CC BY-NC-ND 4.0 · Indian J Radiol Imaging 2025; 35(02): 254-262
DOI: 10.1055/s-0044-1788575
Original Article

Role of Virtual Monoenergetic Images in the Assessment of Vessel Enhancement in Segmental Level in Third-Generation Dual-Source Dual-Energy CT Pulmonary Angiography—A Prospective Study

1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Rajeshkumar Varatharajaperumal
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Venkatesh Kasi Arunachalam
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Abdulla KuruVambath
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Rajesh Shanmugam Punniyakotti
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
Sriman Rajasekaran
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
,
1   Department of Radiology, Kovai Medical Center and Hospital, Coimbatore, Tamil Nadu, India
› Author Affiliations
Funding None.
 

Abstract

Introduction Pulmonary embolism is the third most common cause of cardiovascular death worldwide and imaging plays a pivotal role in establishing the diagnosis. Computed tomography pulmonary angiography (CTPA) scores over other modalities and is the current diagnostic investigation of choice. In this study, we assessed the main pulmonary artery and its corresponding segmental artery attenuation in reconstructed virtual monoenergetic (mono plus) images (VMI-MP) and linear blended images (spectral post processing, SPP) obtained from dual-energy CTPA. The values were compared using contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR).

Materials and Methods Forty patients (mean age group, 53.6 years; 26 females and 14 males) with segmental pulmonary thromboembolism were included in this prospective study. The patients underwent CTPA study using bolus tracking in the dual-source CT-SOMATOM Force, Siemens. Postcontrast datasets (90 kV, 150 kV, and SPP) were used to reconstruct the monoenergetic images using syngo.via software virtually. Comparison was done between bivariate samples using the paired sample t-test.

Results The mean Hounsfield unit (HU) artery in the left lung for VMI-MP and SPP images were 886.9 ± 242 and 356.8 ± 121.3 HU, respectively. Similarly, for the right lung, it was 868.3 ± 243.5 and 336.1 ± 105.5 HU, respectively. The mean attenuation of the arteries in MP images was higher and statistically significant (p-value <0.005). Likewise, the CNR) and SNR were found to have a statistically significant p-value (<0.005). An acceptable increase in image noise was seen in VMI as compared with SPP images.

Conclusion Low-keV VMIs perform more effectively than the conventional polyenergetic spectrum to evaluate vessel attenuation, which in turn increases thrombus detectability. The increased CNR in VMI enables improved lesion conspicuity.


#

Introduction

Pulmonary embolism (PE) is the third most common cause of cardiovascular death worldwide following stroke and myocardial infarction.[1] The vast majority of patients present with nonspecific complaints ranging from tachycardia to hemoptysis, and at times, they may be totally asymptomatic. Hence, a high index of suspicion is necessary to diagnose PE. Imaging plays a pivotal role in establishing the diagnosis of acute pulmonary thromboembolism (PTE). Ventilation (V) and perfusion (Q) scans are considered to surpass conventional radiography because of their higher sensitivity and specificity to detect PE. However, the PIOPED I (Prospective Investigation of Pulmonary Embolism Diagnosis) study by Vreim et al showed that 65% of the V/Q scans were nondiagnostic for PE.[2] Single-photon emission computed tomography (V/Q SPECT) provides better anatomical delineation over planar techniques for VQ imaging and improves diagnostic accuracy in the detection of PE. CT pulmonary angiography (CTPA) scores over other modalities and is the current diagnostic investigation of choice in suspected PE. The faster acquisition times and free availability with its higher diagnostic accuracy make it a far more advantageous investigation in the diagnosis of PE. All these factors paved the way for the widespread use of CTPA as the corner stone for the visualization of pulmonary arteries all the way to the subsegmental level. The key disadvantage that existed in the single source CT has been the visualization of the distal small branches of the pulmonary vessels. The recent innovations in CT include dual-energy and photon counting CT. The virtual monoenergetic images (VMI) is a subset of images generated from the dual-energy dataset. In low-keV VMI, the iodine containing structures showed increased attenuation compared with the rest of the areas. In the study done at our tertiary care center, we utilized this property of VMI and assessed the attenuation in the pulmonary arterial circulation and the image quality based on the contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR).


