Utilizing Indocyanine Green Dye Angiography to Detect Simulated Flap Venous Congestion in a Novel Experimental Rat Model

Ahmed Nasser
1   Department of Surgery, Stony Brook University Medical Center, Stony Brook, New York
,
Mitchell S. Fourman
2   Department of Orthopaedic Surgery, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania
,
Robert P. Gersch
3   Department of Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania
,
Brett T. Phillips
4   Division of Plastic Surgery, Department of Surgery, Duke University Medical Center, Durham, North Carolina
,
Hsingli Kai Hsi
5   School of Medicine, Stony Brook University Medical Center, Stony Brook, New York
,
Sami U. Khan
6   Division of Plastic Surgery, Department of Surgery, Stony Brook University Medical Center, Stony Brook, New York
,
Mark A. Gelfand
6   Division of Plastic Surgery, Department of Surgery, Stony Brook University Medical Center, Stony Brook, New York
,
Alexander B. Dagum
6   Division of Plastic Surgery, Department of Surgery, Stony Brook University Medical Center, Stony Brook, New York
,
Duc T. Bui
6   Division of Plastic Surgery, Department of Surgery, Stony Brook University Medical Center, Stony Brook, New York
› Author Affiliations
Further Information

Publication History

10 February 2015

10 May 2015

Publication Date:
01 September 2015 (online)

Abstract

Background Venous congestion is a leading cause for free flap failure and still relies on clinical observation as the diagnostic gold standard. We sought to characterize blood flow in a variable venous congestion murine hind limb model using indocyanine green (ICG, SPY Pack, LifeCell, Branchburg, NJ) angiography.

Methods Male Sprague–Dawley rats (Charles River, Hudson, NY) underwent bilateral partial amputation at the inguinal ligament, leaving only the femoral vessels and femur intact. Complete unilateral venous occlusion was achieved via suture ligation, while partial occlusion was achieved by surrounding the femoral vein with a synthetic microtube to achieve 25, 75, 85, or 92% occlusion. Relative blood flow of occluded and control limbs was tracked with ICG angiography throughout a 90-minute time course.

Results ICG angiography detected statistically significant (p < 0.05) reductions in limb blood flow 1 and 2 minutes following ICG injection in the 100, 92, and 85% occluded limbs when compared with contralateral control limbs. Dynamic tracking using the slope of ICG inflow for 45 seconds postinjection reflected this same significant difference. No statistically significant change in limb blood flow or dye influx rate was observed in the 25 and 75% occlusion groups.

Conclusions ICG angiography can detect venous congestion in a rat lower extremity model reliably at occlusion rates ≥ 85%. This method may offer surgeons an intraoperative diagnostic tool to identify venous congestion at extremely early time points, allowing for immediate intervention. Further investigation and characterization is warranted in a larger animal model before clinical adaptation.

Note

This article was presented as LOP22: Utilizing indocyanine green angiography in the evaluation of varying levels of venous congestion in a novel rat model. Paper presented at: 6th European Plastic Surgery Research Council; August 21–24, 2014; Hamburg, Germany; 133: Indocyanine green angiography predicts flap pedicle venous congestion within one minute of occlusion in rat models. 58th Annual Meeting of Plastic Surgery Research Council; May 2–4, 2013; Santa Monica, CA; and Can laser assisted indocyanine green dye angiography detect venous congestion more reliably than clinical assessment in a rat venous occlusion model. 30th Annual Meeting of Northeastern Society of Plastic Surgeons; September 20–22, 2013; Washington, DC.


