DOI: 10.1055/a-2131-7180
Global prospective case series of ERCPs using a single-use duodenoscope
Trial Registration: ClinicalTrials.gov Registration number (trial ID): NCT04103749 Type of study: Multi-Center Study
Abstract
Background The first commercialized single-use duodenoscope was cleared by the US Food and Drug Administration in December 2019. Data regarding endoscopic retrograde cholangiopancreatography (ERCP) using a single-use duodenoscope are needed on a broader range of cases conducted by endoscopists with varying levels of experience in a wide range of geographic areas.
Methods 61 endoscopists at 22 academic centers in 11 countries performed ERCP procedures in adult patients aged ≥ 18. Outcomes included ERCP completion for the intended indication, rate of crossover to a reusable endoscope, device performance ratings, and serious adverse events (SAEs).
Results Among 551 patients, 236 (42.8 %) were aged > 65, 281 (51.0 %) were men, and 256 (46.5 %) had their procedure as an inpatient. ERCPs included 196 (35.6 %) with American Society for Gastrointestinal Endoscopy complexity of grades 3–4. A total of 529 ERCPs (96.0 %) were completed: 503 (91.3 %) using only the single-use duodenoscope, and 26 (4.7 %) with crossover to a reusable endoscope. There were 22 ERCPs (4.0 %) that were not completed, of which 11 (2.0 %) included a crossover and 11 (2.0 %) were aborted cases (no crossover). Median ERCP completion time was 24.0 minutes. Median overall satisfaction with the single-use duodenoscope was 8.0 (scale of 1 to 10 [best]). SAEs were reported in 43 patients (7.8 %), including 17 (3.1 %) who developed post-ERCP pancreatitis.
Conclusions In academic medical centers over a wide geographic distribution, endoscopists with varying levels of experience using the first marketed single-use duodenoscope had good ERCP procedural success and reported high performance ratings for this device.
#
Introduction
Post-endoscopic retrograde cholangiopancreatography (ERCP) infections are estimated to occur in approximately 5 % of patients [1] [2], commonly due to enteric organisms [3]. Of greatest concern are infections from multidrug-resistant organisms (MDRO), which account for a disproportionately higher morbidity and mortality compared with infections with antibiotic-susceptible organisms [4]. These can be endogenous infections due to translocation of organisms from the gastrointestinal (GI) tract of the patient into the pancreaticobiliary ducts by the duodenoscope or endoscopic devices, or can be exogenous, due to MDROs that were colonizing the ducts of another patient being harbored inside a reusable duodenoscope. A 2022 systematic review estimated the minimum risk of duodenoscope-associated MDRO infection to be at least 0.01 % per ERCP procedure in the Netherlands, which was substantially higher than reported rates in frequently cited older data [5]. The high morbidity and mortality associated with MDRO infections [6] and the delayed clinical detection of some of these infections [7] warrant improved reprocessing practices and the development of a uniform and practical protocol for duodenoscope microbiological sampling that can be applied in general healthcare facilities [8].
An alternative strategy to address exogenous MDRO infections associated with duodenoscopes is the use of single-use duodenoscopes. Since receiving US Food and Drug Administration (FDA) clearance in 2019, the first single-use duodenoscope has been studied in both US and European settings [9] [10] [11] and by endoscopists with varying levels of experience [12]. A randomized trial [10] and three case series [9] [11] [12] reported comparable safety profiles, technical performance, and endoscopist user satisfaction ratings for the first marketed single-use duodenoscope compared with reusable duodenoscopes. A second brand of single-use duodenoscope was cleared in 2020 [13]. Additional data are warranted, particularly to document the performance of the single-use duodenoscopes when used by less expert endoscopists, in cases with high complexity, and across a broad geographic range.
To characterize the efficacy and safety of the first marketed single-use duodenoscope further, we conducted a large clinical case series of ERCPs of any complexity in academic medical centers.
#
Methods
Study design
This was a large, multinational case series of the performance of a single-use duodenoscope for ERCP. Institutional Review Board and Ethics Committee approvals for the study were obtained at all study sites. All enrolled patients provided written informed consent for study participation before they contributed data. Single-use duodenoscopes were provided without charge to the investigational sites for use in the study procedures.
Boston Scientific Corporation (Marlborough, Massachusetts, USA) sponsored and funded the study. Sixteen endoscopists participated in published research during development of the single-use duodenoscope [9] [12] [14]. A statistician (M.J.R.) who is a full-time employee of Boston Scientific Corporation performed the data management and statistical analysis of this study. Review and input to the analysis of the study data were provided by two Boston Scientific employees (J.A.P. and M.J.R.) and by all participating physicians at Investigator Meetings between 2019 and 2022.
#
Single-use duodenoscope
The device used in this study was the EXALT Model D single-use duodenoscope (Boston Scientific Corporation), a sterile, single-use duodenoscope designed to function similarly to currently marketed reusable duodenoscopes, and to be discarded after use in a single procedure. The single-use duodenoscope received US FDA clearance in December 2019 and gained a CE mark in January 2020 [15] [16].
#
Patient population
The study protocol allowed for recruitment of adult patients scheduled for ERCP per the standard of care at up to 40 participating healthcare centers. Eligible patients were recruited for the study on selected weekdays when participating endoscopists were performing ERCPs. Patients were eligible for inclusion if they were aged ≥ 18 years, willing and able to comply with the study procedures and to provide written informed consent to participate in the study, and were scheduled for a clinically indicated ERCP or other duodenoscope-based procedure. Excluded from the study were potentially vulnerable patients including, but not limited to: pregnant women, patients for whom endoscopic techniques were contraindicated, patients enrolled in another investigational study that would directly interfere with the current study, and patients excluded at investigator discretion. Unlike previous studies of the single-use duodenoscope, the current study included patients with “altered pancreaticobiliary anatomy” because this was not a contraindication in the Directions for Use approved in 2020 [17].
#
Study procedures
Patient assessments
All enrolled participants had a preprocedural study visit for assessment of their demographics and relevant medical history. After the index procedure, participants were evaluated in person or by telephone at 72 hours (−1 day to + 2 days) and again at 30 days (± 3 days) to screen for post-procedure adverse events (AEs) or resolution of any previously reported issues. If the last follow-up visit was not completed, the reason was noted on the study completion form. Participants were considered lost to follow-up if they failed to return for their scheduled follow-up visits and were unable to be contacted by the study site staff after at least three documented attempts.
