The first capsule endoscopy system available in the market was named M2A (M2A Capsule, GivenImaging Ltd, Yokneam, Israel) [1]. This acronym reveals the philosophy behind this technology: the implementation of a miniaturized system allowing to endoscopically evaluate the entire digestive system, ideally from the mouth to the anus (which is what M2A stands for).

The technical barriers that had initially prevented the exploration of the entire surface of the digestive system have been removed over time through the development of batteries with longer life, the optimization of power consumption (e. g., through the adaptive frame rate) and the increase in the number of cameras and in the angle of view (to cover the entire mucosal surface of wider lumen organs). The study by Vuik et al. [2] in the current issue of Endoscopy International Open, as well as several other publications, [3] confirm that the endoscopic study of the entire surface of the digestive system appears to be now technically possible.

However, does “technical feasible” also mean “clinically reliable”? Although the very quick small bowel transit time reported in the present study (median small bowel transit time: 47’) might negatively impact on small bowel capsule diagnostic yield, the key role of capsule endoscopy in the evaluation of the small bowel mucosa has been confirmed in several studies, and current guidelines endorse the use of this device as first-line examination in many clinical scenarios [4]. The clinical reliability concerns, therefore, are related mostly to the evaluation of the colonic, esophageal, and gastric mucosa, which would be the potential added value of panenteroscopy.

In standard colonoscopy, there is a well-documented correlation between the degree of cleanliness and diagnostic performance [5]. This is likely to have even greater relevance in the field of wireless endoscopy, where the capsule is purely passive (e. g., it cannot wash or remove debris) and bowel preparation also plays a key role in promoting device propulsion. In the paper from Vuik et al. [2] the rate of patients with adequate colon cleansing (76.6 %) was below the ≥ 90 % minimum standard recommended for optical colonoscopy [6], although involved subjects were healthy and highly compliant with the preparation schedule (93.6 % of them took the entire bowel preparation). Although this may reflect the specific preparation schedule adopted in the study, a recent systematic review [3] showed that the rate of patients with adequate colon cleansing was highly variable, but ≥ 90 % in only 54 % of studies, even when different schedules were used. Furthermore, although no formal standard quality measure exists, the rate of patients with adequate cleansing of the whole gastrointestinal tract (52.8 %) is far from being acceptable. Finally, as far as the completion rate for colon evaluation is concerned, the 51 % rate reported by Vuik et al. [2] is similar to those reported in other publications [7] [8] [9] and far below the recommended threshold of 90 % [6].

Although relevant gastric findings, missed by previous traditional gastroscopy, were reported in patients undergoing small bowel capsule endoscopy [10], the parameter for assessing the efficacy of panenteroscopy in evaluation of gastric mucosa (observation of more than 90 % of the mucosa) appears to be arbitrary, contrived, and very difficult to measure. In the absence of a traditional examination as comparator, no clinically relevant conclusions can be drawn about the diagnostic yield in the stomach of such a non-steerable capsule. In addition, because the capsule was ingested without a dedicated protocol for esophageal inspection [11], the Z line was identified in fewer than 50 % of the subjects included in the study, underscoring that capsule endoscopy, when administered as a panenteroscopy tool, cannot represent a viable alternative to regular gastroscopy.

Patient acceptability has always been emphasized as one of the major advantages of capsule endoscopy over more invasive and less tolerated procedures, such as gastroscopy and colonoscopy. However, Vuik et al. [2] demonstrated that moving from theory (pre-procedural questionnaire) to practice (post-procedural questionnaire), patient expectations are somewhat disregarded. In line with these results, in the TOPAZ study [12] more subjects who underwent both procedures under evaluation unexpectedly preferred traditional colonoscopy over colon capsule endoscopy. Once again, the major determinant behind this preference was the burden related to colon capsule endoscopy bowel preparation.

Taking into account all these issues, one might wonder if panenteric capsule endoscopy evaluation is a journey worth undertaking and a goal worth pursuing. Taken individually, each traditional endoscopic examination for evaluation of esophagus, stomach and colon seems to be more reliable than panenteroscopy. However, the latter offers the unique opportunity to combine evaluation of several segments of the digestive tract at the same time, in a single examination. This becomes particularly relevant when the “timing of the procedure” or “disease extent” are key factors. Indeed, there is growing evidence supporting the use of a panenteroscopic approach in patients with digestive bleeding, in whom an upper digestive tract lesion has been excluded [13] [14]. Initial data suggest that in these patients, a timely panenteric evaluation has relevant clinical (increased rate of diagnosis, crucial information for therapeutic work-up) as well as organizational/economic consequences (reduction of the number of unnecessary examinations and hospitalization time). Similarly, for staging and surveillance in patients with Crohn's disease, especially in the pediatric setting, panenteroscopy appears to be a reliable tool for assessing disease mucosal activity and extent, compared to more invasive methods [15] [16] [17].

In light of all the pros and cons, with current technology and organizational/practical issues (e. g. the preparation schedule), the window of opportunity for panenteroscopy in everyday clinical practice currently is narrow. However, the most limiting factors appear amenable to significant improvements, and the introduction of automated image analysis systems (e. g. artificial intelligence-based software) [18], which could make the procedure even easier and more reliable, is expected soon. These improvements are needed before implementing use of the capsule for whole-intestine mucosal assessment in clinical practice.


Competing interests

Dr. Rondonotti received speaker honoraria from Fujifilm Co, consultancy agreement with Medtronic Co. Dr. Pennazio received speaker honoraria from Olympus Co, Medtronic Co.

Acknowledgements

The authors thank Dr. Franco Radaelli for support during the writing and editing process.


Corresponding author

Emanuele Rondonotti MD, PhD
Gastroenterology Unit, Valduce Hospital
Via Dante 10, 22100, Como
Italy   
Fax: +0039031324150   

Publication History

Article published online:
14 December 2021

© 2021. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany