Optimization of Image Quality in Pelvis Lymphoscintigraphy SPECT/CT Using Discovery NM/CT 670
Funding This study was part of an MSc thesis supported under grant number 4011153 by Mashhad University of Medical Sciences, Mashhad, Iran.
Abstract
Aim A lymphoscintigraphy is a crucial diagnostic tool for visualizing lymph nodes. This scan plays a significant role in determining the treatment and recovery plan for the patients. Due to the small lymph node size, obtaining high-quality images is important to prevent inaccurate results. We aimed to identify the most effective method for enhancing image quality through postprocessing techniques and altering the image reconstruction process.
Methods Two data sets were utilized in this study. First, National Electrical Manufacturers Association body phantom was filled with [99mTc]Tc-pertechnetate and prepared with and without any activity in the background of the body. Second, the images of 50 patients who underwent single-photon emission computed tomography/computed tomography imaging received [99mTc]Tc-phytate were collected. Discovery 670 GE gamma camera was used for imaging. Preprocessing of all images was performed by Xeleris and 3DSlicer 5.2.2 software was used for quantification. The effect of image reconstruction parameters such as resolution recovery (RR) algorithm, iteration, subsets, cutoff, and power in Butterworth filter, and full width at half maximum (FWHM) of Gaussian filter was assessed. The image quality index was determined based on contrast-to-noise ratio (CNR), contrast, and coefficient of variation.
Results The utilization of the RR algorithm showed notable improvements equal to 74, 35, and 38% of CNR, contrast, and noise reduction, respectively. Significant differences were observed in subiteration of 40 to 112 (p-value < 0.05). The alteration of effective parameters in both smoothing filters yielded statistically significant results, leading to enhanced detectability, reduced noise, and improved contrast simultaneously. Optimum results in terms of noise reduction and CNR were achieved with subiteration (i × s) 4 × 12 using a Gaussian filter with FWHM of 4 or Butterworth filter with power of 10 and cutoff of 1. The highest contrast was observed at subiteration 40 using the Butterworth filter with cutoff of 0.5 and power of 5 or Gaussian filter with 2 mm FWHM. Qualitative analysis by two nuclear medicine specialists validated the quantified image quality.
Conclusion The reconstruction setting involving subiteration 48 with the Butterworth filter using cutoff of 1 and power of 10 or 4 mm FWHM of Gaussian filter produced the highest quality images.
Introduction
Currently, around two-thirds of cancer cases are situated in the lower abdominal and pelvic regions. Cancer cells that metastasize often travel through the bloodstream and tend to settle in lymph nodes and channels.[1] [2] Understanding the extent of involvement and pathways of lymph nodes is crucial for guiding surgical interventions. Accurate diagnosis through imaging and surgery plays a pivotal role in effective cancer treatment.[3] [4] With the development of new tools and techniques, imaging procedures have been accompanied by significant progress, leading to better cancer diagnosis and prevention of additional surgery.[5] These methods are used to detect tumors and other abnormalities to diagnose the presence of disease and determine the effectiveness of treatment. Cancer that starts in the cervix can spread to the lymph nodes in the pelvis.[6] The most common method for examining lymph nodes before surgery is to use a lymphoscintigraphy scan.[7] Because of the small size of lymph nodes, the diagnosis of them in this scan is along with the false positive or false negative.[8] [9] This scan can utilize single-photon emission computed tomography/computed tomography (SPECT/CT) to evaluate lymph nodes before surgery,[10] providing information on the level of involvement and the anatomical location of these. While this method generally has a low false negative rate, errors can still exist due to factors like attenuation, scattering, and noise, which can degrade image quality.[11] [12] Lymphoscintigraphy may not show any lymphatic drainage in many patients. Although the false negative rate in the SPECT/CT technique is low, its probability is not zero.[13] [14] Therefore, the images must have the highest quality to prevent errors and misdiagnosis. Tissue attenuation, scattering, and noise cause the quality of images to decrease.[15] [16] [17] Reconstruction methods affect the quality of images. The number of iteration, subset, and type of filter could alter the image quality.[18] Using the more subiterations and post-smoothing filter like Gaussian and Butterworth make the images smoother with lower detectability of small objects. Also, attenuation correction, scatter correction, and resolution recovery (RR) are often performed during image reconstruction to achieve higher quality.
