Keywords bone conduction - minimally invasive - bone drilling - craniofacial abnormalities
- bone-anchored prosthesis - children
Introduction
The percutaneous bone-anchored hearing system (BAHS) is a valuable treatment option
for patients with conductive hearing loss, mixed hearing loss, or single-sided deafness.
It is comprised of a titanium implant inserted into the calvarium and integrated into
the skull from the growth of bone tissue over its titanium surface, an abutment fixed
to the implant through the skin, and an external sound processor attached to the abutment,
thereby allowing direct bone conduction, that is, sound energy transmission from the
skull vibrations to the cochlea, which ultimately results in wave propagation along
the basilar membrane and stimulation of the cochlear nerve.[1 ]
The BAHS was first introduced in 1977,[2 ] and the U.S. Food and Drug Administration (FDA) approved the surgery in children
with ages over 5 years old in 1999.[3 ] The surgical technique was initially performed in two-stages in both adults and
children, intended for implant installation and abutment-implant attachment associated
to the soft-tissue reduction, both performed separately. Technological advances in
the BAHS design, such as wider diameter implants and different abutment length options,
enabled the performance of installation mostly in one-stage, with soft-tissue preservation.
It reduced the postoperative complications, adverse skin reactions and implant failure,
and thereby becoming the gold standard in adult patients.[4 ] Aiming to reduce even more the invasiveness of the surgical technique, the first
minimally invasive punch-only surgical technique was introduced,[5 ] allowing to install the BAHS implant with the use of a drill guide through a hole
punch incision. The use of this surgical technique showed additional benefits, such
as shorter operative times, faster healing, and less soft-tissue reaction compared
with the standard linear incision technique in adults.[6 ] Although it is one of the least invasive procedures available to install BAHS implants,
there is a lack of studies in the literature regarding its results among the pediatric
population. A recent systematic review[4 ] showed that new surgical techniques for percutaneous BAHS implant installation are
promising for pediatric patients and claimed for new studies on the topic. It concluded
that one-stage surgery and soft-tissue preservation do not provide higher implant
loss or adverse skin reactions to pediatric patients and then, new techniques could
additionally benefit this population.
In 2021, French et al.[7 ] performed a retrospective study on minimally invasive Ponto surgery (MIPS) in 14
pediatric patients and found similar results to those reported for adults. To the
best of our knowledge, this is the only study involving minimally invasive punch-only
technique to install a BAHS implant specifically in pediatric patients, but anesthetics
procedures were not investigated. Sardiwalla et al.[8 ] reported that MIPS under local anesthesia may be performed as an outpatient procedure,
outside of the main operating room, leading to a further reduction in surgical time,
staffing, and running costs, and allowing a higher patient throughput. Then, the use
of MIPS associated to local anesthesia could also reduce the exposure to general anesthesia
and allow earlier (re)habilitation process in pediatric patients.
Based on this, the aim of the present study was to investigate the hearing performance
and soft-tissue outcomes of the MIPS procedure under local anesthesia to install the
BAHS implant in children with unilateral craniofacial malformation (UCM).
Methods
Study Design
The study used a retrospective cohort design. The inclusion criteria comprised subjects
aged up to 17 years old, with UCM, who underwent a BAHS surgery with the one-stage
MIPS and local anesthesia (2% lidocaine with 1:200,000 epinephrine). The exclusion
criteria comprised subjects who showed up to 4 months of postoperative follow-up,
since it was considered the minimal experience period with the bone conduction stimulation
needed to perform speech perception tests. The postoperative follow-up period comprised
the time between the first abutment fitting and the evaluation.
Surgical procedure, implant and abutment lengths, surgical time, and intraoperative
complications (bleeding and implant instability) were investigated. Postoperative
complications such as implant loss and revision surgery, as well as soft-tissue outcomes
were also investigated during the entire postoperative follow-up period for each subject.