#
ERCP procedures
Participating endoscopists agreed to use the single-use duodenoscope in place of other brands of duodenoscope(s) used in the endoscopy unit for ERCP. Endoscopists’ level of experience was categorized as “expert” (> 2000 lifetime ERCPs) and “less expert” (≤ 2000 lifetime ERCPs). Start and stop times of the procedure, post-procedure subjective feedback on performance-related attributes, American Society for Gastrointestinal Endoscopy (ASGE) grade for complexity of the ERCP procedure, and all attempted or completed maneuvers during the ERCP were documented. Device deficiencies were recorded and reported, regardless of whether they led to an AE or inability to complete the ERCP. If the necessary maneuvers could not be completed with the single-use device and the endoscopist crossed over to use a reusable duodenoscope, the reason for noncompletion with the single-use device was recorded, and the ability to complete the ERCP maneuvers with a reusable duodenoscope was documented.
Difficult common bile duct (CBD) cannulation was defined by a modification of the European Society of Gastrointestinal Endoscopy (ESGE) definition [18] as the presence of one or more of the following: more than five contacts with the papilla while attempting to cannulate; more than 5 minutes spent attempting to cannulate following visualization of the papilla; at least one (instead of “more than one”) unintended pancreatic duct cannulation or opacification [18]. To represent all difficult procedures, difficult CBD cannulation was tabulated with failed ERCPs (even if cannulation was achieved).
#
#
Outcomes
The primary end point was the ability to complete the ERCP procedures for the intended indication(s). The secondary end points were: (i) the incidence of crossover from the single-use duodenoscope to a reusable duodenoscope; (ii) comparison of outcomes by ASGE grade, endoscopist level of experience, or prior sphincterotomy; (iii) endoscopist ratings of overall satisfaction with the single-use duodenoscope on a scale of 1 (worst) to 10 (best); (iv) serious AEs (SAEs) related to the device and/or procedure, assessed through to 30 days after the ERCP or other duodenoscope-based procedure.
#
Statistical analysis
Descriptive statistics included: frequency statistics for patient and procedural characteristics and procedure completion rates (completion rates reported separately for cases with and without crossover to a reusable duodenoscope); median and range for overall satisfaction and procedure duration; and mean (SD) and range for age and mean number of cannulation attempts. Endoscopist experience and ASGE complexity grades were assessed for completion time and performance ratings using Wilcoxon rank-sum tests, and for number of cannulation attempts using a negative binomial model. Fisher’s exact test was used to compare the occurrence of crossover to a reusable duodenoscope by level of experience or by ASGE case complexity level. For binary data, 95 %CIs were calculated using the Clopper–Pearson exact methods. Statistical analyses were performed using SAS 9.4 software (SAS Institute Inc., Cary, North Carolina, USA).
#
#
Results
Enrollment and patient characteristics
Of 809 patients who were screened, 551 (68.1 %) were enrolled after providing written informed consent to participate in the study. Reasons for nonparticipation are summarized in [Fig. 1].


Of the 551 patients (mean age 59.9 years) who had a procedure, 281 (51.0 %) were male and 332 (60.3 %) had a prior ERCP ([Table 1]), with 243 patients (44.1 %) having had a prior biliary and/or pancreatic sphincterotomy. The most common prior GI disorders in patients scheduled for ERCP or other duodenoscope-based procedure were: current bile duct stones/gallstones (n = 168; 30.5 %), biliary or pancreatic duct stricture (n = 124; 22.5 %), current immunosuppression for any cause (n = 71; 12.9 %), chronic pancreatitis (n = 68; 12.3 %), pancreatic tumor (n = 63; 11.4 %), abnormal imaging finding (n = 60; 10.9 %), current cholangitis (n = 53; 9.6 %), or cholangiocarcinoma (n = 46; 8.3 %). Two patients (0.4 %) had a known history of MDRO colonization.
ERCP, endoscopic retrograde cholangiopancreatography; MDRO, multidrug-resistant organism.
Of the 332 patients for whom the referral source was documented, 102 (30.7 %) presented directly to the participating study site, 154 (46.4 %) were referred from an academic medical center, and 76 (22.9 %) from a community medical center.
#
General ERCP characteristics
A total of 61 endoscopists (46 expert, 15 less expert) at 22 academic centers in 11 countries (1–6 endoscopists per center) performed 551 procedures: 25 cases at 21 of the study sites and 26 cases in the other. The median number of procedures performed by each endoscopist was seven (range 1–25). All procedures were performed in endoscopy suites, almost all (99.6 %; 549/551) during normal working hours, and most with endoscopy processing staff present (79.1 %; 436 /551) ([Table 2]). Prophylactic antibiotics were used in 272 (49.4 %) patients, and prophylactic nonsteroidal anti-inflammatory drugs were used in 315 (57.2 %). The ERCPs included all ASGE grades of complexity: grade 1 (least complex; 10.2 %; 56/551), grade 2 (51.5 %; 284/551), grade 3 (28.5 %; 157/551), and grade 4 (most complex; 7.1 %, 39/551). There were also 15 cases (2.9 %) that were duodenoscope-based procedures not included in the ASGE grading system. The median (range) procedure completion time was 24.0 (2.0–157.0) minutes, including times of ≤ 20 minutes in 240/534 cases (44.9 %) to ≥ 60 minutes in 39/534 (7.5 %) cases.
