Introduction
Introduction
Video capsule endoscopy (VCE) represents a significant advance in the investigation
of intestinal diseases. The performance of the procedure and the indications for it
are reviewed here in order to establish guidelines for its use, in accordance with
current knowledge based on clinical practice and the published literature. In this
guideline, all of the technical features described refer to the M2A Given Imaging
video capsule (Given Imaging, Ltd., Yoqneam, Israel).
Video capsule endoscopy is a new technique in which a disposable video capsule is
used that is swallowed by the patient. This allows visualization of the segment of
the small bowel that is not within the reach of standard upper and lower endoscopy.
The main current indication for capsule endoscopy is an obscure source of gastrointestinal
bleeding. VCE may be particularly helpful in identifying the cause of bleeding when
standard upper endoscopy and colonoscopy examinations are negative.
Description of the Procedure
Description of the Procedure
The entire VCE procedure can be conducted on an outpatient basis. The patient is instructed
to fast for 12 h before the procedure. The patient wears a shoulder-supported belt
pack holding a power supply containing five nickel-metal 1.2-V batteries (the size
of D cells), and a small 305-gigabyte hard drive for storing the received images.
Before ingestion of the capsule, a sensor array is taped to the patient’s anterior
abdominal wall in a designated pattern and connected to the hard drive. Correlation
between the findings in the gastrointestinal lumen and the general location of the
capsule in the abdomen is attempted by triangulating the signal received from the
three closest sensor ports.
Patients are allowed to drink clear liquids 1 h after ingestion of the capsule and
to eat a light meal 4 h after ingestion. During the recording period, patients are
instructed to record abdominal symptoms in a diary and are requested to check the
blinking light on the belt pack periodically to confirm that the signal is being received.
After 8h, the sensor array and the data recorder are removed, and the recorded digital
information is downloaded to a computer workstation. The capsule is excreted naturally,
usually within 24 to 48 h.
Data are downloaded from the belt pack recorder to a customized PC workstation. This
process takes 2 - 3 h. Proprietary software is then used to review the images in an
adjustable rapid-scanning mode that can display between 1 and 25 frames per second.
For subsequent access, thumbnail video clips of segments incorporating 50 images before
and after an identified finding can be annotated and saved. Reading and interpreting
the images requires an average of 40 - 60 min.
There are a few issues that are still unresolved with regard to the procedure and
patient preparation. The expert consensus recommendation is that the patient should
always receive preparation with bowel cleansing using 2 liters of polyethylene glycol,
administered either on the evening before the recording or on the same day at least
2-3 h before ingestion of the capsule. The advantage of preparation is that it cleanses
the terminal ileum, allowing a better examination.
Reading the Recorded Images
Reading the Recorded Images
Localization
The workstation software used to process the capsule images incorporates a localization
feature. This makes it possible to estimate the location of any pathological findings
within an average range of 3.77 cm. An attempt is made to identify the capsule’s location
in the abdomen by triangulating the three closest sensors, selected according to the
strength of the signals received from the capsule by the eight sensors attached to
the abdominal wall. The location is then calculated from the position of these sensors
and displayed on the computer screen as a two-dimensional image. However, in clinical
practice, the software identification of the location of lesions is often considered
to be too imprecise, and the location is mainly assessed on the basis of the distinct
mucosal patterns visualized in the jejunum and ileum.
Suspected Bleeding Indication
The software application now includes an indicator feature (the ”Suspected Blood Indicator”),
which automatically marks images that correlate with suspected blood or red areas
in the small intestine. This is helpful for identifying bleeding sites during the
reading process. The clinical relevance of this software feature has yet to be evaluated.
Multiview
The software application has also recently had a ”Multiview” feature added for reading
of the VCE recordings. This allows simultaneous display of two images from the recording
in separate, adjacent windows.
Indications
Indications
Obscure Digestive Bleeding
The most frequent indication for VCE is currently to diagnose obscure bleeding. Obscure
bleeding is defined as either recurrent episodes of gastrointestinal bleeding - e.
g., with melena or hematochezia - or a positive fecal occult blood test, or chronic
iron-deficiency anemia, after negative findings on gastroscopy and colonoscopy.
