Subscribe to RSS

DOI: 10.5935/1984-0063.20180040
Patients' preferences and the efficacy of a hybrid model of a minimal contact nasal mask in patients with sleep apnea treated with CPAP
Objectives To evaluate patient's satisfaction, efficacy and adherence to CPAP with a hybrid nasal mask (DW) we tested patients with OSA in unattended setting under real-life conditions.
Material and Methods Prospective, comparative study using DW mask 7 days against habitual mask in patients adapted to CPAP therapy.
Results We analyzed 52 patients: 35 men (67%) with IAH mean; 24.3±12.3 events/hour. At baseline mean compliance of 5.42±1.83 hours/night. After using DW mask, patients reported fewer marks, more comfort, greater partners acceptance, easier to use and was ranked higher to preventing leaks; p<0.05, and adherence (1 more hour per night,p>0.0042).Differences were not found in pressure 90th-95th percentile (9.6±9.2 cm of H2O, p<0.5), leaks (19.8±17 liters/min. p<0.37) or residual AHI (3.38±3.05 events/hour. p<0.93).
Conclusion In an uncontrolled non-randomized study, patients can use DW mask with similar leak level and better adherence as compared to conventional masks.
#
INTRODUCTION
Obstructive Sleep Apnea-Hypopnea Syndrome (OSA) is characterized by repeated obstructions of the upper airway with oxygen desaturation and fragmented sleep[1]. According to recent data, its prevalence in adult population of Latin America is > 32%[2].
If untreated, OSA may be associated with excessive daytime sleepiness, poor quality of life, medium-to-long term cardiovascular and cerebrovascular complications, and higher mortality rates[3].
Continuous positive airway pressure (CPAP) is the current gold standard to treat moderate-to-severe OSA[4] , [5]. When properly administered, CPAP therapy can relieve symptoms and improve quality of life[3] - [5]. CPAP therapy has been proposed to reduce cardiovascular risk, however recent studies have shown negative results in this regard and have shown the low adherence of patients to therapy in the medium term[6].
CPAP efficacy depends largely on proper use. When systematic programs of education and training are implemented in the use of CPAP, individual or group, it has been observed that compliance to treatment at the first year can exceed 80%[7] , [8].
The percentage of patients with medium-to-long term adherence varies considerably and is subject to complex factors, such as OSA severity, symptom perception, level of education and socioeconomic status[5] - [8]. Compliance also depends on a good first experience; therefore, the first week of CPAP therapy becomes crucial[9] - [12]. It is important to objectively estimate CPAP usage rate, since it has been shown that patients overestimate average CPAP use time in more than one hour per night[10]. Currently available CPAP devices have their own internal memory includes information on hours of effective use, pressure levels (per night), leaks, and residual respiratory events.
CPAP therapy success largely depends on selecting the right mask and training patients on proper use. The right size is very important to reduce leaks, red marks, and discomfort. At present, there is a large variety of masks, including oronasal masks, which cover patient's nose and mouth, nasal masks, which cover the area from the bridge of the nose to the upper lip; and nasal pillow masks, which have two nostril inserts. Nasal pillows have emerged as an alternative to nasal masks because they are smaller and have less contact with the face.
Oronasal masks are considered the best option for patients in need of CPAP therapy who suffer some degree of nasal obstruction or documented mouth leaks[13] - [15]. Since most patients prefer nasal masks, several models are available on the market. However, the level of scientific evidence necessary to validate the efficacy of these new models and their impact on patient's adherence to treatment is being questioned[14] - [16].
Borel et al.[17] conducted a descriptive study on 2,311 OSA patients and found statistically significant differences between the 3 types of masks. In terms of effective pressure during CPAP titration, the best results were obtained with oronasal masks, followed by nasal masks, and nasal pillows. Compliance, however, was higher with nasal masks.