The present study aimed to find the image reconstruction protocols that reduce the false negative or positive diagnosis in the SPECT/CT lymphoscintigraphy scan. All image reconstruction parameters were optimized using quantitative analysis and visual assessment.
Methods
Data Collection
The National Electrical Manufacturers Association image quality phantom containing six fillable spheres with diameters of 10, 13, 17, 22, 28, and 37 mm was used in the phantom study part of our research.[19] The spheres were filled with a homogenous [99mTc]Tc-pertechnetate solution at an activity concentration of 70 kBq/mL. Phantom imaging was acquired in two conditions, first, when the background of the body was air and the second when the background had been filled with about 7 KBq/mL of [99mTc]Tc-pertechnetate solution to obtain sphere-to-background ratio 10:1.
Fifty female patients, between 27 and 72 years, an average of 56 years, who underwent lymphoscintigraphy and SPECT/CT of pelvic, were selected. Seventy-five percent of them were postmenopausal. All patients suffered from malignancies affecting the uterus, ovaries, and cervix. Pelvic lymph node involvement was an important factor in choosing the patients and some parameters like age, weight, and stage of disease were neglected. The images of patients were retrospectively collected and retrieved from the hospital's picture archiving and communication system.
Image Acquisition
Data acquisition of the phantom and patients was performed using the dual-head GE Healthcare Discovery NM/CT 670 CZT system with low-energy high-resolution collimators. Imaging parameters of CT images were 80 kV tube voltage, 50 mA tube current, 2.5 mm slice thickness, and 256 × 256 matrix size. SPECT imaging was acquired in 120 projections at 25 seconds per projection, step-and-shoot mode, and 128 × 128 matrix size.
Image Reconstruction
Vendor-provided Xeleris software was used for image reconstruction. According to Xeleris options, all raw data of SPECT images were reconstructed using the ordered subsets expectation-maximization (OSEM) algorithm with a combination of three iterations 2, 4, and 8, and subsets 10, 12, and 14. Gaussian filter with three different full width at half maximum (FWHM) 2, 4, and 6 mm and Butterworth filter with 0.5, 1, and 1.5 cutoff and power of 5, 10, 15, and 20 were separately applied on the images as the post-smoothing filters. Attenuation and scatter corrections were performed in all SPECT images. Additionally, all SPECT images were reconstructed once more applying the RR to assess the RR algorithm. In total, 270 SPECT images were obtained by applying different reconstruction settings. Quantitative and qualitative assessment of phantom images was carried out for the sphere with the 13-mm diameter because of the similarity of the size of lymphatic nodules in clinical situation. The reconstruction settings employed in our study are demonstrated in [Fig. 1].


Image Preprocessing
Reconstructed images should be preprocessed before quantitative analysis. We created a three-dimensional (3D) model for the sphere's phantom segmentation using IGT module in 3DSlicer 5.2.2. The segment editor tool of 3DSlicer 5.2.2 was utilized for lymph node delineation in SPECT images. The volume of interest (VOI) of spheres and lymph nodes represents the cumulative activity in the selected region. The average and maximum count of each VOI, average count of background, and standard deviation of background count were measured using the quantification module and segment statistic tool in 3DSlicer 5.2.2.
Quantitative and Qualitative Analysis
Three important metrics were calculated for evaluation of image quality in the sphere with 13 mm diameter of phantom and involved lymph node in patients; contrast-to-noise ratio (CNR), contrast, and noise (coefficient of variation [CV%]). CNR was calculated using [Eq. (1)]:


where Cmean(sphere/nodule) is the average count in the region of interest (ROI) for the sphere in the phantom and nodule in the patient, Cmean(background) is the average count of the spherical VOI in the background of the body phantom or the tissue surrounding the lymph node, and SD is the standard deviation of the background VOI. Contrast was computed with [Eq. (2)]:


where Cmax(sphere/nodule) is the maximum count in ROI, and Cmean(background) is the average count of the spherical VOI in the background of the body phantom or the tissue surrounding the lymph node.
CV% as a statistical noise of image was calculated as follows [Eq. (3)]:


Two nuclear medicine specialists and one expert medical physicist visually evaluated the overall image quality and detectability of the lymphatic nodule in the pelvis and the sphere in the phantom. All images achieved from different reconstructions were qualitatively assessed and the best one was introduced.
