Keywords
impedance - cochlear implants - telemetry - hearing loss
Cochlear implantation is now a widely available technology used to rehabilitate significant
hearing loss in patients presenting poor benefit with hearing aids, particularly in
regard to speech intelligibility.[1] The most frequent cause of genetic hearing loss is mutations at the Connexin GJB2
gene,[2] responsible for the codification of a gap-junction protein that is essential for
the physiological function of supporting cells in the cochlea.[3] This microscopic structural damage is frequently related to normal macroscopic anatomy
of the inner ear, without signs of progressive ossification or fibrosis, such as in
meningitis or otosclerosis, making these patients the ideal candidates for early surgical
intervention. Patients affected by congenital profound hearing loss undergo (unilateral
or bilateral) cochlear implantation around the first year of life with minimal surgical
risk, good auditory outcome, and good development of communicative skills.[1] These patients are expected to be cochlear implant (CI) users for decades and, while
optimal fitting is frequently obtained within the first year after surgery, they should
undergo regular testing to check the status of the device.
CI status can be checked by using different types of objective measures; among them,
impedance telemetry has a crucial role.[4] Impedance is defined as the measure of opposition to electrical current flow through
an electrode lead and across an electrode contact.[4] It is given by the vector sum of resistance and reactance components; although the
factors that might influence impedance values are still not completely understood,
the role of the physical properties of the electrode lead and contact and of the medium
surrounding them seem to be extremely important.[4]
[5]
[6]
[7]
[8] Consequently, it can give information about the function of the electrode contacts,
and it is a crucial parameter for the implant fitting, as it is used to calculate
the compliance. In addition, impedance of electrode contacts correlates strongly with
the current levels used for CI fitting.[9] The test lasts a few seconds, and it is harmless. Some patients might perceive an
extremely slight auditory stimulus during the impedance telemetry.
Abnormal impedance levels are classified as open circuit (OC) for extremely high values
and short circuit (SC) for extremely low values.[10] A particular case is the partial SC that is a specific pattern of high–low sequence
of impedance values that needs specific attention and strict follow-up of the patient.[11]
[12] These abnormalities should raise clinicians' attention because they may be a sign
of CI dysfunction. Also, in-range variations might be of interest as they might be
precocious signs of electrode dysfunction or inflammatory processes within the cochlea,
progressive fibrosis/ossification, or fluctuations of the array's position.[4]
[5]
[13] Early identification of these anomalies is crucial to minimize the impact on the
hearing performance.[10] In some cases, the identification of major problems affecting the CI functioning
might also lead to explant of the device. For all these reasons, it is recommended
to evaluate impedance telemetry during cochlear implantation and in each programming
session.[4]
Very few studies are available in medical literature describing the trend of impedance
values over time for the Cochlear Nucleus devices.[5]
[13]
[14]
[15]
[16]
[17] All studies had the limitation to be conducted in heterogeneous populations in terms
of hearing loss cause, including etiologies characterized by progressive ossification
processes such as meningitis or otosclerosis. Consequently, it is still unknown if
impedance trend may or may not be influenced by the cause of the hearing loss. In
addition, in most studies the impedance-value trend was investigated in a short period
of time.[14]
[15]
[16] To date, only two studies[13]
[18] considered a 2-year post cochlear implantation follow-up period, and only one[13] considered a 3-year period.
The present retrospective monocentric study has been conducted on a cohort of patients
affected by genetically determined congenital hearing loss who underwent cochlear
implantation with Cochlear Nucleus devices. The primary endpoint of the study was
to describe the trend of CI impedance values in the present cohort at the 2-year follow-up.
The secondary endpoint was to evaluate the trend in a 5-year span considering a subgroup
of the cohort. The assessment of these trends should be considered a valuable tool
that could be further used for the systematic comparison with impedance trends of
other implanted patients affected by different disorders.
Materials and Methods
Type of Study
This is an observational retrospective monocentric study. Data were analyzed in accordance
with Italian privacy and data laws (D. Lgs 196/03).
Sample of the Study
The cohort of the study is composed of 27 consecutive patients who underwent CI from
2010 to 2020 at the Azienda Ospedale - Università Padova (University of Padova, Italy).
All patients underwent cochlear implantation with Cochlear Nucleus (Cochlear Ltd,
Macquarie, NSW, Australia) devices. The type of implant was CI532 in 6 patients, CI512
in 10, and CI24RE(CA) in 12. For all patients, the chosen array was perimodiolar half-banded,
with 22 electrodes. The mean age of the patients was 12.0 ± 7.6 years; range, 4.2–40.4
(18 males and 9 females); and median age 10.6 years.
All patients had genetic testing consistent with Connexin 26 (GJB2) mutation causing
congenital profound hearing loss. The cochlear implantation was performed with the
same surgical procedure (mastoidectomy with posterior tympanotomy and round window
insertion) performed by the same surgeon, with full insertion of the array. All patients
were considered full-time users (more than 10 hours/day checked by means of data logging)
with optimal performance (pure-tone audiometry under 30 dB HL at tested frequencies
of 250–500–1,000–2,000–4,000–6,000 Hz).
Impedance Values Measurements
Impedance measurement data at 5 years after implantation were available for 16 patients
(mean age 13.4 ± 8.3 years, range 7.8–40.4, median age 11.5 years).
Impedance values were measured over time (at activation, 6, 12, 24, and 60 months
after cochlear implantation), for each of the 22 electrodes.
In Cochlear Nucleus devices, impedance is tested at the end of the first phase of
the single biphasic pulse. During the measurement, the software stimulates each electrode
at 80 Current Levels in postoperative and intraoperative testing, using a pulse width
of 25 μs.[19] In Custom Sound/Custom Sound EP clinical software, impedance telemetry is measured
in four coupling modes or configurations: common ground (CG), monopolar 1 (MP1), monopolar
2 (MP2), and monopolar 1 + 2 (MP1 + 2). In CG stimulation, current flows between the
active electrode and all the other electrodes on the array, which are connected together
electronically to form a single indifferent or reference electrode. In monopolar (MP)
mode, the active electrode is inside the cochlea, and the indifferent electrode is
outside the cochlea. There are three MP configurations. In MP1 mode, current flows
between the active intracochlear electrode and the extracochlear electrode MP1 (located
at the tip of a separate lead and placed in the temporal muscle). In MP2 mode, current
flows between the active intracochlear electrode and the extracochlear electrode MP2
(located on the internal receiver/stimulator case). In MP1 + 2 mode, current flows
between the active intracochlear electrode and the extracochlear electrodes (MP1 and
MP2) shorted together. In general, the pattern of impedance values for each electrode
is expected to be similar for all the four testing configurations.[4]
[5]
[6]
[7]
[8]
[9]
[10]
[11]
[12]
[13]
[14]
[15]
[16]
[17]
[18]
[19]
[20] See [Fig. 1] for a schematic explanation of the different coupling modes. Valid impedance values
range from 0.565 to 30 kΩ for half-banded arrays, and from 0.565 to 20 kΩ for full-banded
arrays.[5]
Fig. 1 Schematic representation of the coupling modes used to measure impedance levels for
each electrode in Cochlear Nucleus devices.
Statistical Analysis
Repeated-measures analysis of variance (ANOVA) was applied to the data, and the main
effects (time and electrodes) and interaction effects were evaluated using the Greenhouse–Geisser
method.
It was decided to consider a more robust statistical method to correct for violating
the assumption of sphericity with repeated-measures ANOVA.
Significance level was set at p < 0.05.
All analyses were performed using the Statistical Package for Social Sciences, version
12.0.
Results
Mean impedance values and standard deviations measured in CG are shown in [Table 1].