A recent literature review revealed 2 well-designed studies comparing the 3 types of masks in terms of CPAP efficacy[16] , [17]. Ebben el al.[18] evaluated 55 OSA patients and randomized them to CPAP titration with oronasal mask, nasal mask, or nasal pillow. The last 2 masks were similar in terms of CPAP levels, but consistently lower then oronasal ones. No differences were found in terms of residual AHI. One study[17] reported better adherence with nasal pillows than oronasal masks but another that assessed the combination of efficacy, patient's satisfaction, and adherence found adherence was lower with nasal pillows[19].
During this study, we evaluated patient's satisfaction, efficacy, and adherence to CPAP therapy with a hybrid model similar to a nasal pillow but without nostril inserts in-home unattended setting under real-life conditions.
Objective To evaluate patient satisfaction and preferences as well as the therapeutic efficacy an compliance of a hybrid nasal mask in OSA patients previously adapted to CPAP or automatic CPAP therapy.
#
METHODS
Design
Prospective, comparative, interventional, with unblinded follow-up.
#
Sample
This study enrolled patients diagnosed with OSA under CPAP therapy in 2 sleep units of university hospitals in Buenos Aires city during a 6-month period (from June to December 2017).
#
Inclusion criteria
Our study included > 18 year-old male and female patients who: 1) had been diagnosed with OSA by laboratory polysomnography or home-based respiratory polygraphy at least 6 months before enrollment and already under regular treatment with CPAP or automatic CPAP (> 5 nights per week, regardless of number of hours per night), 2) lived in the metropolitan area of Buenos Aires city and received routine care at the sleep units of participating sites and, 3) owned a CPAP or automatic CPAP machine with internal memory (i.e. SD card or similar) and capacity to record objective use, leaks, and residual events. All the patients included signed an informed consent form.
#
Exclusion Criteria
We excluded patients: 1) with OSA diagnosis and indication for CPAP therapy who do not use the machine, 2) with chronic respiratory failure who need treatment with O2 or other type of home-based mechanical ventilation, 3) with a terminal condition or who receive intensive care at home for chronic illnesses, 4) with psychiatric and/or cognitive disorders and difficulties to understand and take part in study procedures, 5) with OSA who do not have their own CPAP machine and 6) who are already using the hybrid mask model under study.
#
Study Procedures
1st Visit: Patients were recruited during their visits to intervening sleep units. During the first visit, the following information was gathered through interviews and medical histories: anthropometric data, symptoms (daytime sleepiness and sleep apnea symptoms), comorbidities, sleep disorder severity and indicators, and treatment characteristics. The initial visit included the following procedures:
CPAP machine examination and data collection about usage, leaks, and treatment efficacy for the last week (Residual Apnea-Hypopnea Index, AHIr). Data were downloaded using specific software (Encore Pro II Philips-Respironics. USA) and ResScan (ResMed. Sidney, Australia). Patients also reported the type of mask they were using at enrollment.
Self-administered questionnaire about patient's level of satisfaction with their usual mask, perceived results, and treatment efficacy. This questionnaire was designed by study authors and is made up of 13 questions to be answered quantitively (1 to 10 visual numeric scale).
Epworth Sleepiness Scale (ESS).
#
Study nasal Mask
After patients signed the informed consent form, 2 physiotherapists specialized in sleep medicine made a 20-minute demonstration of how to use the new minimal-contact DreamWear™ nasal mask (Philips-Respironics. USA) without nostril inserts (DW). The best fitting model was delivered to each patient. No changes were made in pressure, mode, and settings of patients' CPAP machines.
2nd Visit: The following procedures were conducted after 1 week of use:
CPAP machines were checked, and the following data were collected: use, leaks, and treatment efficacy (AHIr).
The self-administered questionnaire designed by study authors was distributed among patients.
Epworth Sleepiness Scale (ESS).
