CC BY 4.0 · Endoscopy 2023; 55(12): 1103-1114
DOI: 10.1055/a-2131-7180
Original article

Global prospective case series of ERCPs using a single-use duodenoscope

Marco J. Bruno
1   Department of Gastroenterology and Hepatology, Erasmus Medical Center, University Medical Center, Rotterdam, The Netherlands
,
2   Department of Internal Medicine, Evangelisches Krankenhaus Düsseldorf, Düsseldorf, Germany
,
David Carr-Locke
3   Division of Gastroenterology & Hepatology, New York Presbyterian Hospital, Weill Cornell Medicine, New York, New York, USA
,
Prabhleen Chahal
4   Gastroenterology and Hepatology, Digestive Disease and Surgery Institute, Cleveland Clinic, Cleveland, Ohio, USA
,
5   Fondazione Policlinico Universitario Agostino Gemelli IRCCS (Università Cattolica del Sacro Cuore), Rome, Italy
,
Benedict Devereaux
6   Department of Gastroenterology, University of Queensland, Royal Brisbane and Women’s Hospital, Brisbane, Australia
,
Marc Giovannini
 7   Endoscopy Unit, Institut Paoli-Calmettes, Marseille, France
,
Mahesh K. Goenka
 8   Department of Gastroenterology, Apollo Multispecialty Hospitals, Kolkata, India
,
Christopher Khor
 9   Department of Gastroenterology and Hepatology, Singapore General Hospital, Singapore
,
James Lau
10   Prince of Wales Hospital, Hong Kong, China
,
Gary May
11   The Centre for Therapeutic Endoscopy and Endoscopic Oncology, Division of Gastroenterology, St. Michael’s Hospital, Alberta, Canada
,
V. Raman Muthusamy
12   Vatche and Tamar Manoukian Division of Digestive Diseases, David Geffen School of Medicine at UCLA, University of California, Los Angeles, California, USA
,
Sandeep Patel
13   Division of Gastroenterology, UT Health San Antonio, San Antonio, Texas, USA
,
Bret T. Petersen
14   Division of Gastroenterology and Hepatology, Mayo Clinic, Rochester, Minnesota, USA
,
15   Center for Advanced Endoscopy, Division of Gastroenterology, Beth Israel Deaconess Medical Center, and Harvard Medical School, Boston, Massachusetts, USA
,
Isaac Raijman
16   Texas International Endoscopy Center, Houston, Texas, USA
,
D. Nageshwar Reddy
17   Asian Institute of Gastroenterology, Hyderabad, India
,
Alessandro Repici
18   Endoscopy Unit, Humanitas Clinical and Research Center IRCCS, Rozzano, Milan, Italy
19   Department of Biomedical Sciences, Humanitas University, Rozzano, Milan, Italy
,
Andrew S. Ross
20   Department of Gastroenterology, Digestive Disease Institute, Virginia Mason Medical Center, Seattle, Washington
,
Divyesh V. Sejpal
21   Digestive Disease Institute, Dignity/CommonSpirit Health, Creighton School of Medicine, Phoenix, Arizona, USA
,
Stuart Sherman*
22   Division of Gastroenterology and Hepatology, Department of Medicine, Indiana University, Indianapolis, Indiana, USA
,
Uzma D. Siddiqui
23   Center for Endoscopic Research and Therapeutics (CERT), University of Chicago, Chicago, Illinois, USA
,
Christopher Ziady
24   Dr. George Mukhari Academic Medical Center, Pretoria, South Africa
,
Joyce A. Peetermans
25   Endoscopy Division, Boston Scientific Corporation, Marlborough, Massachusetts, USA
,
Matthew J. Rousseau
25   Endoscopy Division, Boston Scientific Corporation, Marlborough, Massachusetts, USA
,
Adam Slivka
26   Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania, USA
,
EXALT Single-use Duodenoscope Study Group › Author Affiliations
Boston Scientific Corporation

Trial Registration: ClinicalTrials.gov Registration number (trial ID): NCT04103749 Type of study: Multi-Center Study
 


Abstract

Background The first commercialized single-use duodenoscope was cleared by the US Food and Drug Administration in December 2019. Data regarding endoscopic retrograde cholangiopancreatography (ERCP) using a single-use duodenoscope are needed on a broader range of cases conducted by endoscopists with varying levels of experience in a wide range of geographic areas.

Methods 61 endoscopists at 22 academic centers in 11 countries performed ERCP procedures in adult patients aged ≥ 18. Outcomes included ERCP completion for the intended indication, rate of crossover to a reusable endoscope, device performance ratings, and serious adverse events (SAEs).

Results Among 551 patients, 236 (42.8 %) were aged > 65, 281 (51.0 %) were men, and 256 (46.5 %) had their procedure as an inpatient. ERCPs included 196 (35.6 %) with American Society for Gastrointestinal Endoscopy complexity of grades 3–4. A total of 529 ERCPs (96.0 %) were completed: 503 (91.3 %) using only the single-use duodenoscope, and 26 (4.7 %) with crossover to a reusable endoscope. There were 22 ERCPs (4.0 %) that were not completed, of which 11 (2.0 %) included a crossover and 11 (2.0 %) were aborted cases (no crossover). Median ERCP completion time was 24.0 minutes. Median overall satisfaction with the single-use duodenoscope was 8.0 (scale of 1 to 10 [best]). SAEs were reported in 43 patients (7.8 %), including 17 (3.1 %) who developed post-ERCP pancreatitis.

