A Novel Free Flap Monitoring System Using Tissue Oximetry with Text Message Alerts

Joseph A. Ricci
1   Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts
,
Christina R. Vargas
1   Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts
,
Samuel J. Lin
1   Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts
,
Adam M. Tobias
1   Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts
,
Amir H. Taghinia
2   Department of Plastic Surgery, Boston Children's Hospital, Boston, Massachusetts
,
Bernard T. Lee
1   Division of Plastic and Reconstructive Surgery, Beth Israel Deaconess Medical Center, Boston, Massachusetts
› Author Affiliations
Further Information

Publication History

31 January 2016

02 March 2016

Publication Date:
02 May 2016 (online)

Abstract

Background The time to detection of vascular compromise is a significant predictor of free flap salvage outcomes as early reexploration improves salvage rates. Continuous transcutaneous near-infrared tissue oximetry is an objective, quantitative method of detecting flap vascular compromise and has been shown to allow earlier reexploration and higher salvage rates than clinical assessment alone. We designed a novel text messaging system to improve communication using tissue oximetry monitoring.

Methods A retrospective review was performed of a prospectively collected database of all microsurgical breast reconstructions from 2008 to 2015. A novel text messaging system was introduced in 2013 and programmed to send text messages alert when the tissue oximetry readings suggested potential flap compromise based on established thresholds. Patient demographics and complications, including rate of reexploration and flap loss were assessed.

Results There were 900 autologous microsurgical breast free flaps during the study period: 614 were monitored with standard clinical monitoring and tissue oximetry compared with 286 flaps with the additional text messaging system. There were 27 unplanned returns to the operating room in the tissue oximetry group and 5 in the text messaging group with 1 complete flap loss in each group. Reexploration occurred sooner as a result of these text message alerts (17.5 vs. 26.6 hours postoperatively), however, it did not achieve statistical significance.

Conclusions We were able to demonstrate the use of a novel text messaging system for tissue oximetry. This alert system shows promise in identifying impending flap loss with rapid notification of the surgical team.

Note

Portions of this article were presented at the WSRM 2015 meeting, 8th World Society for Reconstructive Microsurgery Congress, March 19–22, 2015, Mumbai, India.


 
  • References

  • 1 Schusterman MA, Miller MJ, Reece GP, Kroll SS, Marchi M, Goepfert H. A single center's experience with 308 free flaps for repair of head and neck cancer defects. Plast Reconstr Surg 1994; 93 (3) 472-478 , discussion 479–480
  • 2 Hidalgo DA, Jones CS. The role of emergent exploration in free-tissue transfer: a review of 150 consecutive cases. Plast Reconstr Surg 1990; 86 (3) 492-498 , discussion 499–501
  • 3 Kroll SS, Schusterman MA, Reece GP , et al. Timing of pedicle thrombosis and flap loss after free-tissue transfer. Plast Reconstr Surg 1996; 98 (7) 1230-1233
  • 4 Kubo T, Yano K, Hosokawa K. Management of flaps with compromised venous outflow in head and neck microsurgical reconstruction. Microsurgery 2002; 22 (8) 391-395
  • 5 Irons GB, Wood MB, Schmitt III EH. Experience with one hundred consecutive free flaps. Ann Plast Surg 1987; 18 (1) 17-23
  • 6 Krag C. Experience with transplantation of composite tissues by means of microsurgical vascular anastomoses. I. Indications, techniques and early results. Scand J Plast Reconstr Surg 1985; 19 (2) 135-156
  • 7 Lineaweaver WC, Buncke HJ. Complications of free flap transfers. Hand Clin 1986; 2 (2) 347-351
  • 8 Macnamara M, Pope S, Sadler A, Grant H, Brough M. Microvascular free flaps in head and neck surgery. J Laryngol Otol 1994; 108 (11) 962-968
  • 9 Creech BJMS. Evaluation of Circulation in Skin Flaps. Boston, MA: Little, Brown; 1975. . PMID:17530610
  • 10 Keller A. Noninvasive tissue oximetry for flap monitoring: an initial study. J Reconstr Microsurg 2007; 23 (4) 189-197
  • 11 Cheng X, Mao JM, Bush R, Kopans DB, Moore RH, Chorlton M. Breast cancer detection by mapping hemoglobin concentration and oxygen saturation. Appl Opt 2003; 42 (31) 6412-6421
  • 12 Keller A. A new diagnostic algorithm for early prediction of vascular compromise in 208 microsurgical flaps using tissue oxygen saturation measurements. Ann Plast Surg 2009; 62 (5) 538-543
  • 13 Lin SJ, Nguyen MD, Chen C , et al. Tissue oximetry monitoring in microsurgical breast reconstruction decreases flap loss and improves rate of flap salvage. Plast Reconstr Surg 2011; 127 (3) 1080-1085
  • 14 Jonas R, Schaal T, Krimmel M, Gülicher D, Reinert S, Hoffmann J. Monitoring in microvascular tissue transfer by measurement of oxygen partial pressure: four years experience with 125 microsurgical transplants. J Craniomaxillofac Surg 2013; 41 (4) 303-309
  • 15 Steele MH. Three-year experience using near infrared spectroscopy tissue oximetry monitoring of free tissue transfers. Ann Plast Surg 2011; 66 (5) 540-545
  • 16 Disa JJ, Cordeiro PG, Hidalgo DA. Efficacy of conventional monitoring techniques in free tissue transfer: an 11-year experience in 750 consecutive cases. Plast Reconstr Surg 1999; 104 (1) 97-101
  • 17 Chen KT, Mardini S, Chuang DC , et al. Timing of presentation of the first signs of vascular compromise dictates the salvage outcome of free flap transfers. Plast Reconstr Surg 2007; 120 (1) 187-195
  • 18 Koolen PG, Vargas CR, Ho OA , et al. Does increased experience with tissue oximetry monitoring in microsurgical breast reconstruction lead to decreased flap loss? The learning effect. Plast Reconstr Surg 2016; 137 (4) 1093-1101
  • 19 Smit JM, Zeebregts CJ, Acosta R, Werker PM. Advancements in free flap monitoring in the last decade: a critical review. Plast Reconstr Surg 2010; 125 (1) 177-185
  • 20 Pelletier A, Tseng C, Agarwal S, Park J, Song D. Cost analysis of near-infrared spectroscopy tissue oximetry for monitoring autologous free tissue breast reconstruction. J Reconstr Microsurg 2011; 27 (8) 487-494