#
Indicators and metrics of CPAP machine memory as a study objective
Investigators calculated the average values for usage (hours/night and nights/week), leaks, and residual events for the last week of baseline treatment and the intervention week (study mask). Effective pressure values in the 90th (Philips devices) or 95th (ResMed devices) percentile (P90 and P95, respectively) were obtained using the corresponding software for each device.
#
Statistical Analysis
Categorical and numerical variables are presented as percentages or means ± standard deviation respectively. Differences were compared using Fisher's exact test, Mann-Whitney or χ2 test. Statistical analysis was conducted using GraphPad Prism-5™ software.
#
Ethical Considerations
The protocol was approved by the Institutional Review Board and the Ethics Committee of participating sites (number protocol: #839). All patients signed a written informed consent. The identity of every patient was protected. At the end of the study, patients could decide whether to continue using their previous mask or not.
#
#
RESULTS
55 patients were enrolled. Data were collected and analyzed for the 52 patients who completed the protocol: 35 men (67%) and 17 women (69%). Their characteristics were; Age: 65.2±9.9 years, Body Mass Index (kg/m2); 32.5±6.5, Epworth Sleepiness Scale basal value; 6.9±5.5, and AHI; 24.3±12.3 events/hour.
The most common comorbidities were arterial hypertension (57.6%), diabetes mellitus (15%), and the use of hypnotics or sedatives drugs (21%). 65% of subjects wore glasses, 15% had a beard, and 10% a mustache or long hair.
Baseline treatment with CPAP
At baseline, mean CPAP usage time was 8 months (range; 3-84 months) and mean baseline compliance was of 5.42 ± 1.83 hours/night across the entire population. 33% used fixed CPAP and 67% automatic CPAP. The mean prescribed CPAP pressure was 8.5±1.1 cm of H2O for fixed CPAP machines and ranged from 4±2 (minimum) to 14.4±2.4 (maximum) for automatic CPAP machines. Baseline leaks were 18.3±14.2 liters/min and 5 patients (20%) had leaks above the maximum compensation limit recommended by manufacturer (42.3 ± 11.2 liters/min) during the observation period. The study mask was compared with 12 nasal models. [Table 1] show types of commonly used masks.
Original Interface |
n |
% |
---|---|---|
Nasal |
43 |
82.6 |
Oronasal |
1 |
2 |
Pillow |
8 |
15 |
Most frequently used models |
|
|
Pico |
10 |
19.2 |
Easy Life |
9 |
17 |
Mirage Fx |
9 |
17 |
Comfort Classic |
5 |
9 |
Wisp |
4 |
7.6 |
Comfort Gel |
5 |
9 |
Others |
10 |
19.2 |
#
Subjective Questionnaires
The comparison between patients' previous masks and the DW mask revealed statistically significant differences in the visual numeric scale. After using the DW mask for one week, patients reported fewer red marks in their faces, more comfort, greater partner's acceptance, and stated the DW mask was more appealing ([Figure 1]) and easier to use (7.6 vs. 9.0; p<0.001). The DW model was ranked higher than other models in terms of preventing leaks (7.88 vs. 6.32;p<0.05 - [Figure 2A]).




The most frequents adverse events reported with the study mask were; leak towards the eyes (11.5%), face irritation 7.7% and rhinitis (13.4%). At the end of the study, 8 patients (15.4%) decided to go back to their previous mask.
At baseline, 25% of patients had ESS > 11. After using the study mask for seven days, this percentage fell by 11%.