Conclusions In academic medical centers over a wide geographic distribution, endoscopists with varying levels of experience using the first marketed single-use duodenoscope had good ERCP procedural success and reported high performance ratings for this device.


#

Introduction

Post-endoscopic retrograde cholangiopancreatography (ERCP) infections are estimated to occur in approximately 5 % of patients [1] [2], commonly due to enteric organisms [3]. Of greatest concern are infections from multidrug-resistant organisms (MDRO), which account for a disproportionately higher morbidity and mortality compared with infections with antibiotic-susceptible organisms [4]. These can be endogenous infections due to translocation of organisms from the gastrointestinal (GI) tract of the patient into the pancreaticobiliary ducts by the duodenoscope or endoscopic devices, or can be exogenous, due to MDROs that were colonizing the ducts of another patient being harbored inside a reusable duodenoscope. A 2022 systematic review estimated the minimum risk of duodenoscope-associated MDRO infection to be at least 0.01 % per ERCP procedure in the Netherlands, which was substantially higher than reported rates in frequently cited older data [5]. The high morbidity and mortality associated with MDRO infections [6] and the delayed clinical detection of some of these infections [7] warrant improved reprocessing practices and the development of a uniform and practical protocol for duodenoscope microbiological sampling that can be applied in general healthcare facilities [8].

An alternative strategy to address exogenous MDRO infections associated with duodenoscopes is the use of single-use duodenoscopes. Since receiving US Food and Drug Administration (FDA) clearance in 2019, the first single-use duodenoscope has been studied in both US and European settings [9] [10] [11] and by endoscopists with varying levels of experience [12]. A randomized trial [10] and three case series [9] [11] [12] reported comparable safety profiles, technical performance, and endoscopist user satisfaction ratings for the first marketed single-use duodenoscope compared with reusable duodenoscopes. A second brand of single-use duodenoscope was cleared in 2020 [13]. Additional data are warranted, particularly to document the performance of the single-use duodenoscopes when used by less expert endoscopists, in cases with high complexity, and across a broad geographic range.

To characterize the efficacy and safety of the first marketed single-use duodenoscope further, we conducted a large clinical case series of ERCPs of any complexity in academic medical centers.


#

Methods

Study design

This was a large, multinational case series of the performance of a single-use duodenoscope for ERCP. Institutional Review Board and Ethics Committee approvals for the study were obtained at all study sites. All enrolled patients provided written informed consent for study participation before they contributed data. Single-use duodenoscopes were provided without charge to the investigational sites for use in the study procedures.

Boston Scientific Corporation (Marlborough, Massachusetts, USA) sponsored and funded the study. Sixteen endoscopists participated in published research during development of the single-use duodenoscope [9] [12] [14]. A statistician (M.J.R.) who is a full-time employee of Boston Scientific Corporation performed the data management and statistical analysis of this study. Review and input to the analysis of the study data were provided by two Boston Scientific employees (J.A.P. and M.J.R.) and by all participating physicians at Investigator Meetings between 2019 and 2022.


#

Single-use duodenoscope

The device used in this study was the EXALT Model D single-use duodenoscope (Boston Scientific Corporation), a sterile, single-use duodenoscope designed to function similarly to currently marketed reusable duodenoscopes, and to be discarded after use in a single procedure. The single-use duodenoscope received US FDA clearance in December 2019 and gained a CE mark in January 2020 [15] [16].


#

Patient population

The study protocol allowed for recruitment of adult patients scheduled for ERCP per the standard of care at up to 40 participating healthcare centers. Eligible patients were recruited for the study on selected weekdays when participating endoscopists were performing ERCPs. Patients were eligible for inclusion if they were aged ≥ 18 years, willing and able to comply with the study procedures and to provide written informed consent to participate in the study, and were scheduled for a clinically indicated ERCP or other duodenoscope-based procedure. Excluded from the study were potentially vulnerable patients including, but not limited to: pregnant women, patients for whom endoscopic techniques were contraindicated, patients enrolled in another investigational study that would directly interfere with the current study, and patients excluded at investigator discretion. Unlike previous studies of the single-use duodenoscope, the current study included patients with “altered pancreaticobiliary anatomy” because this was not a contraindication in the Directions for Use approved in 2020 [17].


#

Study procedures

Patient assessments

All enrolled participants had a preprocedural study visit for assessment of their demographics and relevant medical history. After the index procedure, participants were evaluated in person or by telephone at 72 hours (−1 day to + 2 days) and again at 30 days (± 3 days) to screen for post-procedure adverse events (AEs) or resolution of any previously reported issues. If the last follow-up visit was not completed, the reason was noted on the study completion form. Participants were considered lost to follow-up if they failed to return for their scheduled follow-up visits and were unable to be contacted by the study site staff after at least three documented attempts.


#

ERCP procedures

Participating endoscopists agreed to use the single-use duodenoscope in place of other brands of duodenoscope(s) used in the endoscopy unit for ERCP. Endoscopists’ level of experience was categorized as “expert” (> 2000 lifetime ERCPs) and “less expert” (≤ 2000 lifetime ERCPs). Start and stop times of the procedure, post-procedure subjective feedback on performance-related attributes, American Society for Gastrointestinal Endoscopy (ASGE) grade for complexity of the ERCP procedure, and all attempted or completed maneuvers during the ERCP were documented. Device deficiencies were recorded and reported, regardless of whether they led to an AE or inability to complete the ERCP. If the necessary maneuvers could not be completed with the single-use device and the endoscopist crossed over to use a reusable duodenoscope, the reason for noncompletion with the single-use device was recorded, and the ability to complete the ERCP maneuvers with a reusable duodenoscope was documented.

