Geburtshilfe Frauenheilkd 2004; 64(9): 923-929
DOI: 10.1055/s-2004-821146
Originalarbeit

Georg Thieme Verlag KG Stuttgart · New York

Evaluation of a New Cordless CTG Monitoring System

Evaluation eines neuen schnurlosen CTG-ÜberwachungssystemsB. Schauf1 , A. Kontner1 , H. Abele1 , D. Wallwiener1
  • 1Universitäts-Frauenklinik, Tübingen
Further Information

Publication History

Eingang Manuskript: April 14, 2004 Eingang revidiertes Manuskript: June 7, 2004

Akzeptiert: June 20, 2004

Publication Date:
02 September 2004 (online)

Zusammenfassung

Einleitung: Die Studie an der Universitätsfrauenklinik Tübingen untersucht die Vor- und Nachteile eines schnurlosen CTG-Systems (Avalon, Philips, Böblingen).

Material und Methoden: 250 Frauen nahmen insgesamt im Zeitraum von Oktober 2002 bis April 2003 an der Studie teil. 100 Frauen wurden mit dem schnurlosen CTG-Gerät unter der Geburt überwacht, 100 Frauen mit dem konventionellen kabelgestützten System. 50 Schwangere, die zur Überwachung aus verschiedenen Gründen stationär aufgenommen wurden, wurden sowohl mit dem schnurlosen CTG-Gerät als auch mit dem kabelgebundenen CTG-Gerät abwechselnd überwacht. Ärztinnen, Ärzte, Hebammen und die Patientinnen wurden unter anderem bezüglich ihrer Zufriedenheit befragt und das neue schnurlose CTG-Gerät wurde im Vergleich zu den herkömmlichen kabelgebundenen CTG-Geräten auf Überwachungssicherheit hin untersucht.

Ergebnisse/Diskussion: Das schnurlose CTG-Gerät bietet der Schwangeren und vor allem der Kreißenden bezüglich der Bewegungsfreiheit enorme Vorteile. Im Kreißsaal lässt sich die Geburt für Kreißende angenehmer gestalten. Hebammen, Ärztinnen und Ärzte beurteilen das Gerät vor allem im Hinblick auf Komfort überaus positiv bei gleicher Überwachungssicherheit wie bei den herkömmlichen CTG-Geräten.

Zusammenfassung: Das neue kabellose CTG-System bietet sowohl den Patientinnen als auch den Geburtshelfern Vorteile bei gleicher Überwachungssicherheit wie mit den herkömmlichen CTG-Geräten. Die Patientinnen unter der Geburt haben eine wesentlich größere Bewegungsfreiheit. Die Hebammen, Ärztinnen und Ärzte profitieren von einer guten Handhabung und Überwachungssicherheit.

Abstract

Introduction: The study at the University Hospital for Women at Tübingen investigates the advantages and disadvantages of a cordless CTG system (Avalon, Philips, Böblingen).

Material and Methods: A total of 250 women took part in the study between October 2002 and April 2003. 100 women were monitored using cordless CTG during labour and 100 using the conventional wired system. 50 pregnant women who were in-patients under supervision for various reasons were monitored using both the cordless CTG device and wired foetal monitors in alternation. Doctors, midwives and the patients were asked about their satisfaction, and the monitoring reliability of the new cordless CTG device was compared with that of the conventional wired CTG equipment.

Results/Discussion: The freedom of movement provided by the cordless CTG device, for pregnant women and above all for women in labour, is an enormous advantage. In the labour ward, giving birth is more pleasant for the mother. Doctors and midwives give the equipment an extremely positive assessment, particularly regarding comfort; at the same time the cordless device demonstrates the same monitoring reliability as conventional CTG equipment.

Conclusion: The new cordless CTG system offers advantages to both patients and obstetricians, and monitoring is as reliable as with the conventional CTG devices. Patients giving birth have considerably more freedom of movement. Midwives and doctors benefit from ease of operation and reliable monitoring.

References

  • 1 Langer B, Caneva M-P, Schlaeder G. Routine prenatal care in Europe: the comparative experience of nine departments of gynaecology and obstetrics in eight different countries.  Eur J Obstet Gynecol Reprod Biol. 1999;  85 191-198
  • 2 Saling E. Comments on past and present situatuion of intensive monitoring of the fetus during labor.  J Perinat Med. 1996;  24 7-13
  • 3 Boehm F H. Intrapartum fetal heart rate monitoring.  Obstet Gynecol Clin North America. 1999;  26 623-639
  • 4 Kühnert M, Seelbach-Göbel B, Butterwegge M. Kritische Evaluation der fetalen Pulsoxymetrie im klinischen Einsatz.  Geburtsh Frauenheilk. 2001;  61 290-296
  • 5 Deutsche Gesellschaft für Perinatale Medizin . Standards in der Perinatalmedizin - Dopplersonographie in der Schwangerschaft.  Geburtsh Frauenheilk. 2003;  63 21-25
  • 6 Schauf B, Lowery C, Wilson J D, Eswaran H, Birbaumer N, Aydeniz B, Wallwiener D, Preissl H. Neue Perspektiven in der intrauterinen Überwachung mittels fetalem Magnetezephalogramm.  Geburtsh Frauenheilk. 2001;  61 880-887
  • 7 Sinclair M. Midwives' attitudes to the use of the cardiotocograph machine.  J Adv Nurs. 2001;  35 599-606
  • 8 Ringler M, Huber J C, Reinold E. Das Kardiotokogramm im Erleben der Frau: Hilfe für ein schöneres Geburtserlebnis oder technologisches Schreckgespenst?.  Z Geburtsh Perinat. 1981;  185 236-239
  • 9 Lapaire O, Güth U, Holzgreve W, Hösli I. Fetale Pulsoxymetrie: Akzeptanz der Mütter und des geburtshilflichen Teams.  Geburtsh Frauenheilk. 2003;  63 140-144
  • 10 Kirke P N. Mother's views of care in labour.  Br J Obstet Gynaecol. 1980;  87 1034-1038
  • 11 Bodner-Adler B, Bodner K, Joura E A, Husslein P, Wagenbichler P, Kaider A, Mayerhofer K. Einfluß der unterschiedlichen Geburtspositionen auf mütterliche Geburtsverletzungen und kindliche Parameter während spontaner vaginaler Geburt.  Geburtsh Frauenheilk. 2001;  61 766-770
  • 12 Eberhard J, Geissbühler V, Chiffelle C, Stein S. Alternative Gebärmethoden verändern die Geburtsmedizin.  Geburtsh Frauenheilk. 2001;  61 771-777

B. Schauf

Universitäts-Frauenklinik Tübingen

Calwer Straße 7

72076 Tübingen

Email: Burkhard.Schauf@med.uni-tuebingen.de

    >