Soft-tissue outcomes were scored according to the Holger scale,[9 ] and it was determined that a Holger score ≥ 2 was representative of adverse skin
reactions.[10 ]
All subjects were fitted with the Ponto 3 SuperPower sound processor (Oticon Medical,
Smørum, Denmark) and programmed according to the manufacturer's instructions, using
the Genie Medical 2016.1 fitting software (Oticon Medical). The gain was calculated
by the software according to the desired sensation level-bone conduction (DSL-BC)
prescription formula and, finetuned to the subject's preference when required.
The Sentences in Portuguese Test[11 ] was used to assess the subject's speech recognition in quiet and noise, in unaided
and aided conditions, that is, without and with the use of the BAHS, respectively.
The test was composed of 60 sentences distributed into 6 lists of 10 phonetically
balanced sentences with up to 7 words with lexical/semantic or grammatical meaning,
and with similar phonetic content.
The test in quiet was performed using one loudspeaker positioned 1 meter from the
patient, in front (0° azimuth, [Fig. 1A ]). The sentences were first presented at 65 dB SPL, and the presentation level varied
according to the subject's response, using an ascending-descending technique.[12 ] The results were expressed in sentence recognition threshold in quiet (SRTQ), which
corresponded to the presentation level (dB SPL) at which the subject correctly repeated
50% of the sentences. The test in noise was carried out using 2 loudspeakers, positioned
1 meter from the patient, in the S90N90 position: the sentences were presented in
the implanted ear, and the noise was presented in the contralateral ear ([Fig. 1B ]). In this condition, the noise was fixed at 65 dB, and the presentation level of
the sentences varied according to the subject's response. The results were expressed
in sentence recognition in noise (SRTN), which corresponded to the signal-to-noise
level (dB SNR) at which the subject correctly repeated 50% of the sentences. The SRTQ
and SRTN were then compared between unaided and aided (BAHS) conditions in all subjects.
Fig. 1 Positioning of the subjects and loudspeakers to the speech recognition test in quiet
(A ) and noise (B ) situations. Abbreviations: N, noise, “H, normal hearing; S, sentences. Note: The
black circle corresponded to the bone-anchored hearing system (BAHS) device.
The subjective listening effort of the subjects was also investigated in both the
quiet and noise situations. The subjects scored their perceptual listening effort
using a visual-analogue scale ([Fig. 2 ]). Scores from 0 to 2, from 3 to 7, and from 8 to 10 corresponded to the low, moderate,
and high subjective listening effort categories, respectively. The results were compared
between the unaided and aided (BAHS) conditions in all subjects.
Fig. 2 Visual analogue scale used to score the subjective listening effort. Scores 0–2,
3–7 and 8–10 corresponded to low, moderate, and high listening effort perceived by
the subjects, respectively.
All subjects filled the questionnaire Satisfaction with Amplification in Daily Life
(SADL)[13 ] to quantify their overall satisfaction with the device. The subjects who were not
capable of answering the questionnaire by themselves had assistance from their parents
to complete it. The subject's daily use of the device was objectively assessed using
the Genie Medical fitting software, version 2016.1 (Oticon Medical).
The Brazilian Ethics Committee approved this research under protocol number 4.217.145.
The legal guardians of all patients gave their written informed consent to participate
in this study.
Data Analysis
The Wilcoxon signed-rank test was used to compare the speech recognition thresholds
and subjective listening effort of the subjects, between the unaided and aided conditions.
A significant level of 5% was adopted.
Results
Nine subjects aged between 6.5 and 17.1 (median = 12) years met the inclusion criteria.
[Table 1 ] shows the subject demographics. It was possible to perform MIPS under local anesthesia
in 8 of 9 subjects (88.9%). For one subject (#S8), the surgical technique was converted
to a linear incision with tissue preservation (LITP) due to implant instability and
bleeding. Sedation was also administered to this subject. [Table 2 ] shows the surgical and postoperative soft-tissue outcomes of the subjects. The mean
postoperative follow-up period was of 11.4 ± 6.7 months. Two subjects (#S4 and #S8)
showed adverse skin reactions (Holgers score ≥ 2) during the entire postoperative
follow-up period, which were easily solved with local care.
