Successful removal from the esophagus of a self-expandable metal stent that had shriveled up into a tangled ball

Gianfranco Donatelli
1   Therapeutic Endoscopy Unit, Hôpital Privé des Peupliers, Paris, France
,
Parag Dhumane
2   Department of General and Laparoscopic Surgery, Lilavati Hospital and Research Center, Mumbai, India
,
Bertrand M. Vergeau
1   Therapeutic Endoscopy Unit, Hôpital Privé des Peupliers, Paris, France
,
Jean-Loup Dumont
1   Therapeutic Endoscopy Unit, Hôpital Privé des Peupliers, Paris, France
,
Thierry Tuszynski
1   Therapeutic Endoscopy Unit, Hôpital Privé des Peupliers, Paris, France
,
Bruno Meduri
1   Therapeutic Endoscopy Unit, Hôpital Privé des Peupliers, Paris, France
› Author Affiliations
Further Information

Corresponding author

Gianfranco Donatelli, MD
Hôpital Privé des Peupliers – Unité d’Endoscopie Thérapeutique
8 Place de l’Abbé G. Hénocque
Paris 75013
France   
Fax: +33-1-44165212    

Publication History

Publication Date:
27 November 2013 (online)

 

A 39-year-old woman developed a leak from the cardia following a sleeve gastrectomy. She underwent placement of an 18-cm Niti-S covered esophageal self-expandable metal stent (SEMS; TaeWoong Medical, Gimpo, Korea) at the gastroesophageal junction ([Fig. 1 a]).

Zoom Image
Fig. 1 Radiographic views showing: a the 18-cm covered self-expandable metal stent (SEMS) in position at the esophagogastric junction; b a foreign-body forceps being used to grasp the distal lasso stitch of the SEMS during its removal; c the SEMS shriveled up into a tangle of metal wire in the mid-esophagus.

The patient returned to undergo stent removal 4 weeks after its placement. During this procedure the lasso stitch at the distal end was grasped with endoscopic foreign-body forceps and was pulled ([Fig. 1 b]). The stent began to invaginate from within but soon shrank into a tangled ball of metal wires ([Fig. 1 c] and [Fig. 2]). The distal lasso stitch slipped out of the forceps, leaving the now compacted SEMS lying in the mid-esophagus and completely obstructing the lumen.

Zoom Image
Fig. 2 Endoscopic view of the self-expandable metal stent (SEMS) contracted into a knot of metal wire.

The SEMS was gradually eased away from the esophageal mucosa using the proximal lasso stitch. The disjointed mesh wires were slowly pulled off the mucosa, and thereby the whole of the stent was successfully removed under fluoroscopic guidance ([Fig. 3]). Endoscopy confirmed the absence of any esophageal injury, but persistence of the original leak was noted. An over-the-scope clip (OTSC) was used to obliterate the leak ([Fig. 4]).

Zoom Image
Fig. 3 The broken self-expandable metal stent (SEMS) after its removal.
Zoom Image
Fig. 4 Radiographic view showing an over-the-scope clip (OTSC) in position to treat the leak from the cardia after the self-expandable metal stent (SEMS) had been removed.

Fracture of an esophageal SEMS has been previously described, during positioning because of defective material and as a consequence of laser therapy that caused a late spontaneous fracture during palliative treatment [1]. It appears that corrosion is the principal cause of fracture because, once in place and fully expanded, SEMS are not subject to any great mechanical stress [2]. Cases of fracture during removal of a covered biliary SEMS have also been described [3].

Here we report the first case, to our knowledge, of a covered SEMS compacting into a ball at the time of removal and its successful retrieval. We believe that this happened probably because the SEMS was partially fractured and became deeply embedded in the esophageal mucosa; when traction was applied, the now-damaged mesh contracted in on itself because of the inherent forces in its structure. The risk of a SEMS curling up should be considered as a potential complication of SEMS removal, in addition to other well-documented complications [4].

Endoscopy_UCTN_Code_CPL_1AJ_2AF


#

Competing interests: None

  • References

  • 1 Reddy AV, Alwair H, Trewby PN. Fractured esophageal nitinol stent: Report of two fractures in the same patient. Gastrointest Endosc 2003; 57: 138-139
  • 2 Rana SS, Bhasin DK, Sidhu GS et al. Esophageal nitinol stent dysfunction because of fracture and collapse. Endoscopy 2009; 41 (Suppl. 02) E1701-E1702
  • 3 Joo KR, Paik CN, Chung WC et al. Fracture of self-expandable metal stent during endoscopic removal in benign biliary stricture. Clin Endosc 2013; 46: 95-97
  • 4 Donatelli G, Dhumane P, Perretta S et al. Endoscopic placement of fully covered self expanding metal stents for management of post-operative foregut leaks. J Minim Access Surg 2012; 8: 118-124

Corresponding author

Gianfranco Donatelli, MD
Hôpital Privé des Peupliers – Unité d’Endoscopie Thérapeutique
8 Place de l’Abbé G. Hénocque
Paris 75013
France   
Fax: +33-1-44165212    

  • References

  • 1 Reddy AV, Alwair H, Trewby PN. Fractured esophageal nitinol stent: Report of two fractures in the same patient. Gastrointest Endosc 2003; 57: 138-139
  • 2 Rana SS, Bhasin DK, Sidhu GS et al. Esophageal nitinol stent dysfunction because of fracture and collapse. Endoscopy 2009; 41 (Suppl. 02) E1701-E1702
  • 3 Joo KR, Paik CN, Chung WC et al. Fracture of self-expandable metal stent during endoscopic removal in benign biliary stricture. Clin Endosc 2013; 46: 95-97
  • 4 Donatelli G, Dhumane P, Perretta S et al. Endoscopic placement of fully covered self expanding metal stents for management of post-operative foregut leaks. J Minim Access Surg 2012; 8: 118-124

Zoom Image
Fig. 1 Radiographic views showing: a the 18-cm covered self-expandable metal stent (SEMS) in position at the esophagogastric junction; b a foreign-body forceps being used to grasp the distal lasso stitch of the SEMS during its removal; c the SEMS shriveled up into a tangle of metal wire in the mid-esophagus.
Zoom Image
Fig. 2 Endoscopic view of the self-expandable metal stent (SEMS) contracted into a knot of metal wire.
Zoom Image
Fig. 3 The broken self-expandable metal stent (SEMS) after its removal.
Zoom Image
Fig. 4 Radiographic view showing an over-the-scope clip (OTSC) in position to treat the leak from the cardia after the self-expandable metal stent (SEMS) had been removed.