
of spontaneous bradycardia, on-site availability of a defibrillator as well as a programmer
compatible with the respective pacemaker system is absolutely mandatory.
For safety reasons, the presence of a cardiologist during MRI should be requested
for patients with conventional pacemaker systems and higher risk profile ([Table 1]) – particularly for patients with absolute PM dependency, for patients with abandoned
pacemaker leads, as well as for patients with vulnerable myocardium.
For practical reasons, it would be reasonable for conventional pacemaker patients
with a low risk profile ([Table 1], [2]) to have a cardiologist available on an emergency stand-by basis rather than being
continuously present during MRI. In this case, adequate patient monitoring during
the MR examination should be guaranteed by the presence of a physician who is capable
to treat the patient properly until the arrival of a cardiologist.
MRI-related safety precautions
By far, the most numerous clinical experiences are with closed/cylindrical MRI units
with a field strength of 1.5T; thus these MRI systems are preferable for scheduled
MR examinations of patients with conventional pacemakers [6]
[8]
[10]
[11]
[22]
[23]
[24]
[25]. There are studies demonstrating the safety of MRI of the brain using a field strength
of 3 T [20]. Due to the lack of sufficient data, open MRI systems should not be employed.
The built-in body RF coil, a large volume coil integrated permanently in the MR scanner,
should be used as the RF transmitter coil. Local transmit/receive coils are absolutely
contraindicated in the thoracic region. The use of local receiver coils is non-critical
and possible in all regions of the body – including the chest – if the integrated
body RF coil is used as the transmitter coil.
RF-induced heating of the PM leads cannot be predicted in individual cases, as it
is dependent on numerous influencing parameters, including position and configuration
of the PM leads in the patient as well as position relative to the RF transmitter
coil, relation of the wave length of the RF excitation pulse to the effective lead
length, SAR value (specific absorption rate) of the MRI sequence used. The SAR value
(unit W/kg) is a measure of the absorption of electromagnetic field energy in biological
tissue. To limit consecutive tissue heating during an MRI scan, the radio frequency
power irradiated into the body is monitored on the equipment and the corresponding
SAR value is displayed. In keeping with legally-mandated IEC threshold values, [30] the SAR value during an MR examination must not exceed 4 W/kg in the whole body
or 3.2 W/kg in the head independent of the presence or absence of a pacemaker or an
otherwise active or inactive implant.
If all other influential parameters remain constant, there is a linear relationship
between the SAR value of the related MRI sequence and tissue heating at the tip of
the pacemaker lead. This RF-induced tissue heating can thus be simply and effectively
reduced by limiting the SAR value. It is therefore strongly recommended that during
MR examinations of patients with conventional pacemakers, the SAR value of all MRI
sequences should be limited to < 2 W/kg in the whole body (upper threshold of normal
operating mode) and < 3.2 W/kg in the head.
Compliance with these SAR thresholds during a standard sequencing protocol without
relevant decrease of image quality is usually possible in routine clinical situations.
MRI systems made by all manufacturers include standard simple software options that,
by selecting the normal operating mode, ensure that the MRI parameters of the selected
sequence is automatically adapted such that the SAR threshold of 2 W/kg in the whole
body and 3.2 W/kg in the head is not exceeded.
Basically it is also possible to reduce the SAR value manually by modifying individual
sequence parameters in order to remain in the normal operating mode:
-
reduction of number of slices
-
reduction of turbo factor
-
replacement of spin echo or turbo spin echo sequences with gradient echo sequences
-
reduction of excitation angle during gradient echo sequences
-
replacement of steady-state free precession sequences with gradient echo sequences,
particularly in cardiac imaging.
A limitation of the active examination time (cumulative examination time with active
gradient and RF fields) to a maximum of 30 minutes is recommended in order to minimize
the risk of thermal endo-/myocardial damage due to RF-induced lead heating which is
dependent not only on the level but also on the duration of the heating.
Pacemaker-related safety precautions
The pacemaker system should be reprogrammed immediately prior to the MRI scan, and
then reprogrammed again to its original mode immediately after the MR examination.
The attending cardiologist determines and takes responsibility for the pacemaker mode
during MRI. Specific current reprogramming and monitoring guidelines, taking into
account safety as well as practical aspects are currently being developed in collaboration
with the German Cardiac Society. Until then, the following procedure is recommended;
the cardiologist should view it as a framework which can be modified on an individual
basis, depending on the specific electrophysiological situation of the patient.