#

Materials and Methods

This prospective observational study was conducted in the department of radiology in a tertiary care hospital after obtaining ethical and scientific committee approval. The patients who presented to our hospital with chest pain, breathlessness, tachycardia, or desaturation had clinical suspicion of PTE underwent CTPA from December 2020 to June 2022 and were included in the study. A total of 40 patients were included in the study of which all of them had segmental PTE.


#

CT Technique

All the patients were subjected to CTPA study in the third-generation dual-source dual-energy CT (SOMATOM Force, Siemens). Automated tube-current modulation was activated in all examinations. Standard soft tissue kernels in postcontrast images were used to analyze the study. The patient was positioned supine with arms extended above the head. The patients were scanned in the craniocaudal direction with breath held in the maximal inspiratory effort. Ultravist 320 mg/mL (Bayer Schering Pharma, Berlin, Germany) was used and a volume of 1 mL/kg was injected for each study. The bolus tracking method was used in which phase I consists of 30 mL of saline with a flow rate of 4.5 mL/s and was injected using an automated dual syringe power injector through an 18 gauge intravenous access placed in the right median cubital/cephalic vein. This was followed by phase II consisting of contrast with a flow rate of 4.5 mL/s, and finally, phase III saline chase with 30 mL of saline at the same flow rate. Region of interest (ROI) was placed in the main pulmonary trunk and the scan was triggered when +100 Hounsfield unit (HU) was reached, with a delay of 4 seconds. Dual-energy datasets (90 and 150 keV) were obtained. Each patient had three sets of images (90 kV, 150 kV, and SPP) following contrast injection.

Dual-energy postcontrast study was obtained using the parameters as mentioned in [Table 1].

Table 1

CT parameters for acquisition of dual-energy datasets in SOMATOM Force, Siemens

 Pitch

 0.6

 Tube A voltage

 90 kV

 Tube B voltage

 150 kV

 Tube A current

 105 mAs

 Tube B current

 90 mAs

Abbreviation: CT, computed tomography.


After the completion of image acquisition, images were transferred to a dedicated workstation (syngo.via, Siemens) for further analysis. The dual-energy images obtained were postprocessed to obtain the VMIs at 40 keV, in 1 mm slice thickness.


#

Evaluation of CT Findings

We had two subsets of images for each patient, which includes VMI (40 keV) and linear blended images. The vessel attenuation at the segmental level was calculated for all the segments of both lungs in both VMI and linear blended images, which are then extrapolated to the corresponding lobes/lungs. The bronchopulmonary segments were divided into 8 on the left side and 10 on the right side as per anatomical classifications. The vascular attenuation was measured by placing the circular ROIs with a minimum area of 1 mm2 and repeated twice to obtain consistency for which the average of the two values was used. The HU of the muscle was obtained by placing the ROI in the pectoralis muscle with a minimum area of 1 cm2. The image noise is defined as the standard deviation (SD) of the HU of muscle.


#

Quantitative Analysis

The CNR can be derived with the following equation:

CNR = HU (pulmonary artery) − HU (muscle)/image noise

The SNR can be derived with the following equation:

SNR = HU (pulmonary artery)/image noise


#

Statistical Analysis

The data for VMI and linear blended images were collected in a Microsoft Excel sheet. The collected data were analyzed with IBM SPSS Statistics for Windows, Version 23.0 (Armonk, New York, United States: IBM Corp). To describe the data descriptive statistics, frequency analysis and percentage analysis were used for categorical variables, whereas mean and SD were used for continuous variables.

To find the significant difference between bivariate samples in paired groups, the paired sample t-test was used. In the above statistical tool, the probability value of 0.05 was considered a statistically significant level.


#

Results

A total of 40 patients who underwent CTPA in a dual-energy system for clinically indicated causes were included. The sample's mean age group was 53.6 ranging from 21 to 81 years. The majority of the patients were between 51 and 60 years which included 11 patients. Among the 40 patients included in the sample, the majority of them were females which comprised 26. The males constituted 14 out of 40 patients.

