RSS-Feed abonnieren

DOI: 10.1055/s-0045-1809980
Evaluation of the Accuracy of Working Casts Fabricated by Four Implant Impression Techniques in Parallel and Nonparallel Implant Configurations
Autoren
Abstract
Objective
The implant impression technique plays a critical role in determining the accuracy of implant working casts. This laboratory study aimed to evaluate and compare the accuracy of working casts fabricated using four different impression techniques in both parallel and nonparallel implant configurations.
Materials and Methods
An aluminum master model was made to simulate a mandibular dental arch. Impressions of the dental implants were made using four techniques within an incubator: (A) unsplinted tapered impression copings; (B) unsplinted squared impression copings; (C) splinted squared impression copings with dental floss supported with self-curing acrylic resin; and (D) splinted squared impression copings with prefabricated self-curing acrylic resin bars. Measurements were made using a universal measuring microscope in the x- and y-dimensions. The mean deviation index was calculated and compared. Statistical analysis was done using SPSS version 23. Comparisons of the distance deviation index in various groups were conducted using one-way analysis of variance. Furthermore, comparisons of the distance deviation index in nonparallel implants in various techniques were performed using the independent t-test. A p-value of 0.05 was considered a significant difference.
Results
For the parallel implants, the distance deviation index difference between the techniques was not significant. In the case of nonparallel implants, there was statistically significant difference only between the technique C versus technique D (p = 0.019). All the impression techniques showed statistically significant differences between parallel versus nonparallel implant placement in favor of the parallel. Overall, technique D demonstrated the best results for both parallel and nonparallel implant placements when compared with other techniques under the same conditions.
Conclusion
The impression technique affects the accuracy of the implant impression. Within the limitations of this study, impression transfer copings splinted with dental floss and reinforced with self-cure acrylic resin are not recommended, particularly when the implants are not parallel. In the case of parallel implants, there appears to be no statistically significant benefit from splinting transfer impression copings.
Introduction
Currently, dental implants are commonly used for the rehabilitation of patients with complete and partially edentulous arches.[1] [2] Considering this, the accuracy of the impression of dental implants is a critical factor for implant prosthesis success. Numerous factors affect the accuracy of implant restorations, such as impression materials and techniques,[3] [4] [5] implant angulations,[6] abutment types,[7] and prosthetic connection features.[3] In addition, edentulous space type, bone quality, surgical guide, and surgery protocol affect the implant placement.[8] All these factors affect the success of prosthetic restorations.[9]
The implant placement can be either parallel or tilted (nonparallel).[10] [11] [12] It is found that the stress is higher in the tilted implants. Naini et al[10] conducted a study to evaluate stress concentration in peri-implant bone in two different configurations: the distal implants tilted (model A) or four parallel implants (model S) and hybrid superstructures, and they found that during posterior loading, the tilted posterior implants were subjected to higher stresses in every condition but lower stress concentrations around the anterior implants of model A. In addition, Vojdani et al[11] compared the accuracy of implant impressions using three impression materials in both parallel and nonparallel implant positions and they found that in parallel conditions, the type of impression material does not affect the accuracy of the implant impressions; however, in nonparallel conditions, polyvinyl siloxane shows a better choice, followed by vinyl siloxane ether and polyether, respectively.
Furthermore, the implants can be splinted or unsplinted and it shows that the splinted technique demonstrated higher accuracy compared with the unsplinted technique.[13] At present, various digital aids are used in clinical dentistry, including implant placement and planning.[14] [15] The patients prefer the digital impression over the conventional impression due to convenience, time, and comfort.[16] [17] [18] For the implant prosthesis, the accuracy of the implant impression is an important step.[19] [20] The accuracy of digital impressions using intraoral scanners is acceptable for single crowns and short-span implant restorations.[21] [22] [23] But the digital impressions may show distortion for more than two implants in the long-span prosthesis.[24] [25] Hence, while making the impression for long-span implant-supported restorations, the use of digital impression presents some difficulties and subsequently affects the fit of the implant restoration.[26] Various studies show that the conventional impression technique shows more accuracy than the digital impression technique, especially for long-span prostheses.[24] [25] [27] [28] Hence, in full arch impression, the digital impression can show some error, hence, one needs to consider this when making an impression for long-span prosthesis.
In addition, there is limited information about the accuracy of impression techniques in parallel versus nonparallel implants in the complete edentulous arch.[29] This study aimed to evaluate and compare the accuracy of four impression techniques in angulated implants and the effect of the parallelism of the implants on the accuracy of the working model.
Materials and Methods
Metal Model
An aluminum master model was made to simulate a mandibular dental arch ([Fig. 1A]) with four holes to accommodate four implants. The holes were distributed like how implants are placed in the lower jaw. The two holes in the front were made parallel and the two holes at the back were inclined 15 degrees inward (lingually) to simulate the lingual tilt in the posterior regions of the lower jaw ([Fig. 1B]). Then, seven X-shape reference points ([Fig. 1C]) were placed in healing caps with precisely defined centers placed over the implants to make measurements at later stages as described previously.[29] Three individual trays (one closed and two open trays) were made from self-curing acrylic. Stoppers were placed inside the trays to unify the thickness of the impression material across all impressions; the protocol was the one utilized by Herbst et al.[29]


Impression and Working Model Fabrication
The study sample consisted of 64 impressions across four techniques (A, B, C, and D) as shown in [Table 1]. Two impressions were discarded due to deformation, resulting in 62 gypsum models being used for analysis. Technique A was only a closed-tray technique, whereas all other techniques (B, C, and D) were open-tray techniques ([Table 1]). The implants: two anterior parallel implants (implants 1 and 2) and two posterior implants (implants 3 and 4) inclined at 15 degrees inward (lingual). For each gypsum model, 56 readings were recorded and a total of 3,472 readings were recorded for all gypsum models as shown in [Table 2].
Abbreviations: SD, standard deviation; SE, standard error of the mean.