#
Objective data from CPAP machine's memory
After one week of use, no statistically significant differences were found in pressure values between the 90th-95th percentile (9.6±9.2 cm of H2O; p=0.5), leaks (19.8±17 Liters/min; p=0.37) or mean residual AHI (3.38±3.05 events/hour; p=0.93) in patients using automatic CPAP ([Table 2]).
|
|
Routine Mask |
DW Mask |
|
---|---|---|---|---|
|
n |
Mean /SD (±) |
Mean / SD (±) |
Fisher Test (p) |
Effective pressure Pth 90-95% |
35 |
9.6±2.9 |
9.2±2.6 |
0.5 |
Compliance (hours) |
52 |
5.4±1.8 |
6.4±1.3 |
0.0042 |
Mean leak (liters /min) |
52 |
19.8±15.6 |
17±14.8 |
0.37 |
Residual AHI (events/hour) |
52 |
3.38 ± 3.59 |
3.05±2.7 |
0.93 |
Epworth Sleepiness Scale (ESS) |
52 |
6.9±5.5 |
4.7±4.8 |
0.01 |
Statistically significant differences were observed in terms of treatment adherence, after using the study mask for seven days (p>0.004) [Figure 2B], and ESS > 11 (25% vs. 11%; p<0.05).
#
#
DISCUSSION
According to this pilot study conducted in a real-life setting the use of the new minimal-contact nasal mask is feasible and achieves better treatment adherence. In terms of efficacy, the new mask is comparable to other nasal masks available in the market and is well-accepted by patients. In our experience, patients already adapted to CPAP therapy were able to use the new mask properly after a basic demonstration performed at the sleep unit.
Data from other mask designs are not fully comparable and cannot be extrapolated to the new model, which has no nostril inserts. As mentioned before, two studies described the impact of mask types (nasal mask versus nasal pillow with nostril inserts) in CPAP therapy adherence and efficacy with inconsistent results[15] - [19].
According to the literature, oronasal masks may require higher effective titration pressure than smaller models[17] , [20] - [21]. This difference, however, was not observed when comparing nasal masks to this hybrid nasal pillow. In terms of P90 or P95, our study found no difference between the study mask and the home-based automatic CPAP models used for several nights by patients already adapted to CPAP therapy.
Residual AHI (AHIr) has been validated as a metric to infer the efficacy of CPAP therapy as < 5 events/hour correlate well with the results obtained by polysomnography and suggest proper treatment of the disorder[22]. We found that 7 nights using the DW mask did not increase AHIr.
A recent study with a similar design with nasal pillows and found no short/long-term differences[23]. After 7 days, we observed greater usage time with the DW mask (1 hour more on average). This finding, however, calls for a cautious interpretation since it may derive from the effect of the visit to the sleep unit, the training demonstration, and memory card surveillance.
Despite these problems, ESS scores were lower, and most patients seemed willing to continue using the minimal-contact mask after the 7-day trial period. This suggests that it is feasible to use the new mask model in real-life settings.
Minimal-contact masks offer an alternative to conventional nasal masks and seem effective for OSA treatment. In addition, these masks are lighter and could be better accepted by claustrophobic patients, patients with a beard or mustache, and patients who want to wear glasses while using the mask[24].
Limitations
Our study has some limitations. Firstly, we have a heterogeneous control group because patients used different types of masks and CPAP machines. Results refer only to the first week of use and, therefore, it would be necessary to conduct mid-term studies to verify our results. In addition, all the patients knew that they would use a new test mask and this could introduce biases. Patients reporting fewer marks and more comfort may be due to better instructions by the research team, regarding principally the subjective assessments, but also the objective outcomes. Our results cannot be extrapolated to patients not adapted to CPAP treatment and devices. Finally, AHIr and leak metrics are not consistent across manufacturers making result interpretation more difficult.
#
#
CONCLUSION
In an uncontrolled non-randomized study, patients already adapted to CPAP therapy can use the new DW mask properly and effectively treat their disorder with a similar leak level and better adherence (1 hour more after 7 days) as compared to conventional masks.
#
#
Conflict of Interests
The authors have no conflict of interests to declare.
-
REFERENCES
- 1 Young T, Peppard PE, Gottlieb DJ. Epidemiology of obstructive sleep apnea: a population health perspective. Am J Respir Crit Care Med. 2002;165(9):1217-39.
