Dynamics of Distal Tibiofibular Joint in Correlation with the Position of the Syndesmotic Screw in Maisonneuve Fractures of the Ankle - A Cadaver Study
Publikationsverlauf
Publikationsdatum:
24. September 2003 (online)

Abstract
Introduction: Maisonneuve-fractures of the ankle joint are best treated by fixation of the ruptured
tibiofibular syndesmosis with one or two syndesmotic screws. A rigid fixation of
the syndesmosis however, would lead to pain, limitation of range of motion and to
early arthrosis of the ankle joint. The purpose of this study was to investigate
the optimal position of the syndesmotic screw, which allows physiological movements
in the syndesmosis. The influence of application of the screw through 3 or 4 cortical
shells on the distal tibiofibular joints stability was also studied.
Material and Methods: 14 pairs of fresh frozen cadaver lower legs with intact knee joints were divided
into two groups. In all specimens the soft tissue was removed and the distal tibiofibular
ligaments exposed. The specimens were then potted in Araldit-G-bone cement. The syndesmosis
and the interosseus membrane were then sharply dissected, to simulate a Maisonneuve-fracture
situation. In group I (n = 7) the syndesmotic screw was applied in each pair directly
above the upper end of the syndesmosis through 3 cortical shells (lateral and medial
fibular shell, lateral tibia shell) on the one side and through 4 cortical shells
on the other side. In the group II (n = 7) the syndesmotic screw was applied in each
pair through four cortical shells, directly above the syndesmosis and 50 mm above
the upper end of the syndesmosis on the either side. In both groups one pair served
as control. The specimens were tested with an Instron 2000®-material-testing machine.
A medio-lateral translation force was applied to each specimen. All testing was conducted
in the plastic range of specimens and were destructive. Single factor analysis of
variance and Student's t-Test were used for statistical evaluation.
Results: In the control specimen of both groups the mean maximal translation force (mTmax)
was 1 500 N. In the first group mTmax for the screw with 3 cortical shells was 300
N, and for the four cortical screw 312 N. In the second group the mTmax with the
screw directly above the syndesmosis was 380 N and with the screw inserted 50 mm
above the syndesmosis 160 N. The higher the screw-level, the lower was the power
needed to break the fibula in the screw level.
Discussion and Conclusion: There is no significant difference between syndesmotic screws with 3 or 4 fixed
cortical shells at a medio-lateral load-axis: The more proximal the position of the
screw, the fewer is the force needed to widen the distal tibiofibular joint. In both
groups the stiffness achieved with the screws was significantly less than the control
ones. The ankle joint was however, most stable with the syndesmotic screw inserted
directly above the syndesmosis.
Key words
Ruptured tibiofibular syndesmosis - ankle fracture - biomechanic - positioning screw - injury
References
- 1 Burwell H N, Charnley A P. The treatment of the displaced fractures at the ankle by rigid internal fixation and early joint movement. J Bone Joint Surg [Br]. 1965; 47 634-660
- 2 Joy G, Patzakis M J, Harvey JP J r. Precise evaluation of the reduction of severe ankle fractures. Technique and correlation with end results. J Bone Joint Surg [Am]. 1974; 56 979-993
- 3 Leeds H C, Ehrlich M G. Instability of the distal tibiofibular syndesmosis after bimalleolar and trimalleolar ankle fractures. J Bone Joint Surg [Am]. 1984; 66 490-503
- 4 Miller R S, Weinhold P S, Dahners L E. Comparison of tricortical screw fixation versus a modified suture construct for fixation of ankle syndesmosis injury: a biomechanical study. J Orthop Trauma. 1999; 13 39-42
- 5 Purvis G P. Displaced, unstable ankle fractures, classification, incidence and management of a consecutive series. Clin Orthop. 1982; 165 91-98
- 6 Riegels-Nielson P, Christensen J, Greiff J. The stability of the tibiofibular syndesmosis following rigid internal fixation for type C malleolar fractures: an experimental and clinical study. lnjury. 1983; 14 357-360
- 7 Sauer H D, Jungfer E, Jungblut K H. Experimentelle Untersuchungen zur Reißfestigkeit des Bandapparates am menschlichen Sprunggelenk. Hefte Unfallheilkunde. 1987; 131 37-45
- 8 Schenk R. Anatomie des oberen Sprunggelenkes. Hefte Unfallheilkunde. 1978; 131 1-9
- 9 Schmidt H M. Die Artikulationsflächen der menschlichen Sprunggelenke. Adv Anat Embryol Cell Biol. 1981; 66 1-81
- 10 Thordarson D B, Hedman T P, Gross D, Magre G. Biomechanical evaluation of polylactide absorbable screws used for syndesmosis injury repair. Foot Ankle Int. 1997; 18 622-627
- 11 Weber B G. Die Verletzungen des oberen Sprunggelenkes. 2. Auflage Huber Verlag, Bern 1972
- 12 Weinert C R, McMaster J H, Ferguson R J. Dynamic function of the human fibula. Am J Anat. 1973; 138 145-150
- 13 Wirth C J, Küsswetter W, Jäger M. Biomechanik und Pathomechanik des oberen Sprunggelenkes. Hefte Unfallheilkunde. 1978; 131 10-22
Dr. Alexander Egkher
Univ.-Klinik für Unfallchirurgie Wien
Währinger Gürtel 18-20
1090 Wien
Austria
Telefon: +43/1-4 04 00 56 19
Fax: +43/1-4 04 00 59 49
eMail: alexander.egkher@akh-wien.ac.at