Statistical Analysis
All data were statistically analyzed using GraphPad Prism 10 software. CNR, contrast, and CV% obtained from all subiterations were compared using paired t-test analysis. After selecting the subiterations that showed significantly different from others, various parameters of the Butterworth (cutoff and power) and Gaussian (FWHM) filters were compared using the paired t-test in selected subiterations. The significance level of the p-value was set to less than 0.05 for all comparisons. The percentage difference of quantitative metrics was calculated to compare the role of the RR algorithm in image quality using [Eq. (4)]:


Results
Phantom
Quantitative Analysis
It is not feasible to calculate CNR and CV% when the body phantom background is devoid of any material, so only contrast can be calculated. Statistical analysis revealed a significant difference between subiterations 4 × 10, 4 × 12, and 4 × 14 in the calculation of CNR, contrast, and noise. The impact of the RR algorithm on the SPECT images was evaluated by the percentage difference of each index under two conditions; the presence of [99mTc]Tc-pertechnetate in the background and no radioactive material in the background of the body phantom.
[Table 1] represents the image reconstruction sets most affected by the RR algorithm. Also, the percentage difference of calculated metrics for these reconstruction settings is demonstrated. RR algorithm improved CNR and contrast by 74% and more than 30%, respectively. Additionally, image noise considerably decreased by approximately 35% using the RR algorithm.
Abbreviations: CNR, contrast-to-noise ratio; CV%, coefficient of variation; FWHM, full width at half maximum; RR resolution recovery; SPECT, single-photon emission computed tomography.
Based on the results of scientific studies in this area, increasing CNR along with reducing contrast.[8] [20] Comparisons of both metrics in the present study as a function of the reconstruction settings are shown in [Fig. 2]. Regardless of the type of post-smoothing filter, subiterations 40 to 56 (4 × 10, 4 × 12, and 4 × 14 in our study) can produce high-quality SPECT images. As shown in [Fig. 2A–C], while the Butterworth filter is applied, the appropriate parameters to achieve the best overall image quality are recommended to be 1 to 1.5 cutoff and 10 to 15 power. Also, according to [Fig. 2D], a Gaussian filter with 4 mm FWHM seems suitable based on the tradeoff between CNR and contrast.


Qualitative Analysis
Visual assessment was performed by a medical physicist and two nuclear medicine specialists. According to [Fig. 3A], the most appropriate reconstruction setting was subiteration 4 × 12 with the Butterworth filter using cutoff of 0.5 or 1 and power of 5 and 10 in the presence of [99mTc]Tc-pertechnetate in the phantom background. Also, 2 mm FWHM of Gaussian filter can be used for smoothing. As shown in [Fig. 3B], the best reconstruction set was subiteration 4 × 12 with Butterworth filter in the cutoff of 0.5 and 10, 15, and 20 of the power or cutoff of 1 and power of 5 in the phantom without any radioactive material in the background. However, the Gaussian filter with 4 mm FWHM was suitable too.


Patients
Quantitative Analysis
The effect of the RR algorithm on the SPECT images was assessed by calculating the percentage difference of CNR, contrast, and CV%. [Table 2] represents the image reconstruction settings most influenced by the RR algorithm. RR algorithm enhanced overall image quality resulting in a 42.2% reduction of noise, 74% CNR improvement, and 37.2% contrast enhancement.
Abbreviations: CNR, contrast-to-noise ratio; CV%, coefficient of variation; RR resolution recovery; SPECT, single-photon emission computed tomography.
Statistical analysis showed a significant difference between four subiterations 4 × 10, 4 × 12, 8 × 10, and 8 × 14, and the other subiterations with p-values 0.0019, 0.022, 0.069, and ≤ 0.001, respectively. No significant differences were observed between the four mentioned subiterations.
As [Fig. 4] shows, more subiterations create more CNR values. The highest CNR was obtained in power of 20 and cutoff of 1.5 by applying the Butterworth filter. Increasing the power and cutoff raises the CNR value in all subiterations. The difference between the lowest and the highest CNR in our investigated reconstruction sets was approximately 37%.