- 2 Tufik S, Santos-Silva R, Taddei JA, Bittencourt LR. Obstructive sleep apnea syndrome in the Sao Paulo Epidemiologic Sleep Study. Sleep Med. 2010;11(5):441-6.
- 3 Marin JM, Carrizo SJ, Vicente E, Agusti AG. Long-term cardiovascular outcomes in men with obstructive sleep apnoea-hypopnoea with or without treatment with continuous positive airway pressure: an observational study. Lancet. 2005;365(9464):1046-53.
- 4 Kushida CA, Littner MR, Hirshkowitz M, Morgenthaler TI, Alessi CA, Bailey D, et al.; American Academy of Sleep Medicine. Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders. Sleep. 2006;29(3):375-80.
- 5 Giles TL, Lasserson TJ, Smith BH, White J, Wright J, Cates CJ. Continuous positive airway pressure for obstructive sleep apnoea in adults. Cochrane Database Syst Rev. 2006;(3):CD001106.
- 6 McEvoy RD, Antic NA, Heeley E, Luo Y, Ou Q, Zhang X, et al.; SAVE Investigators and Coordinators. CPAP for Prevention of Cardiovascular Events in Obstructive Sleep Apnea. N Engl J Med. 2016;375(10):919-31.
- 7 Haniffa M, Lasserson TJ, Smith I. Interventions to improve compliance with continuous positive airway pressure for obstructive sleep apnoea. Cochrane Database Syst Rev. 2004;(4):CD003531.
- 8 Meurice JC, Ingrand P, Portier F, Arnulf I, Rakotonanahari D, Fournier E, et al.; ANTADIR Working Group “PPC”, CMTS ANTADIR. A multicentre trial of education strategies at CPAP induction in the treatment of severe sleep apnoea-hypopnoea syndrome. Sleep Med. 2007;8(1):37-42.
- 9 Billings ME, Auckley D, Benca R, Foldvary-Schaefer N, Iber C, Redline S, et al.; Race and residential socioeconomics as predictors of CPAP adherence. Sleep. 2011;34(12):1653-8.
- 10 Kohler M, Smith D, Tippett V, Stradling JR. Predictors of long-term compliance with continuous positive airway pressure. Thorax. 2010;65(9):829-32.
- 11 Weaver TE, Grunstein RR. Adherence to continuous positive airway pressure therapy: the challenge to effective treatment. Proc Am Thorac Soc. 2008;5(2):173-8.
- 12 Ayas NT, FitzGerald JM, Fleetham JA, White DP, Schulzer M, Ryan CF, et al. Cost-effectiveness of continuous positive airway pressure therapy for moderate to severe obstructive sleep apnea/hypopnea. Arch Intern Med. 2006;166(9):977-84.
- 13 Kribbs NB, Pack AI, Kline LR, Smith PL, Schwartz AR, Schubert NM, et al. Objective measurement of patterns of nasal CPAP use by patients with obstructive sleep apnea. Am Rev Respir Dis. 1993;147(4):887-95.
- 14 Olsen KD, Kern EB, Westbrook PR. Sleep and breathing disturbances secondary to nasal obstruction. Otolaryngol Head Neck Surg. 1981;89(5):804-10.
- 15 Andrade RG, Piccin VS, Nascimento JA, Viana FM, Genta PR, Lorenzi- Filho G. Impact of the type of mask on the effectiveness of and adherence to continuous positive airway pressure treatment for obstructive sleep apnea. J Bras Pneumol. 2014;40(6):658-68.
- 16 Andrade RGS, Viana FM, Nascimento JA, Drager LF, Moffa A, Brunoni AR, et al. Nasal vs Oronasal CPAP for OSA Treatment: A Meta-Analysis. Chest. 2018;153(3):665-74.
- 17 Borel JC, Tamisier R, Dias-Domingos S, Sapene M, Martin F, Stach B, et al.; Scientific Council of The Sleep Registry of the French Federation of Pneumology (OSFP). Type of mask may impact on continuous positive airway pressure adherence in apneic patients. PLoS One. 2013;8(5):e64382.
