Rofo 2022; 194(04): 400-408
DOI: 10.1055/a-1665-6988
Quality/Quality Assurance

Handling of Incidents in the Clinical Application of Ionizing Radiation in Diagnostic and Interventional Radiology – a Multi-center Study

Article in several languages: English | deutsch
Birgit Sabine Müller
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
,
Julian Singer
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
,
Georg Stamm
2   Institute for Diagnostic and Interventional Radiology, University Medical Center Gottingen, Gottingen, Germany
,
Lukas Pirl
3   Institute for Diagnostic Radiology and Nuclear medicine, Braunschweig Municipal Hospital, Braunschweig, Germany
,
Markus Borowski
3   Institute for Diagnostic Radiology and Nuclear medicine, Braunschweig Municipal Hospital, Braunschweig, Germany
,
Thomas Hertlein
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
,
Eugenia Rerich
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
4   Medizinphysik-Experten für sonstige Einrichtungen, Charite University Hospital Berlin, Germany
,
Sebastian Trinkl
5   External and Internal Dosimetry, Biokinetics, Federal Office for Radiation Protection Neuherberg, Germany
,
Michael Wucherer
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
,
Josefin Ammon
1   Institute of Medical Physics, Klinikum Nurnberg, Paracelsus Medical University, Nurnberg, Germany
› Author Affiliations
Supported by: Bundesamt für Strahlenschutz 3617S42333
 

Abstract

Purpose According to the German legislation and regulation of radiation protection, i. e. Strahlenschutzgesetz und Strahlenschutzverordnung (StrlSchG and StrlSchV), which came into force on 31st December 2018, significant unintended or accidential exposures have to be reported to the competent authority. Furthermore, facilities have to implement measures to prevent and to recognize unintended or accidental exposures as well as to reduce their consequences. We developed a process to register incidents and tested its application in the framework of a multi-center-study.

Materials and Methods Over a period of 12 months, 16 institutions for x-ray diagnostics and interventions, documented their incidents. Documentation of the incidents was conducted using the software CIRSrad, which was developed, released for testing purposes and implemented in the frame of the study. Reporting criteria of the project were selected to be more sensitive compared to the legal criteria specifying “significant incidents”. Reported incidents were evaluated after four, eight, and twelve months. Finally, all participating institutions were interviewed on their experience with the software and the correlated effort.

Results The rate of reported incidents varied between institutions as well as between modalities. The majority of incidents were reported in conventional x-ray imaging, followed by computed tomography and therapeutic interventions. Incidents were attributed to several different causes, amongst others to the technical setup and patient positioning (19 %) and patient movement or insufficient cooperativeness of the patient (18 %). Most incidents were below corresponding thresholds stated in StrlSchV. The workload for documenting the incidents was rated as appropriate.

Conclusion It is possible to monitor and handle incidents complient with legal requirements with an acceptable effort. The number of reported incidents can be increased by frequent trainings on the detection and the processing workflow, on the software and legal regulation as well as by a transparent error handling within the institution.

Key Points:

  • The software CIRSrad was developed to enable the present study and as prototype platform for a future radiological incident management system.

  • 586 exceedances of thresholds were recorded by 16 facilities in a period of one year.

  • Frequent trainings of all users increase the number of reported cases.

Citation Format

  • Müller BS, Singer J, Stamm G et al. Handling of Incidents in the Clinical Application of Ionizing Radiation in Diagnostic and Interventional Radiology – a Multi-center Study. Fortschr Röntgenstr 2022; 194: 400 – 408


#

Introduction

The German Radiation Protection Act of June 27, 2017 which implements the European Directive 2013/59/Euratom in national law, and the new Radiation Protection Ordinance of November 29, 2018, bring “significant incidents” in the use of ionizing radiation in medicine into focus [1]. Paragraf 90 of the act requires both the initiation of measures for the detection, elimination and prevention of reportable incidents as well as the reporting of significant incidents to the supervisory authority. The Radiation Protection Ordinance specifies the duties of those responsible for radiation protection, the tasks and powers of the authorities, and the definition of reportable incidents. According to paragraph 108 of the ordinance, a significant occurrence must be reported to the competent authority, especially if it meets criteria of Annex 14 and 15.