Difficult common bile duct (CBD) cannulation was defined by a modification of the European Society of Gastrointestinal Endoscopy (ESGE) definition [18] as the presence of one or more of the following: more than five contacts with the papilla while attempting to cannulate; more than 5 minutes spent attempting to cannulate following visualization of the papilla; at least one (instead of “more than one”) unintended pancreatic duct cannulation or opacification [18]. To represent all difficult procedures, difficult CBD cannulation was tabulated with failed ERCPs (even if cannulation was achieved).


#
#

Outcomes

The primary end point was the ability to complete the ERCP procedures for the intended indication(s). The secondary end points were: (i) the incidence of crossover from the single-use duodenoscope to a reusable duodenoscope; (ii) comparison of outcomes by ASGE grade, endoscopist level of experience, or prior sphincterotomy; (iii) endoscopist ratings of overall satisfaction with the single-use duodenoscope on a scale of 1 (worst) to 10 (best); (iv) serious AEs (SAEs) related to the device and/or procedure, assessed through to 30 days after the ERCP or other duodenoscope-based procedure.


#

Statistical analysis

Descriptive statistics included: frequency statistics for patient and procedural characteristics and procedure completion rates (completion rates reported separately for cases with and without crossover to a reusable duodenoscope); median and range for overall satisfaction and procedure duration; and mean (SD) and range for age and mean number of cannulation attempts. Endoscopist experience and ASGE complexity grades were assessed for completion time and performance ratings using Wilcoxon rank-sum tests, and for number of cannulation attempts using a negative binomial model. Fisher’s exact test was used to compare the occurrence of crossover to a reusable duodenoscope by level of experience or by ASGE case complexity level. For binary data, 95 %CIs were calculated using the Clopper–Pearson exact methods. Statistical analyses were performed using SAS 9.4 software (SAS Institute Inc., Cary, North Carolina, USA).


#
#

Results

Enrollment and patient characteristics

Of 809 patients who were screened, 551 (68.1 %) were enrolled after providing written informed consent to participate in the study. Reasons for nonparticipation are summarized in [Fig. 1].

Zoom Image
Fig. 1 Patient flow through the study.

Of the 551 patients (mean age 59.9 years) who had a procedure, 281 (51.0 %) were male and 332 (60.3 %) had a prior ERCP ([Table 1]), with 243 patients (44.1 %) having had a prior biliary and/or pancreatic sphincterotomy. The most common prior GI disorders in patients scheduled for ERCP or other duodenoscope-based procedure were: current bile duct stones/gallstones (n = 168; 30.5 %), biliary or pancreatic duct stricture (n = 124; 22.5 %), current immunosuppression for any cause (n = 71; 12.9 %), chronic pancreatitis (n = 68; 12.3 %), pancreatic tumor (n = 63; 11.4 %), abnormal imaging finding (n = 60; 10.9 %), current cholangitis (n = 53; 9.6 %), or cholangiocarcinoma (n = 46; 8.3 %). Two patients (0.4 %) had a known history of MDRO colonization.

Table 1

Characteristics of the 551 patients who were entered into the study and underwent an ERCP procedure that began with a single-use duodenoscope.

Characteristic

n (%), unless otherwise specified

Age, mean (SD) [range], years

59.9 (15.9) [18.0−94.0]

  • Older than 65 years

236 (42.8)

Sex, male

281 (51.0)

American Society of Anesthesiology physical status

  • I

66 (12.0)

  • II

198 (35.9)

  • III

250 (45.4)

  • IV

32 (5.8)

  • V

0 (0.0)

  • Not assessed

5 (0.9)

Relevant gastrointestinal medical history

  • Bile duct stones/gallstones(current)

168 (30.5)

  • Documented biliary or pancreatic stricture, unresolved

124 (22.5)

  • Current immunosuppression (any cause)

71 (12.9)

    • Pharmacologically induced

64 (11.6)

    • Post-transplant

30 (5.4)

    • Post-chemotherapy for cancer other than bile duct

19 (3.4)

    • Post-chemotherapy for bile duct cancer

12 (2.2)

    • Disease induced

11 (2.0)

    • In preparation for bone marrow or other transplantation

1 (0.2)

    • Other

5 (0.9)

  • Chronic pancreatitis

68 (12.3)

  • Pancreatic tumor

63 (11.4)

  • Abnormal image finding

60 (10.9)

  • Cholangitis (current)

53 (9.6)

  • Cholangiocarcinoma

46 (8.3)

  • Primary sclerosing cholangitis

24 (4.4)

  • Other gastrointestinal cancer

22 (4.0)

  • Pancreatic stones (current)

22 (4.0)

  • Recurrent cholangitis

21 (3.8)

  • Hepatic tumor

18 (3.3)

  • Hepatitis

11 (2.0)

  • Pancreatic pseudocyst

9 (1.6)

  • Known history of MDRO colonization

2 (0.4)

  • Known current MDRO colonization

2 (0.4)

  • Current documented bacterial infection, other than cholangitis

8 (1.5)

  • Other

146 (26.5)

ERCP, endoscopic retrograde cholangiopancreatography; MDRO, multidrug-resistant organism.

Of the 332 patients for whom the referral source was documented, 102 (30.7 %) presented directly to the participating study site, 154 (46.4 %) were referred from an academic medical center, and 76 (22.9 %) from a community medical center.