- 18 Ebben MR, Oyegbile T, Pollak CP. The efficacy of three different mask styles on PAP titration night. Sleep Med. 2012;13(6):645-9.
- 19 Bachour A, Vitikainen P, Virkkula P, Maasilta P. CPAP interface: satisfaction and side effects. Sleep Breath. 2013;17(2):667-72.
- 20 Schorr F, Genta PR, Gregório MG, Danzi-Soares NJ, Lorenzi-Filho G. Continuous positive airway pressure delivered by oronasal mask may not be effective for obstructive sleep apnoea. Eur Respir J. 2012;40(2):503-5.
- 21 Genta PR, Eckert DJ, Gregorio MG, Danzi NJ, Moriya HT, Malhotra A, et al. Critical closing pressure during midazolam-induced sleep. J Appl Physiol (1985). 2011;111(5):1315-22.
- 22 Rabec C, Georges M, Kabeya NK, Baudouin N, Massin F, Reybet- Degat O, et al. Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside study. Eur Respir J. 2009;34(4):902-13.
- 23 Lanza A, Mariani S, Sommariva M, Campana C, Rubino A, Nichelatti M, et al. Continuous positive airway pressure treatment with nasal pillows in obstructive sleep apnea: long-term effectiveness and adherence. Sleep Med. 2018;41:94-9.
- 24 Dibra MN, Berry RB, Wagner MH. Treatment of Obstructive Sleep Apnea: Choosing the Best Interface. Sleep Med Clin. 2017;12(4):543-9.
Corresponding author:
Publication History
Received: 05 May 2018
Accepted: 24 August 2018
Article published online:
16 October 2023
© 2023. Brazilian Sleep Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Thieme Revinter Publicações Ltda.
Rua do Matoso 170, Rio de Janeiro, RJ, CEP 20270-135, Brazil
-
REFERENCES
- 1 Young T, Peppard PE, Gottlieb DJ. Epidemiology of obstructive sleep apnea: a population health perspective. Am J Respir Crit Care Med. 2002;165(9):1217-39.
- 2 Tufik S, Santos-Silva R, Taddei JA, Bittencourt LR. Obstructive sleep apnea syndrome in the Sao Paulo Epidemiologic Sleep Study. Sleep Med. 2010;11(5):441-6.
- 3 Marin JM, Carrizo SJ, Vicente E, Agusti AG. Long-term cardiovascular outcomes in men with obstructive sleep apnoea-hypopnoea with or without treatment with continuous positive airway pressure: an observational study. Lancet. 2005;365(9464):1046-53.
- 4 Kushida CA, Littner MR, Hirshkowitz M, Morgenthaler TI, Alessi CA, Bailey D, et al.; American Academy of Sleep Medicine. Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders. Sleep. 2006;29(3):375-80.
- 5 Giles TL, Lasserson TJ, Smith BH, White J, Wright J, Cates CJ. Continuous positive airway pressure for obstructive sleep apnoea in adults. Cochrane Database Syst Rev. 2006;(3):CD001106.
- 6 McEvoy RD, Antic NA, Heeley E, Luo Y, Ou Q, Zhang X, et al.; SAVE Investigators and Coordinators. CPAP for Prevention of Cardiovascular Events in Obstructive Sleep Apnea. N Engl J Med. 2016;375(10):919-31.
- 7 Haniffa M, Lasserson TJ, Smith I. Interventions to improve compliance with continuous positive airway pressure for obstructive sleep apnoea. Cochrane Database Syst Rev. 2004;(4):CD003531.
- 8 Meurice JC, Ingrand P, Portier F, Arnulf I, Rakotonanahari D, Fournier E, et al.; ANTADIR Working Group “PPC”, CMTS ANTADIR. A multicentre trial of education strategies at CPAP induction in the treatment of severe sleep apnoea-hypopnoea syndrome. Sleep Med. 2007;8(1):37-42.