In technique A, the impression copings were fixed to the implants, and the impression was taken using a closed-tray impression technique ([Fig. 1D]). After the set, the impression was removed from the metal cast, leaving the impression copings on the implant. The impression coping was removed from the metal cast (master model), the implant replica was attached to the coping, then it was repositioned in the impression, and the gypsum was poured. In technique B, an open impression was used where the square transfer impression coping was placed on the model using a long fixation screw ([Fig. 1D]). A window was made in the impression at the location of the screw so it can be seen through it. After the material was set, the impression was removed after loosening the screw leaving the impression post in place within the impression material. In technique C, the impression posts were attached with dental floss (Oral B Essential Floss, Procter & Gamble, United States) ([Fig. 1E]), and self-curing acrylic (Dura Lay, Reliance Dental Manufacturing Co, Worth, Illinois, United States) was placed. In technique D, an open-tray impression was used where square impression copings were connected by prefabricated acrylic bars with an average thickness of 5 mm, with a small amount of acrylic ([Fig. 1F]).
The impression was made with polyvinyl siloxane (Dimension Penta H; 3M ESPE, Seefeld, Germany) using the double impression technique (single-stage impression). The heavy body was mixed and placed inside the tray and the light body was injected on the heavy body and around the impression copings, according to the manufacturer's instructions. All impressions were taken in an incubator (Taba l, Tsugami, Japan) at 37°C and 98% humidity to simulate the oral environment. The impressions were poured with gypsum according to the manufacturer's instructions. The metal model was placed in the incubator for 10 minutes before taking the impression to become homogeneous at 37°C and 98% humidity. Then, gypsum for model fabrication (Moldano, Heraeus Kulzer GmbH & Co., Germany) was used to approximate the working conditions in our laboratory. After pouring, all gypsum models were kept at room temperature for 24 hours before doing measurements.
Measurements and Calculations
The healing cap was placed on the implants in the master model, and then the dimensions were measured between the center of each healing cap and the seven points engraved on the master model using a universal measuring microscope with an accuracy of 0.1 micrometer (UM2D, Tsugami, Japan) ([Fig. 2]). Then, these caps were transferred to the gypsum models to take the measurements so that the healing cap no. (1) was placed on implant no. (1), and so on. The gypsum models were fixed on a base with two arms resting on the back corners of the model to prevent its movement. The coordinates (X points) (A, B, C, D, E, F, G) located on the surface of the gypsum model were measured for each of the implants (1, 2, 3, 4) under the constant working conditions of temperature (20 ± 1°C) and relative humidity (55 ± 5%). Then, the direct distance between the points and the implants was estimated based on the Pythagorean theorem. The readings were taken by the researcher with the help of laboratory technicians.


The accuracy of the readings was evaluated by measuring the distance between two points by a single reader more than 30 times, and the maximum difference did not exceed 0.0003, per manufacturer recommendation. The researcher and laboratory technician were asked to take the measurement between two points 30 times, and the maximum recorded difference was 0.0028 mm, which is acceptable per the manufacturer's recommendation. All the measurements were recorded with an accuracy of 0.0001 mm as mentioned by previous studies on different implant elements.[30]
The master model was measured three times and then the mean was calculated. The distance between the seven points on the model was measured ([Fig. 3]). For each of the four implants (after placing the healing cap on it), the distance is calculated as the total distance of the implant as follows:
TD (total distance) of implant 1 = Distance from Implant 1 to point A + Distance from Implant 1 to point B + Distance from Implant 1 to point C + Distance from Implant 1 to point D + Distance from Implant 1 to point E + Distance from Implant 1 to point F + Distance from Implant 1 to point G
Similarly, the healing caps were then transferred to the gypsum model and measurements were made to compare with the values of the master model. By calculating the value of the distance, the deviation index for each implant was calculated from the following equation:
Distance deviation index for implant 1 = [Total distance value in the Master model – value in the gypsum model | Total distance value in the Master model] × 100.[29]
Statistical Analysis
All data were entered in Excel and then into SPSS version 23 analysis software. Descriptive statistics (mean, standard deviation [SD], and standard error of the mean [SE]) were computed to summarize the data. Comparisons of the distance deviation index among the four techniques were done using one-way analysis of variance with post hoc analysis. Furthermore, comparisons of the distance deviation index in nonparallel implants among the four techniques were performed using the independent t-test. A p-value of 0.05 was considered a significant level.
Results
Effect of Different Impression Techniques on Dimensional Accuracy as Measured by Deviation Index
The deformity is defined as the relative movement of a specific point or a group of points relative to a point or several reference points.[31] In this study, the fixed reference points were marked as X (A, B, C, D, E, F, G), while the variable points were the implants, labeled 1, 2, 3, and 4. [Table 2] shows the descriptive statistics, mean, SD, and SE of the distance deviation index according to the impression technique and provides insight into the relative performance of each technique. Technique C showed the highest mean deviation index compared with techniques A, B, and D.
Furthermore, [Table 3] shows the group comparisons of the distance deviation index in various techniques. It showed that there was a significant difference among the groups (p = 0.033). Furthermore, the mean differences of the distance deviation index in various techniques and their multiple comparisons among the four techniques with p-values between each pair are shown in [Table 4]. It was found that statistically significant differences existed for the distance deviation index between various technique pairs, including technique A versus technique C (p-value = 0.027), technique B versus technique C (p-value = 0.027), and technique B versus technique D (p-value = 0.027). The comparisons like technique A versus technique B and technique A versus technique D showed no significant differences (p-value > 0.05). Overall, technique D outperforms the other techniques in most comparisons.
|
Impression techniques |
Sum of squares |
Degree of freedom |
Variance |
F |
p-Value |
|---|---|---|---|---|---|
|
Between groups |
0.165 |
3 |
0.055 |
2.948 |
0.033[a] |
|
Within groups |
4.543 |
244 |
0.019 |
Note: Statistical analysis was done using one-way analysis of variance (ANOVA).
a Significant difference at p-value of 0.05.
|
Impression techniques |
Comparison group |
Mean difference |
SE |
p-Value |
|---|---|---|---|---|
|
Technique A |
Technique B |
0.0183 |
0.02 |
1.000 |
|
Technique C |
−0.0294 |
0.02 |
0.027[a] |
|
|
Technique D |
0.0410 |
0.02 |
0.577 |
|
|
Technique B |
Technique C |
0.0227 |
0.02 |
0.027[a] |
|
Technique D |
0.0703 |
0.02 |
0.027[a] |
|
|
Technique C |
Technique D |
0.0703 |
0.02 |
0.027[a] |
Abbreviation: SE, standard error of the mean difference.
Note: Statistical analysis was done using one-way analysis of variance (ANOVA) with post hoc analysis.
a Significant difference at p-value of 0.05.