According to [Fig. 5], more subiterations lead to less contrast values. The lowest contrast was obtained in the power of 20 and cutoff of 1.5 by applying the Butterworth filter. Increasing the power and cutoff could decrease the contrast value in all subiterations, albeit the role of cutoff is more considerable than power. The difference between the lowest and the highest contrast in our examined reconstruction sets was approximately 55%.


[Fig. 6] illustrates the image noise reduced using more subiterations. The lowest CV% was obtained in subiteration 8 × 14 by applying the Butterworth filter with the power of 10 and cutoff of 1.5. The noise of the image could be reduced by approximately 51% if the appropriate cutoff has been selected.


As shown in [Fig. 7], three quantitative metrics were evaluated in four subiterations with 2, 4, and 6 mm FWHM of the Gaussian filter. Increasing subiterations does not noticeably make a difference in the image noise, CNR, and contrast values, while FWHM of Gaussian filter impressively affects CNR and contrast. On average, CNR increases by 65%, contrast decreases by 36%, and noise reduces by 15% if 6 mm FWHM is used instead of 2 mm FWHM.


Qualitative Analysis
Two nuclear medicine specialists clinically assessed images of all patients obtained from various reconstruction methods. The best image reconstruction protocol between four selected reconstruction sets was used for qualitative assessment. The transaxial images of one patient in different parameters of Butterworth or Gaussian filters in subiteration 4 × 12 are demonstrated in [Fig. 8]. Preferred parameters by physicians in using the Butterworth filter tended to have less cutoff and more power values.


Discussion
Generally, reconstruction sets using the high subiteration numbers, Gaussian filter with more FWHM and Butterworth filter with more cutoff and power, make images smoother.[18] Our findings show that CNR increased and image noise decreased by increasing the number of subiterations using smoother filters, while contrast is reduced in this situation. Also, overall image quality was improved using RR algorithm. RR algorithm could increase CNR and decrease contrast and noise almost 74, 40, and 35%, respectively ([Table 1]). The effect of RR on reducing the imaging time and assessing image quality in cardiac acquisition was evaluated by Ismail and Mansor. He showed that RR gives better quantitative evaluation and results in a good resolution in the myocardial perfusion images.[21] Another study in 2021 demonstrated that using attenuation and scatter correction and RR algorithm in the OSEM reconstruction method could noticeably enhance spatial resolution in SPECT/CT of a Jaszczak phantom.[22] A retrospective study in 2022 showed RR improves spatial resolution and CNR in the bone SPECT images.[23]
According to the inverse relationship between CNR and contrast, the higher the CNR the lower the contrast. This results in a loss of detectability especially in small objects. Therefore, the tradeoff between CNR and contrast in the phantom images filled with [99mTc]Tc-pertechnetate in the background that demonstrated the most appropriate reconstruction setting is sub-iteration 4 × 12 using Butterworth filter with cutoff of 1 and power of 10 or Gaussian filter with 2 or 4 mm FWHM ([Fig. 2]). Based on our quantitative and qualitative analysis, it seems that the appropriate image reconstruction setting be subiteration 4 × 12 using Butterworth filter with cutoff of 0.5 and power of 5/10 or cutoff of 1 and power of 5/10 or 2 mm FWHM of Gaussian filter in the phantom with radioactive material in the background. Also, the appropriate image reconstruction setting in phantom without any radioactivity in the background is subiteration 4 × 12 using the Butterworth filter with cutoff of 0.5 and power of 10/15/20 or cutoff of 1 and power of 5 or 4 mm FWHM of Gaussian filter. As many research papers demonstrate, increasing the number of iterations, subsets, FWHM of Gaussian filter, and the more cutoff and power of Butterworth filter made higher CNR but lower contrast and noise.[24] [25] [26] Increasing the iteration numbers, cutoff frequency in the Butterworth filter, and reducing FWHM of the Gaussian filter leads to enhanced spatial resolution in the SPECT of a Jaszczak phantom.[22] Fukami et al optimized the number of iterations in SPECT images of thoracic spine phantom. They evaluated contrast, noise, and standardized uptake value as a quantitative analysis. Their results showed that increasing iteration numbers caused an increase in CV% and contrast. However, contrast almost converged uniformly in subiterations of more than 50.[27] The difference between this study and our work was the type of phantom used and the data analysis methods.