-
In the case of a pacemaker-dependent patient: reprogram the pacemaker into an asynchronous
mode (D00, V00) with deactivation of the sensing function. This will prevent inhibition
of the pacemaker by pulsed magnetic fields while ensuring continuous cardiac stimulation
of the patient.
-
In the case of a non-pacemaker-dependent patient (representing approx. 80 % of pacemaker
patients): reprogram the pacemaker into a SENSE-only mode (e. g. 0D0) with deactivation
of the stimulation function or, if the PM system allows, complete deactivation of
the pacemaker (000) in order to prevent improperly triggered atrial and/or ventricular
stimulation or asynchronous stimulation after reed switch actuation or during interference
mode.
-
Increase the stimulation pulse (e. g. to 5.0 V/1.0 ms), in order to compensate for
a possible increase of the pacing capture threshold caused by heating of the pacemaker
leads.
-
Reprogram the sensing and stimulation polarity of the pacemaker leads to bipolar (if
technically possible).
-
Deactivate all supplemental stimulation functions (i. e. frequency-adapted stimulation,
anti-tachycardia stimulation).
A complete PM interrogation should be performed both immediately before as well as
after the MR examination as part of the reprogramming procedure, as well as 3 months
afterward. The 3 months follow-up is done to detect any delayed damage/malfunctions
induced during the MRI (e. g. chronic increased pacing capture thresholds resulting
from scar tissue formation).
MRI of patients with MR conditional pacemaker systems
MRI of patients with MR conditional pacemaker systems
MR conditional pacemakers have been tested for an MR examination under specific conditions,
and approved for use by the European Medical Device Directive with CE certification
(“in-label” use). The manufacturer guarantees its safety when properly used following
specific conditions for use. The Medtronic company has published prospective randomized
multi-center studies demonstrating the safety of the Enrhythm and Advisa models [31]
[32]
[33].
Technical modifications of MR conditional pacemaker systems include replacement of
the reed switch with a Hall sensor (unlike a standard reed switch, its behavior is
predictable in a strong static magnetic field), improved protection of the internal
circuits (thereby avoiding electromagnetic interference and voltage drop-induced electrical
resets); modification of input capacities and protective diodes in the pacemaker system
(to reduce voltage induction into the leads), as well as a software-based MRI protective
mode (see below).
Some manufacturers (Medtronic, St. Jude, Boston Scientific) have developed new leads
in which RF-induced heating is significantly reduced. In addition, all manufacturers
have tested their conventional leads and have identified some lead models, which under
certain conditions could be approved as MR conditional (back labeling). This implies
that a lead primarily labeled “conventional” and not “MR-safe” on the pacemaker ID
card during implantation could be retrospectively tested and approved as “MR conditional”.
This emphasizes that data regarding MR safety of the leads in the pacemaker ID card
may not be reliable, and should be currently reviewed with the manufacturer (manual,
hotline, web site, [Table 3]).
Table 3
Manufacturer-specific conditions for use of MR conditional pacemaker systems (as of
March 1, 2015).
|
medtronic
|
biotronik
|
Boston Scientific
|
St. Jude Medical
|
Sorin Group
|
field strength
|
1.5 T
|
1.5 T some pacemaker systems additionally 3 T
|
1.5 T some pacemaker systems additionally 3 T
|
1.5 T
|
1.5 T
|
maximum gradient slew rate
|
< 200 T/m/s
|
< 200 T/m/s Individual models only < 125 T/m/s
|
< 200 T/m/s
|
< 200 T/m/s
|
< 200 T/m/s
|
maximum whole-body SAR (specific absorption rate)
|
for all pacemaker systems < 2 W/kg (head SAR < 3.2 W/kg)
|
< 2 W/kg (head SAR 3.2 W/kg) some pacemaker systems < 4 W/kg (head SAR < 3.2 W/kg)
|
< 2 W/kg (head SAR < 3.2 W/kg) some pacemaker systems < 4 W/kg (head SAR < 3.2 W/kg)
|
both < 4 W/kg (head SAR < 3.2 W/kg) as well as some pacemaker systems < 2 W/kg (head
SAR < 3.2 W/kg)
|
for all pacemaker systems < 2 W/kg (head SAR < 3.2 W/kg)
|
whole-/partial-body approval
|
whole-body approval for all pacemaker systems
|
Pacemaker system-dependent whole-body approval or partial-body approval (isocenter