HU Arterial Attenuation of Lungs

The HU of segmental arteries was obtained which was then extrapolated to the corresponding right and left lungs. The mean HU value of the artery in the left lung for VMI monoplus (MP) images and linear blended (SPP) images were 886.9 ± 242 and 356.8 ± 121.3, respectively. Similarly, for the right lung, it was 868.3 ± 243.5 and 336.1 ± 105.5, respectively. The mean attenuation of the arteries in VMI-MP images was higher and statistically significant (p-value <0.005) ([Tables 2] and [3]).

Table 2

Demonstrating HU of arteries in VMI-MP and SPP for right and left lungs

Mean

Number

 SD

 Left lung—HU artery (MP)

 886.991

 40

 242.0557

 Left lung—HU artery (SPP)

 356.834

 40

 121.3978

 Right lung—HU artery (MP)

 868.388

 40

 243.5138

 Right lung—HU artery (SPP)

 336.179

 40

 105.5403

Abbreviations: HU, Hounsfield unit; MP, monoplus; SD, standard deviation; VMI, virtual monoenergetic image.


Table 3

Statistical comparison of arterial attenuation in MP and SPP for right and left lungs

 HU artery (VMI-MP)–HU artery (SPP)

Paired differences

 t-Value

Degree of freedom

p-Value

Mean

 Standard deviation

Standard error mean

 95% confidence interval of the difference

 Lower

 Upper

Left lung

530.156

140.2236

22.1713

485.3106

575.0019

23.912

39

0.0005

Right lung

532.208

162.3763

25.6740

480.2781

584.1391

20.730

39

0.0005

Abbreviations: HU, Hounsfield unit; MP, monoplus; VMI, virtual monoenergetic image.


HU Noise of Lungs

The mean HU noise of the right lung was 25.7 ± 6.3 in MP and 14.4 ± 4 in SPP images, respectively. Similarly, in the left lung, it was 25.5 ± 6.3 in MP and 14.3 ± 3.9 in SPP. A significant increase in the noise was observed in MP images as compared with linear blended images (SPP). The decrease in energy caused an increase in the noise level of VMIs ([Table 4]).

Table 4

Demonstrating HU of noise in MP and SPP for right and left lungs

 Mean

 Number

 SD

 Left lung—HU noise (MP)

 25.775

 40

 6.3548

 Left lung—HU noise (SPP)

 14.475

 40

 4.0191

 Right lung—HU noise (MP)

 25.550

 40

 6.3244

 Right lung—HU noise (SPP)

 14.325

 40

 3.9574

Abbreviations: HU, Hounsfield unit; MP, monoplus; SD, standard deviation.



#
#

Contrast-to-Noise Ratio

By extrapolating the HU arteries obtained from the segmental arteries (8 segments in the left and 10 segments in the right lung) to the lobewise distribution following CNR was obtained ([Fig. 1]).

Zoom Image
Fig. 1 Bar diagram showing mean CNR values in MP and SPP for all the lobes. CNR, contrast-to-noise ratio; MP, monoplus.

Left Lung

The mean CNR for the upper lobe in MP and SPP were 33.8 ± 10.8 and 22.9 ± 8.9, respectively. Similarly, the mean CNR for the lower lobe was 31.7 ± 10.6 in MP and 21.5 ± 9.4 in SPP datasets ([Tables 5] and [6]). The CNR was found to have a statistically significant p-value of less than 0.005 as mentioned in [Table 7].

Table 5

Demonstrating CNR values of arteries in MP and SPP for left upper and lower lobes

 Left lung

 Mean

Number

 SD

Upper lobe—CNR (MP)

33.894

40

10.8467

Upper lobe—CNR (SPP)

22.948

40

8.9607

Lower lobe—CNR (MP)

31.728

40

10.6772

Lower lobe—CNR (SPP)

21.530

40

9.4512

Abbreviations: CNR, contrast-to-noise ratio; MP, monoplus; SD, standard deviation.


Table 6

Statistical comparison of CNR values in MP and SPP for left upper and lower lobes

 CNR (MP)–CNR (SPP)

Mean

Paired differences

t-Value

Degree of freedom

 p-Value

Standard deviation

Standard error mean

95% confidence interval of the difference

Lower

Upper

Left upper lobe

10.9455

7.2965

1.1537

8.6120

13.2790

9.488

39

0.0005

Left lower lobe

10.1982

6.1393

.9707

8.2348

12.1617

10.506

39

0.0005

Abbreviations: CNR, contrast-to-noise ratio; MP, monoplus.