The proposed image reconstruction parameters based on the evaluation of patient images were subiteration 4 × 12, 4 mm FWHM of Gaussian filter, and Butterworth filter with cutoff of 1 and power of 10/15. Our finding is in agreement with the results of a study by Lanfranchi et al that recommended the optimized image reconstruction setting for SPECT of brain Dat-scan. They applied the Butterworth with a 0.96 cutoff.[9] Another study by Lyra and Ploussi on SPECT of the liver showed that the Butterworth filter with the cutoff of 0.1 to 0.5 could help to achieve the high detectability of small lesions.[28] We found the cutoff of 0.5 in the Butterworth filter as the optimized parameter in the reconstruction protocol, similar to the Lyra and Ploussi research. Two studies utilized the SPECT/CT images of cardiac[10] and bone scan.[14] In the myocardial perfusion scan, Gaussian filter with 12 to 14 mm FWHM was applied to smooth the images.[10] Optimized filters for the assessment of bony metastatic lesions were proposed 10 to 13 mm FWHM of Gaussian filter by Alqahtani et al.[14] These results were in contrast with our findings because of the object sizes. Extra smoothing of the images could lead to misdiagnosis and false negatives when the lesion size is small, like lymph nodes. While in large lesions such as obvious bone metastasis or cardiac muscle, more smoothing makes CNR better, reduces noise, and improves image quality without any false negativity. According to our results, 2 and 4 mm FWHM of Gaussian filter were recommended in lymphoscintigraphy SPECT/CT of pelvic. Choosing the appropriate reconstruction parameters strongly depends on the size of the lesions. Overall image quality does not degrade using the smoother filter in evaluating large objects, while it is influenced on detectability and assessment of small lesions.
Although many studies about the optimization of image reconstruction settings have been performed, none of them simultaneously assessed all parameters. We evaluated iterations, subsets, two different smoothing filters, and the RR algorithm to find the optimized reconstruction sets for small lesions. Nonetheless, our study had some limitations. First, we studied the optimization of reconstruction sets only for patients with lymph nodes involved in pelvic and these results cannot be generalized to other cases with different sizes of lesions. Second, we optimized the reconstruction parameter options that exist in one make and model of gamma camera scanner (GE Discovery 670 CZT), while it can be evaluated for other gamma cameras as a multicenter study similar to harmonization in positron emission tomography/CT scanner.[29] [30]
Conclusion
According to our findings, subiteration 4 × 12 proposed with the Butterworth filter using a power of 10 and a cutoff of 1 or 4 mm FWHM of the Gaussian filter. Furthermore, utilizing the RR algorithm in the Discovery NM/CT 670 GE gamma camera is recommended to achieve the highest image quality in lymphoscintigraphy scans.
Conflict of Interest
None declared.
Acknowledgments
We are grateful to the nuclear medicine department of Ghaem Hospital in Mashhad.
-
References
- 1 Zhang X, Zhao M, Wen L. et al. Sequential SPECT and NIR-II imaging of tumor and sentinel lymph node metastasis for diagnosis and image-guided surgery. Biomater Sci 2021; 9 (08) 3069-3075
- 2 Gien LT, Covens A. Lymph node assessment in cervical cancer: prognostic and therapeutic implications. J Surg Oncol 2009; 99 (04) 242-247
- 3 Lengelé B, Scalliet P. Anatomical bases for the radiological delineation of lymph node areas. Part III: pelvis and lower limbs. Radiother Oncol 2009; 92 (01) 22-33