of FOV above the orbit or below the major trochanter)
|
whole-body approval for all pacemaker systems
|
Pacemaker system-dependent whole-body approval or partial-body approval (model-dependent,
upper isocenter limit above thoracic vertebra 1 or 10 cm above thoracic vertebra 1,
lower isocenter limit below lumbar vertebra 4)
|
partial-body approval for all pacemaker systems (isocenter of FOV above the orbit
or below the major trochanter)
|
maximum MRI examination time
|
no limitations
|
< 30 min per MRI examination < 10 hour maximum cumulative MRI examination time per
life time of pacemaker system
|
no limitations
|
Some pacemaker systems no limitation, some systems < 30 min per MRI examination
|
< 40 min per MRI examination
|
maximum body size
|
no limitations
|
1.40 m
|
no limitations
|
no limitations
|
1.47 m
|
body temperature
|
no limitations
|
< 38.0 °C
|
< 38.0 °C
|
pacemaker systems with whole-body approval: no limitation partial-body approval: < 38.0 °C
|
< 38.0 °C
|
positional requirements
|
supine and prone position
|
only supine position
|
supine and prone position
|
supine and prone position
|
supine and prone position
|
additional cardiac leads or lead components [1]
|
no additional leads or lead components
|
additional coronary sinus leads approved for some CRT systems, otherwise no additional
leads or lead components
|
no additional leads or lead components
|
no additional leads or lead components
|
no additional cardiac leads or lead components
|
other cardiac and extracardiac implants
|
other implants approved as “MR conditional” are permitted within their conditions
for use
|
-
other implants approved as “MR conditional” are permitted within their conditions
for use
-
implants directly adjacent to leads (< 4 cm) may not be longer than 5 cm
|
warning notice in manuals:
“Other implants can impair the safety of MR conditional pacemaker systems”
|
no clear data in the manuals,
|
According to manual “Implanted non MR-safe medical products are contraindicated” authors’
comment: therefore all metallic implants which by definition cannot be classified
as MR-safe are de facto contraindicated
|
contact address
|
Medtronic GmbH
Earl-Bakken-Platz 1 40 670 Meerbusch, Germany
+ 49 2159 -8149- 0
|
BIOTRONIK SE & Co. KG
Woermannkehre 1 12 359 Berlin, Germany
+ 49 30 68 905-0
|
Boston Scientific
Daniel-Goldbach-Strasse 17 – 27
40 880 Ratingen, Germany
+ 49 2102 489 750
|
St. Jude Medical GmbH
Helfmannpark 1, D65 760 Eschborn, Germany
+ 49 6196 77 110
+ 49 6196 7711 177
|
Sorin Group Deutschland GmbH
Lindberghstr. 25
, 80 939 Munich, Germany
+ 49 893 230 10
|
hotline
|
24-hour support
+ 49 159 8149 112
|
24-hour support + 41 415 603 660
|
Germany: 02 102 – 489 770 (24 h)
European hotline: + 32 2 416 7222
24-hour support (800) 505 – 4636
|
24-hour support + 46 – 8-474 – 4147 (Sweden)
08:00 to 18:00 German-language support + 49 (0)619 677 110
|
support during business hours 0172/8222 222
|
internet links / guidance
|
http://www.mrisurescan.com/europe/index.htm
|
http://www.biotronik.com/wps/wcm/connect/en_promri/biotronik/home/promri_system_check/#jump
https://www.biotronik.com/wps/wcm/connect/int_web/emanuals/emanuals/
|
http://www.bostonscientific.com/content/dam/Manuals/eu/current-rev-de/359 259 – 002_Ingenio2_MRI_TG_DE_S.pdf
|
http://www.sjm.de/mrt/Accent-MRI-System
|
https://www.sorinmanuals.com/PDFUSERS/ADEU201A.pdf
|
1 Abandoned leads, epicardial leads, coronary sinus leads, lead adapters or extensions.
The recommended procedural management of patients with MR conditional pacemakers is
shown in Algorithm 2 ([Fig. 2]). The individual steps are explained below.
Fig. 2 Algorithm 2.
It should be noted that all information is only a snapshot of the current situation.
Changes to the conditions for use are possible, and it is necessary to review each
case individually and in a timely manner. Therefore, the telephone hotline numbers
as well as the related Internet addresses of the individual manufacturers are provided
in [Table 3].
Indications and review of the conditions for use
In the cardiac/pacemaker outpatient clinic PM interrogation, and if necessary a review of the medical record of the center
performing the implantation, is performed to determine whether the pacemaker-specific
conditions of use are fulfilled for MR imaging.
