Thromb Haemost 1978; 39(03): 640-645
DOI: 10.1055/s-0038-1646738
Original Article
Schattauer GmbH Stuttgart

Comparison of Reagents for Determining the Activated Partial Thromboplastin Time

J J M L Hoffmann
The Haemostasis Department, Clinical Laboratory, St. Elisabeth’s Hospital, Arnhem The Netherlands
,
P N Meulendijk
The Haemostasis Department, Clinical Laboratory, St. Elisabeth’s Hospital, Arnhem The Netherlands
› Author Affiliations
Further Information

Publication History

Received 12 August 1977

Accepted 04 November 1977

Publication Date:
12 July 2018 (online)

Summary

Six commercially available reagents for the determination of the activated partial thromboplastin time have been evaluated and compared with respect to their sensitivity to the coagulation factors VIII, IX and XI and to their response to heparin. Some variation was observed among the reagents regarding their sensitivity to factor XI and even greater differences were obtained with factors VIII and IX. It was also clear that none of the reagents was sensitive to the same extent to the factors tested. The sensitivity to heparin shows considerable variation, in terms of time as well as mode of response to increasing heparin levels. In four reagents this response is linear, it is logarithmic in one and the remaining one is yet again different.

It seems unlikely that any standardization of the APTT determination is at present possible with the reagents studied.

 
  • References

  • 1 Babson AL, Babson SR. 1974; Comparative evaluation of a partial thromboplastin reagent containing a non-settling, particulate activator. American Journal of Clinical Pathology 62: 856
  • 2 Hoffmann JJML, Meulendijk PN. 1978 Plasma heparin: a simple, one-sample determination of both effect and concentration based on the activated partial thromboplastin time. Clinica chimica acta, in press
  • 3 Shapiro GA, Huntzinger SW, Wilson JE. III 1977; Variation among commercial activated partial thromboplastin time reagents in response to heparin. American Journal of Clinical Pathology 67: 477
  • 4 Sibley C, Singer JW, Wood RJ. 1973; Comparison of activated partial thromboplastin reagents. American Journal of Clinical Pathology 59: 581
  • 5 Soloway HB, Cornett BM, Grayson jr JW. 1973; Comparison of various activated partial thromboplastin reagents in the laboratory control of heparin therapy. American Journal of Clinical Pathology 59: 587