- 9 Billings ME, Auckley D, Benca R, Foldvary-Schaefer N, Iber C, Redline S, et al.; Race and residential socioeconomics as predictors of CPAP adherence. Sleep. 2011;34(12):1653-8.
- 10 Kohler M, Smith D, Tippett V, Stradling JR. Predictors of long-term compliance with continuous positive airway pressure. Thorax. 2010;65(9):829-32.
- 11 Weaver TE, Grunstein RR. Adherence to continuous positive airway pressure therapy: the challenge to effective treatment. Proc Am Thorac Soc. 2008;5(2):173-8.
- 12 Ayas NT, FitzGerald JM, Fleetham JA, White DP, Schulzer M, Ryan CF, et al. Cost-effectiveness of continuous positive airway pressure therapy for moderate to severe obstructive sleep apnea/hypopnea. Arch Intern Med. 2006;166(9):977-84.
- 13 Kribbs NB, Pack AI, Kline LR, Smith PL, Schwartz AR, Schubert NM, et al. Objective measurement of patterns of nasal CPAP use by patients with obstructive sleep apnea. Am Rev Respir Dis. 1993;147(4):887-95.
- 14 Olsen KD, Kern EB, Westbrook PR. Sleep and breathing disturbances secondary to nasal obstruction. Otolaryngol Head Neck Surg. 1981;89(5):804-10.
- 15 Andrade RG, Piccin VS, Nascimento JA, Viana FM, Genta PR, Lorenzi- Filho G. Impact of the type of mask on the effectiveness of and adherence to continuous positive airway pressure treatment for obstructive sleep apnea. J Bras Pneumol. 2014;40(6):658-68.
- 16 Andrade RGS, Viana FM, Nascimento JA, Drager LF, Moffa A, Brunoni AR, et al. Nasal vs Oronasal CPAP for OSA Treatment: A Meta-Analysis. Chest. 2018;153(3):665-74.
- 17 Borel JC, Tamisier R, Dias-Domingos S, Sapene M, Martin F, Stach B, et al.; Scientific Council of The Sleep Registry of the French Federation of Pneumology (OSFP). Type of mask may impact on continuous positive airway pressure adherence in apneic patients. PLoS One. 2013;8(5):e64382.
- 18 Ebben MR, Oyegbile T, Pollak CP. The efficacy of three different mask styles on PAP titration night. Sleep Med. 2012;13(6):645-9.
- 19 Bachour A, Vitikainen P, Virkkula P, Maasilta P. CPAP interface: satisfaction and side effects. Sleep Breath. 2013;17(2):667-72.
- 20 Schorr F, Genta PR, Gregório MG, Danzi-Soares NJ, Lorenzi-Filho G. Continuous positive airway pressure delivered by oronasal mask may not be effective for obstructive sleep apnoea. Eur Respir J. 2012;40(2):503-5.
- 21 Genta PR, Eckert DJ, Gregorio MG, Danzi NJ, Moriya HT, Malhotra A, et al. Critical closing pressure during midazolam-induced sleep. J Appl Physiol (1985). 2011;111(5):1315-22.
- 22 Rabec C, Georges M, Kabeya NK, Baudouin N, Massin F, Reybet- Degat O, et al. Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside study. Eur Respir J. 2009;34(4):902-13.
- 23 Lanza A, Mariani S, Sommariva M, Campana C, Rubino A, Nichelatti M, et al. Continuous positive airway pressure treatment with nasal pillows in obstructive sleep apnea: long-term effectiveness and adherence. Sleep Med. 2018;41:94-9.
- 24 Dibra MN, Berry RB, Wagner MH. Treatment of Obstructive Sleep Apnea: Choosing the Best Interface. Sleep Med Clin. 2017;12(4):543-9.