The Thermal Diffusion Probe (TDP) System allows continuous real-time measurement of
tissue perfusion in flaps. The authors used a TDP with two thermistors, one active,
the other passive, embedded in a 0.9-mm diameter catheter to measure continuous tissue
perfusion in rabbit epigastric flaps. The distal thermistor is heated to 2° C above
the tissue baseline temperature. The power required to maintain this temperature difference
is mathematically related to the tissue perfusion in the volume surrounding the probe
tip.
Central and peripheral TDPs were placed. The TDP effectiveness in detecting and measuring
daily tissue perfusion in buried epigastric flaps was tested. Contralateral epigastric
pedicles were transposed into the flaps prior to ligation of the original pedicle.
Flaps with transposed pedicles showed a progressive and significant increase in tissue
perfusion during the initial 3 weeks of the experiment, compared to flaps without
the pedicle transfer.
The TDP System is a useful experimental method for the continuous and real-time quantification
of flap perfusion and may be helpful in making clinical decisions about prefabricated
flap transfer.
KEYWORDS
Thermal Diffusion Probe - blood perfusion monitoring - prefabricated flaps
REFERENCES
- 1
Pribaz J J, Weiss D D, Mulliken J B, Eriksson E.
Prelaminated free flap reconstruction of complex central facial defects.
Plast Reconstr Surg.
1999;
104
357-365
- 2
Pribaz J J, Fine N, Orgill D P.
Flap prefabrication in the head and neck: a 10-year experience.
Plast Reconstr Surg.
1999;
103
808-820
- 3
Shen T Y.
Microvascular transplantation of prefabricated free thigh flap (Letter).
Plast Reconstr Surg.
1982;
69
568
- 4
Jones B M.
Monitors for the cutaneous microcirculation.
Plast Reconstr Surg.
1984;
73
843-850
- 5 Bowman H F. Estimation of tissue blood flow. In: Eberhart RC, Shitzer A Heat Transfer
in Medicine and Biology: Analysis and Application New York; Plenum Publishing Corporation
1984: 193-230
- 6
Klar E, Kraus T, Bredt M, et al..
First clinical realization of continuous monitoring of liver microcirculation after
transplantation by thermodiffusion.
Transpl Int.
1996;
9
S140-S143
- 7
Vajkoczy P, Horn P, Bauhuf C, et al..
Effect of intra-arterial papaverine on regional cerebral blood flow in hemodynamically
relevant cerebral vasospasm.
Stroke.
2001;
32
498-505
- 8
Newman W H, Bowman H F, Orgill D P, Klar E.
A methodology for in vivo measurements of blood flow in small tissue volumes.
Advances in Biological Heat and Mass Transfer, ASME HTD.
1995;
322
99-105
- 9
Falco N A, Pribaz J J, Eriksson E.
Vascularization of skin following implantation of an arterio-venous pedicle: implications
in flap prefabrications.
Microsurgery.
1992;
13
249-254
- 10
Duffy F J, Pribaz J J, Hergrueter C A, et al..
Flap prefabrication using an exteriorized vascular pedicle in a rabbit ear model.
Br J Plast Surg.
1993;
46
279-284
- 11
Duffy F J, Maitz P KM, Hergrueter C A, et al..
Maximizing flap survival in a prefabrication model using exogenous and endogenous
bFGF: a new approach.
Microsurgery.
1996;
17
176-179
- 12
Maitz P KM, Pribaz J J, Duffy F J, et al..
The value of the delay phenomenon in flap prefabrication: an experimental study in
rabbits.
Br J Plast Surg.
1994;
47
149-154
- 13
Maitz P KM, Pribaz J J, Hergrueter C A.
Impact of tissue expansion on flap prefabrication: an experimental study in rabbits.
Microsurgery.
1996;
17
35-40
- 14
Furnas H, Rosen J M.
Monitoring in microvascular surgery.
Ann Plast Surg.
1991;
26
265-272
- 15
Ono H, Tamai S, Yajima H, et al..
Blood flow through prefabricated flaps-an experimental study in rabbits.
Br J Plast Surg.
1993;
46
449-455
- 16
Clinton M S, Sepka R S, Bristol D, et al..
Establishment of normal ranges of laser Doppler blood flow in autologous tissue transplants.
Plast Reconstr Surg.
1991;
87
299-309
- 17
Marks N J, Trachy R E, Cummings C W.
Dynamic variations in blood flow as measured by laser Doppler velocimetry: a study
in rat skin flaps.
Plast Reconstr Surg.
1984;
73
804-810
- 18
Wardell K, Jakobsson A, Nilsson G E.
Laser Doppler perfusion imaging by dynamic light scattering.
IEEE Trans Biomed Eng.
1993;
40
309-316
- 19
Arnold F, He C F, Jia C Y, Cherry G W.
Perfusion imaging of skin island flap blood flow by a scanning laser-Doppler technique.
Br J Plast Surg.
1995;
48
280-287
- 20
Martin G T, Bowman H F.
Validation of real-time continuous perfusion measurement.
Medical and Biological Engineering and Computing.
2000;
38
319-325
Dennis P OrgillM.D. Ph.D.
Division of Plastic Surgery, Brigham and Women's Hospital
75 Francis Street
Boston, MA 02115