The basic prerequisite for this is the recognition of an incident. This requires constant monitoring of medical-related radiation exposures and addressing any unintended exposures. This is followed by complete documentation and evaluation of the incident with the decision whether there is a reporting obligation according to paragraph 108 of the ordinance. Independent of a report, a systematic investigation as well as the initiation of appropriate countermeasures are necessary to avoid similar incidents in the future.

Critical Incident Reporting Systems (CIRS) are used to record unintentional exposures and near-exposures. So far, CIRS are mainly known in the field of general clinical processes and radiation therapy [2] [3] [4] [5] [6] [7] [8] [9]. Very little has been published on this topic in the area of diagnostic radiology. Possible categorizations of incidents as well as draft criteria based on consequential damages, among others, are available [10] [11] [13] [14] [15]. In this regard, all relevant publications are of the opinion that a radiological CIRS is a useful and valuable quality management tool that can be used to detect gaps in safety measures and initiate necessary changes in process procedures.

Sendlhofer et al. have summarized requirements for a functioning CIRS [16]. For users, sufficient information about the CIRS, the guarantee of anonymity, confidentiality with accompanying freedom from sanctions (“safety culture” instead of “culture of blame”) and a transparent feedback mechanism play an essential role. Current CIRS for general errors in the daily routine of a practice or hospital are mostly web-based, anonymous systems that allow users to make their own comments and provide feedback in the form of regular reports [17] [18] [19]. However, these products and the underlying workflows are of limited use for the special issues and legal requirements in X-ray diagnostics. They also usually do not contain a process step for evaluation by radiation protection officers (or responsible persons) and for reporting to the authorities.

The CIRSrad [22] software was developed for incidents in radiology and nuclear medicine and tested as part of a multicenter study in the course of the departmental research projects initiated by the German Federal Office for Radiation Protection (BfS) “Operational handling of (significant) incidents in medical applications of ionizing radiation in X-ray diagnostics and interventional radiology and practical testing of the reporting criteria” [20] and “Operational handling of (significant) incidents in medical applications of radioactive substances in nuclear medicine and practical testing of the reporting criteria” [21]. Intervention thresholds were defined for the clinical trial, most of which were below the legal criteria of reportable incidents according to the Radiation Protection Ordinance in force since December 31, 2018. In this paper, the term “incidents” describes violations of the intervention thresholds developed here, rather than the legally-defined “significant incidents”.

The aim of this study was the realization of software for recording and processing of incidents and its application in order to obtain initial experiences with incident types as well as the scope of recording in clinical practice.


#

Materials and Methods

Study participants and reporting criteria

Sixteen institutions of varying sizes, including five university hospitals, five non-academic hospitals and six practices (three radiology, one cardiology, two dental), were recruited for project participation. The facilities differed in their technical equipment. Thus, a radiology practice with only computed tomography (CT) scanners participated, as well as clinics possessing multiple systems offering all modalities. The number of examinations and therapies performed per modality ranged from 0 to 55 000 per year for computed tomography scans, from 0 to 130 000 for projection radiography and mammography, and from 0 to 5500 for interventions, depending on the institution. For evaluation purposes, facility sizes were divided into the following categories: University hospitals and large clinics (A), small hospitals and large practices (B), and individual, stand-alone practices (C).

The lower-threshold reporting criteria defined for the project, the intervention thresholds, included more incidents than the legally prescribed criteria of a reportable incident according to the Radiation Protection Ordinance (see [Table 1]). Every repeated exposure as well as confusion of body parts and patients was reportable, regardless of the applied dose. Likewise, projection radiographs and mammograms were included in the criteria.

