The Cut Out Phenomenon: An Experimental Analysis of the Migration Behaviour of Lag Screws for Fixation of Pertrochanteric Fractures

U. Horas1 , R. Schnettler1 , S. K. Tschöke1 , A. Speitling2 , C. Meyer1 , J. Mockwitz3
  • 1Department of Trauma Surgery, Justus-Liebig-University Giessen
  • 2Applied Trauma and Spine Research, Schönkirchen, Germany
  • 3Department of Trauma Surgery, Main-Taunus County Hospital, Hofheim a. Taunus, Germany
Further Information

Publication History

Publication Date:
30 March 2004 (online)

Abstract

The cut out phenomenon of lag screws after internal fixation of intertrochanteric fractures remains an unsolved pathophysiological and biomechanical problem. The purpose of the current study was to analyse the migration behaviour of different hip screw implants in a standardised experimental setting using a specific testing machine and bone substitute material to mimic different bone densities. Migration of lag screws was significantly influenced by bone density and could significantly be improved by a central position of the lag screw in the femoral head adjacent to its cortical bone and a wide collodiaphyseal angle. Larger outside diameter lag screws showed a significantly higher resistance to migration than those with a smaller outside diameter. Self cutting or non-self cutting threads did not show any significant difference. In a double screw system different positions of the screws showed different cut out resistance.

References

  • 1 Alho A, Benterud J G, Solovieva S. Internally fixed femoral neck fractures; early prediction of failure in 203 elderly patients with displaced fractures.  Acta Orthop Scand. 1999;  70 141
  • 2 Baitner A C, Maurer S G, Hickey D G, Jazrawi L M. et al . Vertical shear fractures of the femoral neck. A biomechanical study.  Clin Orthop. 1999;  367 300
  • 3 Baumgaertner M R, Curtin S L, Lindskog D M, Keggl J M. The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip.  J Bone Joint Surg [Am]. 1995;  77 1058
  • 4 Bergmann G, Rohlmann A, Graichen F. In-vivo-Messung der Hüftgelenkbelastung; 1. Teil Krankengymnastik.  Z Orthop. 1989;  127 627
  • 5 Chapman J R, Harrington R M, Lee K M, Anderson P A. et al . Factors affecting the pullout strength of cancellous bone screws.  J Biomech Eng. 1996;  118 391
  • 6 Choueka J, Koval K J, Kummer F J, Crawford G. et al . Biomechanical comparison of the sliding hip screw and the dome plunger; effects of material and fixation design.  J Bone Joint Surg [Br]. 1995;  77 277
  • 7 Davis T R, Sher J L, Horsman A, Simpson M. et al . Intertrochanteric femoral fractures; mechanical failure after internal fixation.  J Bone Joint Surg [Br]. 1990;  72 26
  • 8 Debruner H U. Studien zur Biomechanik des Hüftgelenkes I.  Z Orthop. 1975;  113 377
  • 9 De Laet C E, Pols H A. Fractures in the elderly: Epidemiology and demography.  Baill Best Pract Res Clin Endocr Metab. 2000;  14 171
  • 10 Elder S, Frankenburg E, Cavusoglu T, Sepici B. Biomechanical evaluation of calcium phosphate cement-augmented fixation of unstable intertrochanteric fractures.  J Orthop Trauma. 2000;  14 386
  • 11 Graichen F, Bergmann G. Four-channel telemetry system for in vivo measurement of hip joint forces.  J Biomed Eng. 1991;  13 370
  • 12 Haynes R C, Pöll R G, Miles A W, Weston R B. An experimental study of the failure modes of the gamma locking nail and AO dynamic hip screw under static loading: A cadaveric study.  Med Eng Phys. 1997;  19 446
  • 13 Haynes R C, Pöll R G, Miles A W, Weston R B. Failure of femoral head fixation: A cadaveric analysis of lag screw cut out with the gamma nail and AO dynamic hip screw.  Injury. 1997;  28 337
  • 14 Jenny J Y, Rapp E, Cordey J. Type of screw does not influence holding power in the femoral head.  Acta Orthop Scand. 1999;  70 435
  • 15 Jensen J S. Incidence of hip fractures.  Acta Orthop Scand. 1980;  51 511
  • 16 Jensen J S. Mechanical strength of sliding screw-plate hip implants: A biomechanical study of instable trochanteric fractures.  Acta Orthop Scand. 1980;  51 625
  • 17 Kauffman J I, Simaon J A, Kummer F J, Pearlman C J. et al . Internal fixation of femoral neck fractures with posterior comminution: A biomechanical study.  J Orthop Trauma. 1999;  13 155
  • 18 Kawaguchi S, Sawada K, Nabeta Y. Cutting-out of the lag screw after internal fixation with the Asiatic gamma nail.  Injury. 1998;  29 47
  • 19 Lenzner A, Kaur I, Haviko T, Sögel V. et al . Impaction bone-grafting increases the holding power of cancellous screws in the femoral head: A pull-out study in human cadaver hips.  Acta Orthop Scand. 1999;  70 25
  • 20 Loch D A, Kyle R F, Bechthold J E, Kane M. et al . Forces required to initiate sliding in second-generation intramedullary nails.  J Bone Joint Surg [Am]. 1998;  80 1626
  • 21 Meislin R J, Zuckerman J D, Kummer F J, Frankel V H. A biomechanical analysis of the sliding hip screw: The question of plate angle.  J Orthop Trauma. 1990;  4 130
  • 22 Müller M E, Allgöwer M, Schneider R, Willenegger H. Manual der Osteosynthese AO-Technik. Springer, Berlin, Heidelberg, New York 1993
  • 23 Parker M J, Blundell C. Choice of implant for internal fixation of femoral neck fractures: Meta-analysis of 25 randomised trials including 4 925 patients.  Acta Orthop Scand. 1998;  69 138
  • 24 Sochart D H. Poor results following internal fixation of displaced subcapital femoral fractures: Complacency in fracture reduction.  Arch Orthop Trauma Surg. 1998;  117 379
  • 25 Stürmer K M. Die pertrochantere Fraktur. Leitlinien Unfallchirurgie, Thieme, Stuttgart 1999; 141
  • 26 Tencer A F, Johnson K D. Lower extremity fixation. Biomechanics in orthopaedic trauma. J B Lippencott Comp, Philadelphia 1994
  • 27 Thomas A P. Dynamic hip screws that fail.  Injury. 1991;  22 45
  • 28 Tschöke S K. Das Migrationsverhalten von Schenkelhalsschrauben in pertrochantären Frakturen: Eine experimentelle biomechanische Analyse. Dissertation aus der Klinik und Poliklinik für Unfallchirurgie, Justus-Liebig-Universität Gießen 2001
  • 29 Wang C J, Brown C J, Yettram A L, Procter P. Intramedullary femoral nails: One or two lag screws? A Preliminary Study.  Med Eng Phys. 2000;  22 613
  • 30 Yoshimine F, Latta L L, Milne E L. Sliding characteristics of compression hip screws in the intertrochanteric fracture: A clinical study.  J Orthop Trauma. 1993;  7 348

Dr. med. Uwe Horas

Department of Trauma Surgery

Justus-Liebig-University

Rudolf-Buchheim-Straße 7

35385 Giessen

Germany

Phone: +49/6 41-99-4 46 01

Fax: +49/6 41-99-4 46 09

Email: Uwe.Horas@chiru.med.uni-giessen.de

    >