Table 1
Subject demographics
Subject
Sex
Age at surgery (years)
Implanted ear
Type of HL (implanted ear)
Degree of HL (implanted ear)
Contralateral ear
1
M
12
R
Mixed
Profound
NH
2
F
6.5
R
Conductive
Severe
NH
3
M
7.1
L
Conductive
Moderate
NH
4
M
15
R
Conductive
Moderate-to-severe
NH
5
M
12.7
R
Mixed
Severe
NH
6
F
11.1
L
Conductive
Severe
NH
7
M
16.5
L
Conductive
Moderate-to-severe
NH
8
M
8.3
R
Conductive
Severe
NH
9
M
17.1
L
Conductive
Moderate
NH
Abbreviations: F, female; HL, hearing loss; L, left; M, male; NH, normal hearing; R, right.
Table 2
Surgical and postoperative soft-tissue outcomes
Subject
Surgery time (min)
Surgical technique
Anesthesia
Implant length (mm)
Abutment length (mm)
Postoperative follow-up period (m)
Holgers score
1
20
MIPS
L
4
9
21
0
2
20
MIPS
L
4
6
4
0
3
30
MIPS
L
4
9
15
1
4
40
MIPS
L
4
12
5
3*
5
30
MIPS
L
4
9
4
0
6
27
MIPS
L
4
12
15
1
7
20
MIPS
L
4
9
6
1
8
60
ILTP
L + S
3
9
18
2*
9
29
MIPS
L
4
9
15
1
Abbreviations: ILTP, linear incision with tissue preservation; L, local anesthesia; L + S, local
anesthesia + sedation; m, months; min, minutes; MIPS, minimally invasive Ponto Surgery;
mm, millimeters.
Note: *Adverse skin reaction (Holgers score ≥ 2).
One subject (#S8) showed implant loss ∼ 13 months after surgery due to trauma. For
the remaining subjects, implant extrusion was not observed, and revision surgery was
not required in the entire follow-up period.
[Fig. 3A,B ] show the comparison between SRTQ and SRTN mean scores obtained from subjects in
the unaided and aided (BAHS) conditions, respectively. The SRTQ and SRTN scores significantly
decreased with the use of BAHS, although 1 subject (#S5) showed no changes in the
SRTQ comparing both the unaided and aided conditions ([Table 3 ]). The mean decrease in the SRTQ and SRTN scores with the use of the BAHS were 1.7 dB
and −7.4 dB SNR, respectively.
Fig. 3 Comparison of the mean scores for speech recognition threshold in quiet (A ) and speech recognition threshold in noise (B ) obtained from subjects between unaided and aided (bone-anchored hearing system,
BAHS) conditions, respectively. The Wilcoxon signed-rank test showed that the thresholds
significantly decreased with the use of BAHS, p < 0.05, effect size (r) > 0.8.
Table 3
Comparison of the sentence recognition threshold in quiet and in noise between unaided
and aided (BAHS) conditions across subjects
SRTQ (dB SPL)
SRTN (dB SNR)
Subject
Speech: front (0°)
Quiet
Speech: impaired ear, Noise: NH (S90-N90)
Unaided
Ponto
Unaided
Ponto
1
21.6
21
-2
-10
2
31
27
4
-2
3
32
27.6
0
-6.66
4
39
37.6
2
-5.33
5
25
25
-3.33
-9.33
6
29
27.6
1.33
-5.33
7
29.6
29
2.66
-7.33
8
30.3
28.3
3.33
-4.66
9
31
30
6
-2
Abbreviations: dB SNR, decibel signal-to-noise ratio; dB SPL, decibel sound pressure levels; NH,
normal hearing; Ponto, with the use of the bone-anchored hearing system; SRTN, speech
recognition threshold in noise; SRTQ, speech recognition threshold in quiet; Unaided,
without the use of the bone-anchored hearing system.