Table 1

Reporting criteria for intervention threshold violations for all institutions, with differences in the respective criteria for increased dose with MG: Mammography, TI: Therapeutic Intervention, DI: Diagnostic Intervention, PR: Projection Radiography, DVT: Digital Volume Tomography, CT: Computed Tomography, DTDI: Computed Tomography Dose Index, DAP: Dose Area Product, DRV Diagnostic Reference Value.

CT

DI

TI

DVT

PR/MG

Excess dose (adult)

CTDIhead > 100 mGy,

CTDIbody > 50 mGy

DAP > 15 000 cGycm2

DAP > 50 000 cGycm2

DAP >> in-house reference value

excess dose (child)

CTDI > doubled DRV

DAP > doubled DRV

DAP > doubled DRV

repeat examination

x

x

x

x

x

wrong patient

x

x

x

x

x

incorrect examination/incorrect performance

x

x

x

x

x

device error/failure

x

x

x

x

x

deterministic skin damage

x

x

x

x

x

unintended prenatal exposure

x

x

x

x

x

dose > 1 mSv to support staff

x

x

x

x

x

contrast medium extravasation

x

x

x

x

x

Facilities recorded incidents exceeding the intervention threshold over a one-year period in 2018 and 2019.


#
#

Recording intervention threshold violations

The CIRSrad software developed to record incidents exceeding the intervention threshold [22] replicates the entire process of facility-wide standardized recording, including assessment and documentation of a significant incident relevant to this project.

The user interface of CIRSrad enables the direct entry of an incident or violation of an intervention threshold via drop-down lists for the modality and, based on this, possible types of incidents, such as wrong patient, excess dose, etc. (see [Fig. 1]). In addition to recording radiation protection-relevant incidents, a separate tab can also be used to document “other incidents” that are not significant in terms of radiation protection, but are important for reasons of general quality management. The administrative component of the software contains the corresponding work steps to evaluate and, if necessary, to report a significant incident to the responsible authority. The average initial input time is about two minutes (specifying the modality, incident type and access number), the time required for the subsequent steps depends primarily on the type of incident. The initial mandatory fields (highlighted in orange in [Fig. 1]) could be customized by the respective institutions. The setup of the software was, with uncomplicated configuration, one hour per separately entered setup.

Zoom Image
Fig. 1 German user interface for entering an incident or intervention threshold violation in CIRSrad [22]. First, the intervention threshold violation in this project is distinguished by modality and then by incident type.

Evaluation

After recording periods of four, eight and twelve months, the data collected from the study participants were analyzed by the study management. Intervention threshold violations were assessed according to institution, modality, incident type and cause. Finally, an anonymous online questionnaire was conducted among all participating institutions, which particularly surveyed the effort and ergonomics of processing reports with CIRSrad.


#
#

Results

The one-year recording period in 2018 and 2019 identified 586 intervention threshold violations, including 581 meeting the criteria of this research project, and five near-violations. Near-intervention threshold violations describe almost-occurring violations that were discovered outside of the quality assurance measures. In the first period 269, in the second period 183 and in the third period 134 intervention threshold violations were registered. The facilities differed in the number of reported incidents (see [Fig. 2]). Few institutions reported continuously during all time periods. Three institutions of different sizes recorded no incidents over the entire period of the survey.

Zoom Image
Fig. 2 Number of violations including five near-violations reported by the various participating Institutions. Institutions are categorized by size as A (university hospitals and large clinics), B (small hospitals and large practices), and C (individual practices). Some of the participating institutions reported no incidents.

Most violations (272 reports) were recorded in the area of projection radiography, followed by computed tomography (CT) (147) and therapeutic intervention (104). Broken down by incident type, repeat projection radiographs (224) formed the largest group of reported threshold exceedances (see [Fig. 3]), followed by excess dose exposures during therapeutic interventions (99) and repeated CT examinations (59). No excess doses related to radiation exposure of support staff were reported.

Zoom Image
Fig. 3 Number of intervention threshold violations reported, excluding near-violations, by defined incident; with MG: Mammography, TI: Therapeutic Intervention, DI: Diagnostic Intervention, PR: Projection Radiography, DVT: Digital Volume Tomography, CT: Computed Tomography.

