Introduction and background
Introduction and background
Incidence of obesity is increasing rapidly worldwide. Multiple studies have shown
that it is a major risk factor for cardiovascular morbidity and mortality [1]. Despite the success of bariatric surgery, obese patients are at increased risk
of anesthetic adverse events (AEs) as preoperative comorbidities are relatively common
among this group of patients [2]. Because of this, a minimally-invasive alternative, the intragastric balloon (IGB),
has gained popularity.
Recent studies have demonstrated the safety and efficacy of IGBs [3]. However, most IGBs require liquid to be introduced into the balloon via a catheter,
hence the need for catheter and/or endoscopic assistance for insertion and/or removal
of the balloon. We believe that an ideal IGB should be inflated and deflated without
the need for endoscopic or catheter assistance. We have developed an IMI-BC that can
be inflated via external magnetic control without onboard electronics. We performed
successful proof-of-concept studies on the IMI-BC using a virtual stomach EMS Trainer
and fresh porcine stomachs [4].
The aims of the current study were twofold: (1) to demonstrate efficacy of the original
wireless mechatronic capsule IGB prototype in inducing weight loss in a surviving
animal; and (2) to examine the feasibility of inflating the balloon of the latest
ingestible microcapsule IGB prototype using an external magnet in an animal and a
human volunteer.
Methods
Devices
We initially designed a wireless mechatronic capsule (Ø 19 × 50 mm) that consisted
of an enclosure and a balloon that stored acid and base respectively (U.S. Patent
9375554, 2016) [5]. It also contained on-board electronics including battery and communication unit
([Fig. 1]). When the electronics received command wirelessly, the linear actuator pushed the
acid into the balloon to react with the base, thus inflating the balloon. The wireless
mechatronic balloon capsule is able to inflate to a volume of 500 mL.
Fig. 1 Configuration of the wireless capsule.
The latest prototype has no onboard electronics ([Fig. 2]). The device is housed inside a gelatin casing with a diameter and length of 13 mm
and 35 mm, respectively ([Table 1]). Magnetic sensors are used to confirm that the device has reached the stomach for
safe inflation. Upon triggering the inflation actuator, citric acid and potassium
bicarbonate in separate chambers interact and release carbon dioxide immediately.
The balloon of the IMI-BC is able to inflate to a volume of 120 mL within 6 minutes.
Fig. 2 Configuration of IMI-BC.
Table 1
IMI-BC characteristics.
|
Characteristic
|
Potential benefits
|
|
Compact microscapsule (13 mm × 35 mm)
|
Likely easily ingestible and may not require insertion via endoscopy.
Multiple capsules can be ingested gradually over several days to achieve the desired
space occupying effect while avoiding the side effect of nausea and vomiting due to
sudden inflation
|
|
Inflation mechanism and base chamber
|
Valve is easily opened using a hand-held external magnet, allowing for acid-base interaction
to release carbon dioxide for inflation of the balloon
|
|
Gelatin casing
|
Expanding balloon easily breaks the gelatin casing
|
|
Balloon made of biodegradable materials
|
Upon degradation of the balloon, it will deflate automatically and hence allow self-excretion
of IMI-BC without the need for endoscopy for retrieval
|
Survival animal trial: mechatronic wireless capsule IGB
This animal survival study was designed to evaluate weight loss after placement of
the mechatronic wireless capsule IGB. It involved the use of six live adult female
pigs weighing 35 to 40 kg each. Given the limited funding and time available, we administered
the wireless capsule IGB to only one of the pigs which weighed 37.8 kg. The other
five pigs served as controls.
Procedure
Under general anaesthesia, the wireless capsule IGB was delivered into the stomach
of one of the pigs endoscopically. A radiofrequency signal was sent from an external
unit to activate a linear actuator onboard the capsule, releasing the acetic acid
to interact with the sodium bicarbonate, hence generating carbon dioxide to inflate
the balloon. The wireless balloon capsule was inflated to a volume of 500 mL. The
inflated balloon was left in the pig’s stomach for a week and the pig was fed the
same feed as five other control pigs. The treated pig was observed for a week before
being weighed again and then it was euthanized. Necropsy was performed to check for
treatment-related AEs.
Outcome measures
The main outcome measure was change in weight following administration of the wireless
capsule IGB. We measured the weights of the treated and control pigs prior to the
start of the trial, and again at 1 week post-procedure to determine if any weight
loss had occurred. The secondary outcome measure was AEs related to the procedure.
Non-survival animal trial: magnetically actuated capsule IGB (IMI-BC)
This was a preclinical non-survival animal trial to demonstrate the feasibility of
inflating two IMI-BCs in one pig using an externally-controlled magnet.
Procedure
One adult male pig weighing 31.9 kg was used for the study. The animal was fed a normal
diet until 3 days prior to the procedure and subsequently fasted for 12 hours. The
animal was euthanatized prior to the experiment to save cost of the general anesthesia.
The IMI-BC was delivered endoscopically and activated via an external handheld magnet.
The procedure was repeated with the delivery and inflation of the second IMI-BC. We
dissected the stomach to visualize the two inflated intragastric balloons.
Outcome measures
The main outcome measures were: (1) time required for each of the balloons to start
inflating upon activation by the external hand-held magnet; (2) time required for
the balloons to be fully inflated; (3) volumes of inflation; and (4) presence or absence
of leak of the balloons.