Table 1
Electrode impedance values in kΩ, measured in common ground modality for each interval
time
|
Activation (n = 27)
|
6 months (n = 27)
|
12 months (n = 27)
|
24 months (n = 27)
|
5 years (n = 16)
|
|
Electrode
|
Mean ± SD
|
Mean ± SD
|
Mean ± SD
|
Mean ± SD
|
Mean ± SD
|
|
1
|
11.13 ± 4.75
|
10.21 ± 2.68
|
10.40 ± 2.79
|
10.07 ± 2.98
|
11.23 ± 3.69
|
|
2
|
10.67 ± 4.10
|
9.57 ± 2.39
|
9.43 ± 2.43
|
9.49 ± 2.44
|
10.19 ± 3.53
|
|
3
|
10.17 ± 3.19
|
9.44 ± 2.23
|
9.54 ± 2.45
|
9.22 ± 2.73
|
10.06 ± 3.72
|
|
4
|
9.92 ± 3.3
|
9.11 ± 2.43
|
9.05 ± 2.30
|
8.92 ± 2.65
|
9.33 ± 3.87
|
|
5
|
9.32 ± 3.65
|
8.24 ± 2.10
|
8.28 ± 2.06
|
7.75 ± 2.68
|
8.92 ± 3.54
|
|
6
|
9.48 ± 3.23
|
8.05 ± 2.45
|
7.97 ± 2.34
|
8.01 ± 2.72
|
9.00 ± 3.87
|
|
7
|
9.50 ± 3.41
|
7.78 ± 2.28
|
7.72 ± 2.27
|
7.67 ± 2.61
|
8.80 ± 3.62
|
|
8
|
9.26 ± 3.23
|
7.69 ± 2.19
|
7.91 ± 2.33
|
7.45 ± 2.79
|
9.04 ± 3.80
|
|
9
|
8.85 ± 3.66
|
7.23 ± 2.56
|
7.71 ± 2.36
|
7.18 ± 2.75
|
8.10 ± 3.98
|
|
10
|
8.82 ± 3.84
|
7.71 ± 2.22
|
7.80 ± 2.45
|
7.50 ± 2.71
|
8.14 ± 2.86
|
|
11
|
9.08 ± 3.54
|
7.77 ± 2.09
|
7.99 ± 2.54
|
7.52 ± 2.74
|
8.74 ± 3.61
|
|
12
|
9.05 ± 3.48
|
7.74 ± 2.32
|
7.74 ± 2.29
|
7.67 ± 2.78
|
8.09 ± 3.08
|
|
13
|
9.40 ± 3.89
|
7.99 ± 2.39
|
8.01 ± 2.17
|
7.63 ± 2.54
|
8.34 ± 2.89
|
|
14
|
9.38 ± 3.72
|
7.84 ± 2.39
|
8.22 ± 2.55
|
8.08 ± 3.00
|
8.72 ± 3.38
|
|
15
|
9.39 ± 4.19
|
8.03 ± 2.40
|
7.93 ± 1.86
|
7.70 ± 2.55
|
8.51 ± 3.39
|
|
16
|
9.16 ± 3.78
|
7.59 ± 1.81
|
7.82 ± 1.91
|
7.54 ± 2.55
|
8.11 ± 3.28
|
|
17
|
9.07 ± 3.72
|
7.77 ± 1.97
|
7.70 ± 1.96
|
7.51 ± 2.30
|
7.82 ± 3.08
|
|
18
|
9.36 ± 3.67
|
8.03 ± 1.79
|
8.15 ± 2.29
|
7.72 ± 2.29
|
7.84 ± 2.82
|
|
19
|
9.56 ± 3.84
|
7.94 ± 1.27
|
7.66 ± 1.49
|
7.59 ± 1.86
|
7.97 ± 2.88
|
|
20
|
9.06 ± 4.33
|
7.14 ± 1.36
|
7.09 ± 1.55
|
6.97 ± 1..66
|
7.51 ± 3.33
|
|
21
|
9.45 ± 4.22
|
7.41 ± 1.66
|
7.24 ± 1.77
|
7.04 ± 1.99
|
7.17 ± 2.95
|
|
22
|
10.18 ± 3.96
|
8.09 ± 2.32
|
8.26 ± 2.06
|
7.72 ± 2.22
|
7.87 ± 3.64
|
Abbreviation: SD, standard deviation.
Statistical analysis for repeated measures showed a significant time-linked variation
(F [1.786] = 5.747, p = 0.01) regarding the analysis on CG data available up to 24-month follow-up. This
difference, as can be seen in [Fig. 2], is due to activation-time data. As a matter of fact, the same analysis, considering
only 6, 12, and 24 months of follow-up controls, showed no significant effect of time
factor (F [1.658] = 1.383, p = 0.3) while statistically significant variation was found between CI activation
and 6-month follow-up (F [1.000] = 8.012, p = 0.009).
Fig. 2 Impedance-estimated marginal means differentiated according to time intervals (activation,
6, 12, and 24 months) and related to all 22 electrodes—common ground values.
A significant main effect of electrodes was also detected from ANOVA analysis (F [5.758] = 7.881, p = 0.001).