 
  • References

  • 1 Dagum AB, Dowd AJ. Simple monitoring technique for muscle flaps. Microsurgery 1995; 16 (11) 728-729
  • 2 Arya R, Griffiths L, Figus A, King D, Ramakrishnan V, Griffiths M. Post-operative assessment of perfusion of Deep Inferior Epigastric Perforator (DIEP) free flaps via Pulsatility Index (PI) using a portable colour Doppler sonogram device. J Plast Reconstr Aesthet Surg 2013; 66 (7) 931-936
  • 3 Talbot SG, Pribaz JJ. First aid for failing flaps. J Reconstr Microsurg 2010; 26 (8) 513-515
  • 4 Galanis C, Nguyen P, Koh J, Roostaeian J, Festekjian J, Crisera C. Microvascular lifeboats: a stepwise approach to intraoperative venous congestion in DIEP flap breast reconstruction. Plast Reconstr Surg 2014; 134 (1) 20-27
  • 5 Ali R, Bernier C, Lin YT , et al. Surgical strategies to salvage the venous compromised deep inferior epigastric perforator flap. Ann Plast Surg 2010; 65 (4) 398-406
  • 6 Mirzabeigi MN, Wang T, Kovach SJ, Taylor JA, Serletti JM, Wu LC. Free flap take-back following postoperative microvascular compromise: predicting salvage versus failure. Plast Reconstr Surg 2012; 130 (3) 579-589
  • 7 Chen KT, Mardini S, Chuang DC , et al. Timing of presentation of the first signs of vascular compromise dictates the salvage outcome of free flap transfers. Plast Reconstr Surg 2007; 120 (1) 187-195
  • 8 Beier JP, Horch RE, Arkudas A, Dragu A, Schmitz M, Kneser U. Decision-making in DIEP and ms-TRAM flaps: the potential role for a combined laser Doppler spectrophotometry system. J Plast Reconstr Aesthet Surg 2013; 66 (1) 73-79
  • 9 Sacks JM, Nguyen AT, Broyles JM, Yu P, Valerio IL, Baumann DP. Near-infrared laser-assisted indocyanine green imaging for optimizing the design of the anterolateral thigh flap. Eplasty 2012; 12: e30
  • 10 Fourman MS, McKenna P, Phillips BT , et al. ICG angiography predicts burn scarring within 48h of injury in a porcine vertical progression burn model. Burns 2014;
  • 11 Schols RM, Connell NJ, Stassen LP. Near-Infrared Fluorescence Imaging for Real-Time Intraoperative Anatomical Guidance in Minimally Invasive Surgery: A Systematic Review of the Literature. World J Surg 2015; 39 (5) 1069-1079
  • 12 Nguyen JT, Ashitate Y, Buchanan IA , et al. Face transplant perfusion assessment using near-infrared fluorescence imaging. J Surg Res 2012; 177 (2) e83-e88
  • 13 Ashitate Y, Lee BT, Ngo LH , et al. Quantitative assessment of nipple perfusion with near-infrared fluorescence imaging. Ann Plast Surg 2013; 70 (2) 149-153
  • 14 Polom K, Murawa D, Rho YS, Spychala A, Murawa P. Skin melanoma sentinel lymph node biopsy using real-time fluorescence navigation with indocyanine green and indocyanine green with human serum albumin. Br J Dermatol 2012; 166 (3) 682-683
  • 15 Anderson RR, Parrish JA. The optics of human skin. J Invest Dermatol 1981; 77 (1) 13-19
  • 16 Lee BT, Hutteman M, Gioux S , et al. The FLARE intraoperative near-infrared fluorescence imaging system: a first-in-human clinical trial in perforator flap breast reconstruction. Plast Reconstr Surg 2010; 126 (5) 1472-1481
  • 17 Fourman MS, Phillips BT, Crawford L , et al. Indocyanine green dye angiography accurately predicts survival in the zone of ischemia in a burn comb model. Burns 2014; 40 (5) 940-946
  • 18 Phillips BT, Lanier ST, Conkling N , et al. Intraoperative perfusion techniques can accurately predict mastectomy skin flap necrosis in breast reconstruction: results of a prospective trial. Plast Reconstr Surg 2012; 129 (5) 778e-788e
  • 19 Sood M, Glat P. Potential of the SPY intraoperative perfusion assessment system to reduce ischemic complications in immediate postmastectomy breast reconstruction. Ann Surg Innov Res 2013; 7 (1) 9
  • 20 Matsui A, Lee BT, Winer JH, Vooght CS, Laurence RG, Frangioni JV. Real-time intraoperative near-infrared fluorescence angiography for perforator identification and flap design. Plast Reconstr Surg 2009; 123 (3) 125e-127e
  • 21 Matsui A, Lee BT, Winer JH, Laurence RG, Frangioni JV. Quantitative assessment of perfusion and vascular compromise in perforator flaps using a near-infrared fluorescence-guided imaging system. Plast Reconstr Surg 2009; 124 (2) 451-460
  • 22 Matsui A, Lee BT, Winer JH, Laurence RG, Frangioni JV. Submental perforator flap design with a near-infrared fluorescence imaging system: the relationship among number of perforators, flap perfusion, and venous drainage. Plast Reconstr Surg 2009; 124 (4) 1098-1104
  • 23 Schalch P, Patejunas G, Retuerto M , et al. Homozygous deletion of early growth response 1 gene and critical limb ischemia after vascular ligation in mice: evidence for a central role in vascular homeostasis. J Thorac Cardiovasc Surg 2004; 128 (4) 595-601
  • 24 Liu HL. Microvascular anastomosis of submillimeter vessels-a training model in rats. J Hand Microsurg 2013; 5 (1) 14-17
  • 25 Tenorio X, Mahajan AL, Wettstein R, Harder Y, Pawlovski M, Pittet B. Early detection of flap failure using a new thermographic device. J Surg Res 2009; 151 (1) 15-21
  • 26 Matsui A, Lee BT, Winer JH, Kianzad V, Frangioni JV. Image-guided perforator flap design using invisible near-infrared light and validation with x-ray angiography. Ann Plast Surg 2009; 63 (3) 327-330
  • 27 Matsui A, Lee BT, Winer JH, Laurence RG, Frangioni JV. Predictive capability of near-infrared fluorescence angiography in submental perforator flap survival. Plast Reconstr Surg 2010; 126 (5) 1518-1527