First-in-human feasibility study: magnetically actuated IGB (IMI-BC)
We then conducted a first-in-man feasibility study to evaluate the magnetically-controlled
balloon inflation mechanism of the ingestible weight loss capsule in one human subject.
Eligibility criteria and recruitment of subject
One human volunteer was recruited for the study. Inclusion criteria included being
a healthy subject aged between 21 to 50 years old with no gastrointestinal symptoms
or previous gastrointestinal surgery. Exclusion criteria included being a pregnant
female; having anatomical, congenital or functional abnormalities of the upper gastrointestinal
tract; having bleeding diathesis; and having previous implants or bioenteric devices.
Procedure
The subject, a female aged 41 years, was fasted 6 hours prior to the procedure and
subsequently sedated. The IMI-BC was delivered endoscopically. A magnetic sensor was
used to confirm its location prior to activation via an external handheld magnet.
Once the balloon had been fully inflated, the subject was observed for a further one
minute to identify any immediate adverse events. Subsequently, the balloon was punctured
with an injection needle and retrieved via a snare.
Outcome measures
The main outcome measures were: (1) time required for the balloon to start inflating
upon activation by the external hand-held magnet; and (2) time required for the balloon
to be fully inflated.
Results
Survival animal trial: mechatronic wireless capsule IGB
After 1 week, the pig with the wireless capsule in its stomach lost 1.5 kg, from 37.8 kg
before to 36.3 kg after placement of the wireless capsule. The five control pigs gained
a range of 3 to 5 kg over a period of 2 weeks. There was no sign of distress nor complication
in any of the pigs during the 1-week observation period. After euthanasia, the balloon
was noted to remain inflated with the volume of 110 mL and the capsule was intact.
Non-survival animal trial: magnetically actuated capsule IGB (IMI-BC)
For the non-survival study, the first IMI-BC was successfully delivered. It inflated
within 15 seconds after activation by the hand-held magnet, and was fully inflated
within 6 minutes. ([Fig. 3]). The second IMI-BC was also successfully deployed and activated to inflate within
15 seconds. However, it became deflated before it could achieve full inflation. At
the end of the animal trial, we retrieved the two IMI-BCs by cutting into the pig’s
stomach. There was no leak in the first balloon. Its volume of inflation was 120 mL.
In the second balloon, suboptimal sealing was noted, thus explaining the failure to
achieve full inflation of the balloon.
Fig. 3 a Dimensions of IMI-B. b Insertion of IMI-BC and inflation of the balloon using an externally controlled magnet.
c Inflated intragastric balloon seen via endoscopy within a live euthanized pig’s stomach.
First-in-human feasibility study: magnetically actuated IGB (IMI-BC)
The IMI-BC was successfully deployed endoscopically into the subject’s stomach ([Fig. 4]). The magnetic sensors were able to confirm the device’s location. There was successful
inflation of the microcapsule’s balloon within 1 minute to a full inflation volume
of approximately 120 mL within 3 minutes ([Fig. 5]). For the next 1 minute of observation, the balloon remained fully inflated. The
patient’s vital signs were recorded to be normal throughout the procedure.
Fig. 4 Introduction of IMI-BC via a gastroscope.
Fig. 5 Inflated intragastric balloon seen via endoscopy within a human stomach.
Discussion
For an IGB to be widely used in the clinical setting, it should have efficacy in inducing
weight loss, safety in administration, and ease of use. While current IGBs have proven
themselves to be effective and safe in weight management, most require use of a catheter
to inflate the balloon with liquid and endoscopy to insert and/or retrieve it. In
doing do, the procedures need sedation or anesthesia, thereby increasing the overall
cost of therapy. The IMI-BC that we have developed is able to overcome these usability
issues.
The IMI-BC only requires a simple external hand-held magnet to activate its inflation
valve within the chamber, initiating a chemical reaction to release the carbon dioxide,
which inflates the balloon. This removes the need for a catheter and injection of
saline or water to inflate the balloon. For safety purposes, an external magnetic
sensor is used to confirm that the IMI-BC is within stomach prior to inflation. Absence
of onboard electronics is attractive in terms of cost-effectiveness. These advantages
potentially allow the IMI-BC to be easily administered by doctors in the hospital
setting, and even within the outpatient primary care setting.
There are certain limitations in our trials, which we plan to address in subsequent
studies. First, given the limited funds and time available, we started off by performing
small-scale animal and human studies. In our subsequent studies, we plan to perform
larger-scale animal and human trials to allow us to draw appreciable conclusions regarding
the efficacy and safety of our device.
In our survival animal trial, the pig with the wireless capsule in the stomach lost
a significant amount of weight. With the larger-scale animal trials, we aim to evaluate
mean weight loss and demonstrate sustained weight loss post-procedure. Second, as
the aims of the studies above were mainly to demonstrate feasibility, we will investigate
in future trials the optimal volume and number of capsules required, and duration
of their inflation to achieve sustained weight loss. Third, we aim to further investigate
the safety of our device, ensuring that the balloon can be deflated safely and excreted
through the gastrointestinal tract without causing small bowel obstruction. We aim
to construct the balloon using a biodegradable material, allowing carbon dioxide to
be self-excreted, hence aiding the balloon in deflation. We also aim to enhance our
location sensor to tracking the capsule’s migration along the gastrointestinal system.
Our future human trials will then aim to confirm that the device can be swallowed
without endoscopic assistance for insertion, externally activated to inflate in the
stomach, left inflated in the stomach for a few days, naturally deflated, and excreted
through the gastrointestinal system.