The perceptual listening effort scored by the subjects was much higher in the speech
perception-in-noise task than in quiet. It significantly decreased with the use of
BAHS in both situations. The listening effort category perceived by the subjects decreased
from high to moderate with the use of BAHS in the speech-in-noise task ([Fig. 4 ]). The mean SADL global score of the subjects was of 6.0 ± 0.3, indicating that they
were very satisfied with the device. Still, the daily use of the BAHS by the subjects
ranged from 6 to 11.5 (mean: 8.2 ± 2.0) hours per day.
Fig. 4 Comparison of the perceptual listening effort scores of the subjects in the speech
tasks in quiet and noise, between unaided and aided (bone-anchored hearing system,
BAHS) conditions. The Wilcoxon signed-rank test showed a significant decrease of the
perceptual listening effort scores with the use of BAHS, p < 0.05, effect size (r) > 0.8.
Discussion
The purpose of this study was to investigate the hearing performance and soft tissue
outcomes of the MIPS under local anesthesia in children with unilateral craniofacial
malformation implanted with percutaneous BAHS. It was possible to install the implant
in 8 of 9 subjects using this procedure with no intraoperative complications, thereby
showing to be a viable option for use in the clinical routine.
Adverse skin reactions were minimal, and it occurred only in one subject (11.1%) during
the entire postoperative follow-up period, considering those who had their implant
installed with MIPS. The soft-tissue outcomes found in this study are consistent with
a retrospective study[7 ] on 14 pediatric patients who underwent a MIPS procedure to install the BAHS implant,
and also with those reported for adults with this procedure.[6 ] Based on this, we assume that the soft-tissue outcomes reported with the use of
MIPS in adults seem to be replicable children, that is, the use of MIPS to install
the BAHS implant should not increase the risk of postoperative soft-tissue complications
in pediatric patients. Furthermore, in our study, all subjects underwent the MIPS
procedure under local anesthesia and then, the effects of their exposure to the general
anesthesia were reduced, thereby providing an additional benefit for this population.
The mean surgical time was of 30.7 ± 12.78 minutes, and it comprised the entire time
of the surgical procedure, including preparation and aftercare. Our results are comparable
with those found by French et al.[7 ] since they reported a combined average of 12.4 ± 2.6 minutes for skin-to-skin surgical
time. Still, even considering the entire time of the surgical procedure in our study,
we found shorter surgical time with MIPS than those reported by the authors for Baha
Attract (56 minutes) and Connect (53 minutes) (Cochlear Baha Products and Services,
Mölnlycke, Sweden) procedures in pediatric patients.[7 ]
Implant loss occurred only in 1 patient (#S8), who had the implant installed by LITP,
due to trauma. It is known that one of the risk factors for implant loss in children
is the higher incidence of head trauma.[4 ]
[14 ] Considering MIPS, although children with craniofacial malformation have a higher
risk of implant loss due to the bone abnormalities and problems with peri-implant
hygiene on those with underlying syndromes,[4 ] osseointegration did not fail for any subject during the mean follow-up period of
11.4 months.