Influence of Implant Parallelism on Dimensional Accuracy as Reflected by the Deviation Index
[Table 5] shows the group comparison of the distance deviation index in nonparallel implants in various techniques. It showed that there was a significant difference among the groups in nonparallel implants (p-value = 0.027). In the case of nonparallel implants, there was a statistically significant difference in the distance deviation index only between technique C versus technique D (p = 0.019) as shown in [Table 6].
|
Implant parallelism |
Impression techniques |
Sum of squares |
Degree of freedom |
Variance |
F |
p-Value |
|---|---|---|---|---|---|---|
|
Parallel |
Between groups |
0.009 |
3 |
0.003 |
1.100 |
0.352 |
|
Within groups |
0.324 |
120 |
0.003 |
|||
|
Nonparallel |
Between groups |
0.260 |
3 |
0.087 |
3.154 |
0.027[a] |
|
Within groups |
3.302 |
120 |
0.028 |
Note: Statistical analysis was done using an independent t-test.
a Significant difference at p-value 0.05.
|
Implant parallelism |
Impression techniques |
Comparison group |
Mean difference |
SE |
p-Value |
|---|---|---|---|---|---|
|
Nonparallel |
Technique A |
Technique B |
0.0182 |
0.042 |
1.000 |
|
Technique C |
−0.0661 |
0.041 |
0.680 |
||
|
Technique D |
0.0608 |
0.042 |
0.911 |
||
|
Technique B |
Technique C |
−0.0844 |
0.042 |
0.285 |
|
|
Technique D |
0.0426 |
0.043 |
1.000 |
||
|
Technique C |
Technique D |
0.1269 |
0.042 |
0.019[a] |
Abbreviation: SE, standard error of the mean difference.
Note: Statistical analysis was done using an independent t-test.
a Significant difference at p-value of 0.05.
Similarly, [Table 7] shows the comparison of the distance deviation index in parallel and nonparallel implants across various techniques. All the impression techniques showed statistically significant differences between parallel versus nonparallel implant placement in favor of the parallel. Overall, technique D demonstrated the best results for both parallel and nonparallel implant placements when compared with other techniques under the same conditions.
|
Impression technique |
Implant parallelism |
Mean |
Mean difference |
SE |
p-Value |
|---|---|---|---|---|---|
|
Technique A |
Parallel |
0.0793 |
−0.1052 |
0.0389 |
0.009[a] |
|
Nonparallel |
0.1845 |
||||
|
Technique B |
Parallel |
0.0611 |
−0.1051 |
0.0260 |
< 0.0001[a] |
|
Nonparallel |
0.1662 |
||||
|
Technique C |
Parallel |
0.0720 |
−0.1786 |
0.0360 |
< 0.0001[a] |
|
Nonparallel |
0.2506 |
||||
|
Technique D |
Parallel |
0.0096 |
−0.1141 |
0.0170 |
< 0.0001[a] |
|
Nonparallel |
0.1237 |
Abbreviation: SE, standard error of the mean difference.
Note: Statistical analysis was done using an independent t-test.
a Significant difference at p-value of 0.05.
Discussion
The present laboratory study evaluated the impact of different impression techniques and implant parallelism on the working casts, with the aim of determining the most accurate method for replicating implant positions. The results demonstrated that technique D yielded the best outcomes for both parallel and nonparallel implant placements when compared with other tested techniques. Interestingly, when implants were placed in a parallel configuration, the study was unable to detect any statistically significant differences between the impression techniques, suggesting that implant parallelism may reduce the technique sensitivity and overall influence on cast accuracy.
An implant impression is crucial to fabricate an implant restoration on a working model that precisely replicates the implant position in the patient's mouth.[32] Splinting the impression posts improves the accuracy of the impression by fixing the impression posts and preventing their movement within the impression material during the different stages of cast fabrication.[33] [34] This technique accurately replicates the spatial position of the implant to the model as it exists in the patient's mouth.[35] [36] [37] Furthermore, while making the implant impression, splinting the implant components using the open-tray method increases the impression accuracy.[38] [39] However, Humphries et al[40] evaluated three impression techniques using polyvinyl siloxane and found no significant difference between splinted and nonsplinted open-tray impression methods. The closed-tray impression technique, utilizing tapered impression copings, demonstrated the highest accuracy. However, all three methods were generally capable of producing clinically acceptable impressions. But still, clinicians support that splinting the impression posts improves the accuracy of the implant impression. For the splinting of implant components, Duralay acrylic is commonly used. Additionally, splints made with acrylic resin showed more accuracy than those made with light-cured composite resins.[41] Hariharan et al[42] showed that splinting with acrylic resin is more accurate than the nonsplint method. Furthermore, the concept of attaching impression posts using self-cure acrylic supported on dental floss was recommended previously.[43] However, this technique is relatively complex, requires time and effort, and may be inaccurate. This requires bonding with acrylic to make the impression more accurate with less distortion.[44] Spector et al[44] conducted a study on a model containing six parallel implants and compared the indirect method with and without bonding, and they found that no method was free of distortion. Similarly, Mojon et al[45] conducted a study to estimate the effect of the amount of acrylic on the accuracy of the impression and they found that the larger the amount of acrylic used to bind the impression copings, the greater the distortion. The distortion is due to the shrinkage of the hardened acrylic.[45] [46] Furthermore, Mojon et al[45] reported that the shrinkage within the first 24 hours reaches 7.9%, with approximately 80% of this shrinkage occurring within the first 17 minutes after mixing at room temperature. Hence, need to consider this while splinting, but still, it is a preferred material for chairside splinting.
For the full arch implant prosthesis, conventional impressions are used due to accuracy concerns.
A recent study by Jasim et al[28] compared the accuracy of implant-level conventional and digital impressions for atrophied maxillary ridges: (1) conventional (splinted open-tray) impression technique and (2) digital impression technique. The accuracy was measured between two impression techniques using in vitro (two-dimensional [2D] and three-dimensional [3D]) and in vivo methods. The 2D methods include measurement of the difference in linear distances, whereas 3D deviations were measured by the superimposition of STL files using the Geomagic software. The degree of framework misfits of final restorations was assessed using the single screw in the two impression techniques. It was found that digital impressions showed significantly higher deviation in 2D distances than conventional impressions. Similarly, digital impressions recorded significantly higher 3D deviation for all scan bodies than conventional impressions. And, digital impressions recorded a significantly higher incidence of nonpassive frameworks and framework misfits compared with conventional impressions. And they concluded that the conventional impression technique showed greater accuracy than the digital for the full-arch fixed restorations on inclined implants in the maxillary arch.
