Endoscopy 2003; 35(6): 469-471
DOI: 10.1055/s-2003-39662
Original Article

© Georg Thieme Verlag Stuttgart · New York

Improvement of the Automatic Endoscopic Reprocessor: Self-Cleaning Disinfecting Connectors Between Endoscope and Reprocessor

Y.  Ishino1 , K.  Ido1 , K.  Sugano1
  • 1Department of Gastroenterology, Jichi Medical School, Tochigi, Japan
Further Information

Publication History

Submitted 6 September 2001

Accepted after Revision 11 November 2002

Publication Date:
03 June 2003 (online)

Background and Study Aims: We have previously pointed out a defect of automatic endoscopic reprocessors, i. e. the contamination of the connecting part between the endoscope and the reprocessor. We evaluated a newly designed connector (MH-861; Olympus, Tokyo, Japan) with a self-cleaning and disinfection mechanism, which enabled cleaning and disinfection of both the connector itself and its interface with the suction and air/water valves during a reprocessing cycle, which was not previously possible.
Methods: Ten upper gastrointestinal endscopes were examined in the study. Swabs were taken from the suction and air/water valves for microbiological culture before and after reprocessing by the washer-disinfector. The numbers of contaminated endoscopes before and after reprocessing with the new connector were compared.
Results: Before the procedure there were five contaminated endoscopes and none after the procedure. When the new connector was used, the difference in cleaning and disinfection of the connecting parts was significant (P = 0.0325).
Conclusions: We conclude that the newly developed connector permits effective cleaning and disinfection by automatic reprocessors.

References

  • 1 Urayama S, Kozarek R A, Sumida S. et al . Mycobacteria and glutaraldehyde: is high-level disinfection of endoscopes possible?.  Gastrointest Endosc. 1996;  43 451-456
  • 2 Bradley C R, Babb J R. Endoscope decontamination: automated vs manual.  J Hosp Infect. 1995;  30 537-542
  • 3 Ido K, Ishino Y, Ota Y. et al . Deficiencies of automatic endoscopic reprocessors: a method to achieve high-grade disinfection of endoscopes.  Gastrointest Endosc. 1996;  44 583-586
  • 4 Ishino Y, Ido K, Koiwai H. et al . Pitfalls in endoscope reprocessing: brushing of air and water channels is mandatory for high-level disinfection.  Gastrointest Endosc. 2001;  53 165-168
  • 5 Axon A, Kruse A, Urgell R. et al . Guidelines on cleaning and disinfection in gastrointestinal endoscopy.  Endoscopy. 1995;  27 Newsletter I-II
  • 6 BSG Endoscopy Committee Working Party . Cleaning and disinfection of equipment for gastrointestinal endoscopy. Report of a working party of the British Society of Gastroenterology Endoscopy Committee.  Gut. 1998;  42 585-593
  • 7 Martin M A, Reichelderfer M. APIC guidelines for infection control practice.  Am J Infect Control. 1994;  22 19-38
  • 8 Ogoshi K, Akamatsu T, Iishi H. et al . JGES guidelines for cleaning and disinfection in gastrointestinal endoscopy.  Digestive Endoscopy. 2000;  12 369-382
  • 9 Systchenko R, Marchetti B, Canard J N. et al . Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy.  Endoscopy. 2000;  32 807-818
  • 10 ECRI . Liquid disinfecting and sterilizing reprocessors used for flexible endoscopes.  Health Devices. 1994;  23 214-253
  • 11 Fraser V J, Zuckerman G, Clouse R E. et al . A prospective randomized trial comparing manual and automated endoscope disinfection methods.  Infect Control Hosp Epidemiol. 1993;  14 383-389
  • 12 Muscarella L F. Advantages and limitations of automatic flexible endoscope reprocessors.  Am J Infect Control. 1996;  24 304-309
  • 13 ASGE Technology Committee. Technology status evaluation report: automatic endoscope reprocessors.  Gastrointest Endosc. 1999;  50 925-927
  • 14 ECRI . Contamination of the modified Olympus EW-10 and EW-20 automatic flexible endoscope reprocessors.  Health Devices. 1994;  23 143-145
  • 15 Axon T, Jung M, Kruse A. et al . The European Society of Gastrointestinal Endoscopy (ESGE): checklist for the purchase of washer-disinfectors for flexible endoscopes.  Endoscopy. 2000;  32 914-919
  • 16 Cowen A. Of microbes, men, and machines.  Am J Gastroenterol. 1998;  93 2011-2013

Y. Ishino, M.D.

Department of Gastroenterology · Jichi Medical School

Minamikawachi-machi · Kawachi Gun · Tochigi 329-0498 · Japan

Fax: + 81-285-44-8297

Email: ishino@jichi.ac.jp

    >