- 4 Boughey JC, Suman VJ, Mittendorf EA. et al; Alliance for Clinical Trials in Oncology. Sentinel lymph node surgery after neoadjuvant chemotherapy in patients with node-positive breast cancer: the ACOSOG Z1071 (Alliance) clinical trial. JAMA 2013; 310 (14) 1455-1461
- 5 Szuba A, Shin WS, Strauss HW, Rockson S. The third circulation: radionuclide lymphoscintigraphy in the evaluation of lymphedema. J Nucl Med 2003; 44 (01) 43-57
- 6 Kim SH, Kim SC, Choi BI, Han MC. Uterine cervical carcinoma: evaluation of pelvic lymph node metastasis with MR imaging. Radiology 1994; 190 (03) 807-811
- 7 Giammarile F, Vidal-Sicart S, Paez D. et al. Sentinel lymph node methods in breast cancer. Semin Nucl Med 2022; 52 (05) 551-560
- 8 Rodriguez-Molares A, Rindal OMH, D'hooge J. et al. The generalized contrast-to-noise ratio: a formal definition for lesion detectability. IEEE Trans Ultrason Ferroelectr Freq Control 2020; 67 (04) 745-759
- 9 Lanfranchi F, Arnaldi D, Miceli A. et al. Different z-score cut-offs for striatal binding ratio (SBR) of DaT SPECT are needed to support the diagnosis of Parkinson's disease (PD) and dementia with Lewy bodies (DLB). Eur J Nucl Med Mol Imaging 2023; 50 (04) 1090-1102
- 10 Okuda K, Nakajima K, Yamada M. et al. Optimization of iterative reconstruction parameters with attenuation correction, scatter correction and resolution recovery in myocardial perfusion SPECT/CT. Ann Nucl Med 2014; 28 (01) 60-68
- 11 Kogler AK, Polemi AM, Nair S. et al. Evaluation of camera-based freehand SPECT in preoperative sentinel lymph node mapping for melanoma patients. EJNMMI Res 2020; 10 (01) 139
- 12 Vermeeren L, Valdés Olmos RA, Meinhardt W. et al. Value of SPECT/CT for detection and anatomic localization of sentinel lymph nodes before laparoscopic sentinel node lymphadenectomy in prostate carcinoma. J Nucl Med 2009; 50 (06) 865-870
- 13 Coffey JP, Hill JC. Breast sentinel node imaging with low-dose SPECT/CT. Nucl Med Commun 2010; 31 (02) 107-111
- 14 Alqahtani MM, Willowson KP, Constable C, Fulton R, Kench PL. Optimization of 99m Tc whole-body SPECT/CT image quality: a phantom study. J Appl Clin Med Phys 2022; 23 (04) e13528
- 15 Patton JA, Turkington TG. SPECT/CT physical principles and attenuation correction. J Nucl Med Technol 2008; 36 (01) 1-10
- 16 Xiang H, Lim H, Fessler JA, Dewaraja YK. A deep neural network for fast and accurate scatter estimation in quantitative SPECT/CT under challenging scatter conditions. Eur J Nucl Med Mol Imaging 2020; 47 (13) 2956-2967
- 17 Hulme KW, Kappadath SC. Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging. Med Phys 2014; 41 (04) 042502
- 18 Vosoughi H, Emami F, Momennezhad M, Geramifar P, Rahmim A. Standardization and optimization of Siemens Biograph TruePoint PET/CT acquisition and reconstruction parameters: Simultaneous qualitative and quantitative assessments. . Iran J Nucl Med 2024
- 19 Vosoughi H, Hajizadeh M, Emami F, Momennezhad M, Geramifar P. PET NEMA IQ Phantom dataset: image reconstruction settings for quantitative PET imaging. Data Brief 2021; 37: 107231
- 20 Dieckens D, Lavalaye J, Romijn L, Habraken J. Contrast-noise-ratio (CNR) analysis and optimisation of breast-specific gamma imaging (BSGI) acquisition protocols. EJNMMI Res 2013; 3 (01) 21
- 21 Ismail FS, Mansor S. Impact of resolution recovery in quantitative 99mTc SPECT/CT cardiac phantom studies. J Med Imaging Radiat Sci 2019; 50 (03) 449-453
- 22 Wang M, Dong S, Zhang R. et al. Impact of reconstruction parameters on spatial resolution and its comparison between cadmium-zinc-telluride SPECT/CT and conventional SPECT/CT. Nucl Med Commun 2022; 43 (01) 8-16
- 23 Salkica N, Begic A, Zubovic S. et al. Impact of reduced acquisition time on bone single-photon emission computed tomography images in oncology patients. Acta Inform Med 2022; 30 (01) 36-40
