Rofo 2003; 175(10): 1330-1334
DOI: 10.1055/s-2003-42895
Rapid Communication
© Georg Thieme Verlag Stuttgart · New York

Coronary MR Imaging Using Free-Breathing 3D Steady-State Free Precession with Radial k-space Sampling

MR-Koronarangiographie während freier Atmung mit 3D Steady-State Free Precession und radialer k-Raum AbtastungE.  Spuentrup1, 2 , M.  Katoh1 , M.  Stuber2, 3, 4 , R.  Botnar2, 4 , T.  Schaeffter5 , A.  Buecker1 , R.  W.  Günther1
  • 1Department of Diagnostic Radiology, Aachen Technical University, Aachen, Germany
  • 2Department of Medicine (Cardiovascular Division) Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA
  • 3Johns Hopkins University School of Medicine, Baltimore, MD
  • 4Philips Medical Systems, Best, NL
  • 5Philips Research Laboratories, Hamburg, Germany
Further Information

Publication History

Publication Date:
13 October 2003 (online)

Zusammenfassung

Zielsetzung: Untersuchung einer 3D steady-state free precession (SSFP) MR-Koronarangiographiesequenz während freier Atmung mit radialer k-Raum Abtastung für die MR-Koronarangiographie, Projektions-Angiographie und MR-Koronargefäßwandbildgebung. Material und Methode: Eine navigator-gesteuerte 3D-SSFP-Sequenz (TR = 6,1 ms, TE = 3,0 ms, Flip-Winkel = 120°, Sichtfenster [„field-of-view”] = 360 mm2) mit 384 Radialen wurde für eine MR-Koronarlumen- und Projektions-Angiographie implementiert. Für die Projektionsangiographie erfolgte die Kombination mit einem lokalen Spininversions-Puls in der Aorta ascendens. Die MR-Koronargefäßwandbildgebung wurde mit einer radialen 3D-SSFP-Sequenz (384 Radiale, TR = 5,3 ms, TE = 2,7 ms, Flip-Winkel = 55°, rekonstruierte Ortsauflösung 0,35 × 0,35 × 1,2 mm3) mit Schwarzblut-Kontrast („black-blood”) und lokalem Rückstellpuls („local inversion pulse”) durchgeführt. Sechs gesunde Probanden (jeweils zwei für jede Sequenz) wurden untersucht und die Bewegungsartefakte analysiert. Ergebnisse: In der 3D-SSFP-MR-Koronarangiographie stellte sich das Koronargefäßlumen signalreich dar. In Kombination mit der Projetionstechnik wurde das umgebende Gewebe fast vollständig unterdrückt. In der MR-Koronargefäßwandbildgebung kam die Gefäßwand im Vergleich zum Blut und dem umgebenden Gewebe signalreich zur Darstellung. Es traten keine relevanten Bewegungsartefakte auf. Schlussfolgerung: 3D-SSFP mit radialer k-Raumabtastung kann für die MR-Koronarangiographie, Projektions-Angiographie und MR-Koronargefäßwandbildgebung eingesetzt werden.

Abstract

Purpose: To investigate the potential of free-breathing 3D steady-state free precession (SSFP) imaging with radial k-space sampling for coronary MR-angiography (MRA), coronary projection MR-angiography and coronary vessel wall imaging. Materials and Methods: A navigator-gated free-breathing T2-prepared 3D SSFP sequence (TR = 6.1 ms, TE = 3.0 ms, flip angle = 120°, field-of-view = 360 mm2) with radial k-space sampling (384 radials) was implemented for coronary MRA. For projection coronary MRA, this sequence was combined with a 2D selective aortic spin tagging pulse. Coronary vessel wall imaging was performed using a high-resolution inversion-recovery black-blood 3D radial SSFP sequence (384 radials, TR = 5.3 ms, TE = 2.7 ms, flip angle = 55°, reconstructed resolution 0.35 × 0.35 × 1.2 mm3) and a local re-inversion pulse. Six healthy volunteers (two for each sequence) were investigated. Motion artifact level was assessed by two radiologists. Results: In coronary MRA, the coronary lumen was displayed with a high signal and high contrast to the surrounding lumen. Projection coronary MRA demonstrated selective visualization of the coronary lumen while surrounding tissue was almost completely suppressed. In coronary vessel wall imaging, the vessel wall was displayed with a high signal when compared to the blood pool and the surrounding tissue. No visible motion artifacts were seen. Conclusion: 3D radial SSFP imaging enables coronary MRA, coronary projection MRA and coronary vessel wall imaging with a low motion artifact level.

