Am J Perinatol 1992; 9(4): 239-246
DOI: 10.1055/s-2007-994780
ORIGINAL ARTICLE

© 1992 by Thieme Medical Publishers, Inc.

Assessment of Thermal Conditions in Neonatal Care: Use of a Manikin of Premature Baby Size

Ihsan Sarman, Dan Bolin, Ingvar Holmér, Ragnar Tunell
  • National Institute of Occupational Health, Division of Climate Physiology, and Department of Pediatrics, Huddinge Hospital, Huddinge, Sweden
Further Information

Publication History

Publication Date:
04 March 2008 (online)

ABSTRACT

A thermal manikin of the size of a 1 kg premature baby has been constructed. The nonevaporative heat loss from eight different regions and the total heat loss were measured. The measurements of heat loss have high repeatability and the values are in good agreement with measurements of dry heat loss for premature babies, using indirect calorimetry. The heat losses from the manikin in a single-walled, air-heated incubator and on a recently described heated, water-filled mattress have been compared. The total heat loss was found to be 20 to 30 W/m2 with both methods at ambient temperatures between 15° and 25°C. Treatment on a heated, water-filled mattress provides a means of direct conductive heat input to the baby, with a conductive heat transfer coefficient of 0.4 W/°C or 21 W/m2°C. The thermal manikin appears to provide an accurate method for assessment of the thermal conditions in neonatal care.

    >