Of the reported threshold violations, 19 % were due to setup technique, 18 % were caused by patient movement or lack of cooperation, and 14 % resulted from a complex treatment modality. No cause was mentioned in 20 % of reports.

In the voluntary, anonymous online survey, the effort required to record threshold breaches using CIRSrad was rated as appropriate. Of each participant group, 33 % estimated the proportion of threshold violations recorded to be between 76 % and 100 % and between 51 % and 75 %, respectively (see [Fig. 4]). A recording rate of 0 to 25 % and 26 % to 50 % was assumed by 17 % of the survey participants. The reasons given for the failure to record exceedances included “concerns about the consequences” and “lack of time” or “additional workload”.

Zoom Image
Fig. 4 Results of the online survey to assess the recorded proportion of intervention threshold violations in CIRSrad. Wording of the questionnaire: “Please estimate the proportion of intervention threshold violations occurring at your institution that were actually recorded in CIRS during the study period”.

A guideline [23] was developed as part of the research project for the systematic treatment of threshold violations [20] [21] which is available as a free download for interested parties. This can be consulted to plan the facility-specific process starting with detection, through recording and documentation, to assessment and further handling of incidents. The guideline covers processes for institutions of different sizes and for orientation, lists possible systems to be used as well as the groups of personnel involved.


#

Discussion

In the course of the research project 586 incidents were registered. The institutions differed significantly in the number of reported incidents. The size of the institutions with respect to equipment and number of examinations performed per year, as well as the number of admissions per modality, varied greatly among the institutions. No correlation could be established between facility size and the number of reported threshold violations. Likewise, no correlation between institution size and number of violations could be established.

Given the study experience and the one year recording period, it is unlikely that there were no occurrences of threshold violations, especially at larger institutions. For example, an eight-week recording study of faulty images in projection radiography showed that, averaged over all examination regions, approx. 7 % of the acquired images were faulty [24]. Similarly, it can be assumed that the actual number of threshold violations exceeds the number of recorded incidents, which is also indicated by the data of the estimated proportion of recorded violations in the online survey. The number of reports, which varies and is not correlated with the size of the institution, illustrates the importance of the reporting person. Whether an accidental exposure or near exposure is acquired in the system is primarily personnel-dependent.

Overall, reports decreased over the period of record. A clear decrease was particularly evident in the repeat exposures of the X-ray projection images. Here, a correlation with the entry into force of the Radiation Protection Ordinance during the second recording period is conceivable. In contrast to the study criteria, projection images are explicitly excluded from the reportable significant incidents in this ordinance (Annex 14), which may have led to misunderstandings among participants between project specifications and legal requirements. In contrast, a significant decrease in reports in the area of therapeutic interventions could be due to successful reprocessing from previous reports.

In the course of the study, both the Radiation Protection Act and Radiation Protection Ordinance became effective using criteria other than those specified in the study. The study was deliberately continued with the largely lower-threshold criteria originally developed. The difference is significant with regard to the projection images, which are not considered in Annex 14 of the Radiation Protection Ordinance, and within the study represented a significant proportion of all threshold violations. Similarly, the study criteria included any type of repeat image and confusion of body parts and mistaken individuals, regardless of the dose applied. Intention of the lower thresholds was to record clinically “less serious” incidents as well, in order on the one hand, to reduce the likelihood of a recurrence of the incident through targeted reprocessing as well as to fundamentally promote a more transparent error policy. A reduction in the frequency of less serious incidents will also have a long-term effect on the frequency of serious incidents [25]. Against this background, the Radiation Protection Ordinance also calls for the implementation of a local quality management system to record and process less serious incidents.

Over the entire period, the project management is aware of the reporting of four significant incidents according to the criteria of the ordinance by all participating facilities to the respective competent authorities. Of these, three of the four incidents were subsequently determined by agency assessment to be not significant and non-reportable. Regarding the process between operators and authorities, it would be desirable from a clinical perspective to promote more transparency and exchange, as well as a detailed explanation of the background of why an incident would not be considered significant. The process between regulatory agencies and clinical facilities, analogous to clinical processes, has yet to be established.

