Klin Monbl Augenheilkd 2007; 224(4): 252-254
DOI: 10.1055/s-2007-962849
Originalarbeit

© Georg Thieme Verlag KG Stuttgart · New York

Comparison of Tendency-Oriented Perimetry and Dynamic Strategy in Octopus Perimetry as a Screening Tool in a Clinical Setting: A Prospective Study

Praxisorientierter Vergleich der Strategien TOP und dynamisch in der statischen Perimetrie (Octopus): Eine prospektive StudieM. Scherrer1 , J. C. Fleischhauer1 , H. Helbig1 , K. Johann auf der Heide2 , F. K. Sutter1, 2
  • 1Augenklinik, UniversitätsSpital Zürich, Zürich, Switzerland (Direktorin: Prof. Dr. med. Klara Landau)
  • 2Augenarztpraxis, Herisau, Switzerland
Further Information

Publication History

received: 14.9.2006

accepted: 31.10.2006

Publication Date:
26 April 2007 (online)

Zusammenfassung

Hintergrund: Die statische Perimetrie erfordert je nach Strategie eine unterschiedlich lange Untersuchungszeit, bei dynamischer Strategie etwa 12 Minuten, bei tendenzorientierter Perimetrie (TOP) etwa 3 Minuten pro Auge. Ziel der Arbeit ist, abzuklären, ob sich TOP als Screeningmethode in der klinischen Routine eignet. Patienten und Methoden: Gesichtsfelder (GF) von 171 Augen wurden prospektiv jeweils mit beiden Methoden untersucht. Die GF wurden ohne weitere klinische Angaben nach den gängigen Kriterien von drei maskierten Beurteilern in die Kategorien normal, grenzwertig und pathologisch eingeteilt und verglichen. Ergebnisse: In 83 % der Augen erfolgte eine identische Beurteilung der beiden Untersuchungen. In 14 % erfolgte eine leicht unterschiedliche Beurteilung (ein GF normal bzw. pathologisch, das andere grenzwertig beurteilt). Nur gerade in 3 % (5 Fälle) erfolgte eine kontroverse Beurteilung. In 4 von 5 Fällen war das erste dynamische GF pathologisch, mit TOP normal. Bei 4 der 5 Fälle wurde der Verlauf kontrolliert und es zeigten sich in allen Fällen ein normales zweites dynamisches GF. Schlussfolgerungen: Die TOP Strategie ist eine schnelle, patientenfreundliche und zuverlässige Screeningmethode in der allgemein-ophthalmologischen Praxis. Durch die methodische Ungenauigkeit entstanden in der untersuchten Patientengruppe keine falsch negativen Resultate.

Abstract

Objective: The aim of this study was to compare the results of tendency-oriented perimetry (TOP) and a dynamic strategy in octopus perimetry as screening methods in clinical practice. Design: A prospective single centre observational case series was performed. Participants and Methods: In a newly opened general ophthalmologic practice 89 consecutive patients (171 eyes) with a clinical indication for octopus static perimetry testing (ocular hypertension or suspicious optic nerve cupping) were examined prospectively with TOP and a dynamic strategy. The visual fields were graded by 3 masked observers as normal, borderline or abnormal without any further clinical information. Results: 83 % eyes showed the same result for both strategies. In 14 % there was a small difference (with one visual field being abnormal or normal, the other being borderline). In only 2.9 % of the eyes (5 cases) was there a contradictory result. In 4 out of 5 cases the dynamic visual field was abnormal and TOP was normal. 4 of these cases came back for a second examination. In all 4 the follow-up examination showed a normal second dynamic visual field. Conclusions: Octopus static perimetry using a TOP strategy is a fast, patient-friendly and very reliable screening tool for the general ophthalmological practice. We found no false-negative results in our series.

References

  • 1 Bengtsson B, Heijl A, Olsson J. Evaluation of a new threshold visual field strategy, SITA, in normal subjects. Swedish Interactive Thresholding Algorithm.  Acta Ophthalmol Scand. 1998;  2 165-169
  • 2 Fiorelli V M. et al . Improved automated perimetry performance following exposure to Mozart.  Br J Ophthalmol. 2006;  5 543-545
  • 3 Gonzales de la Rosa M. et al . Multicenter evaluation of tendency-oriented perimetry (TOP) using the G1 grid.  Eur J Ophthalmol. 2003;  1 32-41
  • 4 Gonzalez-Hernandez M. et al . Comparison of diagnostic ability between a fast strategy, tendency-oriented perimetry, and the standard bracketing strategy.  Ophthalmologica. 2005;  6 373-378
  • 5 Gonzalez de la Rosa M. et al . Diagnostic accuracy and reproducibility of tendency oriented perimetry in glaucoma.  Eur J Ophthalmol. 2006;  2 259-267
  • 6 Hudson C, Wild J M, O’Neill E C. Fatigue effects during a single session of automated static threshold perimetry.  Invest Ophthalmol Vis Sci. 1994;  1 268-280
  • 7 Maeda H, Nakaura M, Negi A. New perimetric threshold test algorithm with dynamic strategy and tendency oriented perimetry (TOP) in glaucomatous eyes.  Eye. 2000;  14 Pt 5 747-751
  • 8 Morales J, Weitzman M L, Gonzalez de la Rosa M. Comparison between tendency-oriented perimetry (TOP) and octopus threshold perimetry.  Ophthalmology. 2000;  1 134-142
  • 9 Schimiti R B. et al . Full-threshold versus Swedish Interactive Threshold Algorithm (SITA) in normal individuals undergoing automated perimetry for the first time.  Ophthalmology. 2002;  109 (11) 2084-2092; discussion 2092

Dr. med. Florian K. P. Sutter

Augenklinik, UniversitätsSpital Zürich

Frauenklinikstrasse 24

8091 Zürich, Switzerland

Phone: ++41/44/2 55 11 11

Fax: ++41/44/2 55-44 72

Email: florian.sutter@usz.ch

    >