Optimal Hair Transplant Recipient Site Slit Design: Minimizing Vascular Damage
- Abstract
- Introduction
- Anatomy of Scalp Vasculature
- Shape of Instrument
- Angle of Slit
- Slit Orientation—Coronal Versus Sagittal
- Conclusion
- References
Abstract
An often overlooked aspect of hair transplantation is the art of recipient site design and slit creation. There is also a lack of consensus on which technique provides the optimum coverage while minimizing vascular damage. This paper aims to provide logical arguments to determine the optimal instrument and method of slit creation, in order to ensure maximum density, optimal survival, minimal pop-out, and minimal damage to scalp vascularity.
The use of semiconical blades reduces the damage to the dermis and vascular plexus as compared with rectangular blades and needles, as the depth of penetration required is lower. The use of acute angle reduces the depth of penetration for the same length of slit and decreases damage to deep plexus. Coronal slits produce less vascular damage than that of sagittal slits with the same size blades.
We believe that these recommendations provide the optimum volume slits while causing minimal vascular damage.
#
Introduction
An often underlooked aspect of hair transplantation is the art of recipient site design and slit creation. There have been multiple articles published on optimizing donor site harvest and graft survival.[1] [2] [3] However, it is ultimately the recipient site design that dictates the aesthetics of the result.
The main methods of implantation are using forceps with slit creation or using implanters. The technique using forceps can work with premade slits/recipient sites or simultaneous creation of recipient site and implantation known as stick and place.[4] Premade slit creation is generally favored, especially in high-volume centers for their efficiency and also for the fact that significant part of the surgery (implantation in premade slits) could be delegated to assistants. There are multiple methods of creating slits. The most commonly used instruments include needles and blades, either for a single slit or multislit creation.[5] [6] [7] [8] The orientation can be coronal or sagittal.[9] The angle and depth of the slit itself can also vary.[10]
However, with either of these techniques, the actual mechanics of slit creation has not been analyzed. There is also a lack of consensus on which technique provides the optimum coverage while minimizing vascular damage.
The key factors determining graft survival are graft viability and scalp vascularity. Better scalp vascularity ensures better graft survival. This paper aims to provide logical arguments, in order to determine the optimal instrument and method of slit creation to ensure maximum density, optimal survival, minimal pop-out, and minimal damage to scalp vascularity.
#
Anatomy of Scalp Vasculature
The main vessels supplying the scalp include the supratrochlear, supraorbital, superficial temporal, posterior auricular, and occipital vessels. These form an intricate network in the form of a deep plexus of vessels. This plexus gives out vertical communicating vessels which connect to a superficial subpapillary plexus. The superficial plexus gives out capillary loops into the dermal papilla ([Fig. 1]).[11]


The deep plexus consists of large vessels which supply a correspondingly large area of skin. Any damage here can result in both ischemia and necrosis, adversely affecting the outcome. Hence, during slit making, it is imperative not to damage the deep plexus. This means our slits should not penetrate too deep to cause damage. One study quantified the average scalp thickness as 4.3 mm.[12] One important factor here is that skin thickness reduces with alopecia, which also has to be kept in mind.
The depth of slit required should be slightly shorter than that of the graft. Very deep slits can lead to the burial of the graft, causing folliculitis. Very superficial slits can cause popping out and cobble stoning.
With this understanding, we will analyze the different dimensions of slit creation to ascertain what would cause minimal damage to vascularity and ensure maximal graft survival.
#
Shape of Instrument
The shape of the instrument plays a role in determining incision volume, which is a more important factor in determining vascular damage than incision length, which ignores the three-dimensional nature of the vascularity. The shape of the slit should mimic the shape of the graft for proper accommodation and avoid unnecessary vascular damage. A typical graft is roughly cylindrical with an element of splay at the lower end near the roots. The upper part of the dermis contains a higher degree of collagen and elastin fibers, which are responsible for the elasticity of the skin. The elasticity of the slit depends on this portion of skin. If more volume of these fibers are cut, then the elasticity will reduce and, in turn, pop-out rate will increase. Hypodermis contains loose collagen and elastin fibers, so the elasticity is less, and it can easily accommodate the larger size bottom part of a hair follicle. Features of different instruments used to make slits are as follows:
Needles
Needles have a bevel that is used to make slits. The bevel has a wide upper segment and a narrow lower segment. The length of the bevel varies with the size of the needle ([Table 1]).
Needle size |
Diameter |
Bevel size |
---|---|---|
18 G |
1.28 mm |
5.25 mm |
19G |
1.06 mm |
4.75 mm |
20G |
0.90 mm |
4.25 mm |
The bevel of the needle has two edges—inner and outer. The lower conical part of the bevel of the needle helps in penetration. The outer edges are responsible for cutting till it starts curving inward, and then as the needle penetrates further, there will be more dilatation than cutting. The needle will create a bigger slit than its width, but the amount of the collagen and elastin fibers cut is less in the upper dermis, which helps in maintaining better elasticity and, in turn, less pop-out rate.
The slit created by the conical part of the bevel is narrow and hence not useful to accommodate the lower end of the graft. Hence, the needle has to penetrate deeper, causing unnecessary vascular damage. If this is not done, then the graft ends up in the upper part of the slit and might pop-out or cause cobble stoning ([Fig. 2]).