Additionally, a post-hoc analysis with the Tukey method was performed. A comparison
between the average of the mean impedance values from electrode number 5 to electrode
number 22 and the mean impedance values of electrodes 1, 2, 3, and 4, for each follow-up,
were performed. No significant difference was found with regard to the activation
follow-up (F [4] = 0.9, p = 0.5), while significant differences were observed for the other follow-ups (6 months:
(F [4] = 4.151, p = 0.003); 12 months: (F [4] = 4.162, p = 0.003); 24 months: (F [4] = 2.898, p = 0.025).
Multiple comparisons, except for the activation follow-up, also show significant differences
(p-value ranging from 0.001 to 0.013) between the mean impedance values from number
5 to number 22 and the other four (1, 2, 3, and 4), while no difference was observed
between the latter four electrodes (all p-value greater than 0.05 for each comparison).
Electrodes 1, 2, 3, and 4 (which are located in the basal cochlear region, next to
the round window) were found to have higher impedance values compared with all other
electrodes, in each time interval (see [Fig. 3]).
Fig. 3 Impedance-estimated marginal means differentiated according to time intervals (activation,
6, 12, and 24 months and 5 years) and related to all 22 electrodes—common ground values.
In [Fig. 3], it can also be observed that all 22 electrodes followed the same trend over time.
Instead, no significant interaction effect was found between electrode and time interval
(F [8.960] = 0.988, p = 0.45). Comparable results were obtained from the analysis assessed up to 5 years
of follow-up ([Figs. 4] and [5]). Finally, very similar results have been also found in MP1, MP2, and MP1 + 2 test
modalities.
Fig. 4 Impedance-estimated marginal means differentiated according to electrodes and related
to time intervals (activation, 6, 12, and 24 months)—common ground values.
Fig. 5 Impedance-estimated marginal means differentiated according to electrodes and related
to time intervals (activation, 6, 12, and 24 months and 5 years)—common ground values.
A comparison between all four test modalities has also been performed, and the CG
mean impedance values across all electrodes were lower than the other modalities in
all analyzed follow-ups, while impedance levels were found to be higher in MP1 modality.
The paired sample t-test, in particular, showed statistically significant differences between CG and
the other modalities (all p-values less than 0.005). To note, the difference is never greater than 2 kΩ.
The other plots related to all repeated-measures ANOVA are available in [Supplementary Material].
Discussion
The present study analyzed the impedance values over time in a sample of CI users.
The results highlighted the long-term stability of these values in a 5-year span,
including follow-up recordings at activation, 6-month, and 1-, 2-, and 5-year post-CI
switch on. These data were observed retrospectively in a cohort of patients all affected
by congenital profound hearing loss with genetically diagnosed GJB2 mutation. All
patients were operated by the same surgeon with the same surgical technique. All patients
were implanted with Cochlear devices, all with a perimodiolar Contour Advance electrode.
Impedance telemetry is considered an unskippable test that should always be performed
in CI users' follow-up controls at the beginning of every programming session.[4] Being a harmless and extremely rapid test, which requires no patient collaboration,
it is useful both in the intrasurgical testing and in postsurgical fitting. Normal
range of impedance values has been established by each CI brand. In patients who can
be considered good users with good auditory outcome, out-of-range values are clearly
reported by the fitting software, and they are a sign of electrode dysfunction or
loss of its integrity.[10]
Nonetheless, the analysis of in-range unexpected variations should be considered essential
since those variations could also be an early sign of electrode dysfunction or possible
changes in the environment surrounding the CI array.[5] Even in these cases, variations of impedance may be a predictor of inflammatory
processes within the cochlea, of progressive fibrosis/ossification, or of fluctuations
of the position of the electrode.[12]
[21] In these situations, not only speech perception is at risk,[19]
[22] but also sometimes patient health.
What can be understood from available literature is that impedance values of the whole
array were low when measured in intrasurgical setting.[23] In some cases, air bubbles could be adherent to the surface of the electrode determining
short-lasting, high-impedance spikes that underwent spontaneous reduction until the
bubbles were reabsorbed.[24] This can be prevented by a slow insertion dynamic.
