Thromb Haemost 1995; 74(05): 1383-1384
DOI: 10.1055/s-0038-1649948
Letter To The Editor
Schattauer GmbH Stuttgart

Response of the Activated Partial Thromboplastin Time (APTT) to Heparin Is Influenced by Coagulometers

A M H P van den Besselaar
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
,
J Meeuwisse-Braun
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
,
A Strebus
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
,
H Schaefer-van Mansfeld
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
,
E Witteveen
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
,
F J M van der Meer
Haemostasis and Thrombosis Research Center, Dept. of Haematology, Leiden University Hospital, Leiden, The Netherlands
› Author Affiliations
Further Information

Publication History

Received 20 April 1995

Accepted after revision 13 July 1995

Publication Date:
10 July 2018 (online)

 
  • References

  • 1 Hirsh J. Heparin. N Engl J Med 1991; 324: 1565-1574
  • 2 Velde EA Van der, Poller L. The APTT monitoring of heparin - The ISTH/ICSH collaborative study. Thromb Haemost 1995; 73: 73-81
  • 3 Reed SV, Haddon ME, Denson KW E. An attempt to standardize the APTT for heparin monitoring, using the P.T. ISI/INR system of calibration Results of a 13 centre study. Thromb Res 1994; 74: 515-522
  • 4 D’Angelo A, Seveso MP, D’Angelo SV, Gilardoni F, Dettori AG, Bonini P. Effect of clot-detection methods and reagents on activated partial thromboplastin time (APTT). Implications in heparin monitoring by APTT Am J Clin Pathol 1990; 94: 297-306
  • 5 van den Besselaar AM H P, Meeuwisse-Braun J. Enzymatic elimination of heparin from plasma for activated partial thromboplastin time and prothrombin time testing. Blood Coag Fibrinol 1993; 4: 635-638
  • 6 Poggio M, Van den Besselaar AM P H, Van der Velde EA, Bertina RM. The effect of some instruments for prothrombin time testing on the International Sensitivity Index (ISI) of two rabbit tissue thromboplastin reagents. Thromb Haemost 1989; 62: 868-874
  • 7 Poller L, Thomson JM, Taberner DA, Clarke DK. The correction of coagulometer effects on international normalized ratios: a multicentre evaluation. Br J Haematol 1994; 86: 112-117
  • 8 Van den Besselaar AM P H, Meeuwisse-Braun J, Bertina RM. Monitoring heparin therapy: relationships between the activated partial thromboplastin time and heparin assays based on ex vivo heparin samples. Thromb Haemost 1990; 63: 16-23