#

General ERCP characteristics

A total of 61 endoscopists (46 expert, 15 less expert) at 22 academic centers in 11 countries (1–6 endoscopists per center) performed 551 procedures: 25 cases at 21 of the study sites and 26 cases in the other. The median number of procedures performed by each endoscopist was seven (range 1–25). All procedures were performed in endoscopy suites, almost all (99.6 %; 549/551) during normal working hours, and most with endoscopy processing staff present (79.1 %; 436 /551) ([Table 2]). Prophylactic antibiotics were used in 272 (49.4 %) patients, and prophylactic nonsteroidal anti-inflammatory drugs were used in 315 (57.2 %). The ERCPs included all ASGE grades of complexity: grade 1 (least complex; 10.2 %; 56/551), grade 2 (51.5 %; 284/551), grade 3 (28.5 %; 157/551), and grade 4 (most complex; 7.1 %, 39/551). There were also 15 cases (2.9 %) that were duodenoscope-based procedures not included in the ASGE grading system. The median (range) procedure completion time was 24.0 (2.0–157.0) minutes, including times of ≤ 20 minutes in 240/534 cases (44.9 %) to ≥ 60 minutes in 39/534 (7.5 %) cases.

Table 2

Characteristics of the 551 endoscopic procedures carried out.

Characteristic

n (%), unless otherwise specified

Inpatient procedure

256 (46.5)

ERCP procedure

545 (98.9)

Prophylactic NSAID used

315 (57.2)

Prophylactic antibiotics used

272 (49.4)

Patient intubated

253 (45.9)

Medications for anesthesia

  • Propofol

477 (86.6)

  • Opioid

178 (32.3)

  • Neuromuscular blocker

137 (24.9)

  • Inhalation agent

74 (13.4)

  • Etomidate

13 (2.4)

  • Other

163 (29.6)

Medications for maintenance of anesthesia

  • Propofol

391 (71.0)

  • Opioid

153 (27.8)

  • Sevoflurane

124 (22.5)

  • Benzodiazepine

94 (17.1)

  • Neuromuscular blocker

54 (9.8)

  • Other

74 (13.4)

Procedure time with single-use duodenoscope, median (range) [n], minutes[1]

22.0 (0.0−157.0) [433]

Reusable scope used

37 (6.7)

Procedure time with reusable duodenoscope, median (range) [n], minutes

22.0 (6.0−75.0) [37]

Patient position

  • Prone

379 (68.8)

  • Supine

133 (24.1)

  • Left lateral decubitus

52 (9.4)

Location of procedure

  • Endoscopy suite

551 (100.0)

ERCP during normal hours

549 (99.6)

Endoscope reprocessing staff present

436 (79.1)

Total procedure time, median (range) [n], minutes[1]

24.0 (2.0−157.0) [548]

Difficult cannulation or failed ERCP[2]

150 (29.2)

  • Difficult cannulation without failed ERCP[2]

140 (27.2)

  • Failed ERCP without difficult cannulation

4(0.8) (514)

  • Failed ERCP with difficult cannulation

6(1.2) (514)

All intended maneuvers completed

524 (95.1)

ERCP completion

  • Completed, no crossover

503 (91.3)

  • Completed with crossover

26 (4.7)

  • Not completed, without crossover

11 (2.0)

  • Not completed, with crossover

11 (2.0)

Pancreatic maneuvers performed

128 (23.2)

Cholangioscopy

24 (4.4)

Pancreatoscopy

9 (1.6)

ASGE complexity grade[3]

  • Grade 1

56 (10.2)

  • Grade 2

284 (51.5)

  • Grade 3

157 (28.5)

  • Grade 4

39 (7.1)

  • Procedures not included in ASGE grading system

15 (2.7)

ERCP, endoscopic retrograde cholangiopancreatography; NSAID, nonsteroidal anti-inflammatory drug; ASGE, American Society for Gastrointestinal Endoscopy.

1 One case was aborted (because patient had food in the stomach) so the procedure time was 0 and was not included in the “total procedure time” calculation.


2 Denominator = 514 procedures.


3 ASGE grade for complexity of ERCPs: grade 1, deep cannulation of duct of interest, main papilla, sampling; grade 2, biliary stone extraction < 10 mm, treat biliary leaks, treat extrahepatic benign and malignant strictures, placed prophylactic stents; grade 3, biliary stone extraction ≥ 10 mm, minor papilla cannulation in divisum and therapy, removal of internally migrated biliary stents, intraductal imaging/biopsy/fine needle aspiration, management of acute or recurrent pancreatitis, treat pancreatic strictures, remove pancreatic stones mobile and < 5 mm, treat hilar tumors, treat benign biliary strictures hilum and above, manage suspected sphincter of Oddi dysfunction (with or without manometry); grade 4, remove internally migrated pancreatic stents, intraductal image-guided therapy (e. g. photodynamic therapy, electrohydraulic lithotripsy), pancreatic stones impacted and/or ≥ 5 mm, intrahepatic stones, pseudocyst drainage/necrosectomy, ampullectomy, ERCP after Whipple or Roux-en-Y bariatric surgery.


Investigators reported that after the study procedures, most of the single-use duodenoscopes were discarded in standard medical waste (35.4 %; 195 /551), or medical grade recycling (29.9 %; 165/551), regulated medical waste (24.9 %; 137/551), a sharps disposal container (4.9 %; 27/551), or other (4.9 %; 27/551). A high proportion of patients completed the 72-hour (92.9 %; 512/551) and 30-day (89.8 %; 495/551) study follow-up.


#

Cannulation details

Difficult CBD cannulation was reported for 150 /514 cases (29.2 %) using the single-use duodenoscope (10 of which were failed ERCPs), and in 15 cases in which a reusable duodenoscope was used after crossover.

