Endoscopy 2007; 39: E274-E275
DOI: 10.1055/s-2007-966615
Unusual cases and technical notes

© Georg Thieme Verlag KG Stuttgart · New York

A new device for the introducer technique for percutaneous endoscopic gastrostomy placement

J.  H.  Giordano-Nappi1 , S.  Ishioka1 , F.  Maluf-Filho1 , F.  Y.  Hondo1 , S.  E.  Matuguma1 , P.  Sakai1
  • 1Department of Gastrointestinal Endoscopy, Division of Gastroenterology, Hospital of Clinics of São Paulo University Medical School, São Paulo, Brazil
Further Information

J. H. Giordano-Nappi, MD 

Department of Gastrointestinal Endoscopy, Division of Gastroenterology, Hospital of Clinics of São Paulo University Medical School

R. Teodoro Sampaio 498 Apto. 147

Pinheiros

05406-000

São Paulo

Brazil

Email: drgiordano@hotmail.com

Publication History

Publication Date:
24 October 2007 (online)

Table of Contents

Percutaneous endoscopic gastrostomy (PEG) is the preferred method for achieving enteral access for patients requiring long-term nutritional support [1]. The popular PEG pull-through technique is not adequate for patients with obstructive head and neck or esophageal cancer because of the higher morbidity associated with the technique, including the risk of malignant implantation at the stoma. The introducer technique is then indicated as described by Russell et al. [2]. This involves direct puncture of an inflated stomach through the anterior abdominal wall, followed by the introduction of a guide wire and dilation of the newly created tract. The introduction system is removed by peeling it off.

It is believed that gastropexy provides safer conditions for direct puncture of the stomach, in particular for the insertion of larger gastrostomy feeding tubes [3] [4] [5] [6]. We believe that there is a simple and cheaper alternative to the use of a guide wire, dilator, and “peeling off” devices. This is a development of a trocar system, consisting of an external sheath with a lateral cleft. The usual three-sided, sharply pointed end of the internal shaft of the trocar system is replaced by a cone-shaped end, in order to reduce the cutting trauma to the abdominal and gastric walls ([Fig. 1]).

Zoom Image

Fig. 1 Schematic representation of the trocar system with a cone-shaped internal shaft and a lateral cleft for the insertion of the feeding tube.

Two gastropexy stitches are placed under endoscopic guidance using a double-needle puncturing device [3] [4]. The trocar system is then advanced through a 10-mm skin incision between the fixation sutures and into the transilluminated and inflated stomach under endoscopic guidance. The internal shaft is removed and a 20-Fr ballooned tube is introduced through the lateral cleft. Once the tube is in place the balloon is inflated and the external sheath of the trocar is pulled out as the feeding tube is slid out through the lateral cleft ([Fig. 2], [3]).

Zoom Image

Fig. 2 External and endoscopic (inset) views just after the puncture of the inflated and transilluminated gastric cavity by the trocar system. Note that a gastropexy has been performed previously.

Zoom Image

Fig. 3 As the balloon is inflated the tube can be kept in the stomach by sliding it through the lateral cleft while the external shaft of the trocar is pulled out.

Another advantage of this technique is the potential availability of the device in different diameters, allowing insertion gastrostomy tubes of various sizes.

Endoscopy_UCTN_Code_TTT_1AO_2AK

#

References

  • 1 Kobayashi K, Cooper G S, Chak A. et al . A prospective evaluation of outcome in patients referred for PEG placement.  Gastrointest Endosc. 2002;  55 500
  • 2 Russell T R, Brotman M, Norris F. Percutaneous gastrostomy: a new simplified and cost-effective technique.  Am J Surg. 1984;  148 132-137
  • 3 Dormann A J, Welda B, Kahl S. et al . Long-term result with a new introducer method with gastropexy for percutaneous endoscopic gastrostomy.  Am J Gastroenterol. 2006;  101 1229-1234
  • 4 Dormann A J, Glosemeyer R, Leistner U. et al . Modified percutaneous endoscopic gastrostomy with gastropexy: early experience with a new introducer technique.  Z Gastroenterol. 2000;  38 933-938
  • 5 Hashiba K. Endoscopic gastrostomy.  Endoscopy. 1987;  19 23-24
  • 6 Shinoda M, Kojima M, Fukase T. et al . Percutaneous transgastric intestinal decompression: the management of malignant bowel obstruction without nasal intubation.  Surg Today. 1994;  24 937-939

J. H. Giordano-Nappi, MD 

Department of Gastrointestinal Endoscopy, Division of Gastroenterology, Hospital of Clinics of São Paulo University Medical School

R. Teodoro Sampaio 498 Apto. 147

Pinheiros

05406-000

São Paulo

Brazil

Email: drgiordano@hotmail.com

#

References

  • 1 Kobayashi K, Cooper G S, Chak A. et al . A prospective evaluation of outcome in patients referred for PEG placement.  Gastrointest Endosc. 2002;  55 500
  • 2 Russell T R, Brotman M, Norris F. Percutaneous gastrostomy: a new simplified and cost-effective technique.  Am J Surg. 1984;  148 132-137
  • 3 Dormann A J, Welda B, Kahl S. et al . Long-term result with a new introducer method with gastropexy for percutaneous endoscopic gastrostomy.  Am J Gastroenterol. 2006;  101 1229-1234
  • 4 Dormann A J, Glosemeyer R, Leistner U. et al . Modified percutaneous endoscopic gastrostomy with gastropexy: early experience with a new introducer technique.  Z Gastroenterol. 2000;  38 933-938
  • 5 Hashiba K. Endoscopic gastrostomy.  Endoscopy. 1987;  19 23-24
  • 6 Shinoda M, Kojima M, Fukase T. et al . Percutaneous transgastric intestinal decompression: the management of malignant bowel obstruction without nasal intubation.  Surg Today. 1994;  24 937-939

J. H. Giordano-Nappi, MD 

Department of Gastrointestinal Endoscopy, Division of Gastroenterology, Hospital of Clinics of São Paulo University Medical School

R. Teodoro Sampaio 498 Apto. 147

Pinheiros

05406-000

São Paulo

Brazil

Email: drgiordano@hotmail.com

Zoom Image

Fig. 1 Schematic representation of the trocar system with a cone-shaped internal shaft and a lateral cleft for the insertion of the feeding tube.

Zoom Image

Fig. 2 External and endoscopic (inset) views just after the puncture of the inflated and transilluminated gastric cavity by the trocar system. Note that a gastropexy has been performed previously.

Zoom Image