Hintergrund: Zur Evaluierung des Effektes von weichen Kontaktlinsen mit verschiedenen Dioptrienstärken auf den gemessenen Intraokulardruck mit dem Non-contact-Pneumotonometer. Methode: 120 Augen (80 gesunde Freiwillige: 50 Frauen, 30 Männer, im Alter von 22 - 61 Jahren) wurden in dieser Studie einbezogen. Der Intraokulardruck wurde mit dem Pneumotonometer Canon X-10 vor und nach Einsetzen einer Ciba Vision Focus Night&Day soft contact lens gemessen. Es wurden Kontaktlinsen mit den Dioptrienstärken von + 0,25 D, + 1,00 D, + 4,00 D, - 1,00 D und - 4,00 D eingesetzt. Der Durchschnitt von drei Messungen wurde als repräsentativer Intraokulardruck für die statistische Auswertung herangezogen. Ergebnisse: Der gemessene Intraokulardruck bei myopen Kontaktlinsen von - 1,00 und - 4,00 D zeigte geringere Werte mit einem mittleren Bereich von 1 mm Hg. Die Differenz zwischen den Messungen bei myopen Kontaktlinsen war meist kleiner als ± 2 mm Hg (78 % bei - 1,00 D und 90 % bei - 4,00 D Kontaktlinsen). Alle Differenzen bei + 0,25 D Kontaktlinsen waren geringer als ± 2 mm Hg. Die Differenz war beträchtlich höher bei + 1,00 und + 4,00 D hyperopen Kontaktlinsen und zeigte eine strenge Zunahme in Abhängigkeit von der Dioptrienstärke. Schlussfolgerungen: In unserer Studie konnten wir zeigen, dass der intraokulare Druck verlässlich bei eingesetzen myopen und hyperopen Kontaktlinsen mit geringer Dioptrienstärke mit dem Non-contact-Pneumotonometer gemessen werden kann. Die Non-contact-Pneumotonometrie kann eine hilfreiche Methode bei Patienten mit therapeutischen Kontaktlinsen sein, ebenso müssen weiche Kontaktlinsen bei der Messung des Intraokulardrucks nicht unbedingt entfernt werden.

Abstract

Purpose: To evaluate the effect of soft contact lenses with different lens power on the measured value of the intraocular pressure with non-contact pneumotonometry. Methods: 120 eyes (80 healthy volunteers: 50 women, 30 men, aged 22 to 61 years) were included in this study. Intraocular pressure was measured with a pneumotonometer Canon X-10 before and after insertion of the Ciba Vision Focus Night&Day soft contact lens. We used contact lenses with different lens power of + 0.25 D, + 1.00 D, + 4.00 D, - 1.00 D, and - 4.00 D. The averages of three measurements were taken as representative IOP values that were statistically evaluated. Results: IOP measured over myopic lenses of - 1.00 and - 4.00 D showed lower values within the mean range of 1 mm Hg. The difference between the measurements over the myopic lenses was mostly smaller than ± 2 mm Hg (78 % when using - 1.00 D and 90 % when using - 4.00 D contact lens). All the differences using + 0.25 D contact lens were smaller than ± 2 mm Hg. The difference was considerably higher in measurements over + 1.00 and + 4.00 hyperopic contact lenses and showed strong increase with the lens power. Conclusions: In our study we showed that intraocular pressure can be reliably measured with non-contact pneumotonometry over myopic lenses or hypermetropic lenses with small lens power. This suggests that non-contact pneumotonometry is a useful method in patients wearing therapeutic contact lenses and contact lens wearers who, when measuring the eye pressure, would not need to remove the contact lenses before the examination.

References

  • 1 Brethfeld V. Die Behandlung von Hornhauterkrankungen mit weichen hydrophilen Kontaktlinsen.  Med Welt. 1975;  26 762-764
  • 2 Carel R S, Korczyn A D, Rock M, Goya I. Association between ocular pressure and certain health parameters.  Ophthalmol. 1984;  91 311-314
  • 3 Draeger J. Applanationstonometrie auf Kontaktlinsen mit hohem Wassergehalt. Probleme, Ergebnisse und Korrekturfaktoren.  Klin Monatsbl Augenheilkd. 1980;  176 38-43
  • 4 Insler M S, Robbins R G. Intraocular pressure by noncontact tonometry with and without soft contact lenses.  Arch Ophthalmol. 1987;  105 1358-1359
  • 5 Krieglstein G K, Waller W K, Reimer H, Langham M E. Augeninnendruckmessung auf weichen Kontaktlinsen.  Graefes Arch Clin Exp Ophthalmol. 1976;  199 223-229
  • 6 McMonnies C W. Noncontact tonometry through soft contact lenses.  Am J Optom Physiol Opt. 1986;  63 948-951
  • 7 Meyer R F, Stanifer R M, Bobb K C. Mackay-Marg tonometry over therapeutic soft contact lenses.  Am J Ophthalmol. 1978;  86 19-23
  • 8 Myers K J, Scott C A. The non-contact („air puff”) tonometer: variability and corneal staining.  Am J Optom Physiol Opt. 1975;  52 36-46
  • 9 Rubenstein J B, Deutsch T A. Pneumatonometry through bandage contact lenses.  Arch Ophthalmol. 1985;  103 1660-1661

Petra Schollmayer