Advanced cannulation techniques performed included: precut (access) papillotomy (n = 42; 7.6 %), double wire (n = 31; 5.6 %), pancreatic duct plastic stent (n = 24; 4.4 %), rendezvous with EUS (n = 3; one also had precut but ERCP still failed and went to rendezvous), cholangioscopy (n = 2), papillectomy (n = 1), transpancreatic sphincterotomy (n = 1), clip used to bring hidden papilla out from fold (n = 1), and accessory papilla cannulation (n = 1).


#

Ability to complete ERCP for intended indication

Among the 551 study cases, 529 (96.0 %, 95 %CI 94.0 %–97.5 %) were completed for the intended indication, including 503 (91.3 %, 95 %CI 88.6 %–93.5 %) using only the single-use duodenoscope, and 26 (4.7 %, 95 %CI 3.1 %–6.8 %) including crossover from the single-use to a reusable duodenoscope. Of the 22 cases that were not completed, 11 included crossover to a reusable duodenoscope ([Table 2]).

A total of 514 patients (93.3 %) underwent an ERCP procedure that used the single-use duodenoscope only ([Fig. 1]). Of these, 324 were ASGE grade 1 or 2 cases (98.8 % completed), 181 were ASGE grade 3 or 4 cases (97.8 % completed), and nine did not have an ASGE grade (66.7 % completed). The remaining 37 patients (6.7 %) underwent an ERCP procedure with crossover from single-use to reusable duodenoscope. Of these, 16 were ASGE grade 1 or 2 cases (68.8 % completed), 15 were ASGE grade 3 or 4 cases (73.3 % completed), and six did not have an ASGE grade (66.7 % completed).

For 524 patients (95.1 %), all of the intended maneuvers were completed. Over 20 different types of maneuvers were performed using the single-use duodenoscope, including sphincterotomy, papillectomy/ampullectomy, cannulation, mechanical lithotripsy, clearance of bile duct or pancreatic duct stones, biliary or pancreatic stent placement or removal, balloon dilation, cholangioscopy (n = 24), pancreatoscopy (n = 9), cytology brushing, biopsy, and others. Pancreatic maneuvers were performed in 128 patients (23.2 %).


#

Comparison of outcomes by ASGE grade of complexity

The median time of completion was significantly shorter (20.0 vs. 35.0 minutes; P < 0.001), and the mean number of cannulation attempts was significantly higher (3.0 vs. 2.4; P = 0.01) for ASGE grade 1 or 2 procedures compared with grade 3 or 4 procedures, while ERCP completion (97.4 % [331/340] vs. 95.9 % [188/196]; P = 0.44) and crossover rate (4.7 % [16/340] vs. 7.7 % [15/196]; P = 0.18) were similar ([Table 3]).

Table 3

Comparison of outcomes by ASGE grade of complexity (n = 536)[1].

Outcome

Grade 1 or 2
(n = 340)

Grade 3 or 4
(n = 196)

P value

ERCP completed, n (%)

331 (97.4)

188 (95.9)

0.44

Crossover rate, n (%)[2]

16 (4.7)

15 (7.7)

0.18

Overall satisfaction, median (range) [n]

8.0 (2.0−10.0) [340]

8.0 (1.0−10.0) [195]

0.68

Serious adverse event rate, n (%)

47 (13.8)

36 (18.4)

0.17

Procedural completion time, median (range) [n], minutes

20.0 (2.0−132.0) [338]

35.0 (2.0−157.0) [196]

< 0.001

Number of cannulation attempts, mean (SD)
(range) [n]

3.0 (4.3)
(1.0−28.0) [323]

2.4 (2.8)
(1.0−20.0) [179]

0.01

ASGE, American Society for Gastrointestinal Endoscopy; ERCP, endoscopic retrograde cholangiopancreatography.

1 15 cases (2.9 %) were duodenoscope-based procedures not included in the ASGE grading system.


2 Excludes one crossover case that did not have an ASGE grade.


The median rating of overall satisfaction with the performance of the single-use duodenoscope was 8.0 for cases with both ASGE grades 1 or 2 and ASGE grades 3 or 4 (P = 0.68).


#

Comparison of outcomes for expert versus less expert endoscopists

ERCPs performed by expert and less expert endoscopists were similar with respect to completion rate (96.3 % [444/461] vs. 94.4 % [85/90], respectively; P = 0.38), median procedural completion time (23.0 vs. 27.0 minutes; P = 0.15), mean number of cannulation attempts (2.7 vs. 3.1; P = 0.24), crossover rate (6.5 % vs. 7.8 %; P = 0.65), and proportion of cases with high complexity (36.4 % [168/461] vs. 31.1 % [28/90]) ([Table 4]). On a scale of 1 (worst) to 10 (best), the median rating of overall satisfaction with the performance of the single-use duodenoscope was 8.0 for expert endoscopists and 7.0 for less expert endoscopists (P < 0.001).

Table 4

Comparison of outcomes by endoscopist level of experience.

Outcome

Expert (n = 461)

Less expert (n = 90)

P value

ERCP completed, n (%)

444 (96.3)

85 (94.4)

0.38

Crossover rate, n (%)

30 (6.5)

7 (7.8)

0.65

Median overall satisfaction (range)

8.0 (0.0−10.0)

7.0 (2.0−10.0)

< 0.001

Serious adverse event rate, n (%)

77 (16.7)

8 (8.9)

0.08

Procedural completion time, median (range) [n]

23.0 (2.0−157.0) [458]

27.0 (3.0−100.0)

0.15

Number of cannulation attempts, mean (SD)
(range) [n]

2.7 (3.7)
(1.0−25.0) [428]

3.1 (4.3)
(1.0−28.0) [76]

0.24

ERCP, endoscopic retrograde cholangiopancreatography.