- 24 Miyaji N, Miwa K, Tokiwa A. et al. Phantom and clinical evaluation of bone SPECT/CT image reconstruction with xSPECT algorithm. EJNMMI Res 2020; 10 (01) 71
- 25 Goenka AH, Herts BR, Dong F. et al. Image noise, CNR, and detectability of low-contrast, low-attenuation liver lesions in a phantom: effects of radiation exposure, phantom size, integrated circuit detector, and iterative reconstruction. Radiology 2016; 280 (02) 475-482
- 26 Işikci NI. Comparison between planar, SPECT and SPECT/computed tomography systems in 99mTc- nanocolloid sentinel lymph node imaging: phantom design and clinical investigations. Nucl Med Commun 2019; 40 (08) 786-791
- 27 Fukami M, Matsutomo N, Yamamoto T. Optimization of number of iterations as a reconstruction parameter in bone SPECT imaging using a novel thoracic spine phantom. J Nucl Med Technol 2021; 49 (02) 143-149
- 28 Lyra M, Ploussi A. Filtering in SPECT image reconstruction. Int J Biomed Imaging 2011; 2011: 693795
- 29 Vosoughi H, Momennezhad M, Emami F, Hajizadeh M, Rahmim A, Geramifar P. Multicenter quantitative 18F-fluorodeoxyglucose positron emission tomography performance harmonization: use of hottest voxels towards more robust quantification. Quant Imaging Med Surg 2023; 13 (04) 2218-2233
- 30 Aide N, Lasnon C, Veit-Haibach P, Sera T, Sattler B, Boellaard R. EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies. Eur J Nucl Med Mol Imaging 2017; 44 (Suppl. 01) 17-31
Address for correspondence
Publikationsverlauf
Artikel online veröffentlicht:
26. September 2024
© 2024. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Zhang X, Zhao M, Wen L. et al. Sequential SPECT and NIR-II imaging of tumor and sentinel lymph node metastasis for diagnosis and image-guided surgery. Biomater Sci 2021; 9 (08) 3069-3075
- 2 Gien LT, Covens A. Lymph node assessment in cervical cancer: prognostic and therapeutic implications. J Surg Oncol 2009; 99 (04) 242-247
- 3 Lengelé B, Scalliet P. Anatomical bases for the radiological delineation of lymph node areas. Part III: pelvis and lower limbs. Radiother Oncol 2009; 92 (01) 22-33
- 4 Boughey JC, Suman VJ, Mittendorf EA. et al; Alliance for Clinical Trials in Oncology. Sentinel lymph node surgery after neoadjuvant chemotherapy in patients with node-positive breast cancer: the ACOSOG Z1071 (Alliance) clinical trial. JAMA 2013; 310 (14) 1455-1461
- 5 Szuba A, Shin WS, Strauss HW, Rockson S. The third circulation: radionuclide lymphoscintigraphy in the evaluation of lymphedema. J Nucl Med 2003; 44 (01) 43-57
- 6 Kim SH, Kim SC, Choi BI, Han MC. Uterine cervical carcinoma: evaluation of pelvic lymph node metastasis with MR imaging. Radiology 1994; 190 (03) 807-811
- 7 Giammarile F, Vidal-Sicart S, Paez D. et al. Sentinel lymph node methods in breast cancer. Semin Nucl Med 2022; 52 (05) 551-560
- 8 Rodriguez-Molares A, Rindal OMH, D'hooge J. et al. The generalized contrast-to-noise ratio: a formal definition for lesion detectability. IEEE Trans Ultrason Ferroelectr Freq Control 2020; 67 (04) 745-759
- 9 Lanfranchi F, Arnaldi D, Miceli A. et al. Different z-score cut-offs for striatal binding ratio (SBR) of DaT SPECT are needed to support the diagnosis of Parkinson's disease (PD) and dementia with Lewy bodies (DLB). Eur J Nucl Med Mol Imaging 2023; 50 (04) 1090-1102
- 10 Okuda K, Nakajima K, Yamada M. et al. Optimization of iterative reconstruction parameters with attenuation correction, scatter correction and resolution recovery in myocardial perfusion SPECT/CT. Ann Nucl Med 2014; 28 (01) 60-68
- 11 Kogler AK, Polemi AM, Nair S. et al. Evaluation of camera-based freehand SPECT in preoperative sentinel lymph node mapping for melanoma patients. EJNMMI Res 2020; 10 (01) 139