#
Rectangular Blades
A straight edges blade, on the other hand, is wide at its lower end; hence, it has to penetrate less to accommodate the graft, thereby causing less vascular damage. In addition, it cuts more fibers in the upper dermis. So, the retaining power/elasticity of the slit is decreased, leading to an increased pop-out rate.
#
Conical Blades
These have to penetrate more to create the required size slit (similar to a needle). In addition, it cuts more fibers in the upper part, which leads to a greater pop-out rate.
#
Semiconical Blades
A semiconical knife has a narrow cutting edge at the bottom which cuts the entire slit, but as it penetrates deeper, the blunt tapered sides dilates more than it cuts. So, upper part of slit has less amount of damage to fibers, which will reduce the pop-out rate. The depth of penetration required for an optimum size slit is also less, which reduces vascular damage. Further, the narrow lower part of the knife also damages the vascularity less than a rectangular knife. Lower dermis and hypodermis containing less collagen and elastin fibers will easily accommodate the lower portion of the graft, while the upper dermis will help in retaining the graft ([Figs. 3] and [4]).




Inference—Semiconical blades offer the ideal slit with minimal vascular damage.
#
#
Angle of Slit
The next factor in slit making is the angle at which the slit is made. As the angle of entry becomes more acute, the depth of penetration for the same length of slit made reduces, as explained below ([Fig. 5]).


We know that sin theta = opposite side/hypotenuse. In our situation, sin (angle of entry) = depth/length of slit. Using this formula, the depths of penetration for a slit of 5-mm length at various angles is shown in ([Table 2]).
Angle (in degree) |
Depth (in mm) |
Depth reduction (in percentage) |
---|---|---|
90 |
5 |
0 |
75 |
4.8 |
4 |
60 |
4.3 |
14 |
45 |
3.5 |
30 |
30 |
2.5 |
50 |
15 |
1.3 |
74 |
This equation shows that the lower the entry angle, the lesser the depth of penetration and lesser the chance of deep vascular injury. Hence, we should not use vertical angle and make our angle acute as far as possible, mimicking the natural exit of hairs angling from 15 to 60 degrees.
Hence, we infer that the more acute the angle, the less the depth of penetration ([Fig. 6]).


#
Slit Orientation—Coronal Versus Sagittal
The next factor to consider in slit making is whether the surface orientation of the slit is coronal or sagittal. Coronal versus sagittal is an old debate. Coronal slit making was recommended as a better way to provide a better orientation of grafted hairs with respect to light reflectance; thereby, creating a better visual impact. However, the sagittal slit was preferred by many as the more natural orientation and less likely to interfere with the vasculature.
We would like to consider a different factor—the surface incision, and whether it is the same as the width of the instrument.
In sagittal slits, the width of the blade is in the same axis as the angle of the slit. Hence, the length of the surface incision varies with the angle of the slit, similar to the above equation. Therefore, the following table shows the variation of surface incision for a blade of 1-mm width with angle ([Fig. 7] and [Table 3])
Angle |
Surface incision |
Increase (%) |
---|---|---|
90 |
1 |
0 |
75 |
1.035 |
3.5 |
60 |
1.155 |
16 |
45 |
1.415 |
41 |
30 |
2 |
100 |
15 |
3.85 |
285 |