References

  • 1 Kim W Y, Danias P G, Stuber M. et al . Coronary magnetic resonance angiography for the detection of coronary stenoses.  N Engl J Med. 2001;  345 1863-1869
  • 2 Sommer T, Hofer U, Hackenbroch M. et al . [Submillimeter 3D coronary MR angiography with real-time navigator correction in 107 patients with suspected coronary artery disease].  Fortschr Röntgenstr. 2002;  174 459-466
  • 3 Stuber M, Börnert P, Spuentrup E, Botnar R M, Manning W J. Selective three-dimensional visualization of the coronary artery lumen using arterial spin tagging.  Magn Reson Med. 2002;  47 322-329
  • 4 Botnar R M, Stuber M, Kissinger K V, Kim W Y, Spuentrup E, Manning W J. Noninvasive coronary vessel wall and plaque imaging with magnetic resonance imaging.  Circulation. 2000;  102 2582-2587
  • 5 Fayad Z A, Fuster V, Fallon J T. et al . Noninvasive In Vivo Human Coronary Artery Lumen and Wall Imaging Using Black-Blood Magnetic Resonance Imaging.  Circulation. 2000;  102 506-510
  • 6 Kim W Y, Stuber M, Bornert P, Kissinger K V, Manning W J, Botnar R M. Three-dimensional black-blood cardiac magnetic resonance coronary vessel wall imaging detects positive arterial remodeling in patients with nonsignificant coronary artery disease.  Circulation. 2002;  106 296-299
  • 7 Barkhausen J, Hunold P, Jochims M. et al . [Comparison of gradient-echo and steady state free precession sequences for 3D-navigator MR angiography of coronary arteries].  Fortschr Röntgenstr. 2002;  174 725-730
  • 8 Spuentrup E, Bornert P, Botnar R M, Groen J P, Manning W J, Stuber M. Navigator-gated free-breathing three-dimensional Balanced FFE (TrueFISP) coronary MRA.  Invest Radiol. 2002;  37 637-642
  • 9 Glover G H, Pauly J M. Projection reconstruction techniques for reduction of motion effects in MRI.  Magn Reson Med. 1992;  28 275-289
  • 10 Fleckenstein J L, Archer B T, Barker B A, Vaughan J T, Parkey R W, Peshock R M. Fast short-tau inversion-recovery MR imaging.  Radiology. 1991;  179 499-504
  • 11 Botnar R M, Kim W Y, Bornert P, Stuber M, Spuentrup E, Manning W J. 3D coronary vessel wall imaging utilizing a local inversion technique with spiral image acquisition.  Magn Reson Med. 2001;  46 848-854
  • 12 Stuber M, Botnar R M, Danias P G, Kissinger K V, Manning W J. Submillimeter three-dimensional coronary MR angiography with real-time navigator correction: comparison of navigator locations.  Radiology. 1999;  212 579-587
  • 13 Stuber M, Botnar R M, Spuentrup E, Kissinger K V, Manning W J. Three-dimensional high-resolution fast spin echo coronary magnetic resonance angiography.  Magn Reson Med. 2001;  45 206-211
  • 14 Spuentrup E, Buecker A, Karassimos E, Gunther R W, Stuber M. Navigator-gated and real-time motion corrected free-breathing MR imaging of myocardial late enhancement.  Fortschr Röntgenstr. 2002;  174 562-567
  • 15 Larson A C, Simonetti O P, Li D. Coronary MRA with 3D undersampled projection reconstruction TrueFISP.  Magn Reson Med. 2002;  48 594-601
  • 16 Wang S J, Hu B S, Macovski A, Nishimura D G. Coronary angiography using fast selective inversion recovery.  Magn Reson Med. 1991;  18 417-423
  • 17 Spuentrup E, Buecker A, Meyer J, Gunther R W, Stuber M. Navigator-gated free-breathing 3D balanced FFE projection renal MRA: Comparison with contrast-enhanced breath-hold 3D MRA in a swine model.  Magn Reson Med. 2002;  48 739-743

Elmar Spuentrup MD 

Department of Diagnostic Radiology, University Hospital, Aachen University of Technology

Pauwelsstraße 30

52057 Aachen

Germany

Phone: +49/241/8088332

Fax: +49/241/8082411

Email: spuenti@rad.rwth-aachen.de