Characteristic |
n (%), unless otherwise specified |
Inpatient procedure |
256 (46.5) |
ERCP procedure |
545 (98.9) |
Prophylactic NSAID used |
315 (57.2) |
Prophylactic antibiotics used |
272 (49.4) |
Patient intubated |
253 (45.9) |
Medications for anesthesia |
|
|
477 (86.6) |
|
178 (32.3) |
|
137 (24.9) |
|
74 (13.4) |
|
13 (2.4) |
|
163 (29.6) |
Medications for maintenance of anesthesia |
|
|
391 (71.0) |
|
153 (27.8) |
|
124 (22.5) |
|
94 (17.1) |
|
54 (9.8) |
|
74 (13.4) |
Procedure time with single-use duodenoscope, median (range) [n], minutes[1] |
22.0 (0.0−157.0) [433] |
Reusable scope used |
37 (6.7) |
Procedure time with reusable duodenoscope, median (range) [n], minutes |
22.0 (6.0−75.0) [37] |
Patient position |
|
|
379 (68.8) |
|
133 (24.1) |
|
52 (9.4) |
Location of procedure |
|
|
551 (100.0) |
ERCP during normal hours |
549 (99.6) |
Endoscope reprocessing staff present |
436 (79.1) |
Total procedure time, median (range) [n], minutes[1] |
24.0 (2.0−157.0) [548] |
Difficult cannulation or failed ERCP[2] |
150 (29.2) |
|
140 (27.2) |
|
4(0.8) (514) |
|
6(1.2) (514) |
All intended maneuvers completed |
524 (95.1) |
ERCP completion |
|
|
503 (91.3) |
|
26 (4.7) |
|
11 (2.0) |
|
11 (2.0) |
Pancreatic maneuvers performed |
128 (23.2) |
Cholangioscopy |
24 (4.4) |
Pancreatoscopy |
9 (1.6) |
ASGE complexity grade[3] |
|
|
56 (10.2) |
|
284 (51.5) |
|
157 (28.5) |
|
39 (7.1) |
|
15 (2.7) |
ERCP, endoscopic retrograde cholangiopancreatography; NSAID, nonsteroidal anti-inflammatory drug; ASGE, American Society for Gastrointestinal Endoscopy.
1 One case was aborted (because patient had food in the stomach) so the procedure time was 0 and was not included in the “total procedure time” calculation.
2 Denominator = 514 procedures.
3 ASGE grade for complexity of ERCPs: grade 1, deep cannulation of duct of interest, main papilla, sampling; grade 2, biliary stone extraction < 10 mm, treat biliary leaks, treat extrahepatic benign and malignant strictures, placed prophylactic stents; grade 3, biliary stone extraction ≥ 10 mm, minor papilla cannulation in divisum and therapy, removal of internally migrated biliary stents, intraductal imaging/biopsy/fine needle aspiration, management of acute or recurrent pancreatitis, treat pancreatic strictures, remove pancreatic stones mobile and < 5 mm, treat hilar tumors, treat benign biliary strictures hilum and above, manage suspected sphincter of Oddi dysfunction (with or without manometry); grade 4, remove internally migrated pancreatic stents, intraductal image-guided therapy (e. g. photodynamic therapy, electrohydraulic lithotripsy), pancreatic stones impacted and/or ≥ 5 mm, intrahepatic stones, pseudocyst drainage/necrosectomy, ampullectomy, ERCP after Whipple or Roux-en-Y bariatric surgery.
Investigators reported that after the study procedures, most of the single-use duodenoscopes were discarded in standard medical waste (35.4 %; 195 /551), or medical grade recycling (29.9 %; 165/551), regulated medical waste (24.9 %; 137/551), a sharps disposal container (4.9 %; 27/551), or other (4.9 %; 27/551). A high proportion of patients completed the 72-hour (92.9 %; 512/551) and 30-day (89.8 %; 495/551) study follow-up.
#
Cannulation details
Difficult CBD cannulation was reported for 150 /514 cases (29.2 %) using the single-use duodenoscope (10 of which were failed ERCPs), and in 15 cases in which a reusable duodenoscope was used after crossover.
Advanced cannulation techniques performed included: precut (access) papillotomy (n = 42; 7.6 %), double wire (n = 31; 5.6 %), pancreatic duct plastic stent (n = 24; 4.4 %), rendezvous with EUS (n = 3; one also had precut but ERCP still failed and went to rendezvous), cholangioscopy (n = 2), papillectomy (n = 1), transpancreatic sphincterotomy (n = 1), clip used to bring hidden papilla out from fold (n = 1), and accessory papilla cannulation (n = 1).
#
Ability to complete ERCP for intended indication
Among the 551 study cases, 529 (96.0 %, 95 %CI 94.0 %–97.5 %) were completed for the intended indication, including 503 (91.3 %, 95 %CI 88.6 %–93.5 %) using only the single-use duodenoscope, and 26 (4.7 %, 95 %CI 3.1 %–6.8 %) including crossover from the single-use to a reusable duodenoscope. Of the 22 cases that were not completed, 11 included crossover to a reusable duodenoscope ([Table 2]).
A total of 514 patients (93.3 %) underwent an ERCP procedure that used the single-use duodenoscope only ([Fig. 1]). Of these, 324 were ASGE grade 1 or 2 cases (98.8 % completed), 181 were ASGE grade 3 or 4 cases (97.8 % completed), and nine did not have an ASGE grade (66.7 % completed). The remaining 37 patients (6.7 %) underwent an ERCP procedure with crossover from single-use to reusable duodenoscope. Of these, 16 were ASGE grade 1 or 2 cases (68.8 % completed), 15 were ASGE grade 3 or 4 cases (73.3 % completed), and six did not have an ASGE grade (66.7 % completed).
For 524 patients (95.1 %), all of the intended maneuvers were completed. Over 20 different types of maneuvers were performed using the single-use duodenoscope, including sphincterotomy, papillectomy/ampullectomy, cannulation, mechanical lithotripsy, clearance of bile duct or pancreatic duct stones, biliary or pancreatic stent placement or removal, balloon dilation, cholangioscopy (n = 24), pancreatoscopy (n = 9), cytology brushing, biopsy, and others. Pancreatic maneuvers were performed in 128 patients (23.2 %).
#
Comparison of outcomes by ASGE grade of complexity
The median time of completion was significantly shorter (20.0 vs. 35.0 minutes; P < 0.001), and the mean number of cannulation attempts was significantly higher (3.0 vs. 2.4; P = 0.01) for ASGE grade 1 or 2 procedures compared with grade 3 or 4 procedures, while ERCP completion (97.4 % [331/340] vs. 95.9 % [188/196]; P = 0.44) and crossover rate (4.7 % [16/340] vs. 7.7 % [15/196]; P = 0.18) were similar ([Table 3]).
Outcome |
Grade 1 or 2 |
Grade 3 or 4 |
P value |
ERCP completed, n (%) |
331 (97.4) |
188 (95.9) |
0.44 |
Crossover rate, n (%)[2] |
16 (4.7) |
15 (7.7) |
0.18 |
Overall satisfaction, median (range) [n] |
8.0 (2.0−10.0) [340] |
8.0 (1.0−10.0) [195] |
0.68 |
Serious adverse event rate, n (%) |
47 (13.8) |
36 (18.4) |
0.17 |
Procedural completion time, median (range) [n], minutes |
20.0 (2.0−132.0) [338] |
35.0 (2.0−157.0) [196] |
< 0.001 |
Number of cannulation attempts, mean (SD) |
3.0 (4.3) |
2.4 (2.8) |
0.01 |
ASGE, American Society for Gastrointestinal Endoscopy; ERCP, endoscopic retrograde cholangiopancreatography.