In cases of doubt an X-ray of the chest may be performed, especially to exclude points
7 and 8 (see below).
-
Verification of a complete and approved MR conditional pacemaker system consisting
of a pacemaker generator and pacemaker leads
-
Left or right pectoral implantation site of the pacemaker system
-
Implantation time > 6 weeks
-
Electrically intact pacemaker leads
-
Pacing capture thresholds within the normal range
-
Sufficient battery capacity based on the manufacturer’s specification
-
No additional cardiac leads (particularly no abandoned pacemaker leads), no additional
components such as lead adapters or extensions
-
Exclusion of other cardiac implants, depending on the manufacturer’s specification
(e. g. coronary stents, prosthetic heart valves),[1] [Table 3]
-
Written documentation (check list) by the attending cardiologist indicating that the electrophysiological
conditions for use of the pacemaker have been fulfilled.
Radiology performs verification of the MR-related conditions for use:
-
Design and field strength of the MRI system: The MR conditional pacemaker systems of all manufacturers are approved for an MRI
scan in closed/cylindrical MRI systems with a field strength of 1.5 T. For some of
their pacemaker systems, Biotronik and Boston Scientific also have additional approval
for MRI systems with a field strength of 3 T. Currently no MR conditional pacemaker
system has approval for MRI examinations with a field strength 1 T or lower nor is
there approval for open MRI systems. Based on manufacturers’ information, there are
no plans for such in the future.
-
Gradient system: Most MR conditional pacemaker systems permit MRI sequences up to a maximum gradient
field slew rate of 200 T/m/s per axis. Clinical MRI gradient systems currently in
use do not achieve or exceed this value. An important exception are some pacemaker
systems made by Biotronik, for which the maximum slew rate of the gradient fields
is limited to only 125 T/m/s per axis.
Unlike the SAR value, the slew rate of the gradient fields of specific MRI sequences
cannot be quantified and controlled or is difficult to quantify and control in clinical
applications. Therefore, for the examination of a patient with a pacemaker, it is
recommended to identify an MRI scanner with known technical specifications and maximum
output of the gradient system lying below the approved limit range of the relevant
pacemaker system. It should be mentioned that some manufacturers’ brochures occasionally
indicate higher slew rates of the gradient fields up to 350 T/m/s which then correspond
to cumulative slew rates of the x, y, z gradients. However, the safety-related reference
parameter is the slew rate of the gradient fields per axis.
-
Whole-body vs. partial-body approval: The manufacturers Medtronic and Boston Scientific have whole-body approval for all
their MR conditional pacemaker systems, i. e. all regions of the body can in principle
be examined using MRI. Biotronik and St. Jude, in addition to pacemaker systems with
whole-body approval, also have systems with only partial-body approval. All Sorin
Group pacemaker systems currently have only partial-body approval. For systems with
only partial-body approval MRI examinations are limited to scans outside of the chest
to reduce RF-induced heating of the pacemaker leads. For pacemaker systems with partial-body
approval made by Biotronik and Sorin Group, MRI scans are specifically limited to
those body regions in which the isocenter of the scan field (field of view) is above
the orbital cavities or below the greater trochanter. Pacemaker systems made by St.
Jude with partial-body approval also include model series with MRI scans limited to
examinations of body regions with an isocenter of the scan field above cervical vertebra
1 and below lumbar vertebra 4, as well as model series more stringently limited to
an upper isocenter position 10 cm above cervical vertebra 1 with an identical lower
isocenter position below lumbar vertebra 4.
-
Exclusion of other extracardiac implants, depending on the manufacturer’s specification (e. g. spondylodesis material in the
thoracic spine), 1 [Table 3]
-
Written documentation (check list) by the attending radiologist indicating that the MR-related conditions
for use of the pacemaker have been met.
If the manufacturer-specific electrophysiological and MRI-specific conditions for
use are not met or compliance with these conditions for use cannot be guaranteed the
PM device must be treated like a conventional pacemaker.
Patient information
Unlike patients with conventional pacemaker systems, performance of MRI scans on patients
with MR conditional pacemaker systems are an approved application (“in-label” use).
In principle, however, patients must be informed of the same risks as those with conventional
systems (see Algorithm 2). With respect to MR conditional systems, however, these
risks have been widely minimized, so that the remaining theoretical risk according
to the criteria of the approval authorities is considered reasonably low, taking into
account the anticipated benefit of the MR examination. In individual cases, these
risks cannot be ruled out with absolute certainty.