We found significant improvements in the auditory performance of the subjects in quiet
and in noise with BAHS. A decrease in the SRTQ and SRTN scores in the aided (BAHS)
condition reflected on lower sentence recognition thresholds with the device and indicated
a positive outcome. The mean improvement was stronger in noise compared with in quiet,
and all subjects included in our study showed lower speech recognition thresholds
in noise with the use of BAHS compared with unaided condition. It is known that speech
recognition performance in noise can be problematic for patients with unilateral hearing
loss, particularly when the noise source is positioned near the better ear and the
speech stimulus is presented at or near the poorer ear, with this being the most challenging
situation for them. Then, wearing a BAHS on the affected side would improve that situation,[15 ] and the best speech perception scores in the aided (BAHS) situation would then be
expected. Willenborg et al.[16 ] found a significant improvement of −2.1 dB SNR in speech perception in noise, in
the S0N90 situation (speech coming from the front and noise applied to the normal
hearing ear), with the use of active osteointegrated implants in 6 adult patients
with unilateral hearing loss. They found an even stronger improvement in speech perception
of −4.2 dB SNR, when the patients were evaluated in the S90N90 situation. The authors
mentioned that the results could be even better with the use of the device in the
last test situation (S90N90), but this was the last test performed, and fatigue and
a lack of concentration might have led to a decrement in their performance. In our
study, we found a significant mean improvement in speech perception in noise of −7.2 dB
SNR with the use of BAHS in children with unilateral craniofacial malformation in
the S90N90 test situation. Considering that this was the second test performed and
no fatigue occurred, stronger improvements could be expected when comparing our results
to those found by Willenborg et al.[16 ] A previous study showed that, for speech-in-noise tests, 1dB decrease in S/N ratio
might result in up to 15% better speech recognition.[15 ] Based on this, our results suggest a significant improvement in auditory performance
of these patients with the use of BAHS in their most challenging daily life situation.[17 ]
In quiet, although the mean scores significantly decreased with the use of the BAHS
(mean improvement of 1.7 dB SPL), 1 subject showed equal SRTQ when comparing unaided
and aided (BAHS) conditions. Kunst et al.[18 ] found significant improvements in the speech perception scores of adult patients
with UCMs implanted with BAHS, but some patients showed equal scores when comparing
unaided and aided (BAHS) conditions. Children were not tested in quiet in their study.
The perceptual listening effort reported by the subjects was higher in the speech-in-noise
test, and it significantly decreased with the use of BAHS in both the quiet and noise
conditions. However, the listening effort category perceived by the subjects decreased
with BAHS in the noise condition while it remained equal in quiet, suggesting that
the device was more helpful in the most demanding listening situation to them. Listening
effort was defined as the deliberate allocation of mental resources to overcome obstacles
in goal pursuit when carrying out a listening task.[19 ] The study by Hornsby[20 ] showed that individuals with hearing loss experience greater effort than individuals
with normal hearing to complete intelligibility tests. Lunner et al.[21 ] investigated the listening effort through the ability to recall words in patients
with BAHS connected to both, abutment and soft band, and they found that listeners'
recall ability increased when the sound was directly transmitted to the cochlea via
abutment. Bianchi et al.[22 ] investigated the listening effort via pupillary responses in 21 adult BAHS users
with conductive or mixed conductive-sensorineural hearing loss. They showed that the
overall effort to process a moderate-to-loud speech signal was significantly reduced
by using a sound processor with a higher maximum force output (MFO), due to fewer
saturation artifacts. In our study, all subjects used a SuperPower sound processor
(Oticon Medical), with a high MFO of 135 dBµN, which may have reflected in the significant
reduction of the subjective listening effort, including in the speech test in quiet.
The subjects were very satisfied with the device in daily life. High levels of satisfaction
and increased quality of life with BAHS were previously reported.[23 ]
[24 ] The subjects were using their BAHS daily for more than 8.2 hours, on average, showing
high compliance with the device use. It is known that long daily use of the device
suggests patient satisfaction,[25 ] and these results are comparable with those found in the SADL questionnaire.
Conclusion
Minimally invasive Ponto surgery under local anesthesia showed to be a viable option
for percutaneous BAHS implant installation in children with UCM. The hearing performance
of the subjects improved with BAHS, and they were globally satisfied with the device,
using it for more than 8 hours a day, on average. Soft-tissue complications were minimal,
and our results are comparable to those reported in the literature regarding MIPS
in adults.
Bibliographical Record Andrea Caruso Leone, Arthur Menino Castilho, Fabiana Danieli, Daniela Bortoloti Calil,
Katia de Almeida. Hearing Performance and Soft-Tissue Outcomes of Minimally Invasive
Ponto Surgery and Local Anesthesia in Children with Unilateral Craniofacial Malformation.
Int Arch Otorhinolaryngol 2025; 29: s00441788910. DOI: 10.1055/s-0044-1788910