#

Comparison of outcomes by prior sphincterotomy

On comparison with cases with at least one prior sphincterotomy, cases with no prior sphincterotomy had a higher median total procedural time (25.0 vs. 21.0 minutes; P = 0.01), higher mean (SD) number of cannulation attempts (3.6 [4.7] vs. 1.6 [1.9]; P < 0.001), and a lower likelihood of having high complexity (P = 0.001). The baseline sphincterotomy status did not correlate with the crossover rate (P = 0.09), ERCP completion rate (P = 0.52), or overall satisfaction with the single-use duodenoscope (P = 0.36).


#

Related serious adverse events

Of the 551 enrolled patients, 43 (7.8 %, 95 %CI 5.7 %–10.4 %) experienced at least one SAE related to the device or procedure within a median (range) of 1 (0–28) days post-procedure ([Table 5]). There were 18 patients (3.3 %, 95 %CI 2.0 %–5.1 %) who developed pancreatitis (mild or moderate [n = 16], severe [n = 2]; post-ERCP [n = 17] reported 0–4 days after the ERCP, acute pancreatitis [n = 1]), six (1.1 %) who experienced pain, five (0.9 %) who had a GI bleed, three (0.5 %) who developed sepsis, two each (0.4 %) who developed cholangitis, cholecystitis, fever, or bacteremia, or had a perforation (0.4 %, 95 %CI 0.04 %–1.3 %; one gastric, one associated with sphincteroplasty during ERCP). The following additional SAEs occurred in one patient each: pneumoretroperitoneum, pneumonia, reintervention for biliary stones, exacerbation of COPD. SAE rates were similar for expert and less expert endoscopists (16.7 % [77/461] vs. 8.9 % [8/90], respectively; P = 0.08) and ASGE grade 1 or 2 cases vs. ASGE grade 3 or 4 levels of complexity (13.8 % [47/340] vs. 18.4 % [36/196], respectively; P = 0.17).

Table 5

Serious adverse events that occurred by 30 days after ERCP (n = 551 total).

Events, n[1]

Patients, n (%)

95 %CI

Any serious adverse event

46

43 (7.8)

5.7 %−10.4 %

Pancreatitis

18

18 (3.3)

1.9 %−5.1 %

  • PEP

17

17 (3.1)

1.8 %−4.9 %

  • Fatal acute pancreatitis

1

1 (0.2)

0.0 %−1.0 %

Pain

6

6 (1.1)

0.4 %−2.4 %

Gastrointestinal bleeding (fatal in one case)

5

5 (0.9)

0.3 %−2.1 %

Sepsis

3

3 (0.5)

0.1 %−1.6 %

Cholangitis

2

2 (0.4)

0.0 %−1.3 %

Cholecystitis

2

2 (0.4)

0.0 %−1.3 %

Fever

2

2 (0.4)

0.0 %−1.3 %

Bacteremia

2

2 (0.4)

0.0 %−1.3 %

Perforation (fatal in one case)

2

2 (0.4)

0.0 %−1.3 %

Pneumoretroperitoneum

1

1 (0.2)

0.0 %−1.0 %

Reintervention with ERCP under general anesthesia for biliary stone removal

1

1 (0.2)

0.0 %−1.0 %

Post-ERCP pneumonia

1

1 (0.2)

0.0 %−1.0 %

Exacerbation of COPD

1

1 (0.2)

0.0 %−1.0 %

ERCP, endoscopic retrograde cholangiopancreatography; PEP, post-ERCP pancreatitis; COPD, chronic obstructive pulmonary disease.

1 Each patient had one or more of the listed serious adverse events; rows are not mutually exclusive.


The overall rate of related SAEs that were fatal in our study was 0.5 % (3/551; 95 %CI 0.11 %–1.6 %). These included one case of acute pancreatitis occurring 16 days after ERCP that was fatal on day 70, one fatal hemorrhage from the papilla after balloon dilation, and one fatal perforation associated with sphincteroplasty.


#
#

Discussion

In this largest global study of a single-use duodenoscope, endoscopists with varying levels of experience had high procedural success in ERCPs, one-third of which were high complexity cases. The median performance ratings were high for this device. SAEs were in the expected range, based on the published safety data for reusable duodenoscopes. Consistent with two previously published clinical studies [9] [12], the current study supports the safe and effective performance of the first single-use duodenoscope among endoscopists with varying levels of experience in multiple countries.

In an effort to eliminate the potential for contamination or infection associated with ineffective reprocessing of reusable duodenoscopes, the FDA have cleared duodenoscopes with disposable components [19] [20] and two fully disposable duodenoscopes [13] [15]. Single-use duodenoscopes may be especially useful in certain challenging logistical settings, such as during evening and weekend hours, and in various emergent situations, including when procedures are done outside of the endoscopy department [21]. The disadvantages of single-use duodenoscopes include their current high cost compared with duodenoscopes that are reusable [22] or partially disposable (with disposable endpieces [23] [24]), and their potential adverse impacts on the environment, while recycling programs currently are not yet widely available [25]. Although a 2023 RCT found that duodenoscopes with a disposable elevator cap exhibited reduced contamination following high level disinfection compared with standard scope designs [23], the status of disposable endoscopic endpieces is uncertain after the FDA issued a warning letter to a duodenoscope manufacturer regarding its slow response to complaints about the distal end cover “dropping out” or cracking [26].