Similarly, a study by Elashry et al[47] studied the accuracy of the digital impression compared with the conventional implant impression techniques in bilateral distal extension cases in the mandibular arch using 32 implants. The patients underwent two implant-level impression techniques: digital implant impressions with TRIOS 3 Shape intraoral scanner and conventional open-tray impressions (splinted pick-up) and the accuracy of impressions was evaluated using the 3D superimposition analysis of STL files. Subsequently, the scan bodies were segmented using the Gom Inspect software to measure 3D deviations. It showed that higher positional and angular deviations were shown toward distal scan bodies compared with mesial ones for both impression techniques, but the difference was not significant. They concluded that splinted open-tray conventional impression and intraoral scanning implant impression techniques showed comparable accuracy. Similarly, another study by Palantza et al[48] compared the accuracy of full-arch conventional implant impressions using two different materials (silicone and polyether) to full-arch digital implant impressions produced from two intraoral scanners (Trios vs. Heron). They found that the accuracy of mandibular implant impressions was influenced both by the impression technique (conventional vs. digital) and the impression material used (silicone vs. polyether), or the intraoral scanner. In terms of “trueness,” silicone showed the highest accuracy, followed by polyether and Trios, but the differences between the three groups were not significant. Heron showed lower accuracy results in all measurements compared with the other groups. In terms of “precision,” conventional impressions with silicone or polyether were significantly superior to digital impressions with either scanner. Silicone and polyether showed no statistically significant difference. Hence, while making the impression for implant prostheses, these factors need to be considered.
González Menéndez et al[49] compared the angular deviation of two, four, and six dental implants placed with and without the use of parallel copings, using cone-beam computed tomography scans and 3D implant-planning software. They reported that the use of parallelization copings resulted in lower deviation when two or four implants were included in the impression. However, this technique was not effective in cases involving a greater number of implants. Furthermore, Osman et al[50] evaluated the accuracy of impression of a simulated area restored with an implant-retained fixed partial denture, using the open- and closed-tray implant impression techniques and studied the effect of implant position angulation, parallelism, and implant systems (Straumann, SIC Invent, and Osstem). They found that there was no significant difference between open and closed, although better results were obtained for the open-tray techniques. With nonparallel implants, the open-tray technique provided a better result compared with the closed-tray technique. Papaspyridakos et al[13] compared the 3D accuracy of splinted versus unsplinted impression techniques and determined the optimal clinically undetectable misfit. They mentioned that for the external connection, an acceptable 3D misfit in implant-fixed prosthesis ranges from 59 to 72 μm, resulting in an acceptable fit clinically. In addition, they found that the splinted technique showed more accurate master casts than the unsplinted one in edentulous jaws, which is similar to our study. In our research, technique D with parallel implant placement showed the best results. In addition, a recent study shows that the dynamic computer-assisted implant surgery provides similar accuracy of the 3D implant position and parallelism between two implants compared with static and dynamic computer-assisted implant surgery.[51] [52]
With the rise of digital dentistry, digital impressions using intraoral scanners have been adopted in implant dentistry, but the accuracy is a major issue in long-span arches.[21] In addition, the difficulty in scanning the soft tissue between implants in edentulous arches remains a major issue. There are various methods to increase the accuracy of the full arch impression. Greater accuracy is seen when the implants are placed in bounded edentulous areas with guided surgery using computer-aided design/computer-aided manufacturing-manufactured surgical guides.[8] In digital impression, the successful stitching of images is important for accuracy, and this process is directly related to both the scanning scope and the interimplant distance.[53] [54] A study by Kim et al[55] mentioned that an artificial landmark in the long edentulous area can improve the trueness and precision of the intraoral scanners. The auxiliary geometric devices attached to implant scan bodies via flowable composite can increase 3D spatial recognition of soft tissues.[56] Similarly, Arikan et al[19] suggested that the use of auxiliary geometric appliances yields increased scanning accuracy. In addition, they suggested that the frameworks fabricated using the traditional splinted open-tray technique were more reliable compared with those frameworks from digital impressions.
There are some limitations in this research study. This study investigated only conventional impressions. Second, the limited sample of dental implants due to time and financial constraints, and the findings should be interpreted carefully. In the future, the conventional impression technique can be compared with the digital impression techniques.
Conclusion
The dental implants can be placed parallel or tilted and the implant impression can be made via splinted versus unsplinted impression techniques. The impression technique affects the accuracy of the implant impression. Within the limitations of this study, the impression copings splinted with dental floss and self-cure acrylic resin showed the highest distortion. There seems to be no advantage clinically in splinting transfer impression copings with self-cure acrylic resin in the case of parallel implants. Further clinical studies are needed.


Conflict of Interest
None declared.
Acknowledgments
We would like to acknowledge Prof. Dr. Iyad Al-Shaaran for his research guidance and the National Standards & Calibration Laboratory, Damascus, Syria.
Data Availability Statement
The data sets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.
Authors' Contributions
A.A.J. designed the study, participated in methodology, data analysis, and wrote the initial draft. D.R. helped in designing the study, methodology, interpreted the data, and critically revised the original. All authors approved the final manuscript.
