Minim Invasive Neurosurg 2009; 52(4): 201-203
DOI: 10.1055/s-0029-1239588
Technical Note

© Georg Thieme Verlag KG Stuttgart · New York

Technique for Shaping Microcatheter Tips in Coil Embolization of Paraclinoid Aneurysms using Full-Scale Volume Rendering Images of 3D Rotational Angiography

S. Toyota1 , Y. Fujimoto1 , F. Iwamoto1 , A. Wakayama1 , T. Yoshimine2
  • 1Center for Endovascular Neurosurgery, Osaka Neurological Institute, Osaka, Japan
  • 2Department of Neurosurgery, Osaka University Medical School, Osaka, Japan
Further Information

Publication History

Publication Date:
16 October 2009 (online)

Abstract

Introduction: In coil embolization of paraclinoid aneurysms, it is sometimes difficult to introduce and stabilize microcatheter tips in the aneurysms. We report a new technique for shaping microcatheter tips in the coil embolization of paraclinoid aneurysms.

Methods: From May 2007 to May 2008, this new technique was applied to 10 paraclinoid aneurysms undergoing coil embolization. Before coil embolization, 3D rotational angiography was performed, and volume-rendering images were reconstructed. Vinyl-coated handicraft wire was shaped 3-dimensionally to fit full-scale volume-rendering images on the monitor, from the C5 portion of the internal carotid artery to the center of the dome of the aneurysm from various angles. The microcatheter tip was then shaped with steam to fit the vinyl-coated wire. Thereafter, the microcatheter tip was introduced into the aneurysm and coil embolization was performed.

Results: Microcatheter tips could be easily shaped and could be introduced smoothly into aneurysms, and were stable during coil embolization.

Conclusion: This technique is feasible for shaping microcatheter tips precisely for coil embolization of paraclinoid aneurysms.

References

  • 1 Thornton J, Aletich VA, Debrun GM. et al . Endovascular treatment of paraclinoid aneurysms.  Surg Neurol. 2000;  54 288-299
  • 2 Park HK, Horowitz M, Jungreis C. et al . Endovascular treatment of paraclinoid aneurysms: experience with 73 patients.  Neurosurgery. 2003;  53 14-23
  • 3 Abe T, Hirohata M, Tanaka N. et al . Distal-tip shape-consistency testing of steam-shaped microcatheters suitable for cerebral aneurysm coil placement.  AJNR Am J Neuroradiol. 2004;  25 1058-1061
  • 4 Pakbaz RS, Kerber CW. Complex curve microcatheters for berry aneurysm endovascular therapy.  AJNR Am J Neuroradiol. 2007;  28 179-180
  • 5 Kiyosue H, Hori Y, Matsumoto S. et al . Shapability, memory, and luminal changes in microcatheters after steam shaping: a comparison of 11 different microcatheters.  AJNR Am J Neuroradiol. 2005;  26 2610-2616
  • 6 Lefkowitz MA, Gobin YP, Akiba Y. et al . Balloon-assisted Guglielmi detachable coiling of wide-necked aneurysma: Part II − clinical results.  Neurosurgery. 1999;  45 531-537
  • 7 Kwon OK, Han MH. Marker alignment for Guglielmi detachable coil embolization: practical considerations.  AJNR Am J Neuroradiol. 2002;  23 1276-1283

Correspondence