Ecological sustainability is an important issue for both single-use and reusable devices. For example, a 2023 publication focused on manufacturing-associated carbon dioxide emissions reported that performing ERCP with single-use duodenoscopes releases a carbon dioxide equivalent that is 24–47 times greater than with a reusable duodenoscope or a reusable duodenoscope with disposable endcaps [27]. This report made some assumptions that were not accurate for the EXALT duodenoscope and did not include selective use or recycling in the analysis. Furthermore, the report did not mention ethylene oxide, which is used for approximately 50 % of sterile devices sold in the USA and is regulated by the US Environmental Protection Agency owing to its carcinogenicity [28]. Improvements in recycling efforts, and control and mitigation of the chemicals used in reprocessing are necessary to optimize patient and employee safety as single-use and reusable duodenoscopes are both used in clinical practice.

The results from the current study are consistent with those from previous studies of the single-use duodenoscope [9] [10] [11] [12], but from a much larger study population. The device was successfully used for procedures of all ASGE grades of complexity, with a low (6.7 %) incidence of crossover to a reusable duodenoscope and high (96 %) completion rate by both expert and less expert operators. Study practices met the published ERCP quality indicators [29] [30] [31] for most items that had study documentation. Consistent with the ESGE quality indicators for ERCP [30], the rate of post-ERCP pancreatitis was < 10 % in our study. Although the precise role and place of single-use devices is the subject of ongoing discussion, the current study provides added evidence that there are no arguments against their use from the perspective of procedural outcome and success, which is a prerequisite for considering their use in clinical practice.

All ERCP-associated SAEs are concerning and continual efforts must be made to minimize them. We compared the safety results in the current study with the published safety data for procedures performed with reusable duodenoscopes. The largest analysis of ERCP safety available is from Andriulli et al. [32], who estimated the post-ERCP complication rates among 21 prospective studies published between January 1977 and May 2006. They reported an estimated ERCP-specific mortality rate of 0.33 % (55/16 855 patients; 95 %CI 0.24 %–0.42 %), which is comparable to the rate of fatal related SAEs in our study (0.5 %; 3/551). Andriulli’s estimated rates of perforation (0.60 %, 95 %CI 0.48 %–0.72 %), pancreatitis (585/16 855 patients; 3.47 %, 95 %CI 3.19 %–3.75 %), and infection (1.44 %; 95 %CI 1.26 %–1.62 %) were also similar to the rates in our study (0.4 % [2/551], 3.3 % [18/551], and 1.63 % [9/551; 95 %CI 0.75 %–3.08 %], respectively).

These findings suggest that, when used by a group of endoscopists with a predominantly expert level of experience, the single-use duodenoscope has a comparable safety profile to reusable duodenoscopes. The comparable rate of infection also shows that endogenous infection can still occur with when using a single-use duodenoscope, so the procedure-associated infection rate is not guaranteed to be lower than the rate for reusable duodenoscopes.

Our study has several strengths, limitations, and considerations for study interpretation. This was a multinational observational study including a large number of scheduled ERCP cases with a range of complexity performed by endoscopists at varying levels of experience, including 30-day safety follow-up. Limitations include the lack of randomization and the absence of a control group. Although patients were consecutively screened, the 68 % who were eligible to participate in this study may not be typical of patients at all endoscopy centers. Specific reasons why 10 screened patients were excluded for “concern over safety of subject if study duodenoscope was used for procedure’’ were not documented. Of note, several investigators participated in the development of the single-use duodenoscope [9] [14] and have also received research funding from the study sponsor and from manufacturers of reusable duodenoscopes. Finally, this study evaluated the efficacy and safety of a single-use duodenoscope but was not designed to address the associated costs or impact on reprocessing services or the environment.

In conclusion, academic endoscopists used the first marketed single-use duodenoscope to successfully complete a large number of scheduled ERCP procedures with a range of complexity in a diverse patient population. Consistent with past studies, the device showed good performance and an SAE rate comparable with the published estimates for reusable duodenoscopes.


#

Data sharing

The data, analytic methods, and study materials for this study may be made available to other researchers in accordance with the Boston Scientific Data Sharing Policy (https://www.bostonscientific.com/en-US/data-sharing-requests.html