As we can see, at more acute angles, the surface incision increases in length significantly. This increases the trauma to the superficial plexus, thereby offsetting the benefits of the lower depth of penetration achieved at more acute angles. Also, the wider surface incision would decrease the density of packing.
In coronal slits, the width of the blade is perpendicular to the axis of the angle of the slit. Hence, the length of the surface incision does not change and is always the same as the width of the slit. This reduces the trauma to the superficial plexus, while also preserving the deep plexus, as the angle becomes more acute. Hence, coronal slit would have this advantage of remaining the same width even after angulation.
We infer that the coronal slits do not increase the length of surface incision, while sagittal slits increase the surface incision as the angle reduces.
#
Conclusion
An ideal slit is one that allows easy implantation at an optimum depth without much pop-out, with minimal vascular damage at optimal density.
The slit technique which causes minimal vascular damage includes the following:
-
Use of semiconical blades will reduce the damage to the dermis and vascular plexus as compared with rectangular blades and needles, as the depth of penetration required is lower.
-
Use of acute angle to reduce the depth of penetration for the same length of slit and reduce damage to deep plexus.
-
Use smaller blades to create desired size slits in sagittal slit making.
-
Coronal slits will have lesser vascular damage than that of sagittal slits, regardless of whether one is using a single knife or multiple knives.
We believe that these recommendations provide the optimum volume slits while causing minimal vascular damage.
#
#
Conflict of Interest
All the authors wish to report no conflicts of interest.
-
References
- 1 Rose PT. Advances in hair restoration. Dermatol Clin 2018; 36 (01) 57-62
- 2 Garg AK, Garg S. Donor harvesting: follicular unit excision. J Cutan Aesthet Surg 2018; 11 (04) 195-201
- 3 Sharma R, Ranjan A. Follicular unit extraction (FUE) hair transplant: curves ahead. J Maxillofac Oral Surg 2019; 18 (04) 509-517
- 4 Marwah MK, Mysore V. Recipient Area. J Cutan Aesthet Surg 2018; 11 (04) 202-210
- 5 Wells A. “Use of blades of 0.7 to 1mm in diameter and coronal incisions in hair transplantation: Surgical approach to improve density and natural results”. 2014
- 6 Atodaria P. An innovative instrument for reducing the surgeon's time in making recipient sites. Hair Transplant Forum International 2014; 24 (04) 131-132
- 7 Atodaria PR, Atodaria KP. Modifications of Multi Slit Knife for enhanced versatility in balding and vertex area hair transplant. J Cutan Aesthet Surg 2020; 13 (03) 247-250
- 8 Atodaria PR, Atodaria KP. Reducing surgical time: an innovative and versatile instrument for creating multiple brick-pattern sagittal recipient sites in a single stroke. Hair Transplant Forum International. 2020; 30 (05) 180-181
- 9 Ginzburg A, Agrawal S. Alternate coronal and sagittal slits in recipient area of hair transplantation to ensure the correct graft implantation by assisting surgeon. J Am Acad Dermatol 2020; 83 (03) e189-e190
- 10 Nakatsui T, Wong J, Groot D. Survival of densely packed follicular unit grafts using the lateral slit technique. Dermatol Surg 2008; 34 (08) 1016-1022 , discussion 1022–1025
- 11 Cormia FE. Vasculature of the normal scalp. Arch Dermatol 1963; 88: 692-701
- 12 Hori H, Moretti G, Rebora A, Crovato F. The thickness of human scalp: normal and bald. J Invest Dermatol 1972; 58 (06) 396-399
Address for correspondence
Publikationsverlauf
Artikel online veröffentlicht:
31. Dezember 2021
© 2021. Association of Plastic Surgeons of India. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Rose PT. Advances in hair restoration. Dermatol Clin 2018; 36 (01) 57-62
- 2 Garg AK, Garg S. Donor harvesting: follicular unit excision. J Cutan Aesthet Surg 2018; 11 (04) 195-201
- 3 Sharma R, Ranjan A. Follicular unit extraction (FUE) hair transplant: curves ahead. J Maxillofac Oral Surg 2019; 18 (04) 509-517
- 4 Marwah MK, Mysore V. Recipient Area. J Cutan Aesthet Surg 2018; 11 (04) 202-210
- 5 Wells A. “Use of blades of 0.7 to 1mm in diameter and coronal incisions in hair transplantation: Surgical approach to improve density and natural results”. 2014
- 6 Atodaria P. An innovative instrument for reducing the surgeon's time in making recipient sites. Hair Transplant Forum International 2014; 24 (04) 131-132
- 7 Atodaria PR, Atodaria KP. Modifications of Multi Slit Knife for enhanced versatility in balding and vertex area hair transplant. J Cutan Aesthet Surg 2020; 13 (03) 247-250
- 8 Atodaria PR, Atodaria KP. Reducing surgical time: an innovative and versatile instrument for creating multiple brick-pattern sagittal recipient sites in a single stroke. Hair Transplant Forum International. 2020; 30 (05) 180-181
- 9 Ginzburg A, Agrawal S. Alternate coronal and sagittal slits in recipient area of hair transplantation to ensure the correct graft implantation by assisting surgeon. J Am Acad Dermatol 2020; 83 (03) e189-e190
- 10 Nakatsui T, Wong J, Groot D. Survival of densely packed follicular unit grafts using the lateral slit technique. Dermatol Surg 2008; 34 (08) 1016-1022 , discussion 1022–1025
- 11 Cormia FE. Vasculature of the normal scalp. Arch Dermatol 1963; 88: 692-701
- 12 Hori H, Moretti G, Rebora A, Crovato F. The thickness of human scalp: normal and bald. J Invest Dermatol 1972; 58 (06) 396-399