Table 7

Statistical significance of SNR values in left upper and lower lobes

 SNR (MP)–SNR (SPP)

 Mean

 Paired differences

t-Value

 Degree of freedom

p-Value

 Standard deviation

 Standard error mean

 95% Confidence interval of the difference

Lower

 Upper

Left upper lobe

10.3469

7.6892

1.2158

7.8878

12.8060

8.511

39

0.0005

Left lower lobe

9.5996

6.5387

1.0339

7.5085

11.6908

9.285

39

0.0005

Abbreviations: MP, monoplus; SNR, signal-to-noise ratio.



#

Right Lung

Similarly, a statistically significant p-value was obtained by calculating the CNR for all the lobes in both lungs as mentioned in [Tables 8] and [9].

Table 8

Demonstrating CNR values of arteries in MP and SPP for right upper middle and lower lobes

 Right lung

 Mean

 Number

 SD

 Upper lobe—CNR (MP)

 32.683

 40

 12.3655

 Upper lobe—CNR (SPP)

 21.243

 40

 8.9702

 Middle lobe—CNR (MP)

 31.874

 40

 12.0122

 Middle lobe—CNR (SPP)

 20.824

 40

 9.1450

 Lower lobe—CNR (MP)

 32.683

 40

 12.3655

 Lower lobe—CNR (SPP)

 21.243

 40

 8.9702

Abbreviations: CNR, contrast-to-noise ratio; MP, monoplus; SD, standard deviation.


Table 9

Statistical significance of CNR values in right upper middle and lower lobes

 CNR (MP)–CNR (SPP)

Mean

Paired differences

 t-Value

Degree of freedom

p-Value

Standard deviation

Standard error mean

 95% confidence interval of the difference

 Lower

 Upper

Right upper lobe

11.4400

8.6638

1.3699

8.6692

14.2108

8.351

39

0.0005

Right middle lobe

11.0496

7.8122

1.2352

8.5512

13.5481

 8.946

39

0.0005

Right lower lobe

11.4400

8.6638

1.3699

8.6692

14.2108

 8.351

39

0.0005

Abbreviations: CNR, contrast-to-noise ratio; MP, monoplus.



#
#

Signal-to-Noise Ratio

By extrapolating the HU arteries obtained from the segmental arteries (8 segments in the left and 10 segments in the right lung) to the lobewise distribution following SNR was obtained ([Fig. 2]).

Zoom Image
Fig. 2 Bar diagram showing mean SNR values in MP and SPP for all the lobes. SNR, signal-to-noise ratio; MP, monoplus.

Left Lung

The mean SNR for the upper lobe in MP and SPP were 36.6 ± 11 and 26.2 ± 9.2, respectively. Similarly, the mean SNR for the lower lobe was 34.4 ± 10.9 in MP and 24.8 ± 9.8 in SPP datasets. The SNR was found to have a statistically significant p-value of less than 0.005 as mentioned in [Tables 7] and [10].

Table 10

Demonstrating SNR values of arteries in MP and SPP for left upper and lower lobes

Left lung

Mean

Number

SD

Upper lobe—SNR (MP)

36.638

40

11.0968

Upper lobe—SNR (SPP)

26.291

40

9.2658

Lower lobe—SNR (MP)

34.472

40

10.9139

Lower lobe—SNR (SPP)

24.873

40

9.8054

Abbreviations: MP, monoplus; SD, standard deviation; SNR, signal-to-noise ratio.



#

Right Lung

Similarly, a statistically significant p-value was obtained by calculating the SNR for all the lobes in both lungs as mentioned in [Tables 11] and [12].