#
#

Competing interests

M.J. Bruno has provided consultancy and had support for industry- and investigator-initiated studies for Boston Scientific, Cook Medical, and Pentax Medical, and support for investigator-initiated studies from Mylan, InterScope, and ChiRhoStim. T. Beyna has provided consultancy for Boston Scientific, Olympus, Medtronic, and Microtech Endoscopy. D. Carr-Locke receives royalties from Steris Corporation. P. Chahal has provided consultancy and is on the advisory council for Boston Scientific, and is on the advisory council for Medtronic. G. Costamagna has received grants from Cook Endoscopy, Olympus, and Boston Scientific and has been on advisory committees or review panels for Boston Scientific and Olympus. B. Devereaux has provided consultancy to Olympus, Boston Scientific, and Mylan. M. Giovannini has consultancy agreements with Pentax and Taewoong. C. Khor has provided consultancy for Boston Scientific, Erbe, and Fujifilm. J. Lau is a research consultant for Boston Scientific. G. May is a consultant for Olympus and is a speaker for/on the advisory board for Vantage Endoscopy, Fujifilm, Pentax, Medtronic, and Boston Scientific. V.R. Muthusamy is a consultant for and has received research support from Boston Scientific, is a consultant for Medtronic, and on the advisory board for Motus GI and Endogastric Solutions; he holds stock options in CapsoVision; he was previously a consultant to Medivators and Interpace Diagnostics. S. Patel is a consultant for Boston Scientific, Olympus, Conmed, and AbbVie. B.T. Petersen is an investigator for Boston Scientific and Ambu and has provided consultancy for Olympus and Pentax. D.K. Pleskow has provided consultancy to Boston Scientific, Fuji, and Olympus. I. Raijman is a speaker for/on the advisory board of Boston Scientific, Microtech, and Pentax, has provided consultancy to Olympus, Conmed, Endosound, and Cryo, and is co-owner of EndoRx. A. Repici has received consultancy and speaker's fees from Boston Scientific Corporation. A.S. Ross has provided consultancy to Boston Scientific. D.V. Sejpal has received consultancy fees and research funding frrom Boston Scientific and Olympus. S. Sherman has provided consultancy for Olympus, Boston Scientific, and Cook. U.D. Siddiqui has received research funding from Boston Scientific, consultancy and speaker's fees from Boston Scientific, Cook, Olympus, Medtronic, and ConMed, and speaker's fees from Pinnacle Biologics and Ovesco. J.A. Peetermans and M.J. Rousseau are full-time employees of Boston Scientific. A. Slivka has received research support for IRB-approved studies from Boston Scientific and Olympus. M. Goenka, N. Reddy, and C. Zaidy declare that they have no conflict of interest.

Acknowledgments

Members of the EXALT Single-use Duodenoscope Study Group who participated in this study are: Jennifer S. Chennat and Adam Slivka at the University of Pittsburgh Medical Center (Pittsburgh, Pennsylvania, USA); Divyesh V. Sejpal at Dignity/CommonSpirit Health, Creighton School of Medicine (Phoenix, Arizona, USA); Calvin Lee at North Shore University Hospital (Manhasset, New York, USA); Shayan Irani, Rajesh Krishnamoorthi, Michael Larsen, and Andrew S. Ross at Virginia Mason Medical Center (Seattle, Washington, USA); V. Raman Muthusamy and Adarsh Thaker at University of California, Los Angeles (Los Angeles, California, USA); Vinay Chandrasekhara, Ryan Law, Michael Levy, John A. Martin, Bret T. Petersen, and Andrew C. Storm at Mayo Clinic (Rochester, Minnesota, USA); Tyler M. Berzin, Douglas K. Pleskow, and Mandeep S. Sawhney at Beth Israel Deaconess Medical Center (Boston, Massachusetts, USA); Marco J. Bruno and P.J.F. de Jonge at Erasmus Medical Center (Rotterdam, The Netherlands); Gary May and Jeffrey Mosco at St. Michael’s Hospital (Alberta, Canada); Alessandro Fugazza and Anthony Repici at Instituto Clinical Humanitas (Milan, Italy); Prabhleen Chahal and John Vargo at Cleveland Clinic (Cleveland, Ohio, USA); Benjamin Bick, Mark A. Gromski, Aditya Gutta, and Stuart Sherman at Indiana University Medical Center (Indianapolis, Indiana, USA); Danny Issa, Stephen Kim, V. Raman Muthusamy, Alireza Sedarat, and Adarsh Thaker at Ronald Reagan UCLA Medical Center (Los Angeles, California, USA); Torsten Beyna and Horst Neuhaus at Evangelisches Krankenhaus Dusseldorf (Dusseldorf, Germany); Sundeep Lakhtakia, Mohan Ramchandani, and D. Nageshwar Reddy at Asian Institute of Gastroenterology (Hyderabad, India); Uzma D. Siddiqui at University of Chicago Hospital (Chicago, Illinois, USA); Fabrice Caillol and Marc Giovannini at Institut Paoli-Calmettes (Marseille, France); James Y.W. Lau at Prince of Wales Hospital (Hong Kong, China); David Leslie Carr-Locke, Srihari Mahadev, Kartik Sampath, and Reem Sharaiha at New York Presbyterian Hospital & Weill Cornell Medicine (New York City, New York, USA); Mark Appleyard, Benedict Devereaux, and Florian Grimpen at Royal Brisbane and Women's Hospital (Brisbane, Australia); Mahar Kumar Goenka at Apollo Multispecialty Hospital (Kolkata, India); Christopher Khor and Damien Tan at Singapore General Hospital (Singapore); Moses A. Balabyeki, Nkomba Christophe Kalenga, Oleh Matsevych, and Christopher Charles Michael Ziady at Dr. George Mukhari Provincial (Pretoria, South Africa); Sandeep N. Patel, Isaac Raijman, Shreyas Saligram, and Hari Sayana at Texas International Endoscopy Center (Houston, Texas, USA); Joyce A. Peetermans, Matthew J. Rousseau, Steve Fodem, and Jeff Insull at Boston Scientific Corporation (Marlborough, Massachusetts, USA).

The authors acknowledge Boston Scientific employees Steve Fodem, MS for his project management, Victoria Lazzari for her data management assistance, and Margaret Gourlay, MD, MPH for her assistance in manuscript preparation.

* Deceased; this paper is dedicated to our friend and eminent colleague gastroenterologist Dr. Stuart Sherman.



Corresponding author

Marco J. Bruno, MD, PhD
Department of Gastroenterology and Hepatology
Erasmus University Medical Center
Rotterdam
The Netherlands   

Publication History

Received: 13 March 2023

Accepted after revision: 14 July 2023

Accepted Manuscript online:
18 July 2023

Article published online:
20 September 2023

© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany


Zoom Image
Fig. 1 Patient flow through the study.