The Journal of Hip Surgery 2020; 4(04): 155-157
DOI: 10.1055/s-0040-1716575
Original Article

Assessment of an Orthopaedic Templating Software Package for Uncemented Total Hip Arthroplasty

1   Trauma and Orthopaedic Department, Royal Hampshire County Hospital, Winchester, Hampshire, United Kingdom
,
David Hartwright
1   Trauma and Orthopaedic Department, Royal Hampshire County Hospital, Winchester, Hampshire, United Kingdom
› Author Affiliations

Abstract

The aim of this study was to assess a large, single-surgeon dataset of uncemented total hip arthroplasties that had been templated using software. This assessment compared preoperative templated predictions with what was implanted with regard to (1) femoral prosthesis size and (2) acetabular cup size. The operation notes for two types of uncemented total hip arthroplasty performed by a single surgeon between January 9, 2008, and March 21, 2019, were assessed. The data were refined so that only those that were fully templated on a correctly calibrated pelvic radiograph were included. This provided a total of 153 cases. The predicted software template was compared with actual implanted component sizes. With regard to the femoral stems, the templating software was exactly correct in 45.75% of cases and was accurate to within one size in 87.59% of cases. With regard to the acetabular component, the templating software was exactly correct in 52.94% of cases and was accurate to within one size (2 mm) in 86.94% of cases. Templating software offers an accurate prediction of the size of uncemented femoral and acetabular components. In addition, the benefits of templating include cognitive rehearsal, insight into potentially challenging aspects of surgery, and highlighting intraoperative issues where there is a large intraoperative deviation from what is templated. This study demonstrates the congruence and reproducibility of templating software when compared with other smaller studies performed in the literature.



Publication History

Received: 25 February 2020

Accepted: 14 May 2020

Article published online:
16 September 2020

© 2020. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

 
  • References

  • 1 Newman M, Shaw G, Kane T. Assessment of an orthopaedic templating software package for cemented total hip arthroplasty. J Hip Surg. 2019; 3 (02) 59-61
  • 2 Vidalain JP, Selmi TAS, Beverland D. et al. The CORAIL® Hip System: A Practical Approach Based on 25 Years of Experience. Berlin: Springer; 2011
  • 3 Parwaiz H, Aliaga Crespo B, Filer J. et al. Improving the accuracy of digital templating: achieving success through stakeholder management. BMJ Open Qual 2019; 8 (02) e000378
  • 4 Si HB, Zeng Y, Cao F, Pei FX, Shen B. Accuracy of a simple digital templating in primary uncemented total hip arthroplasty. Chin Med Sci J 2015; 30 (03) 150-155
  • 5 Strøm NJ, Pripp AH, Reikerås O. Templating in uncemented total hip arthroplasty-on intra- and interobserver reliability and professional experience. Ann Transl Med 2017; 5 (03) 43
  • 6 Kovacevic D, Hodgins JL, Lowe DT. et al. Development and validation of cognitive rehearsal as a training strategy for arthroscopic surgery. Orthop J Sports Med 2016; 4 (07) (Suppl. 04) 2325967116S00130