Table 11

Demonstrating SNR values of arteries in MP and SPP for right upper middle and lower lobes

Right lung

Mean

Number

SD

Upper lobe—SNR (MP)

35.428

40

12.5499

 Upper lobe—SNR (SPP)

24.633

40

9.2524

 Middle lobe—SNR (MP)

34.619

40

12.1218

 Middle lobe—SNR (SPP)

24.214

40

9.4035

 Lower lobe—SNR (MP)

34.939

40

10.2806

 Lower lobe—SNR (SPP)

24.010

40

7.8846

Abbreviations: MP, monoplus; SD, standard deviation; SNR, signal-to-noise ratio.


Table 12

Statistical significance of SNR values in right upper middle and lower lobes

 SNR (MP)–SNR (SPP)

 Mean

Paired differences

 t-Value

 Degree of freedom

 p-Value

 Standard deviation

 Standard error mean

95% confidence interval of the difference

 Lower

Upper

Right upper lobe

10.7953

9.0239

1.4268

7.9093

13.6813

7.566

39

0.0005

Right middle lobe

10.4049

8.1872

1.2945

7.7865

13.0233

8.038

39

0.0005

Right lower lobe

10.9289

7.8967

1.2486

8.4034

13.4544

8.753

39

0.0005

Abbreviations: MP, monoplus; SNR, signal-to-noise ratio.



#
#
#

Discussion

Conventional CTPA images show decreased vessel attenuation in the distal pulmonary arteries, which could lead to less effective detection of PTE at this level. The attenuation in the distal arteries can be enhanced for optimal assessment by utilizing lower keV. It is based on the concept that lower energy approaching the K-edge value of iodine produces greater attenuation of the vessels. These low-keV VMIs are generated from a dual-energy dataset. The contrast of these VMI is high, but there is a slight increase in background noise. It is possible to increase the detectability of PTE in this subset of images because of an increase in attenuation of the density of vessels.

In our study, we reconstructed the VMI-MP at 40 keV and compared them with the standard linear blended images for assessing the vessel attenuation at different segmental arteries. The mean HU artery in the left lung for VMI-MP and linear blended (SPP) images were 886.9 ± 242 and 356.8 ± 121.3 HU, respectively ([Fig. 3]). Similarly, for the right lung, it was 868.3 ± 243.5 and 336.1 ± 105.5 HU, respectively ([Fig. 4]). The arteries in MP images had a higher mean attenuation which was statistically significant (p-value < 0.005).

Zoom Image
Fig. 3 Axial contrast-enhanced pulmonary angiography shows a higher attenuation value in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B) measured left apicoposterior segmental branch. MP, monoplus.
Zoom Image
Fig. 4 Axial contrast-enhanced pulmonary angiography shows a higher attenuation value in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B) measured right posterior basal segmental branch. MP, monoplus.

By extrapolating the HU values for segmental arteries of the left upper lobe, the mean CNR in MP and SPP were 33.8 ± 10.8 and 22.9 ± 8.9, respectively. Likewise, the mean CNR for the lower lobe was 31.7 ± 10.6 in MP and 21.5 ± 9.4 in SPP datasets. The CNR was found to have a statistically significant p-value of less than 0.005. The CNR values obtained from MP images in the upper, middle, and lower lobes of the right side were statistically significant when compared with the SPP images. Despite the increase in background noise in VMI, the contrast attenuation was significantly higher, which had an impact on the overall image quality.

Delasalle et al showed conventional monoenergetic reconstructions at 60 keV provided adequate attenuation without significant artifacts in the majority of patients, with the highest SNR and CNR, the lowest level of subjective noise using a dual-source dual-energy approach at 80/140 kVp. In comparison to typical single-energy CTPA images, Delasalle et al demonstrated that virtual monoenergetic reconstructed image sets at 60 keV provide the greatest image quality both objectively (SNR and CNR) and subjectively (reduced artifacts and subjective noise).[3]

Past studies conducted by Matsumoto et al in 2011 evaluated the fast-switching dual-energy technique to assess the virtual monochromatic images for contrast assessment. They concluded that VMI images acquired at around 70 keV and reconstructed from split 80 and 140 kVp data displayed reduced image noise and greater CNR than typical 120 kV CT images.[4]

Recent studies by Yuan et al concluded that VMIs enhance the image quality of dual-energy CTPA with the 50 keV dataset offering the best outcomes for imaging of the pulmonary artery circulation, utilizing single-source dual-energy CT with rapid switching between 80 and 140 kVp.[5]

Assessment of image quality on dual-energy CTPA VMIs done by Dane et al proved that monoenergetic image data from dual-energy CTPA can deliver optimum image quality at 40 keV without considerable noise. The mean attenuation ranged from 914.83 HU for 40 keV images.[6] These results were quite similar to our study where the mean attenuation ranged from 878 ± 242 HU.