According to some authors, impedance values increased in the first hours after surgery,[22]
[25] reaching the highest values at the activation (when it was performed some weeks
after surgery), probably because of the deposition of a layer of fibrous tissue around
the electrode. Inflammatory processes, exudation of proteins, or deposition of macrophages/fibroblasts
on the surface of the array were supposed to be the cause.[11]
[14]
[18]
[23]
[26]
[27] The constant electric stimulation (determined by the use of the CI) induced the
decrease in impedance values in the following weeks.[14]
[18]
[24]
[26] Those values were reported to be stable in the first year.[14]
[15] Activation within hours from surgery seems to not have a significant effect on impedance
values after 1 month post-surgery.[28] In addition, it is known that basal electrodes almost in all cases present in-range
but higher impedance values than those in the middle and apical electrodes of the
array.[10]
[29] This is supposed to be the consequence of the histologically documented chronic
inflammatory/fibrotic reaction involving inflammatory cells, fibrosis, and neo-ossification
affecting the part of the cochlea next to the site of the array surgical insertion.[30]
[31]
The present study confirmed these findings showing higher impedance values in the
four most basal electrodes. To note, all patients of our cohort had all electrodes
active and impedance telemetry was measured for each of them. As all surgeries were
performed by the same surgeon, it is possible that this may be due to the surgical
technique. Another explanation can be the use of low current levels frequently applied
to basal electrodes (responsible for representing high-frequency sounds, frequently
less tolerated by patients). Only one study,[32] conducted on patients with a different CI brand, showed increased impedance values
in the apical part of the array, which could be due either to the different characteristics
of that array or to the trauma of the surgical insertion.
Early activation may have an impact on the inflammatory processes involving the array's
surface, thus determining different patterns of impedance-value trend.[24] Soft surgery technique seems to have a positive influence in reducing postsurgical
impedance levels; according to some authors,[13] this can be related to the reduced production of fibrotic tissue due to the atraumatic
insertion of the array within the cochlea. Also, the choice of the electrode may have
an influence on impedance levels. It is likely that different electrode arrays cause
varied micro-damage of the intracochlear structures and different distances of the
array from the modiolus, which may impact impedance levels.[33] For example, perimodiolar and lateral wall electrodes may vary in distance from
the modiolus.[17] In addition, the use of a dexamethasone-eluting Cochlear Contour Advance electrode
has been shown to determine different impedance values than standard electrodes.[34]
[35] These data suggest that intracochlear inflammation is present, and it can be modulated
and it may influence impedance levels.
As previously highlighted,[24] the present study confirms that, among all the modalities used to measure impedance,
CG provides the lowest values. This should be considered the standard result during
the fitting of the CI when impedance values in all the four modalities are measured
to check the integrity of the system.
In literature, long-term data are still missing, even if their importance can be considered
as crucial. The considered population was composed of congenitally deaf patients.
These patients currently undergo cochlear implantation around the first year of life
and they are expected to live for several decades with their devices. In addition,
the optimization of the CI fitting is frequently reached within the first year after
implantation and subsequent follow-ups are spread over time, with minimal need for
further variations in terms of current levels required for each electrode. Moreover,
minor changes of hearing performance are frequently difficult to be detected by patients
and impedance variations may be an early sign of pathologic processes that can go
unnoticed until clinical symptoms arise. Consequently, follow-ups should be performed
even in long-term CI recipients. In cases without symptoms and with self-perceived
positive outcome, in the future, impedance telemetry may be performed remotely via
telehealth.[36] According to the new long-term data from the present study, even in-range but small
variations should draw clinician attention. In these cases, telemetric data need to
be considered in the context of the patient's medical history, otoscopic evaluation,
and hearing tests results; in addition, close follow-up controls should be considered.
In the case of further variations, even imaging (by means of computed tomography)
and/or surgical inspection should be considered.
The novelties provided by the present study are the relatively large number of patients,
the homogeneity of the cohort in terms of etiology of the hearing loss, and the long
time-span considered in comparison with the existing studies in literature. On the
contrary, the limitations are the heterogeneity in terms of type of implants (even
if all patients had a perimodiolar electrode) and the limited number of patients with
a 5-year follow-up.
In conclusion, our findings show that impedance values are extremely stable at least
in the first 5 years after implantation in CI recipients with genetically determined
congenital profound hearing loss. These data seem to confirm that minimal modifications
occur in the fluid and tissue around the electrode in absence of specific external
stimuli known to have an impact on impedance telemetry (medications, autoimmune disorders,
etc.).
In absence of these factors, even minimal differences in terms of impedance telemetry
should be considered with caution. A strict follow-up might be advisable to prevent
a late diagnosis of electrode dysfunction or disorder affecting the cochlea.
Future studies are necessary to investigate the association between different hearing-loss
etiologies and trends of impedance values in CI users.