A 24-year-old man had suffered from epigastralgia and acid regurgitation for many
years. After local anesthesia, upper gastrointestinal endoscopy was performed without
difficulty. Some hematin-coated spots were noted at the gastric body, and loosening
of the gastroesophageal junction. During the retroflexion procedure, the endoscope
became impacted in the distal esophagus when the patient gave a large belch. Although
the lumen was clearly visible, attempts at external traction were unsuccessful because
resistance was encountered. In addition, it was not possible to release the impacted
loop by advancing the scope.
After several failed attempts, a second flexible endoscope was inserted without fluoroscopic
assistance. The first endoscope was found to be within the distal esophagus as expected
(Figure [1 ]), as a U-shaped loop. It was then released using a double-endoscopy technique (Figure
[2 ]). The patient has been followed up, with no untoward events.
Figure 1 Endoscopic visualization of a portion of the first impacted endoscope above the gastroesophageal
junction.
Figure 2 Double-endoscopy removal technique. a A second endoscope is used to visualize the impaction. b The impacted scope is slowly pushed into the stomach using a second endoscope. c The two endoscopes are removed after advancement into the stomach.
The currently used panendoscope has a shaft diameter of around 9 mm. The esophagus,
whose lumen can distend up to 20 - 30 mm laterally [1 ], is at least the size of two endoscope diameters. However, when there is impaction,
the endoscope remains stuck within the esophagus because simple retrograde traction
on the endoscope results only in some flattening of the proximally created loop. The
Olympus GIF XQ240 endoscope, has a minimal radius of curvature of about 38 - 40 mm,
below which it cannot be further flattened. The curved portion pushes against the
esophagus when traction is applied. However, further advancement of the endoscope
also cannot solve the impaction because movement of the endoscope is still restricted
to above the gastroesophageal junction.
Impaction of an endoscope at retroflexion in the distal esophagus, in a patient with
hiatal hernia or total gastrectomy, has been previously described [2 ]
[3 ]. The greatest risk of impaction of an endoscope occurs during the retroflexion procedure
in patients with a loose gastroesophageal junction.
This example makes it clear that the use of a second endoscope is crucial for identifying
the impacted loop and resolving this potential hazard, if the first endoscope could
not be extracted easily.