1 15 cases (2.9 %) were duodenoscope-based procedures not included in the ASGE grading system.
2 Excludes one crossover case that did not have an ASGE grade.
The median rating of overall satisfaction with the performance of the single-use duodenoscope was 8.0 for cases with both ASGE grades 1 or 2 and ASGE grades 3 or 4 (P = 0.68).
#
Comparison of outcomes for expert versus less expert endoscopists
ERCPs performed by expert and less expert endoscopists were similar with respect to completion rate (96.3 % [444/461] vs. 94.4 % [85/90], respectively; P = 0.38), median procedural completion time (23.0 vs. 27.0 minutes; P = 0.15), mean number of cannulation attempts (2.7 vs. 3.1; P = 0.24), crossover rate (6.5 % vs. 7.8 %; P = 0.65), and proportion of cases with high complexity (36.4 % [168/461] vs. 31.1 % [28/90]) ([Table 4]). On a scale of 1 (worst) to 10 (best), the median rating of overall satisfaction with the performance of the single-use duodenoscope was 8.0 for expert endoscopists and 7.0 for less expert endoscopists (P < 0.001).
ERCP, endoscopic retrograde cholangiopancreatography.
#
Comparison of outcomes by prior sphincterotomy
On comparison with cases with at least one prior sphincterotomy, cases with no prior sphincterotomy had a higher median total procedural time (25.0 vs. 21.0 minutes; P = 0.01), higher mean (SD) number of cannulation attempts (3.6 [4.7] vs. 1.6 [1.9]; P < 0.001), and a lower likelihood of having high complexity (P = 0.001). The baseline sphincterotomy status did not correlate with the crossover rate (P = 0.09), ERCP completion rate (P = 0.52), or overall satisfaction with the single-use duodenoscope (P = 0.36).
#
Related serious adverse events
Of the 551 enrolled patients, 43 (7.8 %, 95 %CI 5.7 %–10.4 %) experienced at least one SAE related to the device or procedure within a median (range) of 1 (0–28) days post-procedure ([Table 5]). There were 18 patients (3.3 %, 95 %CI 2.0 %–5.1 %) who developed pancreatitis (mild or moderate [n = 16], severe [n = 2]; post-ERCP [n = 17] reported 0–4 days after the ERCP, acute pancreatitis [n = 1]), six (1.1 %) who experienced pain, five (0.9 %) who had a GI bleed, three (0.5 %) who developed sepsis, two each (0.4 %) who developed cholangitis, cholecystitis, fever, or bacteremia, or had a perforation (0.4 %, 95 %CI 0.04 %–1.3 %; one gastric, one associated with sphincteroplasty during ERCP). The following additional SAEs occurred in one patient each: pneumoretroperitoneum, pneumonia, reintervention for biliary stones, exacerbation of COPD. SAE rates were similar for expert and less expert endoscopists (16.7 % [77/461] vs. 8.9 % [8/90], respectively; P = 0.08) and ASGE grade 1 or 2 cases vs. ASGE grade 3 or 4 levels of complexity (13.8 % [47/340] vs. 18.4 % [36/196], respectively; P = 0.17).
Events, n[1] |
Patients, n (%) |
95 %CI |
|
Any serious adverse event |
46 |
43 (7.8) |
5.7 %−10.4 % |
Pancreatitis |
18 |
18 (3.3) |
1.9 %−5.1 % |
|
17 |
17 (3.1) |
1.8 %−4.9 % |
|
1 |
1 (0.2) |
0.0 %−1.0 % |
Pain |
6 |
6 (1.1) |
0.4 %−2.4 % |
Gastrointestinal bleeding (fatal in one case) |
5 |
5 (0.9) |
0.3 %−2.1 % |
Sepsis |
3 |
3 (0.5) |
0.1 %−1.6 % |
Cholangitis |
2 |
2 (0.4) |
0.0 %−1.3 % |
Cholecystitis |
2 |
2 (0.4) |
0.0 %−1.3 % |
Fever |
2 |
2 (0.4) |
0.0 %−1.3 % |
Bacteremia |
2 |
2 (0.4) |
0.0 %−1.3 % |
Perforation (fatal in one case) |
2 |
2 (0.4) |
0.0 %−1.3 % |
Pneumoretroperitoneum |
1 |
1 (0.2) |
0.0 %−1.0 % |
Reintervention with ERCP under general anesthesia for biliary stone removal |
1 |
1 (0.2) |
0.0 %−1.0 % |
Post-ERCP pneumonia |
1 |
1 (0.2) |
0.0 %−1.0 % |
Exacerbation of COPD |
1 |
1 (0.2) |
0.0 %−1.0 % |
ERCP, endoscopic retrograde cholangiopancreatography; PEP, post-ERCP pancreatitis; COPD, chronic obstructive pulmonary disease.
1 Each patient had one or more of the listed serious adverse events; rows are not mutually exclusive.
The overall rate of related SAEs that were fatal in our study was 0.5 % (3/551; 95 %CI 0.11 %–1.6 %). These included one case of acute pancreatitis occurring 16 days after ERCP that was fatal on day 70, one fatal hemorrhage from the papilla after balloon dilation, and one fatal perforation associated with sphincteroplasty.
#
#
Discussion
In this largest global study of a single-use duodenoscope, endoscopists with varying levels of experience had high procedural success in ERCPs, one-third of which were high complexity cases. The median performance ratings were high for this device. SAEs were in the expected range, based on the published safety data for reusable duodenoscopes. Consistent with two previously published clinical studies [9] [12], the current study supports the safe and effective performance of the first single-use duodenoscope among endoscopists with varying levels of experience in multiple countries.
In an effort to eliminate the potential for contamination or infection associated with ineffective reprocessing of reusable duodenoscopes, the FDA have cleared duodenoscopes with disposable components [19] [20] and two fully disposable duodenoscopes [13] [15]. Single-use duodenoscopes may be especially useful in certain challenging logistical settings, such as during evening and weekend hours, and in various emergent situations, including when procedures are done outside of the endoscopy department [21]. The disadvantages of single-use duodenoscopes include their current high cost compared with duodenoscopes that are reusable [22] or partially disposable (with disposable endpieces [23] [24]), and their potential adverse impacts on the environment, while recycling programs currently are not yet widely available [25]. Although a 2023 RCT found that duodenoscopes with a disposable elevator cap exhibited reduced contamination following high level disinfection compared with standard scope designs [23], the status of disposable endoscopic endpieces is uncertain after the FDA issued a warning letter to a duodenoscope manufacturer regarding its slow response to complaints about the distal end cover “dropping out” or cracking [26].
