Endoscopy 2004; 36(12): 1130
DOI: 10.1055/s-2004-825980
Unusual Cases and Technical Notes
© Georg Thieme Verlag KG Stuttgart · New York

Endoscopy-Assisted Ferromagnetic Foreign-Body Removal with a Novel Magnetic Instrument

S.  Nijhawan1 , A.  Joshi1 , A.  Shende1 , N.  Agarwal1 , D.  Kumar1 , A.  Mathur1 , R.  R.  Rai1
  • 1Dept. of Gastroenterology, SMS Hospital and Medical College, Jaipur, India
Further Information

S. Nijhawan, M. D.

Dept. of Gastroenterology, SMS Hospital and Medical College

112, Panchsheel Enclave
Gokulbhai Bhatt Marg Durgapura
Jaipur 302018
India

Fax: +91-141-252-0163

Email: dr_nijhawan@yahoo.com

Publication History

Publication Date:
09 May 2006 (online)

Table of Contents

Some metallic foreign bodies have magnetic characteristics and can be retrieved if an instrument with a magnet at its tip can be used. This report describes the development of an instrument of this type and the use of the instrument to remove a coin.

Two magnetic disks 1.5 cm in diameter were joined and attached with adhesive to a 5-cm steel spring. A steel wire 200 cm long and 0.75 mm thick was attached with adhesive to the other end of the spring. A Teflon tube (160 cm, 8 Fr) was used as a sleeve for this instrument (Figure [1]).

Zoom Image

Figure 1 The instrument for retrieving magnetic foreign bodies, with the Teflon tube sleeve and a coin.

A 12-year-old boy had ingested a one-rupee coin (diameter 2.5 cm) 5 days previously. Abdominal radiography identified the coin in the epigastric area, and it was found to be located in the fundus at gastroscopy. After informed consent had been obtained, the patient underwent removal of the coin using this novel instrument. The sleeve was passed through the suction channel up to the tip of the endoscope. The wire end of the retrieval instrument was then covered with the sleeve from the tip of the endoscope until the magnetic disk reached the tip. The whole assembly was then introduced to reach as far as the fundus; the wire was pushed for 5 cm to advance the magnet further into the fundus. The coin attached itself to the magnetic disk (Figure [2]). The wire was then pulled in order to bring the magnetic disk, with the coin attached to it, back to the tip of endoscope. The endoscope was withdrawn along with the assembly. It took 3.5 min to complete the procedure.

Zoom Image

Figure 2 Endoscopic images of the removal of the coin using the magnetic foreign-body retrieval instrument.

To the best of our knowledge, this is the first report of endoscope-assisted retrieval of a ferromagnetic foreign body. A few previous case reports [1] [2] [3] on the use of magnets to remove foreign bodies, with or without fluoroscopic assistance, have described the procedure as cumbersome and involving a high failure rate. The method of removing blunt ferromagnetic foreign bodies under direct visualization described here is precise, safe, and quick.

Endoscopy_UCTN_Code_TTT_1AO_2AL

Endoscopy_UCTN_Code_TTT_1AQ_2AH

#

References

  • 1 Paulson E K, Jaffe R B. Metallic foreign bodies in the stomach: fluoroscopic removal with a magnetic orogastric tube.  Radiology. 1990;  174 191-194
  • 2 Himadi G M, Fischer G J. Magnetic removal of foreign bodies from the upper gastrointestinal tract.  Radiology. 1977;  123 226-227
  • 3 McDermott V G, Taylor T, Wyatt J P. et al . Orogastric magnet removal of ingested disc batteries.  J Pediatr Surg. 1995;  30 29-32

S. Nijhawan, M. D.

Dept. of Gastroenterology, SMS Hospital and Medical College

112, Panchsheel Enclave
Gokulbhai Bhatt Marg Durgapura
Jaipur 302018
India

Fax: +91-141-252-0163

Email: dr_nijhawan@yahoo.com

#

References

  • 1 Paulson E K, Jaffe R B. Metallic foreign bodies in the stomach: fluoroscopic removal with a magnetic orogastric tube.  Radiology. 1990;  174 191-194
  • 2 Himadi G M, Fischer G J. Magnetic removal of foreign bodies from the upper gastrointestinal tract.  Radiology. 1977;  123 226-227
  • 3 McDermott V G, Taylor T, Wyatt J P. et al . Orogastric magnet removal of ingested disc batteries.  J Pediatr Surg. 1995;  30 29-32

S. Nijhawan, M. D.

Dept. of Gastroenterology, SMS Hospital and Medical College

112, Panchsheel Enclave
Gokulbhai Bhatt Marg Durgapura
Jaipur 302018
India

Fax: +91-141-252-0163

Email: dr_nijhawan@yahoo.com

Zoom Image

Figure 1 The instrument for retrieving magnetic foreign bodies, with the Teflon tube sleeve and a coin.

Zoom Image

Figure 2 Endoscopic images of the removal of the coin using the magnetic foreign-body retrieval instrument.