Klinische Neurophysiologie 1997; 28(2): 80-88
DOI: 10.1055/s-2008-1060159
Originalarbeit

© Georg Thieme Verlag Stuttgart · New York

Moderner Stand des Einzelfaser- und des Makro-EMG

New techniques of single fiber EMG and macro EMGW. A. Nix, Th. Vogt
  • Neurologische Universitätsklinik Mainz
Further Information

Publication History

Publication Date:
18 March 2008 (online)

Summary

New techniques permit to study quantitatively action potentials of the motor unit at the single fibre level and at large. With new needle designs, EMG-machines and software it is today also possible to do single fibre-EMG and macro-EMG recordings in all EMG laboratories. Yet, these techniques are usually restricted to solve problems in specific situations. The knowledge of details on recording, technical pitfalls and anatomy, which are discussed here, provide a better and modern understanding of the dynamic changes within the motor unit in its healthy and diseased state.

Zusammenfassung

Neue Techniken erlauben, an der Muskulatur Aktionspotentiale von einzelnen oder allen Fasern einer motorischen Einheit quantitativ zu untersuchen. Einzelfaser- und Makro-EMG können heute mit speziell entwickelten Registriernadeln zusammen mit Auswertungsprogrammen in EMG-Geräten in jedem Labor ausgeführt werden. Diese Methoden bleiben jedoch der Problemlösung von speziellen Situationen vorbehalten. Die Kenntnis der Registrierbedingungen, der technischen Fallstricke und anatomischen Gegebenheiten, die hier dargestellt werden, führen zu einem besseren und moderneren Verständnis der dynamischen Veränderungen in der gesunden wie der pathologisch veränderten motorischen Einheit.

    >