Meier et al performed a retrospective study to assess the contrast of VMI in dual-energy CTPA at lobar pulmonary branches.[7] MP 40 keV images showed a higher SNR and CNR in the pulmonary trunk and right lower lobe pulmonary artery compared with conventional images (p < 0.001).

Multiple studies have demonstrated that 40 keV monoenergetic datasets are optimal for reconstructing and assessing vascular attenuation for optimal imaging. Although the image noise was increased in monoenergetic images, a substantial increase in the HU value was observed even in the subsegmental arteries ([Fig. 5]). Furthermore, numerous recent studies were performed in noise-optimized VMI-MP technique and found that quantitative image quality may be enhanced even further and exhibit the highest contrast attenuation.[8] [9] [10] [11] [12]

Zoom Image
Fig. 5 Axial contrast-enhanced pulmonary angiography with maximal intensity projection shows optimal visualization of the segmental and subsegmental branches in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B). MP, monoplus.

The SNR of the MP images during 40 keV reconstruction was compared with the SPP images in a similar manner. The SNR was found to have a statistically significant p-value of less than 0.005.

An acceptable increase in image noise was observed in 40 keV VMI-MP as compared with linear blended images in our study ([Fig. 6]). The mean HU noise of the right lung was 25.7 ± 6.3 in MP and 14.4 ± 4 in linear blended (SPP) images, respectively. Similarly, in the left lung, mean HU was 25.5 ± 6.3 in MP and 14.3 ± 3.9 in SPP. In a study conducted by Leithner et al, when compared with linearly blended images, VMI reconstructions at 40 keV showed significantly higher attenuation in the pulmonary trunk (mean attenuation: 718.1 and 229.4 HU, respectively, p = 0.001), as well as significantly higher noise (average noise: 35.7 and 19.9 HU, respectively, p = 0.001).[13]

Zoom Image
Fig. 6 Axial contrast-enhanced pulmonary angiography shows the measurement of noise in the form of the standard deviation of HU in the right pectoralis muscle. Significant increase in the background noise in the 40-keV (MP) images (A) was noted. However as compared to VMI images, the noise is lesser in Linear blended image (B). HU, Hounsfiled unit; MP, monoplus.

Beyond the various study designs that unite the aforementioned studies, a common finding is that low-keV VMIs perform more effectively than the conventional polyenergetic spectrum for the evaluation of vessel attenuation. This in turn reflects the detectability of thrombus. VMI's improved diagnostic performance is likely due to the increase in CNR, which allows for improved lesion conspicuity. In our study, low-keV images were able to detect the PE in some instances when it was not detected in blended images during the first evaluation. On the repeated review, it was visualized in linear blended images. Therefore, the diagnostic value of acute PE in doubtful or disregarded standard mixed images can be improved by using low-energy VMI+ images. In a study conducted by Leithner et al, where they assessed the diagnostic accuracy of VMI and iodine perfusion maps of dual-energy CTPA, they showed that when compared linear blended images in dual-energy CTPA with inadequate contrast attenuation, a reconstruction strategy using the 40-keV VMI-MP series and dual-energy CT-MP enhances reader confidence and diagnostic accuracy for segmental PE identification. Therefore, if we are not able to detect PE in linear blended images, we can review low-keV images to confirm that it is truly negative for embolism. Although many studies have evaluated the role of VMI in assessing the density of pulmonary vessels, most of these studies have focused on analyzing larger vessels such as trunks and main pulmonary arteries.[7] [13] [14] To our knowledge, there are only very few studies that have been conducted to assess the pulmonary vascular attenuation at the segmental arterial level using the VMI-MP as we did.