-
References
- 1 Witoonkitvanich P, Amornsettachai P, Panyayong W, Rokaya D, Vongsirichat N, Suphangul S. Comparison of the stability of immediate dental implant placement in fresh molar extraction sockets in the maxilla and mandible: a controlled, prospective, non-randomized clinical trial. Int J Oral Maxillofac Implants 2025; 54 (04) 365-373
- 2 Alkhames HM, Ali RMM, Alzouri SS, Bayome M. Assessment of posterior maxillary alveolar bone for immediate implant placement: a quantitative and qualitative analysis. Eur J Dent 2024; 18 (03) 877-882
- 3 Lee YJ, Heo SJ, Koak JY, Kim SK. Accuracy of different impression techniques for internal-connection implants. Int J Oral Maxillofac Implants 2009; 24 (05) 823-830
- 4 Barrett MG, de Rijk WG, Burgess JO. The accuracy of six impression techniques for osseointegrated implants. J Prosthodont 1993; 2 (02) 75-82
- 5 Albanchez-González MI, Brinkmann JC, Peláez-Rico J, López-Suárez C, Rodríguez-Alonso V, Suárez-García MJ. Accuracy of digital dental implants impression taking with intraoral scanners compared with conventional impression techniques: a systematic review of in vitro studies. Int J Environ Res Public Health 2022; 19 (04) 2026
- 6 Conrad HJ, Pesun IJ, DeLong R, Hodges JS. Accuracy of two impression techniques with angulated implants. J Prosthet Dent 2007; 97 (06) 349-356
- 7 Suphangul S, Pujarern P, Rokaya D, Kanchanasobhana C, Rungsiyakull P, Chaijareenont P. Comparison of plaque accumulation between titanium and PEEK healing abutments. J Funct Biomater 2024; 15 (11) 334
- 8 Putra RH, Yoda N, Astuti ER, Sasaki K. The accuracy of implant placement with computer-guided surgery in partially edentulous patients and possible influencing factors: a systematic review and meta-analysis. J Prosthodont Res 2022; 66 (01) 29-39
- 9 Parameshwari G, Chittaranjan B, Sudhir N, Anulekha-Avinash CK, Taruna M, Ramureddy M. Evaluation of accuracy of various impression techniques and impression materials in recording multiple implants placed unilaterally in a partially edentulous mandible- an in vitro study. J Clin Exp Dent 2018; 10 (04) e388-e395
- 10 Naini RB, Nokar S, Borghei H, Alikhasi M. Tilted or parallel implant placement in the completely edentulous mandible? A three-dimensional finite element analysis. Int J Oral Maxillofac Implants 2011; 26 (04) 776-781
- 11 Vojdani M, Torabi K, Ansarifard E. Accuracy of different impression materials in parallel and nonparallel implants. Dent Res J (Isfahan) 2015; 12 (04) 315-322
- 12 V. M, Chirom B, Gunadhar K, Priyadarshini S, Nongthombam RS, P. M. The accuracy of casts obtained using different impression techniques and impression materials in combined parallel and angulated implants: an in vitro study. Cureus 2024; 16 (04) e59193
- 13 Papaspyridakos P, Benic GI, Hogsett VL, White GS, Lal K, Gallucci GO. Accuracy of implant casts generated with splinted and non-splinted impression techniques for edentulous patients: an optical scanning study. Clin Oral Implants Res 2012; 23 (06) 676-681
- 14 Humagain M, Rokaya D. Integrating digital technologies in dentistry to enhance the clinical success. Kathmandu Univ Med J 2019; 17 (68) 256-257
- 15 Alsulimani O, Alhaddad A, Altassan M, Bukhari A, Munshi L, Sabir G. The precision of all-on-four implant position recorded from three different CBCT machines. Eur J Dent 2025; 19 (02) 337-345
- 16 de Oliveira NRC, Pigozzo MN, Sesma N, Laganá DC. Clinical efficiency and patient preference of digital and conventional workflow for single implant crowns using immediate and regular digital impression: a meta-analysis. Clin Oral Implants Res 2020; 31 (08) 669-686
- 17 Pan S, Guo D, Zhou Y, Jung RE, Hämmerle CHF, Mühlemann S. Time efficiency and quality of outcomes in a model-free digital workflow using digital impression immediately after implant placement: a double-blind self-controlled clinical trial. Clin Oral Implants Res 2019; 30 (07) 617-626
- 18 Guo DN, Liu YS, Pan SX. et al. Clinical efficiency and patient preference of immediate digital impression after implant placement for single implant-supported crown. Chin J Dent Res 2019; 22 (01) 21-28
- 19 Arikan H, Muhtarogullari M, Uzel SM. et al. Accuracy of digital impressions for implant-supported complete-arch prosthesis when using an auxiliary geometry device. J Dent Sci 2023; 18 (02) 808-813
- 20 Alkindi S, Hamdoon Z, Aziz AM. Effect of different impression coping and scan body designs on the accuracy of conventional versus digital implant impressions: an in vitro study. J Dent 2024; 146: 105045
- 21 Asar NV, Yun S, Schwartz S, Turkyilmaz I. Analysis of the relationship between the surface topography of prepared tooth surfaces and data quality of digital impressions from an intraoral scanner. J Dent Sci 2022; 17 (01) 545-550
- 22 Marghalani A, Weber HP, Finkelman M, Kudara Y, El Rafie K, Papaspyridakos P. Digital versus conventional implant impressions for partially edentulous arches: an evaluation of accuracy. J Prosthet Dent 2018; 119 (04) 574-579
- 23 Schmidt A, Wöstmann B, Schlenz MA. Accuracy of digital implant impressions in clinical studies: a systematic review. Clin Oral Implants Res 2022; 33 (06) 573-585
- 24 Abduo J, Palamara JEA. Accuracy of digital impressions versus conventional impressions for 2 implants: an in vitro study evaluating the effect of implant angulation. Int J Implant Dent 2021; 7 (01) 75
- 25 El Osta N, Drancourt N, Auduc C, Veyrune JL, Nicolas E. Accuracy of conventional impressions and digital scans for implant-supported fixed prostheses in maxillary free-ended partial edentulism: an in vitro study. J Dent 2024; 143: 104892
- 26 Rutkunas V, Gedrimiene A, Akulauskas M, Fehmer V, Sailer I, Jegelevicius D. In vitro and in vivo accuracy of full-arch digital implant impressions. Clin Oral Implants Res 2021; 32 (12) 1444-1454