Monitoring during the MR examination
In principle, the presence of a cardiologist during the MRI scan of a patient with
an MR conditional pacemaker system is not required as long as complications which
may occur independent of the presence of a pacemaker can be detected and treated by
attending medical staff until the arrival of an emergency team. The situation is comparable
to the management of contrast agent-induced events during radiological examinations.
All manufacturers of MR conditional pacemakers require patient monitoring using the
following methods during an MRI scan: pulse oximetry, ECG or blood pressure measurement.
The user may select the monitoring modality. The authors of this position paper strongly
recommend monitoring using an MRI-compatible pulse oximeter. The validity of ECG registrations
(even with newer MRI-compatible ECG systems) is still frequently impaired by overlaying
artifacts. Blood pressure measurement does not provide continuous monitoring of the
patient. Pulse oximetry alone is considered sufficient for monitoring patients with
MR conditional pacemaker systems.
MRI-related conditions for use / safety precautions
SAR value: To reduce RF-induced lead heating, the whole-body SAR value is limited for most MR
conditional systems to the normal operating mode at an upper threshold of 2 W/kg (local
head SAR value 3.2 W/kg); see also the comment “SAR value” in the section “Conventional
pacemakers”. Some pacemaker systems made by Biotronik, Boston Scientific and St. Jude
are also approved for MR examinations with a maximum SAR value of 4 W/kg (local head
SAR value 3.2 W/kg).
MRI coils: All currently commercially available receiver coils, including local surface receiver
coils, can be used in all body regions for all pacemaker manufacturers. The integrated
body-RF coil is approved for all pacemaker manufacturers as transmit coil. Some pacemaker
manufacturers (Medtronic, Biotronik and Boston Scientific) also allow the use of combined
transmit/receive coils outside the thoracic region which in relatively rare cases
are available for some MRI systems as knee or head coils, for example.
Positioning: With the exception of the Biotronik company, MR conditional pacemaker systems of
all manufacturers are approved for MR examinations in supine and prone position.
All Biotronik models are tested and approved only for MRI scans in supine position;
consequently an MR examination of the breasts, which is performed in prone position,
for example, cannot be accomplished as “in label” use on a patient with an MR conditional
Biotronik pacemaker system as the specific MR-related conditions for use are not met.
[Table 3] contains a summary of additional manufacturer specifications for MR conditional
pacemaker systems, such as MRI examination time, cumulative examination time over
the service life time of the pacemaker pulse generator, minimum patient body size
as well as exclusion of fever or impaired patient body temperature regulation.
Pacemaker-related conditions for use/safety precautions
Analogously to the procedure with conventional pacemakers, prior to an examination,
MR conditional systems should be programmed into an MRI-protection mode which implies:
asynchronous mode or deactivation of the pacing function, bipolar lead configuration,
increased stimulation output, deactivation of supplementary stimulation functions
such as frequency-adapted stimulation or anti-tachycardia overstimulation. This MRI-protection
mode is stored in the software options of the pacemaker model. This implies that before
and after an MR examination, patients must always be seen by a cardiologist. Although
the manufacturers do not pose explicit requirements, the authors of this paper suggest
that reprogramming be performed as close in place and time to the MR examination as
possible, ideally in the same or adjacent building, and within a time window of < 2 h.
This procedure, in the estimation of the attending cardiologist, can be modified in
individual cases. This applies, for example, to patients for whom no pacemaker activity
has been documented for an extended period.
Unlike conventional pacemaker systems, follow-up after 3 months to rule out long-term
damage is optional and not obligatory.
Summary: MRI and cardiac pacemakers
Summary: MRI and cardiac pacemakers
-
Conventional pacemaker systems are no longer an absolute but rather a relative contraindication
for performing an MR examination. The procedural management includes the assessment
of the individual risk/benefit ratio, comprehensive patient informed consent about
specific related risks and “off label” use, extensive PM- and MRI-related safety precautions
as well as adequate monitoring techniques during the MRI exam
-
MR conditional pacemakers have been tested and approved under specific conditions
(“in-label” use). Decisive for patient safety are precise understanding of, and compliance
with, the terms of use for the specific pacemaker system. If the electrophysiological
and MRI-specific conditions for use are not met or compliance with these conditions
for use cannot be guaranteed, the device must be treated like a conventional pacemaker.