#

Conclusion

To sum up, low-keV VMIs were superior to the conventional polyenergetic spectrum in assessing vessel attenuation, which in turn reveals the detectability of thrombus. Hence, utilizing VMI has the advantage of better CNR and SNR of the pulmonary arteries even at segmental levels, leading to an exceptional assessment with an acceptable increase in background noise. This can be recommended for optimal visualization of the segmental arteries for better detection of PTE. This prospective study patient group was rather limited to a small sample size; therefore, a larger scale study is required to support our findings. Currently, our results only apply to dual-source CT technology and cannot be immediately applied to dual-energy CT solutions by other vendors.


#
#

Conflict of Interest

None declared.

  • References

  • 1 Goldhaber SZ, Bounameaux H. Pulmonary embolism and deep vein thrombosis. Lancet 2012; 379 (9828) 1835-1846
  • 2 Vreim CE, Saltzmann HA, Alavai A. et al. Value of the Ventilation/Perfusion Scan in Acute Pulmonary Embolism. Results of the Prospective Investigation of Pulmonary Embolism Diagnosis (PIOPED). The PIOPED investigators. JAMA 1990; 263 (20) 2753-2759
  • 3 Delesalle M-A, Pontana F, Duhamel A. et al. Spectral optimization of chest CT angiography with reduced iodine load: experience in 80 patients evaluated with dual-source, dual-energy CT. Radiology 2013; 267 (01) 256-266
  • 4 Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S. Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 2011; 259 (01) 257-262
  • 5 Yuan R, Shuman WP, Earls JP. et al. Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography–a prospective randomized trial. Radiology 2012; 262 (01) 290-297
  • 6 Dane B, Patel H, O'Donnell T. et al. Image quality on dual-energy CTPA virtual monoenergetic images: quantitative and qualitative assessment. Acad Radiol 2018; 25 (08) 1075-1086
  • 7 Meier A, Wurnig M, Desbiolles L, Leschka S, Frauenfelder T, Alkadhi H. Advanced virtual monoenergetic images: improving the contrast of dual-energy CT pulmonary angiography. Clin Radiol 2015; 70 (11) 1244-1251
  • 8 Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B. Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Invest Radiol 2014; 49 (09) 586-592
  • 9 Albrecht MH, Trommer J, Wichmann JL. et al. Comprehensive comparison of virtual monoenergetic and linearly blended reconstruction techniques in third-generation dual-source dual-energy computed tomography angiography of the thorax and abdomen. Invest Radiol 2016; 51 (09) 582-590
  • 10 Albrecht MH, Scholtz JE, Hüsers K. et al. Advanced image-based virtual monoenergetic dual- energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 2016; 26 (06) 1863-1870
  • 11 Beeres M, Trommer J, Frellesen C. et al. Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging. Int J Cardiovasc Imaging 2016; 32 (01) 137-144
  • 12 Albrecht MH, Scholtz JE, Kraft J. et al. Assessment of an advanced monoenergetic reconstruction technique in dual-energy computed tomography of head and neck cancer. Eur Radiol 2015; 25 (08) 2493-2501
  • 13 Leithner D, Wichmann JL, Vogl TJ. et al. Virtual monoenergetic imaging and iodine perfusion maps improve diagnostic accuracy of dual-energy computed tomography pulmonary angiography with suboptimal contrast attenuation. Invest Radiol 2017; 52 (11) 659-665
  • 14 Murphy A, Cheng J, Pratap J, Redman R, Coucher J. Dual-energy computed tomography pulmonary angiography: comparison of vessel enhancement between linear blended and virtual monoenergetic reconstruction techniques. J Med Imaging Radiat Sci 2019; 50 (01) 62-67

Address for correspondence

Rajeshkumar Varatharajaperumal, DMRD, DNB
Department of Radiology, Kovai Medical Center and Hospital
Avanashi Road, Coimbatore 641014, Tamil Nadu
India   