Ecological sustainability is an important issue for both single-use and reusable devices. For example, a 2023 publication focused on manufacturing-associated carbon dioxide emissions reported that performing ERCP with single-use duodenoscopes releases a carbon dioxide equivalent that is 24–47 times greater than with a reusable duodenoscope or a reusable duodenoscope with disposable endcaps [27]. This report made some assumptions that were not accurate for the EXALT duodenoscope and did not include selective use or recycling in the analysis. Furthermore, the report did not mention ethylene oxide, which is used for approximately 50 % of sterile devices sold in the USA and is regulated by the US Environmental Protection Agency owing to its carcinogenicity [28]. Improvements in recycling efforts, and control and mitigation of the chemicals used in reprocessing are necessary to optimize patient and employee safety as single-use and reusable duodenoscopes are both used in clinical practice.
The results from the current study are consistent with those from previous studies of the single-use duodenoscope [9] [10] [11] [12], but from a much larger study population. The device was successfully used for procedures of all ASGE grades of complexity, with a low (6.7 %) incidence of crossover to a reusable duodenoscope and high (96 %) completion rate by both expert and less expert operators. Study practices met the published ERCP quality indicators [29] [30] [31] for most items that had study documentation. Consistent with the ESGE quality indicators for ERCP [30], the rate of post-ERCP pancreatitis was < 10 % in our study. Although the precise role and place of single-use devices is the subject of ongoing discussion, the current study provides added evidence that there are no arguments against their use from the perspective of procedural outcome and success, which is a prerequisite for considering their use in clinical practice.
All ERCP-associated SAEs are concerning and continual efforts must be made to minimize them. We compared the safety results in the current study with the published safety data for procedures performed with reusable duodenoscopes. The largest analysis of ERCP safety available is from Andriulli et al. [32], who estimated the post-ERCP complication rates among 21 prospective studies published between January 1977 and May 2006. They reported an estimated ERCP-specific mortality rate of 0.33 % (55/16 855 patients; 95 %CI 0.24 %–0.42 %), which is comparable to the rate of fatal related SAEs in our study (0.5 %; 3/551). Andriulli’s estimated rates of perforation (0.60 %, 95 %CI 0.48 %–0.72 %), pancreatitis (585/16 855 patients; 3.47 %, 95 %CI 3.19 %–3.75 %), and infection (1.44 %; 95 %CI 1.26 %–1.62 %) were also similar to the rates in our study (0.4 % [2/551], 3.3 % [18/551], and 1.63 % [9/551; 95 %CI 0.75 %–3.08 %], respectively).
These findings suggest that, when used by a group of endoscopists with a predominantly expert level of experience, the single-use duodenoscope has a comparable safety profile to reusable duodenoscopes. The comparable rate of infection also shows that endogenous infection can still occur with when using a single-use duodenoscope, so the procedure-associated infection rate is not guaranteed to be lower than the rate for reusable duodenoscopes.
Our study has several strengths, limitations, and considerations for study interpretation. This was a multinational observational study including a large number of scheduled ERCP cases with a range of complexity performed by endoscopists at varying levels of experience, including 30-day safety follow-up. Limitations include the lack of randomization and the absence of a control group. Although patients were consecutively screened, the 68 % who were eligible to participate in this study may not be typical of patients at all endoscopy centers. Specific reasons why 10 screened patients were excluded for “concern over safety of subject if study duodenoscope was used for procedure’’ were not documented. Of note, several investigators participated in the development of the single-use duodenoscope [9] [14] and have also received research funding from the study sponsor and from manufacturers of reusable duodenoscopes. Finally, this study evaluated the efficacy and safety of a single-use duodenoscope but was not designed to address the associated costs or impact on reprocessing services or the environment.
In conclusion, academic endoscopists used the first marketed single-use duodenoscope to successfully complete a large number of scheduled ERCP procedures with a range of complexity in a diverse patient population. Consistent with past studies, the device showed good performance and an SAE rate comparable with the published estimates for reusable duodenoscopes.
#
Data sharing
The data, analytic methods, and study materials for this study may be made available to other researchers in accordance with the Boston Scientific Data Sharing Policy (https://www.bostonscientific.com/en-US/data-sharing-requests.html
#
#
Competing interests
M.J. Bruno has provided consultancy and had support for industry- and investigator-initiated studies for Boston Scientific, Cook Medical, and Pentax Medical, and support for investigator-initiated studies from Mylan, InterScope, and ChiRhoStim. T. Beyna has provided consultancy for Boston Scientific, Olympus, Medtronic, and Microtech Endoscopy. D. Carr-Locke receives royalties from Steris Corporation. P. Chahal has provided consultancy and is on the advisory council for Boston Scientific, and is on the advisory council for Medtronic. G. Costamagna has received grants from Cook Endoscopy, Olympus, and Boston Scientific and has been on advisory committees or review panels for Boston Scientific and Olympus. B. Devereaux has provided consultancy to Olympus, Boston Scientific, and Mylan. M. Giovannini has consultancy agreements with Pentax and Taewoong. C. Khor has provided consultancy for Boston Scientific, Erbe, and Fujifilm. J. Lau is a research consultant for Boston Scientific. G. May is a consultant for Olympus and is a speaker for/on the advisory board for Vantage Endoscopy, Fujifilm, Pentax, Medtronic, and Boston Scientific. V.R. Muthusamy is a consultant for and has received research support from Boston Scientific, is a consultant for Medtronic, and on the advisory board for Motus GI and Endogastric Solutions; he holds stock options in CapsoVision; he was previously a consultant to Medivators and Interpace Diagnostics. S. Patel is a consultant for Boston Scientific, Olympus, Conmed, and AbbVie. B.T. Petersen is an investigator for Boston Scientific and Ambu and has provided consultancy for Olympus and Pentax. D.K. Pleskow has provided consultancy to Boston Scientific, Fuji, and Olympus. I. Raijman is a speaker for/on the advisory board of Boston Scientific, Microtech, and Pentax, has provided consultancy to Olympus, Conmed, Endosound, and Cryo, and is co-owner of EndoRx. A. Repici has received consultancy and speaker's fees from Boston Scientific Corporation. A.S. Ross has provided consultancy to Boston Scientific. D.V. Sejpal has received consultancy fees and research funding frrom Boston Scientific and Olympus. S. Sherman has provided consultancy for Olympus, Boston Scientific, and Cook. U.D. Siddiqui has received research funding from Boston Scientific, consultancy and speaker's fees from Boston Scientific, Cook, Olympus, Medtronic, and ConMed, and speaker's fees from Pinnacle Biologics and Ovesco. J.A. Peetermans and M.J. Rousseau are full-time employees of Boston Scientific. A. Slivka has received research support for IRB-approved studies from Boston Scientific and Olympus. M. Goenka, N. Reddy, and C. Zaidy declare that they have no conflict of interest.