- 27 Drancourt N, Auduc C, Mouget A. et al. Accuracy of conventional and digital impressions for full-arch implant-supported prostheses: an in vitro study. J Pers Med 2023; 13 (05) 832
- 28 Jasim AG, Abo Elezz MG, Altonbary GY, Elsyad MA. Accuracy of digital and conventional implant-level impression techniques for maxillary full-arch screw-retained prosthesis: a crossover randomized trial. Clin Implant Dent Relat Res 2024; 26 (04) 714-723
- 29 Herbst D, Nel JC, Driessen CH, Becker PJ. Evaluation of impression accuracy for osseointegrated implant supported superstructures. J Prosthet Dent 2000; 83 (05) 555-561
- 30 Phillips KM, Nicholls JI, Ma T, Rubenstein J. The accuracy of three implant impression techniques: a three-dimensional analysis. Int J Oral Maxillofac Implants 1994; 9: 533-540
- 31 Nicholls JI. The measurement of distortion: theoretical considerations. J Prosthet Dent 1977; 37 (05) 578-586
- 32 Osman M, Ziada H, Suliman A, Abubakr NH. A prospective clinical study on implant impression accuracy. Int J Implant Dent 2019; 5 (01) 38
- 33 Del Acqua MA, Chavez AM, Castanharo SM, Compagnoni MA, Mollo Jr FdeA. The effect of splint material rigidity in implant impression techniques. Int J Oral Maxillofac Implants 2010; 25 (06) 1153-1158
- 34 Elshenawy EA, Alam-Eldein AM, Abd Elfatah FA. Cast accuracy obtained from different impression techniques at different implant angulations (in vitro study). Int J Implant Dent 2018; 4 (01) 9
- 35 Tabesh M, Alikhasi M, Siadat H. A Comparison of implant impression precision: different materials and techniques. J Clin Exp Dent 2018; 10 (02) e151-e157
- 36 Nateghi M, Negahdari R, Molaei S, Barzegar A, Bohlouli S. Comparison of the accuracy of fixture-level implant impression making with different splinting techniques. Int J Dent 2021; 2021: 2959055
- 37 Bhakta S, Vere J, Calder I, Patel R. Impressions in implant dentistry. Br Dent J 2011; 211 (08) 361-367
- 38 Negahdari R, Barzegar A, Mortazavi Milani A, Sheikh Ahmadi Y, Rahbar M. Comparison of the accuracy of fixture-level implant impressions using two different materials for splinting. J Adv Periodontol Implant Dent 2022; 14 (02) 134-140
- 39 Beyabanaki E, Shamshiri AR, Alikhasi M, Monzavi A. Effect of splinting on dimensional accuracy of impressions made of implants with different subgingival alignments. J Prosthodont 2017; 26 (01) 48-55
- 40 Humphries RM, Yaman P, Bloem TJ. The accuracy of implant master casts constructed from transfer impressions. Int J Oral Maxillofac Implants 1990; 5 (04) 331-336
- 41 Ongül D, Gökçen-Röhlig B, Şermet B, Keskin H. A comparative analysis of the accuracy of different direct impression techniques for multiple implants. Aust Dent J 2012; 57 (02) 184-189
- 42 Hariharan R, Shankar C, Rajan M, Baig MR, Azhagarasan NS. Evaluation of accuracy of multiple dental implant impressions using various splinting materials. Int J Oral Maxillofac Implants 2010; 25 (01) 38-44
- 43 Assif D, Marshak B, Nissan J. A modified impression technique for implant-supported restoration. J Prosthet Dent 1994; 71 (06) 589-591
- 44 Spector MR, Donovan TE, Nicholls JI. An evaluation of impression techniques for osseointegrated implants. J Prosthet Dent 1990; 63 (04) 444-447
- 45 Mojon P, Oberholzer JP, Meyer JM, Belser UC. Polymerization shrinkage of index and pattern acrylic resins. J Prosthet Dent 1990; 64 (06) 684-688
- 46 Ness EM, Nicholls JI, Rubenstein JE, Smith DE. Accuracy of the acrylic resin pattern for the implant-retained prosthesis. Int J Prosthodont 1992; 5 (06) 542-549
- 47 Elashry WY, Elsheikh MM, Elsheikh AM. Evaluation of the accuracy of conventional and digital implant impression techniques in bilateral distal extension cases: a randomized clinical trial. BMC Oral Health 2024; 24 (01) 764
- 48 Palantza E, Sykaras N, Zoidis P, Kourtis S. In vitro comparison of accuracy between conventional and digital impression using elastomeric materials and two intra-oral scanning devices. J Esthet Restor Dent 2024; 36 (08) 1179-1198
- 49 González Menéndez H, Lorrio Castro J, Rodríguez Torres P. et al. Influence of parallel pins on the angle deviation for placement of dental implants: an in vitro study. BMC Oral Health 2024; 24 (01) 134
- 50 Osman MS, Ziada HM, Abubakr NH, Suliman AM. Implant impression accuracy of parallel and non-parallel implants: a comparative in-vitro analysis of open and closed tray techniques. Int J Implant Dent 2019; 5 (01) 4
- 51 Yimarj P, Subbalekha K, Dhanesuan K, Siriwatana K, Mattheos N, Pimkhaokham A. Comparison of the accuracy of implant position for two-implants supported fixed dental prosthesis using static and dynamic computer-assisted implant surgery: a randomized controlled clinical trial. Clin Implant Dent Relat Res 2020; 22 (06) 672-678
- 52 Kaewsiri D, Panmekiate S, Subbalekha K, Mattheos N, Pimkhaokham A. The accuracy of static vs. dynamic computer-assisted implant surgery in single tooth space: a randomized controlled trial. Clin Oral Implants Res 2019; 30 (06) 505-514
- 53 Knechtle N, Wiedemeier D, Mehl A, Ender A. Accuracy of digital complete-arch, multi-implant scans made in the edentulous jaw with gingival movement simulation: an in vitro study. J Prosthet Dent 2022; 128 (03) 468-478
- 54 Kim KR, Seo KY, Kim S. Conventional open-tray impression versus intraoral digital scan for implant-level complete-arch impression. J Prosthet Dent 2019; 122 (06) 543-549
- 55 Kim JE, Amelya A, Shin Y, Shim JS. Accuracy of intraoral digital impressions using an artificial landmark. J Prosthet Dent 2017; 117 (06) 755-761
- 56 Iturrate M, Eguiraun H, Etxaniz O, Solaberrieta E. Accuracy analysis of complete-arch digital scans in edentulous arches when using an auxiliary geometric device. J Prosthet Dent 2019; 121 (03) 447-454
Address for correspondence
Publikationsverlauf
Artikel online veröffentlicht:
