Automated Registration of Intraoperative CT Image Data for Navigated Skull Base Surgery

G. Eggers 1 , B. Kress 2 , J. Mühling 1
  • 1Department of Oral and Cranio-Maxillofacial Surgery, Heidelberg University Hospital, Heidelberg, Germany
  • 2Department of Neuroradiology, Heidelberg University Hospital, Heidelberg, Germany
Further Information

Publication History

Publication Date:
28 February 2008 (online)

Abstract

Objectives: With a new intraoperative computed tomography (CT) imaging system, patient-to-image registration without any invasive registration markers is possible. Furthermore, registration can be performed fully automatically. The accuracy of this method for skull base surgery was investigated in this study.

Methods: We employed a phantom study design. A phantom skull was equipped with 33 target markers in the regions of the anterior and lateral skull base. CT image data were acquired with an intraoperative CT suite. Image data were transferred as DICOM data to the navigation system, and registration was performed automatically. For registration, the position of the patient and the position of the CT gantry were monitored in the imaging process, using the infrared camera of a navigation system. Using the pointing device of the navigation system, the target markers were identified. The accuracy was measured as the spatial difference of the target markers in image space and on the phantom.

Results: Accuracy was always sufficient for image-guided surgery of any region of the skull base, with an average target registration error of below 1.2 mm. In contrast to traditional non-invasive registration methods, there was no difference in registration accuracy between the anterior skull base and the lateral skull base.

Conclusions: Fully automated registration based on a tracked CT gantry is a robust and accurate registration method for skull base surgery.

References

  • 1 Eggers G, Mühling J, Marmulla R. Image-to-patient registration techniques in head surgery.  Int J Oral Maxillofac Surg. 2006;  35 1081-1095
  • 2 Heiland M, Habermann CR, Schmelzle R. Indications and limitations of intraoperative navigation in maxillofacial surgery.  J Oral Maxillofac Surg. 2004;  62 1059-1063
  • 3 Kremer P, Tronnier V, Steiner HH, Metzner R, Ebinger F, Rating D, Hartmann M, Seitz A, Unterberg A, Wirtz CR. Intraoperative MRI for interventional neurosurgical procedures and tumor resection control in children.  Childs Nerv Syst. 2006;  22 674-678
  • 4 Tronnier VM, Bonsanto MM, Staubert A, Knauth M, Kunze S, Wirtz CR. Comparison of intraoperative MR imaging and 3D-navigated ultrasonography in the detection and resection control of lesions.  Neurosurg Focus. 2001;  10 E3
  • 5 Pohlenz P, Blessmann M, Blake F, Heinrich S, Schmelzle R, Heiland M. Clinical indications and perspectives for intraoperative cone-beam computed tomography in oral and maxillofacial surgery.  Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;  103 412-417
  • 6 Mascott CR, Sol JC, Bousquet P, Lagarrigue J, Lazorthes Y, Lauwers-Cances V. Quantification of true in vivo (application) accuracy in cranial image-guided surgery: influence of mode of patient registration.  Neurosurgery. 2006;  59 ((Suppl 1)) 146-156
  • 7 Fitzpatrick JM, West JB, Maurer Jr CR. Predicting error in rigid-body point-based registration.  IEEE Trans Med Imaging. 1998;  17 694-702
  • 8 Hassfeld S, Mühling J, Zöller J. Intraoperative navigation in oral and maxillofacial surgery.  Int J Oral Maxillofac Surg. 1995;  24 111-119
  • 9 Schlaier J, Warnat J, Brawanski A. Registration accuracy and practicability of laser-directed surface matching.  Comput Aided Surg. 2002;  7 284-290
  • 10 Eggers G, Mühling J, Marmulla R. Template-based registration for image guided maxillofacial surgery.  J Oral Maxillofac Surg. 2005;  63 1330-1336
  • 11 Howard  3rd  MA, Dobbs MB, Simonson TM, LaVelle WE, Granner MA. A noninvasive, reattachable skull fiducial marker system. Technical note.  J Neurosurg. 1995;  83 372-376
  • 12 Fenlon MR, Jusczyzck AS, Edwards PJ, King AP. Locking acrylic resin dental stent for image-guided surgery.  J Prosthet Dent. 2000;  83 482-485
  • 13 Mann W, Klimek L. Indications for computer-assisted surgery in otorhinolaryngology.  Comput Aided Surg. 1998;  3 202-204
  • 14 Caversaccio M, Bächler R, Lädrach K, Schroth G, Nolte LP, Häusler R. Frameless computer-aided surgery system for revision endoscopic sinus surgery.  Otolaryngol Head Neck Surg. 2000;  122 808-813
  • 15 Sindwani R, Metson R. Impact of image guidance on complications during osteoplastic frontal sinus surgery.  Otolaryngol Head Neck Surg. 2004;  131 150-155
  • 16 Cartellieri M, Vorbeck F. Endoscopic sinus surgery using intraoperative computed tomography imaging for updating a three-dimensional navigation system.  Laryngoscope. 2000;  110 292-296
  • 17 Cartellieri M, Kremser J, Vorbeck F. Comparison of different 3D navigation systems by a clinical “user”.  Eur Arch Otorhinolaryngol. 2001;  258 38-41
  • 18 Miller RS, Hashisaki GT, Kesser BW. Image-guided localization of the internal auditory canal via the middle cranial fossa approach.  Otolaryngol Head Neck Surg. 2006;  134 778-782

Correspondence

PD Dr. med. Dr. med. dent G. Eggers

Department of Oral and Cranio-Maxillofacial Surgery

Heidelberg University Hospital

Im Neuenheimer Feld 400

69120 Heidelberg

Germany

Phone: +49/6221/567 33 2

Fax: +49/6221/563 39 33

Email: georg.eggers@med.uni-heidelberg.de

    >