Acknowledgments
Members of the EXALT Single-use Duodenoscope Study Group who participated in this study are: Jennifer S. Chennat and Adam Slivka at the University of Pittsburgh Medical Center (Pittsburgh, Pennsylvania, USA); Divyesh V. Sejpal at Dignity/CommonSpirit Health, Creighton School of Medicine (Phoenix, Arizona, USA); Calvin Lee at North Shore University Hospital (Manhasset, New York, USA); Shayan Irani, Rajesh Krishnamoorthi, Michael Larsen, and Andrew S. Ross at Virginia Mason Medical Center (Seattle, Washington, USA); V. Raman Muthusamy and Adarsh Thaker at University of California, Los Angeles (Los Angeles, California, USA); Vinay Chandrasekhara, Ryan Law, Michael Levy, John A. Martin, Bret T. Petersen, and Andrew C. Storm at Mayo Clinic (Rochester, Minnesota, USA); Tyler M. Berzin, Douglas K. Pleskow, and Mandeep S. Sawhney at Beth Israel Deaconess Medical Center (Boston, Massachusetts, USA); Marco J. Bruno and P.J.F. de Jonge at Erasmus Medical Center (Rotterdam, The Netherlands); Gary May and Jeffrey Mosco at St. Michael’s Hospital (Alberta, Canada); Alessandro Fugazza and Anthony Repici at Instituto Clinical Humanitas (Milan, Italy); Prabhleen Chahal and John Vargo at Cleveland Clinic (Cleveland, Ohio, USA); Benjamin Bick, Mark A. Gromski, Aditya Gutta, and Stuart Sherman at Indiana University Medical Center (Indianapolis, Indiana, USA); Danny Issa, Stephen Kim, V. Raman Muthusamy, Alireza Sedarat, and Adarsh Thaker at Ronald Reagan UCLA Medical Center (Los Angeles, California, USA); Torsten Beyna and Horst Neuhaus at Evangelisches Krankenhaus Dusseldorf (Dusseldorf, Germany); Sundeep Lakhtakia, Mohan Ramchandani, and D. Nageshwar Reddy at Asian Institute of Gastroenterology (Hyderabad, India); Uzma D. Siddiqui at University of Chicago Hospital (Chicago, Illinois, USA); Fabrice Caillol and Marc Giovannini at Institut Paoli-Calmettes (Marseille, France); James Y.W. Lau at Prince of Wales Hospital (Hong Kong, China); David Leslie Carr-Locke, Srihari Mahadev, Kartik Sampath, and Reem Sharaiha at New York Presbyterian Hospital & Weill Cornell Medicine (New York City, New York, USA); Mark Appleyard, Benedict Devereaux, and Florian Grimpen at Royal Brisbane and Women's Hospital (Brisbane, Australia); Mahar Kumar Goenka at Apollo Multispecialty Hospital (Kolkata, India); Christopher Khor and Damien Tan at Singapore General Hospital (Singapore); Moses A. Balabyeki, Nkomba Christophe Kalenga, Oleh Matsevych, and Christopher Charles Michael Ziady at Dr. George Mukhari Provincial (Pretoria, South Africa); Sandeep N. Patel, Isaac Raijman, Shreyas Saligram, and Hari Sayana at Texas International Endoscopy Center (Houston, Texas, USA); Joyce A. Peetermans, Matthew J. Rousseau, Steve Fodem, and Jeff Insull at Boston Scientific Corporation (Marlborough, Massachusetts, USA).
The authors acknowledge Boston Scientific employees Steve Fodem, MS for his project management, Victoria Lazzari for her data management assistance, and Margaret Gourlay, MD, MPH for her assistance in manuscript preparation.
* Deceased; this paper is dedicated to our friend and eminent colleague gastroenterologist Dr. Stuart Sherman.