08. September 2025
© 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Witoonkitvanich P, Amornsettachai P, Panyayong W, Rokaya D, Vongsirichat N, Suphangul S. Comparison of the stability of immediate dental implant placement in fresh molar extraction sockets in the maxilla and mandible: a controlled, prospective, non-randomized clinical trial. Int J Oral Maxillofac Implants 2025; 54 (04) 365-373
- 2 Alkhames HM, Ali RMM, Alzouri SS, Bayome M. Assessment of posterior maxillary alveolar bone for immediate implant placement: a quantitative and qualitative analysis. Eur J Dent 2024; 18 (03) 877-882
- 3 Lee YJ, Heo SJ, Koak JY, Kim SK. Accuracy of different impression techniques for internal-connection implants. Int J Oral Maxillofac Implants 2009; 24 (05) 823-830
- 4 Barrett MG, de Rijk WG, Burgess JO. The accuracy of six impression techniques for osseointegrated implants. J Prosthodont 1993; 2 (02) 75-82
- 5 Albanchez-González MI, Brinkmann JC, Peláez-Rico J, López-Suárez C, Rodríguez-Alonso V, Suárez-García MJ. Accuracy of digital dental implants impression taking with intraoral scanners compared with conventional impression techniques: a systematic review of in vitro studies. Int J Environ Res Public Health 2022; 19 (04) 2026
- 6 Conrad HJ, Pesun IJ, DeLong R, Hodges JS. Accuracy of two impression techniques with angulated implants. J Prosthet Dent 2007; 97 (06) 349-356
- 7 Suphangul S, Pujarern P, Rokaya D, Kanchanasobhana C, Rungsiyakull P, Chaijareenont P. Comparison of plaque accumulation between titanium and PEEK healing abutments. J Funct Biomater 2024; 15 (11) 334
- 8 Putra RH, Yoda N, Astuti ER, Sasaki K. The accuracy of implant placement with computer-guided surgery in partially edentulous patients and possible influencing factors: a systematic review and meta-analysis. J Prosthodont Res 2022; 66 (01) 29-39
- 9 Parameshwari G, Chittaranjan B, Sudhir N, Anulekha-Avinash CK, Taruna M, Ramureddy M. Evaluation of accuracy of various impression techniques and impression materials in recording multiple implants placed unilaterally in a partially edentulous mandible- an in vitro study. J Clin Exp Dent 2018; 10 (04) e388-e395
- 10 Naini RB, Nokar S, Borghei H, Alikhasi M. Tilted or parallel implant placement in the completely edentulous mandible? A three-dimensional finite element analysis. Int J Oral Maxillofac Implants 2011; 26 (04) 776-781
- 11 Vojdani M, Torabi K, Ansarifard E. Accuracy of different impression materials in parallel and nonparallel implants. Dent Res J (Isfahan) 2015; 12 (04) 315-322
- 12 V. M, Chirom B, Gunadhar K, Priyadarshini S, Nongthombam RS, P. M. The accuracy of casts obtained using different impression techniques and impression materials in combined parallel and angulated implants: an in vitro study. Cureus 2024; 16 (04) e59193
- 13 Papaspyridakos P, Benic GI, Hogsett VL, White GS, Lal K, Gallucci GO. Accuracy of implant casts generated with splinted and non-splinted impression techniques for edentulous patients: an optical scanning study. Clin Oral Implants Res 2012; 23 (06) 676-681
- 14 Humagain M, Rokaya D. Integrating digital technologies in dentistry to enhance the clinical success. Kathmandu Univ Med J 2019; 17 (68) 256-257
- 15 Alsulimani O, Alhaddad A, Altassan M, Bukhari A, Munshi L, Sabir G. The precision of all-on-four implant position recorded from three different CBCT machines. Eur J Dent 2025; 19 (02) 337-345
- 16 de Oliveira NRC, Pigozzo MN, Sesma N, Laganá DC. Clinical efficiency and patient preference of digital and conventional workflow for single implant crowns using immediate and regular digital impression: a meta-analysis. Clin Oral Implants Res 2020; 31 (08) 669-686
- 17 Pan S, Guo D, Zhou Y, Jung RE, Hämmerle CHF, Mühlemann S. Time efficiency and quality of outcomes in a model-free digital workflow using digital impression immediately after implant placement: a double-blind self-controlled clinical trial. Clin Oral Implants Res 2019; 30 (07) 617-626
- 18 Guo DN, Liu YS, Pan SX. et al. Clinical efficiency and patient preference of immediate digital impression after implant placement for single implant-supported crown. Chin J Dent Res 2019; 22 (01) 21-28
- 19 Arikan H, Muhtarogullari M, Uzel SM. et al. Accuracy of digital impressions for implant-supported complete-arch prosthesis when using an auxiliary geometry device. J Dent Sci 2023; 18 (02) 808-813
- 20 Alkindi S, Hamdoon Z, Aziz AM. Effect of different impression coping and scan body designs on the accuracy of conventional versus digital implant impressions: an in vitro study. J Dent 2024; 146: 105045
- 21 Asar NV, Yun S, Schwartz S, Turkyilmaz I. Analysis of the relationship between the surface topography of prepared tooth surfaces and data quality of digital impressions from an intraoral scanner. J Dent Sci 2022; 17 (01) 545-550
- 22 Marghalani A, Weber HP, Finkelman M, Kudara Y, El Rafie K, Papaspyridakos P. Digital versus conventional implant impressions for partially edentulous arches: an evaluation of accuracy. J Prosthet Dent 2018; 119 (04) 574-579
- 23 Schmidt A, Wöstmann B, Schlenz MA. Accuracy of digital implant impressions in clinical studies: a systematic review. Clin Oral Implants Res 2022; 33 (06) 573-585
- 24 Abduo J, Palamara JEA. Accuracy of digital impressions versus conventional impressions for 2 implants: an in vitro study evaluating the effect of implant angulation. Int J Implant Dent 2021; 7 (01) 75
- 25 El Osta N, Drancourt N, Auduc C, Veyrune JL, Nicolas E. Accuracy of conventional impressions and digital scans for implant-supported fixed prostheses in maxillary free-ended partial edentulism: an in vitro study. J Dent 2024; 143: 104892
- 26 Rutkunas V, Gedrimiene A, Akulauskas M, Fehmer V, Sailer I, Jegelevicius D. In vitro and in vivo accuracy of full-arch digital implant impressions. Clin Oral Implants Res 2021; 32 (12) 1444-1454
