Background: A new prosthesis was designed for reconstruction of the bifurcation of the trachea
and used for experimental carinal replacement in dogs. Methods: The main Frame of the new prosthesis consists of Y-shaped Marlex mesh tube reinforced
with polypropylene spirals. It is coated with collagen extracted from porcine skin
to provide biocompatibility and airtightness. Carinal replacement with omentopexy
was performed in 17 dogs. Results: Seven dogs survived the postoperative period, and 10 dogs died within 11 days after
the operation. The main causes of early postoperative death were fistula and air leakage
from the prosthesis. Causes of late postoperative death were obstruction of the main
bronchus (two) and lung abscess (one). The four long-term survivors had no stenosis
or dehiscence until they were sacrificed 15 months after the operation. Histological
examination of these dogs revealed that the luminal surface was covered either with
ciliated columnar epithelium or non-ciliated squamous epithelium. Conclusions: The study results suggest the possibility of successful prosthetic carinal reconstruction
with epithelial regrowth using this Y-shaped prosthesis.
Key words:
Tracheal prosthesis - Collagen - Carinal reconstruction - Tissue regeneration - Artificial
trachea
References
- 1
Grillo H C.
Carinal reconstruction.
Ann Thorac Surg..
1982;
34
356-73
- 2
Pearson F G, Todd T R, Cooper J D.
Experience wich primary neoplasms of the trachca and carina.
J Thorac Cardiovasc Surg..
1984;
88
511-8
- 3
Tsuchiya R, Goya T, Naruke T, Suemasu K J.
Resection of tracheal carina for lung cancer.
Thorac Cardiovasc Surg..
1990;
99
779-87
- 4
Mathisen D J, Grillo H C.
Carinal resection for bronchogenic carcinoma.
J Thorac Cardiovasc Surg..
1991;
102
16-23
- 5
Jacobs J.
Investigations into tracheal prosthetic reconstruction.
Laryngoscope..
1988;
98
1239-45
- 6
Neville W E, Bolanowski P J, Koita G C.
Clinical experience with the silicone trachcal prosthesis.
J Thorac Cardiovasc Surg..
1990;
99
604-13
- 7
Borrie J, Redshow N, Dobbinson T.
Silastic tracheal bifurcation prosthesis with subterminal Dacron suture cuffs.
J Thorac Cardiovasc Surg..
1973;
65
956-62
- 8
Okumura N, Nakamura, Natsume T, Tomihata K, Ikada Y, Shimizu Y.
Experimental study on a new tracheal prosthesis made from collagen-conjugated mesh.
J Thorac Cardiovasc Surg..
1994;
108
337-45
- 9
Teramachi M, Okumura N, Nakamura T. et al .
Intrathoracic tracheal reconstruction with a collagen-conjugated prosthesis - evaluation
of the efficacy of omental wrapping -.
J Thorac Cardiovasc Surg..
1997;
113
701-11
- 10
Grillo H C.
Circumferential resection and reconstruction of the mediastinal and cervical trachea.
Annals of surgery..
1965;
162
374-88
- 11
Mulliken J B, Grillo H C.
The limits of tracheal resection with primary anastmosis.
J Thorac Cardiovasc Surg..
1968;
55
418-21
- 12
Theman T E, Kerr J H, Nelems J M, Pearson F G.
Carinal resection - A report of two cases and a description of the anesthetic technique.
J Thorac Cardiovasc Surg..
1976;
71
314-20
- 13
Ishihara T, Ikeda T, Inoue H, Fukai S.
Resection of cancer of lung and carina.
J Thorac Cardiovasc Surg..
1977;
73
936-43
- 14
Mogan E, Lima O, Goldberg M, Ferdman A, Luk S K, Cooper J D.
Successful revascularization of totally ischemic bronchial autografts with omental
pedicle flaps in dogs.
J Thorac Cardiovasc Surg..
1982;
84
204-10
- 15
Nakanishi R, Shirakusa T, Takachi T.
Omentopexy for tracheal autografts.
Ann Thorac Surg..
1994;
57
841-5
- 16
Natsume T, Ike O, Okada T, Takimoto N, Shimizu Y, Ikada Y.
Porous collagen sponge for esophageal replacement.
J Biomed Mater Res..
1993;
27
867-75
- 17
Matsubara Y, Kosaba S, Ikeda S. et al .
Experimental and clinical results of tracheal prosthesis.
Kyobugeka..
1990;
43
368-74
- 18
Toomes H, Mickisch G, Vogt-Moykopf I.
Experiences with prosthetic reconstruction of the trachea and bifurcation.
Thorax..
1985;
40
32-7
Dr. T. Sekine
Department of Bioartificial Organs Institute for Frontier Medical Sciences
Kyoto University
53 Kawahara-cho, Shogoin
Sakyo-ku, Kyoto 606-8397, Japan
Phone: +81-75-751-4149
Fax: +81-75-751-4844