-
References
- 1 Kuduva Rajan S, Madireddy S, Jaladi PR. et al. Burdens of postoperative infection in endoscopic retrograde cholangiopancreatography inpatients. Cureus 2019; 11: e5237
- 2 Salminen P, Laine S, Gullichsen R. Severe and fatal complications after ERCP: analysis of 2555 procedures in a single experienced center. Surg Endosc 2008; 22: 1965-1970
- 3 Rerknimitr R, Fogel EL, Kalayci C. et al. Microbiology of bile in patients with cholangitis or cholestasis with and without plastic biliary endoprosthesis. Gastrointest Endosc 2002; 56: 885-889
- 4 US Health and Human Services. Centers for Disease Control and Prevention. Antibiotic Resistance Threats in the United States, 2013 (April 2013). Accessed: 21 July 2023. Available from: www.cdc.gov/drugresistance/threat-report-2013/pdf/ar-threats-2013-508.pdf
- 5 Kwakman JA, Erler NS, Vos MC. et al. Risk evaluation of duodenoscope-associated infections in the Netherlands calls for a heightened awareness of device-related infections: a systematic review. Endoscopy 2022; 54: 148-155
- 6 Barakat MT, Ghosh S, Banerjee S. Cost utility analysis of strategies for minimizing risk of duodenoscope-related infections. Gastrointest Endosc 2022; 95: 929-938 e922
- 7 Rauwers AW, Troelstra A, Fluit AC. et al. Independent root-cause analysis of contributing factors, including dismantling of 2 duodenoscopes, to investigate an outbreak of multidrug-resistant Klebsiella pneumoniae. Gastrointest Endosc 2019; 90: 793-804
- 8 Heuvelmans M, Wunderink HF, van der Mei HC. et al. A narrative review on current duodenoscope reprocessing techniques and novel developments. Antimicrob Resist Infect Control 2021; 10: 171
- 9 Muthusamy VR, Bruno MJ, Kozarek RA. et al. Clinical evaluation of a single-use duodenoscope for endoscopic retrograde cholangiopancreatography. Clin Gastroenterol Hepatol 2020; 18: 2108-2117 e2103
- 10 Bang JY, Hawes R, Varadarajulu S. Equivalent performance of single-use and reusable duodenoscopes in a randomised trial. Gut 2020; 70: 838-844
- 11 Napoleon B, Gonzalez JM, Grandval P. et al. Evaluation of the performances of a single-use duodenoscope: Prospective multi-center national study. Dig Endosc 2022; 34: 215-221
- 12 Slivka A, Ross AS, Sejpal DV. et al. Single-use duodenoscope for ERCP performed by endoscopists with a range of experience in procedures of variable complexity. Gastrointest Endosc 2021; 94: 1046-1055
- 13 US Food and Drug Administration. 510(k) number K201098: Ambu Duodeno System. Accessed: 21 July 2023. Available from: www.accessdata.fda.gov/cdrh_docs/pdf20/K201098.pdf
- 14 Ross AS, Bruno MJ, Kozarek RA. et al. Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP: a randomized bench-model comparison. Gastrointest Endosc 2020; 91: 396-403
- 15 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first fully disposable duodenoscope, eliminating the potential for infections caused by ineffective reprocessing (December 13, 2019). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-fully-disposable-duodenoscope-eliminating-potential-infections-caused-ineffective
- 16 US Food and Drug Administration. FDA News. Boston Scientific Gains CE Mark for Single-Use Duodenoscope (January 24, 2020). Accessed: 21 July 2023. Available from: www.fdanews.com/articles/195625-boston-scientific-gains-ce-mark-for-single-use-duodenoscope
- 17 Boston Scientific Corporation. EXALT Model D Directions for Use. Accessed: 21 July 2023. Available from: www.bostonscientific.com/content/dam/Manuals/us/current-rev-en/50866130-01B_EXALT_Model_D_DFU_en-USA_s.pd
- 18 Testoni PA, Mariani A, Aabakken L. et al. Papillary cannulation and sphincterotomy techniques at ERCP: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. Endoscopy 2016; 48: 657-683
- 19 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first duodenoscope with disposable distal cap (March 26, 2018). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-duodenoscope-disposable-distal-cap
- 20 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first duodenoscope with disposable elevator piece, reducing the number of parts needing disinfection (November 15, 2019). Accessed: 12 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-duodenoscope-disposable-elevator-piece-reducing-number-parts-needing-disinfection
- 21 Rivallin P, Lorenzo D, Hedjoudje A. et al. Single-use duodenoscopes are an efficient tool for emergency ERCP in real life. Endosc Int Open 2022; 10: E1497-E1500
- 22 Bang JY, Sutton B, Hawes R. et al. Concept of disposable duodenoscope: at what cost?. Gut 2019; 68: 1915-1917
- 23 Forbes N, Elmunzer BJ, Allain T. et al. Effect of disposable elevator cap duodenoscopes on persistent microbial contamination and technical performance of endoscopic retrograde cholangiopancreatography: the ICECAP Randomized Clinical Trial. JAMA Intern Med 2023; 183: 191-200
- 24 Ofosu A, Ramai D, Mozell D. et al. Analysis of reported adverse events related to single-use duodenoscopes and duodenoscopes with detachable endcaps. Gastrointest Endosc 2022; 96: 67-72
- 25 Trindade AJ, Copland A, Bhatt A. et al. Single-use duodenoscopes and duodenoscopes with disposable end caps. Gastrointest Endosc 2021; 93: 997-1005
- 26 US Food and Drug Administration. Center for Devices and Radiological Health United States: Inspections, Compliance, Enforcement, and Criminal Investigations; Compliance Actions and Activities; Warning Letters. Warning Letter, Olympus Medical Systems Corp. MARCS-CMS 654013 (March 15, 2023). Accessed: 21 July 2023. Available from: www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/olympus-medical-systems-corp-654013-03152023
- 27 Le NNT, Hernandez LV, Vakil N. et al. Environmental and health outcomes of single-use versus reusable duodenoscopes. Gastrointest Endosc 2022; 96: 1002-1008
- 28 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA Continues Efforts to Support Innovation in Medical Device Sterilization (August 3, 2022). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-continues-efforts-support-innovation-medical-device-sterilization
- 29 Adler DG, Lieb 2nd JG. et al. Quality indicators for ERCP. Gastrointest Endosc 2015; 81: 54-66
- 30 Domagk D, Oppong KW, Aabakken L. et al. Performance measures for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound: A European Society of Gastrointestinal Endoscopy (ESGE) Quality Improvement Initiative. United European Gastroenterol J 2018; 6: 1448-1460
- 31 Alberca de Las Parras F, Lopez-Picazo J, Perez Romero S. et al. Quality indicators for endoscopic retrograde cholangiopancreatography. The procedure of endoscopic retrograde cholangiopancreatography. . Rev Esp Enferm Dig 2018; 110: 658-666
- 32 Andriulli A, Loperfido S, Napolitano G. et al. Incidence rates of post-ERCP complications: a systematic survey of prospective studies. Am J Gastroenterol 2007; 102: 1781-1788
Corresponding author
Publication History
Received: 13 March 2023
Accepted after revision: 14 July 2023
Accepted Manuscript online:
18 July 2023
Article published online:
20 September 2023
© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Kuduva Rajan S, Madireddy S, Jaladi PR. et al. Burdens of postoperative infection in endoscopic retrograde cholangiopancreatography inpatients. Cureus 2019; 11: e5237
- 2 Salminen P, Laine S, Gullichsen R. Severe and fatal complications after ERCP: analysis of 2555 procedures in a single experienced center. Surg Endosc 2008; 22: 1965-1970
- 3 Rerknimitr R, Fogel EL, Kalayci C. et al. Microbiology of bile in patients with cholangitis or cholestasis with and without plastic biliary endoprosthesis. Gastrointest Endosc 2002; 56: 885-889
- 4 US Health and Human Services. Centers for Disease Control and Prevention. Antibiotic Resistance Threats in the United States, 2013 (April 2013). Accessed: 21 July 2023. Available from: www.cdc.gov/drugresistance/threat-report-2013/pdf/ar-threats-2013-508.pdf
- 5 Kwakman JA, Erler NS, Vos MC. et al. Risk evaluation of duodenoscope-associated infections in the Netherlands calls for a heightened awareness of device-related infections: a systematic review. Endoscopy 2022; 54: 148-155
- 6 Barakat MT, Ghosh S, Banerjee S. Cost utility analysis of strategies for minimizing risk of duodenoscope-related infections. Gastrointest Endosc 2022; 95: 929-938 e922
- 7 Rauwers AW, Troelstra A, Fluit AC. et al. Independent root-cause analysis of contributing factors, including dismantling of 2 duodenoscopes, to investigate an outbreak of multidrug-resistant Klebsiella pneumoniae. Gastrointest Endosc 2019; 90: 793-804
- 8 Heuvelmans M, Wunderink HF, van der Mei HC. et al. A narrative review on current duodenoscope reprocessing techniques and novel developments. Antimicrob Resist Infect Control 2021; 10: 171
- 9 Muthusamy VR, Bruno MJ, Kozarek RA. et al. Clinical evaluation of a single-use duodenoscope for endoscopic retrograde cholangiopancreatography. Clin Gastroenterol Hepatol 2020; 18: 2108-2117 e2103
- 10 Bang JY, Hawes R, Varadarajulu S. Equivalent performance of single-use and reusable duodenoscopes in a randomised trial. Gut 2020; 70: 838-844
- 11 Napoleon B, Gonzalez JM, Grandval P. et al. Evaluation of the performances of a single-use duodenoscope: Prospective multi-center national study. Dig Endosc 2022; 34: 215-221
- 12 Slivka A, Ross AS, Sejpal DV. et al. Single-use duodenoscope for ERCP performed by endoscopists with a range of experience in procedures of variable complexity. Gastrointest Endosc 2021; 94: 1046-1055
- 13 US Food and Drug Administration. 510(k) number K201098: Ambu Duodeno System. Accessed: 21 July 2023. Available from: www.accessdata.fda.gov/cdrh_docs/pdf20/K201098.pdf
- 14 Ross AS, Bruno MJ, Kozarek RA. et al. Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP: a randomized bench-model comparison. Gastrointest Endosc 2020; 91: 396-403
- 15 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first fully disposable duodenoscope, eliminating the potential for infections caused by ineffective reprocessing (December 13, 2019). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-fully-disposable-duodenoscope-eliminating-potential-infections-caused-ineffective
- 16 US Food and Drug Administration. FDA News. Boston Scientific Gains CE Mark for Single-Use Duodenoscope (January 24, 2020). Accessed: 21 July 2023. Available from: www.fdanews.com/articles/195625-boston-scientific-gains-ce-mark-for-single-use-duodenoscope
- 17 Boston Scientific Corporation. EXALT Model D Directions for Use. Accessed: 21 July 2023. Available from: www.bostonscientific.com/content/dam/Manuals/us/current-rev-en/50866130-01B_EXALT_Model_D_DFU_en-USA_s.pd
- 18 Testoni PA, Mariani A, Aabakken L. et al. Papillary cannulation and sphincterotomy techniques at ERCP: European Society of Gastrointestinal Endoscopy (ESGE) Clinical Guideline. Endoscopy 2016; 48: 657-683
- 19 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first duodenoscope with disposable distal cap (March 26, 2018). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-duodenoscope-disposable-distal-cap
- 20 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA clears first duodenoscope with disposable elevator piece, reducing the number of parts needing disinfection (November 15, 2019). Accessed: 12 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-clears-first-duodenoscope-disposable-elevator-piece-reducing-number-parts-needing-disinfection
- 21 Rivallin P, Lorenzo D, Hedjoudje A. et al. Single-use duodenoscopes are an efficient tool for emergency ERCP in real life. Endosc Int Open 2022; 10: E1497-E1500
- 22 Bang JY, Sutton B, Hawes R. et al. Concept of disposable duodenoscope: at what cost?. Gut 2019; 68: 1915-1917
- 23 Forbes N, Elmunzer BJ, Allain T. et al. Effect of disposable elevator cap duodenoscopes on persistent microbial contamination and technical performance of endoscopic retrograde cholangiopancreatography: the ICECAP Randomized Clinical Trial. JAMA Intern Med 2023; 183: 191-200
- 24 Ofosu A, Ramai D, Mozell D. et al. Analysis of reported adverse events related to single-use duodenoscopes and duodenoscopes with detachable endcaps. Gastrointest Endosc 2022; 96: 67-72
- 25 Trindade AJ, Copland A, Bhatt A. et al. Single-use duodenoscopes and duodenoscopes with disposable end caps. Gastrointest Endosc 2021; 93: 997-1005
- 26 US Food and Drug Administration. Center for Devices and Radiological Health United States: Inspections, Compliance, Enforcement, and Criminal Investigations; Compliance Actions and Activities; Warning Letters. Warning Letter, Olympus Medical Systems Corp. MARCS-CMS 654013 (March 15, 2023). Accessed: 21 July 2023. Available from: www.fda.gov/inspections-compliance-enforcement-and-criminal-investigations/warning-letters/olympus-medical-systems-corp-654013-03152023
- 27 Le NNT, Hernandez LV, Vakil N. et al. Environmental and health outcomes of single-use versus reusable duodenoscopes. Gastrointest Endosc 2022; 96: 1002-1008
- 28 US Food and Drug Administration. News & Events; FDA Newsroom; Press Announcements. FDA Continues Efforts to Support Innovation in Medical Device Sterilization (August 3, 2022). Accessed: 21 July 2023. Available from: www.fda.gov/news-events/press-announcements/fda-continues-efforts-support-innovation-medical-device-sterilization
- 29 Adler DG, Lieb 2nd JG. et al. Quality indicators for ERCP. Gastrointest Endosc 2015; 81: 54-66
- 30 Domagk D, Oppong KW, Aabakken L. et al. Performance measures for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound: A European Society of Gastrointestinal Endoscopy (ESGE) Quality Improvement Initiative. United European Gastroenterol J 2018; 6: 1448-1460
- 31 Alberca de Las Parras F, Lopez-Picazo J, Perez Romero S. et al. Quality indicators for endoscopic retrograde cholangiopancreatography. The procedure of endoscopic retrograde cholangiopancreatography. . Rev Esp Enferm Dig 2018; 110: 658-666
- 32 Andriulli A, Loperfido S, Napolitano G. et al. Incidence rates of post-ERCP complications: a systematic survey of prospective studies. Am J Gastroenterol 2007; 102: 1781-1788