- 27 Drancourt N, Auduc C, Mouget A. et al. Accuracy of conventional and digital impressions for full-arch implant-supported prostheses: an in vitro study. J Pers Med 2023; 13 (05) 832
- 28 Jasim AG, Abo Elezz MG, Altonbary GY, Elsyad MA. Accuracy of digital and conventional implant-level impression techniques for maxillary full-arch screw-retained prosthesis: a crossover randomized trial. Clin Implant Dent Relat Res 2024; 26 (04) 714-723
- 29 Herbst D, Nel JC, Driessen CH, Becker PJ. Evaluation of impression accuracy for osseointegrated implant supported superstructures. J Prosthet Dent 2000; 83 (05) 555-561
- 30 Phillips KM, Nicholls JI, Ma T, Rubenstein J. The accuracy of three implant impression techniques: a three-dimensional analysis. Int J Oral Maxillofac Implants 1994; 9: 533-540
- 31 Nicholls JI. The measurement of distortion: theoretical considerations. J Prosthet Dent 1977; 37 (05) 578-586
- 32 Osman M, Ziada H, Suliman A, Abubakr NH. A prospective clinical study on implant impression accuracy. Int J Implant Dent 2019; 5 (01) 38
- 33 Del Acqua MA, Chavez AM, Castanharo SM, Compagnoni MA, Mollo Jr FdeA. The effect of splint material rigidity in implant impression techniques. Int J Oral Maxillofac Implants 2010; 25 (06) 1153-1158
- 34 Elshenawy EA, Alam-Eldein AM, Abd Elfatah FA. Cast accuracy obtained from different impression techniques at different implant angulations (in vitro study). Int J Implant Dent 2018; 4 (01) 9
- 35 Tabesh M, Alikhasi M, Siadat H. A Comparison of implant impression precision: different materials and techniques. J Clin Exp Dent 2018; 10 (02) e151-e157
- 36 Nateghi M, Negahdari R, Molaei S, Barzegar A, Bohlouli S. Comparison of the accuracy of fixture-level implant impression making with different splinting techniques. Int J Dent 2021; 2021: 2959055
- 37 Bhakta S, Vere J, Calder I, Patel R. Impressions in implant dentistry. Br Dent J 2011; 211 (08) 361-367
- 38 Negahdari R, Barzegar A, Mortazavi Milani A, Sheikh Ahmadi Y, Rahbar M. Comparison of the accuracy of fixture-level implant impressions using two different materials for splinting. J Adv Periodontol Implant Dent 2022; 14 (02) 134-140
- 39 Beyabanaki E, Shamshiri AR, Alikhasi M, Monzavi A. Effect of splinting on dimensional accuracy of impressions made of implants with different subgingival alignments. J Prosthodont 2017; 26 (01) 48-55
- 40 Humphries RM, Yaman P, Bloem TJ. The accuracy of implant master casts constructed from transfer impressions. Int J Oral Maxillofac Implants 1990; 5 (04) 331-336
- 41 Ongül D, Gökçen-Röhlig B, Şermet B, Keskin H. A comparative analysis of the accuracy of different direct impression techniques for multiple implants. Aust Dent J 2012; 57 (02) 184-189
- 42 Hariharan R, Shankar C, Rajan M, Baig MR, Azhagarasan NS. Evaluation of accuracy of multiple dental implant impressions using various splinting materials. Int J Oral Maxillofac Implants 2010; 25 (01) 38-44
- 43 Assif D, Marshak B, Nissan J. A modified impression technique for implant-supported restoration. J Prosthet Dent 1994; 71 (06) 589-591
- 44 Spector MR, Donovan TE, Nicholls JI. An evaluation of impression techniques for osseointegrated implants. J Prosthet Dent 1990; 63 (04) 444-447
- 45 Mojon P, Oberholzer JP, Meyer JM, Belser UC. Polymerization shrinkage of index and pattern acrylic resins. J Prosthet Dent 1990; 64 (06) 684-688
- 46 Ness EM, Nicholls JI, Rubenstein JE, Smith DE. Accuracy of the acrylic resin pattern for the implant-retained prosthesis. Int J Prosthodont 1992; 5 (06) 542-549
- 47 Elashry WY, Elsheikh MM, Elsheikh AM. Evaluation of the accuracy of conventional and digital implant impression techniques in bilateral distal extension cases: a randomized clinical trial. BMC Oral Health 2024; 24 (01) 764
- 48 Palantza E, Sykaras N, Zoidis P, Kourtis S. In vitro comparison of accuracy between conventional and digital impression using elastomeric materials and two intra-oral scanning devices. J Esthet Restor Dent 2024; 36 (08) 1179-1198
- 49 González Menéndez H, Lorrio Castro J, Rodríguez Torres P. et al. Influence of parallel pins on the angle deviation for placement of dental implants: an in vitro study. BMC Oral Health 2024; 24 (01) 134
- 50 Osman MS, Ziada HM, Abubakr NH, Suliman AM. Implant impression accuracy of parallel and non-parallel implants: a comparative in-vitro analysis of open and closed tray techniques. Int J Implant Dent 2019; 5 (01) 4
- 51 Yimarj P, Subbalekha K, Dhanesuan K, Siriwatana K, Mattheos N, Pimkhaokham A. Comparison of the accuracy of implant position for two-implants supported fixed dental prosthesis using static and dynamic computer-assisted implant surgery: a randomized controlled clinical trial. Clin Implant Dent Relat Res 2020; 22 (06) 672-678
- 52 Kaewsiri D, Panmekiate S, Subbalekha K, Mattheos N, Pimkhaokham A. The accuracy of static vs. dynamic computer-assisted implant surgery in single tooth space: a randomized controlled trial. Clin Oral Implants Res 2019; 30 (06) 505-514
- 53 Knechtle N, Wiedemeier D, Mehl A, Ender A. Accuracy of digital complete-arch, multi-implant scans made in the edentulous jaw with gingival movement simulation: an in vitro study. J Prosthet Dent 2022; 128 (03) 468-478
- 54 Kim KR, Seo KY, Kim S. Conventional open-tray impression versus intraoral digital scan for implant-level complete-arch impression. J Prosthet Dent 2019; 122 (06) 543-549
- 55 Kim JE, Amelya A, Shin Y, Shim JS. Accuracy of intraoral digital impressions using an artificial landmark. J Prosthet Dent 2017; 117 (06) 755-761
- 56 Iturrate M, Eguiraun H, Etxaniz O, Solaberrieta E. Accuracy analysis of complete-arch digital scans in edentulous arches when using an auxiliary geometric device. J Prosthet Dent 2019; 121 (03) 447-454