In the course of the study and the discussion of reporting criteria, as well as in conversations with various groups of study participants, it became clear that the currently defined legal thresholds cause difficulties in some circumstances and create ambiguity and a lack of understanding. Thus, any mistaking of persons and body parts, in CT as well as in fluoroscopy, appears relevant only when the legal dose of the corresponding imaging category is exceeded. In light of the fact that conventional projection radiography is not included in the regulatory criteria, the question arises as to whether the inadvertent failure to switch off a system and the resulting high exposures do not constitute an incident. Diagnostic interventions involving DAP > 20 000 cGycm2 also raise the question of the extent to which a dose was intentional and whether there was a particular clinical challenge and justification, or the extent to which clinical protocols need revision. Likewise, the calculation of the moving average of different dose parameters is currently only reasonably possible via dose management systems (DMS). Not all commercially-available dose management systems contain this function. As a result of the online survey, some participants commented that it was “somewhat unclear to them when an exposure is unintentional” and others stated that there are “too many different conditions, some of which are unintelligibly or misleadingly interdependent of one another”. It should also be noted that the reporting thresholds from Annex 14 and 15 of the Radiation Protection Ordinance are to be understood “in particular” as significant incidents. Accordingly, even if an incident is not explicitly listed in Annexes 14 and 15, it may well be considered an incident.

Dose management systems provide a useful basis for the detection of threshold violations or incidents and, if well integrated, allow filtering according to certain criteria, so that automated monitoring of dose values is possible. Increasingly, larger institutions are planning to implement a DMS, but especially for smaller practices it must be assumed that the detection of an incident must also be feasible with a simple DMS or other evaluation method. It is imperative to note that a DMS with dose threshold violations represents only a small fraction of all occurring incidents. For further information on the use of DMS and related problems, please refer to the German Federal Office for Radiation Protection research project “Evaluation of the use of dose management systems for the optimization of X-ray applications in various X-ray facilities” [26], the guide developed from it [27] as well as the publication by Loose et al. [28].

Dose management systems can facilitate incident handling and analysis. However, the personnel operating the equipment are crucial for the reliable recording and documentation of incidents. Reporting behavior and thus also the reduction of critical incidents can only function with attentive and motivated employees as well as through an established “error culture”. Regular training, joint evaluations and discussions within the staff team, for example, during the annual briefings in accordance with paragraph 63 of the Radiation Protection Ordinance, ensure that the relevance of the topic is made clear. Especially in the phase of introducing reporting systems and incident disclosure, it is important to provide continuous training on the reporting process and to repeat the training regularly. Regular training and the thematization of incidents are of great importance with respect to complete and fully meaningful reports. A prerequisite for the functioning of a reporting system is an open approach to errors within the facilities as well as the constant awareness of the topic. In addition to the legally required reporting of significant incidents, the added value for the facility's own quality and risk management should also be mentioned. For example, regularly occurring deviations can be detected and corrected by analyzing the internally reported incidents, medical physics experts can use the documented incidents to optimize radiation protection, and the database of incidents can provide tips on activities and procedures.


#

Conclusions

It is possible to record incidents in a way that is not very time-consuming, meets the requirements of radiation protection law, and makes sense in terms of risk management. The software CIRSrad can record, analyze and follow up on incidents. During the one-year observation period, 586 intervention threshold violations were recorded by 16 facilities. The number of reports can be increased by regular training of all those involved in recognizing and handling incidents, as well as openly dealing with errors in the operation.

A guideline for handling reportable incidents was prepared and published by the Federal Office for Radiation Protection [23]; this was based on experience gained, together with the research project “Operational handling of significant incidents in medical applications of radioactive substances in nuclear medicine”. After completion of the project, the CIRSrad software was made available for free download for interested institutions [22].