Before VCE, patients with obscure gastrointestinal bleeding should undergo one more
upper gastrointestinal endoscopy examination, conducted by an expert using a lateral-viewing
endoscope, to exclude missed lesions within the reach of the endoscope (such as postpyloric
or postbulbar ulcers), as well as lesions in the third and fourth parts of the duodenum.
The reported diagnostic yield of VCE has ranged from 55 % to 81 %. In comparative
studies, VCE has been found to be consistently superior to other examination methods
for diagnosing the source of gastrointestinal bleeding.
Other Indications Currently Accepted
Investigation of Crohn’s disease
There is growing acceptance of the use of VCE in patients with Crohn’s disease. Several
comparative studies have evaluated the diagnostic yield of VCE and compared it with
various radiographic modalities. VCE is able to detect intestinal lesions in a large
subset of patients with known Crohn’s disease, and it can also often detect intestinal
lesions in patients in whom there is a clinical and/or biological suspicion of Crohn’s.
The diagnostic yield ranges from 43 % to 71 %. VCE frequently extends the area of
the bowel diagnosed as being affected by the disease, or detects lesions compatible
with the diagnosis. However, these results need to be carefully analyzed, for several
reasons: firstly, in most reports VCE has been compared with small-bowel follow-through
examinations, which are not the most sensitive radiographic method; secondly, the
patients included in the various studies may have different clinical situations in
which the outcome of intestinal investigations needs to be assessed in different ways;
thirdly, most studies have included limited numbers of patients; fourthly, it is not
clear from these reports whether the diagnosis of Crohn’s disease was confirmed afterwards
with biopsies. Clinical trials are currently being conducted to evaluate these indications.
On the other hand, it might be worth using the M2A Patency Capsule (Given Imaging,
Ltd., Yoqneam, Israel) in patients with a higher risk of intestinal stenosis.
Evaluation of the side effects of nonsteroidal anti-inflammatory drugs (NSAIDs)
Several studies have shown that intestinal lesions are frequent in patients treated
with NSAIDs. The clinical picture may include overt or occult bleeding, abdominal
pain, or intestinal obstruction in cases of stenosis. The lesions most frequently
found are ulcers and erosions.
Hereditary polyposis syndromes
The use of VCE may be considered for the surveillance of hereditary polyposis syndromes,
such as familial adenomatous polyposis (FAP), in which a biopsy is needed in order
to assess the risk, and familial juvenile polyposis.
Prospective Indications
Complicated celiac disease
VCE may be of interest in the diagnostic work-up patients with celiac disease who
present with atypical symptoms or chronic iron-deficiency anemia. In patients with
a known diagnosis of celiac disease, VCE is indicated in cases of symptomatic relapse
despite compliance with a gluten-free diet. In this situation, VCE may reveal a pattern
of ulcerative jejunitis or T-cell lymphoma.
Pediatric capsule endoscopy
A few reports so far published have evaluated the feasibility and clinical usefulness
of VCE in children. VCE is safe in children approximately 9 years of age or older.
In younger patients, there have been a few tests involving endoscopic release of the
capsule in the stomach.
Tolerance and Safety
Tolerance and Safety
Overall Tolerance and Contraindications
The overall tolerance of VCE by patients has been impressively good in all studies.
The capsule should be used with caution in patients with known or suspected gastrointestinal
obstruction, fistulas, and swallowing disorders. Recent data suggest that the radiofrequency
signal in VCE does not interfere with cardiac pacemakers. Safety in pregnancy has
not yet been established.
Obstruction of Capsule Progression
A tight stenosis or large polyp may obstruct the progression of the capsule and require
treatment. In all situations in which there is thought to be a risk, such as polyposis
syndromes, NSAID complications, or Crohn’s disease, it should be regarded as mandatory
to have the patient complete a questionnaire regarding any intestinal symptoms that
might represent contraindications. Studies are currently being conducted to assess
the clinical value of an initial sham test with the M2A Patency System (which uses
a dissolvable capsule) before the real capsule examination.
Conclusion
Conclusion
VCE is an exciting new diagnostic facility for assessing the lumen of the digestive
tract. However, at present it is clearly only an adjunct to upper and lower gastrointestinal
endoscopy. It is generally preferable to time-consuming diagnostic enteroscopy, and
may lead to further developments in therapeutic enteroscopy. It is currently the preferred
test for mucosal imaging of the entire small intestine. The absence of reimbursement
for the procedure in major European countries is still a major drawback.