- 12 Vermeeren L, Valdés Olmos RA, Meinhardt W. et al. Value of SPECT/CT for detection and anatomic localization of sentinel lymph nodes before laparoscopic sentinel node lymphadenectomy in prostate carcinoma. J Nucl Med 2009; 50 (06) 865-870
- 13 Coffey JP, Hill JC. Breast sentinel node imaging with low-dose SPECT/CT. Nucl Med Commun 2010; 31 (02) 107-111
- 14 Alqahtani MM, Willowson KP, Constable C, Fulton R, Kench PL. Optimization of 99m Tc whole-body SPECT/CT image quality: a phantom study. J Appl Clin Med Phys 2022; 23 (04) e13528
- 15 Patton JA, Turkington TG. SPECT/CT physical principles and attenuation correction. J Nucl Med Technol 2008; 36 (01) 1-10
- 16 Xiang H, Lim H, Fessler JA, Dewaraja YK. A deep neural network for fast and accurate scatter estimation in quantitative SPECT/CT under challenging scatter conditions. Eur J Nucl Med Mol Imaging 2020; 47 (13) 2956-2967
- 17 Hulme KW, Kappadath SC. Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging. Med Phys 2014; 41 (04) 042502
- 18 Vosoughi H, Emami F, Momennezhad M, Geramifar P, Rahmim A. Standardization and optimization of Siemens Biograph TruePoint PET/CT acquisition and reconstruction parameters: Simultaneous qualitative and quantitative assessments. . Iran J Nucl Med 2024
- 19 Vosoughi H, Hajizadeh M, Emami F, Momennezhad M, Geramifar P. PET NEMA IQ Phantom dataset: image reconstruction settings for quantitative PET imaging. Data Brief 2021; 37: 107231
- 20 Dieckens D, Lavalaye J, Romijn L, Habraken J. Contrast-noise-ratio (CNR) analysis and optimisation of breast-specific gamma imaging (BSGI) acquisition protocols. EJNMMI Res 2013; 3 (01) 21
- 21 Ismail FS, Mansor S. Impact of resolution recovery in quantitative 99mTc SPECT/CT cardiac phantom studies. J Med Imaging Radiat Sci 2019; 50 (03) 449-453
- 22 Wang M, Dong S, Zhang R. et al. Impact of reconstruction parameters on spatial resolution and its comparison between cadmium-zinc-telluride SPECT/CT and conventional SPECT/CT. Nucl Med Commun 2022; 43 (01) 8-16
- 23 Salkica N, Begic A, Zubovic S. et al. Impact of reduced acquisition time on bone single-photon emission computed tomography images in oncology patients. Acta Inform Med 2022; 30 (01) 36-40
- 24 Miyaji N, Miwa K, Tokiwa A. et al. Phantom and clinical evaluation of bone SPECT/CT image reconstruction with xSPECT algorithm. EJNMMI Res 2020; 10 (01) 71
- 25 Goenka AH, Herts BR, Dong F. et al. Image noise, CNR, and detectability of low-contrast, low-attenuation liver lesions in a phantom: effects of radiation exposure, phantom size, integrated circuit detector, and iterative reconstruction. Radiology 2016; 280 (02) 475-482
- 26 Işikci NI. Comparison between planar, SPECT and SPECT/computed tomography systems in 99mTc- nanocolloid sentinel lymph node imaging: phantom design and clinical investigations. Nucl Med Commun 2019; 40 (08) 786-791
- 27 Fukami M, Matsutomo N, Yamamoto T. Optimization of number of iterations as a reconstruction parameter in bone SPECT imaging using a novel thoracic spine phantom. J Nucl Med Technol 2021; 49 (02) 143-149
- 28 Lyra M, Ploussi A. Filtering in SPECT image reconstruction. Int J Biomed Imaging 2011; 2011: 693795
- 29 Vosoughi H, Momennezhad M, Emami F, Hajizadeh M, Rahmim A, Geramifar P. Multicenter quantitative 18F-fluorodeoxyglucose positron emission tomography performance harmonization: use of hottest voxels towards more robust quantification. Quant Imaging Med Surg 2023; 13 (04) 2218-2233
- 30 Aide N, Lasnon C, Veit-Haibach P, Sera T, Sattler B, Boellaard R. EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies. Eur J Nucl Med Mol Imaging 2017; 44 (Suppl. 01) 17-31