Clinical Relevance of the Study
  • The Radiation Protection Act requires that incidents be addressed and significant incidents in the use of ionizing radiation be reported.

  • As part of the study, incident recording software was developed and its use was tested in the clinical practice of various institutions.

  • As part of the departmental research project, a workflow for recording, assessing, processing and reporting incidents was developed and its application tested in the daily clinical routine of various institutions. The guidance provided can assist in the development of facility-specific processes related to incident reporting.

Funding

Bundesamt für Strahlenschutz (3617S42333)


#
#
#

Conflict of Interest

The authors declare that they have no conflict of interest.

Acknowledgments

We would like to thank the German Federal Office for Radiation Protection for initiating the departmental research project 3617S42 333 and supporting the study “Operational handling of (significant) incidents in medical applications of ionizing radiation in X-ray diagnostics and interventional radiology and practical testing of the reporting criteria”.


Correspondence

Dr. Josefin Ammon
Institute of Medical Physics, Klinikum Nurnberg Nord
Prof.-Ernst-Nathan-Str. 1
90419 Nurnberg
Germany   
Phone: +49/9 11/3 98 11 47 81   

Publication History

Received: 28 March 2021

Accepted: 30 September 2021

Article published online:
21 December 2021

© 2021. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany


Zoom Image
Fig. 1 German user interface for entering an incident or intervention threshold violation in CIRSrad [22]. First, the intervention threshold violation in this project is distinguished by modality and then by incident type.
Zoom Image
Fig. 2 Number of violations including five near-violations reported by the various participating Institutions. Institutions are categorized by size as A (university hospitals and large clinics), B (small hospitals and large practices), and C (individual practices). Some of the participating institutions reported no incidents.
Zoom Image
Fig. 3 Number of intervention threshold violations reported, excluding near-violations, by defined incident; with MG: Mammography, TI: Therapeutic Intervention, DI: Diagnostic Intervention, PR: Projection Radiography, DVT: Digital Volume Tomography, CT: Computed Tomography.
Zoom Image
Fig. 4 Results of the online survey to assess the recorded proportion of intervention threshold violations in CIRSrad. Wording of the questionnaire: “Please estimate the proportion of intervention threshold violations occurring at your institution that were actually recorded in CIRS during the study period”.
Zoom Image
Abb. 1 Benutzeroberfläche zur Eingabe eines Vorkommnisses bzw. einer Interventionsschwellenüberschreitung (IntS-Überschreitung) in CIRSrad [22]. Zunächst wird die IntS-Überschreitung in diesem Projekt nach Modalität und darauf basierend nach der Ereignisart unterschieden.
Zoom Image
Abb. 2 Anzahl gemeldeter IntS-Überschreitungen inklusive 5 Beinahe-IntS-Überschreitungen der verschiedenen teilnehmenden Einrichtungen. Die Einrichtungen werden nach Größe in Kategorie A (Universitätskliniken und große Kliniken), B (kleine Krankenhäuser und große Praxen) und C (einzelne Praxen) unterschieden. Von einigen der teilnehmenden Einrichtungen wurde kein Vorkommnis gemeldet.
Zoom Image
Abb. 3 Anzahl erfolgter Meldungen von Interventionsschwellenüberschreitungen (IntS-Überschreitungen), exklusive Beinahe-IntS-Überschreitungen, nach definiertem Ereignis; mit MG: Mammografie, TI: Therapeutische Intervention, DI: Diagnostische Intervention, Rö: Projektionsradiografie, DVT: Digitale Volumentomografie, CT: Computertomografie.
Zoom Image
Abb. 4 Ergebnis der Online-Umfrage zur Einschätzung des erfassten Anteils an Interventionsschwellenüberschreitungen (IntS-Überschreitungen) in CIRSrad. Wortlaut der Umfrage: „Bitte schätzen Sie, welcher Anteil der aufgetretenen IntS-Überschreitungen in Ihrer Institution während der Studiendauer tatsächlich in CIRS erfasst wurde.“