Publication History

Article published online:
03 September 2024

© 2024. Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India

  • References

  • 1 Goldhaber SZ, Bounameaux H. Pulmonary embolism and deep vein thrombosis. Lancet 2012; 379 (9828) 1835-1846
  • 2 Vreim CE, Saltzmann HA, Alavai A. et al. Value of the Ventilation/Perfusion Scan in Acute Pulmonary Embolism. Results of the Prospective Investigation of Pulmonary Embolism Diagnosis (PIOPED). The PIOPED investigators. JAMA 1990; 263 (20) 2753-2759
  • 3 Delesalle M-A, Pontana F, Duhamel A. et al. Spectral optimization of chest CT angiography with reduced iodine load: experience in 80 patients evaluated with dual-source, dual-energy CT. Radiology 2013; 267 (01) 256-266
  • 4 Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S. Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 2011; 259 (01) 257-262
  • 5 Yuan R, Shuman WP, Earls JP. et al. Reduced iodine load at CT pulmonary angiography with dual-energy monochromatic imaging: comparison with standard CT pulmonary angiography–a prospective randomized trial. Radiology 2012; 262 (01) 290-297
  • 6 Dane B, Patel H, O'Donnell T. et al. Image quality on dual-energy CTPA virtual monoenergetic images: quantitative and qualitative assessment. Acad Radiol 2018; 25 (08) 1075-1086
  • 7 Meier A, Wurnig M, Desbiolles L, Leschka S, Frauenfelder T, Alkadhi H. Advanced virtual monoenergetic images: improving the contrast of dual-energy CT pulmonary angiography. Clin Radiol 2015; 70 (11) 1244-1251
  • 8 Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B. Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Invest Radiol 2014; 49 (09) 586-592
  • 9 Albrecht MH, Trommer J, Wichmann JL. et al. Comprehensive comparison of virtual monoenergetic and linearly blended reconstruction techniques in third-generation dual-source dual-energy computed tomography angiography of the thorax and abdomen. Invest Radiol 2016; 51 (09) 582-590
  • 10 Albrecht MH, Scholtz JE, Hüsers K. et al. Advanced image-based virtual monoenergetic dual- energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 2016; 26 (06) 1863-1870
  • 11 Beeres M, Trommer J, Frellesen C. et al. Evaluation of different keV-settings in dual-energy CT angiography of the aorta using advanced image-based virtual monoenergetic imaging. Int J Cardiovasc Imaging 2016; 32 (01) 137-144
  • 12 Albrecht MH, Scholtz JE, Kraft J. et al. Assessment of an advanced monoenergetic reconstruction technique in dual-energy computed tomography of head and neck cancer. Eur Radiol 2015; 25 (08) 2493-2501
  • 13 Leithner D, Wichmann JL, Vogl TJ. et al. Virtual monoenergetic imaging and iodine perfusion maps improve diagnostic accuracy of dual-energy computed tomography pulmonary angiography with suboptimal contrast attenuation. Invest Radiol 2017; 52 (11) 659-665
  • 14 Murphy A, Cheng J, Pratap J, Redman R, Coucher J. Dual-energy computed tomography pulmonary angiography: comparison of vessel enhancement between linear blended and virtual monoenergetic reconstruction techniques. J Med Imaging Radiat Sci 2019; 50 (01) 62-67

Zoom Image
Fig. 1 Bar diagram showing mean CNR values in MP and SPP for all the lobes. CNR, contrast-to-noise ratio; MP, monoplus.
Zoom Image
Fig. 2 Bar diagram showing mean SNR values in MP and SPP for all the lobes. SNR, signal-to-noise ratio; MP, monoplus.
Zoom Image
Fig. 3 Axial contrast-enhanced pulmonary angiography shows a higher attenuation value in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B) measured left apicoposterior segmental branch. MP, monoplus.
Zoom Image
Fig. 4 Axial contrast-enhanced pulmonary angiography shows a higher attenuation value in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B) measured right posterior basal segmental branch. MP, monoplus.
Zoom Image
Fig. 5 Axial contrast-enhanced pulmonary angiography with maximal intensity projection shows optimal visualization of the segmental and subsegmental branches in 40 keV (MP) (A) as compared with the linear blended images (SPP) (B). MP, monoplus.
Zoom Image
Fig. 6 Axial contrast-enhanced pulmonary angiography shows the measurement of noise in the form of the standard deviation of HU in the right pectoralis muscle. Significant increase in the background noise in the 40-keV (MP) images (A) was noted. However as compared to VMI images, the noise is lesser in Linear blended image (B). HU, Hounsfiled unit; MP, monoplus.